← Product Code [LZA](/productcode/LZA) · K171616

# Xingyu Powder Free Nitrile Patient Examination Gloves, Blue color (K171616)

_Shandong Xingyu Gloves Co., Ltd. · LZA · Jul 20, 2017 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K171616

## Device Facts

- **Applicant:** Shandong Xingyu Gloves Co., Ltd.
- **Product Code:** [LZA](/productcode/LZA.md)
- **Decision Date:** Jul 20, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

Xingyu Powder Free Nitrile Patient Examination Gloves, Blue Color is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

## Device Story

Nitrile examination glove; acts as physical barrier to prevent contamination between patient and examiner. Device worn on hand/finger during medical procedures. Nitrile material provides water-tight barrier; tensile properties allow hand movement. Used in clinical settings by healthcare providers. Output is a protective barrier; affects clinical decision-making by maintaining hygiene standards and reducing infection risk. No electronic or software components.

## Clinical Evidence

No clinical data provided. Substantial equivalence demonstrated via bench testing including dimensions, tensile strength, elongation, water leak (pinhole) testing, powder residual, and biocompatibility (skin irritation and sensitization).

## Technological Characteristics

Nitrile rubber material. Non-sterile. Dimensions and physical properties conform to ASTM D6319-10 (Reapproved 2015). Freedom from pinholes per ASTM D5151-06 (Reapproved 2015) and 21 CFR 800.20. Powder residual per ASTM D6124-06 (Reaffirmation 2011). Biocompatibility per ISO 10993-10:2010.

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Powder Free Nitrile Patient Examination Glove, Blue Color ([K120970](/device/K120970.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features the department's emblem, which consists of a stylized caduceus-like symbol with three intertwined figures, representing health and well-being. The emblem is surrounded by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 20, 2017

Shandong Xingyu Gloves Co., Ltd % Chu Xiaoan Official Correspondent Beijing Easylink Co,Ltd Rm. F302 Bldg., 41, Jing Cheng Ya Ju, Courtyard 6 of Southern Dou Ge Zhuang Beijing, 100121 Cn

Re: K171616

Trade/Device Name: Xingyu Powder Free Nitrile Patient Examination Gloves, Blue color Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LZA Dated: May 18, 2017 Received: June 1, 2017

Dear Chu Xiaoan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

# Tara A. Ryan -S

for Lori Wiggins Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

510(k) Number (if known)

#### K171616

Device Name

XingYu Powder Free Nitrile Patient Examination Gloves, Blue Color

Indications for Use (Describe)

Xing Yu Powder Free Nitrile Patient Examination Gloves, Blue Color is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)

] Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{3}------------------------------------------------

# 510(K) Summary

"The assigned 510(k) number is:

### Premarket Notification [510(k)] Summary

#### 1.0 Submitter:

| Submitter's name :      | Shandong Xingyu Gloves Co.,Ltd                                                         |
|-------------------------|----------------------------------------------------------------------------------------|
| Submitter's address :   | No.2158, Yaoqian Road,Economic<br>Development Zone,Gaomi<br>City,Shandong,261502,China |
| Phone number :          | 0086-536-2588123                                                                       |
| Fax number :            | 0086-536-2586328                                                                       |
| Name of contact person: | Xu Qiang                                                                               |
| Date of preparation :   | 2017-07-18                                                                             |

#### 2.0 Name of the Device

| Proprietary/Trade name: | Xingyu Powder Free Nitrile Patient<br>Examination Gloves, Blue Color |
|-------------------------|----------------------------------------------------------------------|
| Common Name:            | Exam gloves                                                          |
| Classification Name:    | Patient examination glove                                            |
| Device Classification:  | I                                                                    |
| Regulation Number:      | 21 CFR 880.6250                                                      |
| Panel:                  | General Hospital                                                     |
| Product Code:           | LZA                                                                  |

### 3.0 Predicate device

| Device Name:   | Powder Free Nitrile Patient Examination<br>Glove, Blue Color |
|----------------|--------------------------------------------------------------|
| Company name:  | Tangshan Zhonghong Pulin Plastic Co., Ltd.                   |
| 510(K) Number: | K120970                                                      |

### 4.0 Device Description:

#### 4.1 How the device functions:

Nitrile films form a barrier to prevent contamination between patient and examiner.

