To be used for temporary relief of pain associated with sore and aching muscles in the back due to strain from exercise or normal household and work activities and relief of pain of the upper and lower back associated with arthritis.
Device Story
Battery-powered TENS device; provides electrical stimulation to peripheral nerves for back pain relief. Comprised of TENS generator with integral electrodes, replaceable hydrogel pads, and optional wireless remote. Controlled via onboard buttons or smartphone app (iOS/Android) via Bluetooth. User adheres device to lower back; selects from four 30-minute preprogrammed stimulation modes (WiTouch Pro Exclusive, High Frequency, Low Frequency, High-Low Combo) and intensity level. Output is asymmetrical biphasic waveform. Device intended for OTC use; provides non-invasive analgesic benefit. Healthcare provider role limited; patient self-operates.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including verification and validation of hardware and software performance, and adherence to recognized consensus standards.
Technological Characteristics
Battery-powered TENS; asymmetrical biphasic waveform (0 net DC charge); 0-80mA amplitude; 120-250μs pulse width; 2-120 Hz pulse rate. Materials: replaceable electroconductive hydrogel pads. Connectivity: Bluetooth (mobile app) and RF (remote). Standards: IEC 62304, ANSI/AAMI ES60601-1, IEC 60601-1-11, IEC 60601-1-2, IEC 60601-2-10, ISO 14971. Software-controlled via microprocessor.
Indications for Use
Indicated for temporary relief of back pain associated with sore/aching muscles from exercise, household, or work activities, and relief of upper/lower back pain associated with arthritis.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, featuring three human profiles in a row.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 22, 2017
Hollywog LLC Michael Treas Chief Compliance Officer 2830 Amnicola Highway Chattanooga, Tennessee 37406
Re: K171599
Trade/Device Name: WiTouch Pro, WiTouch, Neubac Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: NUH, NYN Dated: August 22, 2017 Received: August 23, 2017
Dear Michael Treas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
William J. Heetderks -S 2017.09.22 11:06:27 -04'00'
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K171599
Device Name WiTouch Pro, WiTouch, Neubac
Indications for Use (Describe)
To be used for temporary relief of pain associated with sore and aching muscles in the back due to strain from exercise or normal household and work activities and relief of pain of the upper and lower back associated with arthritis.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 5. 510(k) Summary
## 5.1
| 510(k) NUMBER: | K171599 |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------|
| 510(k) SUBMITTER: | Hollywog, LLC<br>2830 Amnicola Highway<br>Chattanooga, TN 37406 |
| ESTABLISHMENT REGISTRATION: | 3008585473 |
| CONTACT: | Michael W Treas<br>Chief Compliance Officer,<br>Regulatory Affairs and Quality Assurance |
| TELEPHONE: | (423) 305-7778 |
| DATE PREPARED: | July 20, 2017 |
| PROPRIETARY NAMES: | WiTouch Pro / WiTouch / Neubac |
| PANEL: | Neurology |
| REGULATION NUMBER: | CFR Title 21, 882.5890 |
| CLASSIFICATION: | Class II |
| PRODUCT CODES: | NUH, NYN |
| COMMON NAME: | Transcutaneous electrical nerve stimulator for pain<br>relief |
| CLASSIFICATION NAME: | Stimulator, Nerve, Transcutaneous, Over-The-<br>Counter.<br>Stimulator, Electrical, Transcutaneous, For<br>Arthritis |
| PREDICATE DEVICE: | WiTouch Pro K120398 |
{4}------------------------------------------------
#### 5.2 Device Description
The second-generation WiTouch Pro (a.k.a. WiTouch/Neubac) device is a battery powered transcutaneous electrical nerve stimulator (TENS) device for relieving back pain. The device is comprised of a TENS generator with integral electrodes, two replaceable "AAA" size batteries, replaceable electroconductive hydrogel pads (Gel Pads), a Mobile App to control the TENS device via Bluetooth® connection, which installs to Apple® iOS® 9.0 or higher or Android® 4.4 or higher smartphone platforms. Additionally, the TENS intensity may be increased or decreased using onboard mechanical buttons on the TENS unit enclosure. The TENS device offers an optional handheld wireless remote control via Radio Frequency (RF) connection which is sold separate and comes with one replaceable CR2032 Lithium-ion coin battery.