#### 4.2 Scientific concepts that form the basis for the device

The nitrile rubber is water tight under normal conditions of use. It's tensile properties cause it to conform to the hand, allowing movements necessary for a medical procedure.

#### 4.3 Physical and performance characteristics such as design, materials and physical properties:

{4}------------------------------------------------

The Nitrile glove acts as a barrier to prevent contamination between patient and examiner. The glove is manufactured in accordance with the requirements of ASTM D6319 and ASTM D5151.

### 5.0 Device Indication for use:

Xingyu Powder Free Nitrile Patient Examination Gloves, Blue Color is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

### 6.0 Summary of the Technological Characteristics of the Device:

Xingyu Powder Free Nitrile Patient Examination Gloves, Blue Color, non-sterile are summarized with the following technological characteristics compared to ASTM or equivalent standard.

| Characteristics          | Standard                                                                                                                                                                                                     |                                                                                                |
|--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
| Dimension                | ASTM standard D 6319-10(Reapproved 2015)                                                                                                                                                                     |                                                                                                |
|                          | Length                                                                                                                                                                                                       | $\geq$ 230mm                                                                                   |
|                          | Width                                                                                                                                                                                                        | Small 70-90 mm                                                                                 |
|                          |                                                                                                                                                                                                              | Medium 85-105mm                                                                                |
|                          |                                                                                                                                                                                                              | Large 100-120mm<br>X large 110-130 mm                                                          |
| Thickness                | Fingertip $\geq$ 0.05mm                                                                                                                                                                                      |                                                                                                |
|                          | Palm $\geq$ 0.05mm                                                                                                                                                                                           |                                                                                                |
| Physical<br>Properties   | ASTM standard D 6319-10(Reapproved 2015)<br>Tensile strength(Before aging) $\geq$ 14MPa<br>Tensile strength(After aging)<br>Elongated rate(Before aging) $\geq$ 500%<br>Elongated rate(After aging) $>$ 400% |                                                                                                |
| Freedom from<br>pinholes | 21 CFR 800.20                                                                                                                                                                                                |                                                                                                |
|                          | ASTM standard D 6319-10(Reapproved 2015).<br>Test method in accordance with ASTM D5151-06(Reapproved 2015)                                                                                                   | Passed Standard Acceptance Criteria                                                            |
| Powder Residual          | ASTM standard D 6319-10(Reapproved 2015)<br>Test method in accordance with D6124-06(Reaffirmation 2011)                                                                                                      | Meets 0.1mg/glove                                                                              |
| Biocompatibility         | Primary Skin Irritation in rabbits ISO 10993-10: Third Edition 2010-08-01                                                                                                                                    | Passes<br>Under the conditions of the study, the subject device is not a primary skin irritant |
|                          | Dermal sensitization in the guinea pig ISO 10993-10: Third Edition 2010-08-01                                                                                                                                | Passes<br>Under the conditions of the study, the subject device is not a skin sensitizer       |

### 7.1 Substantial Equivalent Based on Assessment of Non-Clinical Performance Data:

The following bench testing was conducted for design elements and performance characteristics deemed appropriate to demonstrate equivalence to the predicate device. The Powder Free Nitrile Patient Examination Gloves, Blue Color made by Shandong

{5}------------------------------------------------

Xingyu Gloves Co.,Ltd met the predetermined acceptance criteria ensuring substantial equivalence to the predicate device. No new safety or performance issues were raised during testing.

- Dimension per ASTM D6319-10(Reapproved 2015),
- . Tensile strength (Before aging/ After aging) and Elongated rate (Before aging/ After aging) per ASTM D6319-10(Reapproved 2015),
- . Water leak test on pinhole per ASTM D6319-10(Reapproved 2015) and per 21 CFR 800.20,
- . Powder Residual tests per ASTM D6319-10(Reapproved 2015)
- . Biocompatibility test per ISO 10993-10: Third Edition 2010-08-01.

#### 8.0 Substantial Equivalent Based on Assessment of Clinical Performance Data:

Clinical data was not needed to demonstrate that the subject glove is substantially equivalent to the predicate glove. So determination of substantial equivalence is not based on an assessment of clinical performance data.

| Features &<br>Description                   | Predicate Device                                                                                                                                                                                                                            | Subject Device                                                                                                                                                                                                                                  | Result of<br>Comparison |
|---------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|
| Company                                     | Tangshan Zhonghong Pulin<br>Plastic Co., Ltd.                                                                                                                                                                                               | Shandong Xingyu Gloves<br>Co.,Ltd                                                                                                                                                                                                               | --                      |
| 510(K) Number                               | K120970                                                                                                                                                                                                                                     | K171616                                                                                                                                                                                                                                         |                         |
| Product name                                | Powder Free Nitrile Patient<br>Examination Glove, Blue<br>Color                                                                                                                                                                             | Xingyu Powder Free<br>Nitrile Patient<br>Examination Gloves, Blue                                                                                                                                                                               | similar                 |
| Product Code                                | LZA                                                                                                                                                                                                                                         | LZA                                                                                                                                                                                                                                             | same                    |
| Size                                        | Small/ Medium/<br>Large/X large                                                                                                                                                                                                             | Small/ Medium/<br>Large/X large                                                                                                                                                                                                                 | same                    |
| Indications for<br>Use                      | Powder Free Nitrile Patient<br>Examination Glove, Blue<br>Color is a disposable device<br>intended for medical<br>purposes that is worn on the<br>examiner's hand or finger to<br>prevent contamination<br>between patient and<br>examiner. | Powder Free Nitrile<br>Patient Examination<br>Gloves, Blue Color is a<br>disposable device<br>intended for medical<br>purposes that is worn on<br>the examiner's hand or<br>finger to prevent<br>contamination<br>between patient and examiner. | similar                 |
| Device<br>Description and<br>Specifications | Meets ASTM D6319-10                                                                                                                                                                                                                         | Meets ASTM<br>D6319-10(Reapproved<br>2015)                                                                                                                                                                                                      | similar                 |
| Dimensions<br>--Length<br>ILS-2<br>AQL4.0   | Meets ASTM<br>D6319-10<br>≥230mm min                                                                                                                                                                                                        | 232 mm min for all sizes                                                                                                                                                                                                                        | similar                 |
| Dimensions<br>-- Width<br>IL S-2<br>AQL4.0  | Meets ASTM D6319-10<br>Small 70-90 mm<br>Medium 85-105mm<br>Large 100-120mm<br>X large 110-130 mm                                                                                                                                           | Small 76-88 mm<br>Medium 89-102 mm<br>Large 108-119mm<br>X large 115-128 mm                                                                                                                                                                     | similar                 |

### 9.0 Substantial Equivalence Comparison:

{6}------------------------------------------------

| Dimensions<br>--Thickness<br>IL S-2<br>AQL4.0                           | Meets ASTM D6319-10                                                                                   |                                                                                                                            | Thickness (mm) min.      |         | similar |
|-------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|--------------------------|---------|---------|
|                                                                         | Finger 0.05mm min.<br>Palm 0.05mm min.                                                                |                                                                                                                            | Finger 0.08<br>Palm 0.08 |         |         |
| Physical<br>Properties<br>IL S-2<br>AQL4.0                              | Meets ASTM D D6319-10                                                                                 | Aging                                                                                                                      | Before                   | After   | similar |
|                                                                         | Before aging/after aging<br>Tensile Strength≥ 14MPa                                                   | Elongation<br>(%)                                                                                                          | 530-600                  | 460-580 |         |
|                                                                         | Before aging<br>Elongation ≥500%<br>After aging<br>Elongation ≥400%                                   | Tensile<br>Strength<br>(MPa)                                                                                               | 18-25                    | 17-22   |         |
| Freedom from<br>Pinholes                                                | Meets<br>• 21 CFR 800.20<br>• ASTM D6319-10                                                           | Meets<br>• 21 CFR 800.20<br>• ASTM D<br>6319-10(Reapproved<br>2015)                                                        |                          |         | same    |
| Inspection<br>Level I AQL2.5                                            | Test method in accordance<br>with ASTM D5151-06<br>(Reapproved 2015)                                  | Tested in accordance with<br>ASTM D5151<br>(Reapproved 2015) with<br>acceptable results                                    |                          |         |         |
| Residual<br>Powder                                                      | Meets ASTM<br>D 6124-06<br>(Reaffirmation 2011)                                                       | Meets ASTM<br>D 6124-06<br>(Reaffirmation 2011)                                                                            |                          |         | same    |
|                                                                         | below 2mg of residual<br>powder                                                                       | Results generated values<br>below 2mg of residual<br>powder                                                                |                          |         |         |
| Materials used<br>to fabricate the<br>devices                           | Nitrile                                                                                               | Nitrile                                                                                                                    |                          |         | same    |
| Dusting or<br>Donning<br>Powder: name                                   | Surface Coating Agent                                                                                 | Surface Coating Agent                                                                                                      |                          |         | same    |
| Compare<br>performance<br>data supporting<br>substantial<br>equivalence | Meets<br>ASTM D5151-06<br>(Reapproved 2015)<br>ASTM D6319-10<br>ASTM D6124-06<br>(Reaffirmation 2011) | Meets<br>ASTM D5151-06<br>(Reapproved 2015)<br>ASTM D6319-10<br>(Reapproved 2015)<br>ASTM D6124-06<br>(Reaffirmation 2011) |                          |         | similar |
| Single Patient<br>Use                                                   | Single Patient Use                                                                                    | Single Patient Use                                                                                                         |                          |         | same    |
| Biocompatibility                                                        | Under the conditions of<br>this study, the test article<br>was a non- irritant or non-<br>sensitizer. | Under the conditions of<br>this study, the test article<br>was a non- irritant or<br>non- sensitizer.                      |                          |         | same    |
|                                                                         | SKIN IRRITATION<br>DERMAL and<br>SENSITIZATION<br>STUDIES Meets ISO<br>10993-10:2002/Amd.1:20<br>06   | SKIN IRRITATION<br>DERMAL and<br>SENSITIZATION<br>STUDIES Meets ISO<br>10993-10: Third Edition<br>2010.08.01               |                          |         |         |

{7}------------------------------------------------

| Labeling for the | -Powder Free               | -Powder Free               | same |
|------------------|----------------------------|----------------------------|------|
| legally          | -Patient Examination Glove | -Patient Examination Glove |      |
| marketed device  | -Single Use Only           | -Single Use Only           |      |
| to which         | - Manufactured For:        | - Manufactured For:        |      |
| substantial      | - Lot                      | - Lot                      |      |
| equivalence is   | -Blue color                | -Blue color                |      |
| claimed.         | - Non sterile              | - Non sterile              |      |
|                  |                            |                            |      |

#### Conclusion: 10.0

Based on the nonclinical tests data, it can be concluded that the Xingyu Powder Free Nitrile Patient Examination Gloves, Blue Color is as safe, as effective, and performs as well as the predicate device, Powder Free Nitrile Patient Examination Glove, Blue Color by Tangshan Zhonghong Pulin Plastic Co., Ltd., K120970.

---

**Source:** [https://fda-staging.innolitics.com/device/K171616](https://fda-staging.innolitics.com/device/K171616)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).