The device accessories are the replaceable Gel Pads, the Mobile App, and the optional wireless handheld remote control. The Gel Pads serve to gently adhere the TENS unit to a user's skin on the lower back, and serves to dissipate electrical stimulation across the surface area of the electrode. The Gel Pads are intended for single person use with the number of uses dependent upon skin type, oils, and pH levels.
The TENS unit adheres to the user's lower back across the spine in the area where pain is perceived. Once placed, a user can choose from four 30-minute preprogrammed stimulation output modes, and the level of intensity that is most comfortable. The preprogrammed modes are identified as WiTouch Pro Exclusive, High Frequency, Low Frequency, and High-Low Combo.
Accessories:
- Pairs of replaceable hydrogel pads (Gel Pads) ●
- Mobile Medical Application control using a smartphone via Bluetooth® ● connection (Apple® iOS® or Android® platforms)
- Handheld wireless Remote Control.
#### 5.3 Intended Use
The intended use is to provide approximately thirty minutes of analgesic electrical stimulus to reduce the perception of pain by electrically stimulating peripheral nerves across healthy intact skin of the lower back.
#### 5.4 Indications for Use
To be used for temporary relief of pain associated with sore and aching muscles in the back due to strain from exercise or normal household and work activities and relief of pain of the upper and lower back associated with arthritis.
{5}------------------------------------------------
#### 5.5 Non-Clinical Testing
The verification, validation, performance, and human factors data presented in this 510(k) submission demonstrate the second-generation WiTouch Pro TENS device is substantially equivalent to the predicate device.
#### 5.6 Predicate Device
The predicate device is the first-generation WiTouch Pro TENS device which cleared to market in 510(k) K120398. The predicate device is substantially equivalent to the subject device identified in this 510(k) Summary which is the second-generation WiTouch Pro device. The differences between the predicate device and the subject device do not raise new questions about safety or effectiveness. The subject device is as safe and effective as the predicate device. The following table is a characteristics comparison between the subject device and the predicate device.
#### 5.6.1 Characteristics
| Characteristic | Subject Device – K# TBD<br>Hollywog, LLC<br>WiTouch®<br>Pro/WiTouch®/Neubac®<br>Models | Predicate Device - 510(k)<br>K120398<br>Hollywog, LLC<br>WiTouch® Pro<br>Model 11.1500 |
|-----------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | To be used for temporary relief<br>of pain associated with sore and<br>aching muscles in the back due<br>to strain from exercise or<br>normal household and work<br>activities and relief of pain of<br>the upper and lower back<br>associated with arthritis. | The WiTouch® Pro<br>Transcutaneous Electrical<br>Nerve Stimulator Device is<br>used for the symptomatic relief<br>and management of chronic<br>intractable back pain and relief<br>of pain of the upper and lower<br>back associated with arthritis.<br>It is also used for adjunctive<br>treatment for post-surgical and<br>post-trauma acute back pain. |
| Product Code | NUH, NYN<br>(OTC Use) | GZJ, NYN<br>(Prescription Use) |
| Anatomical Location | Upper and lower back | Upper and lower back |
| Power Source(s) | TENS Unit<br>(2) "AAA" Alkaline Batteries<br><br>Remote Control<br>(1) "CR 2032" Lithium-ion<br>coin battery | TENS Unit<br>(2) "AAA" Alkaline Batteries<br><br>Remote Control<br>(1) "CR 2032" Lithium-ion<br>coin battery |
| Low battery indicator | Yes | Yes |
| Average DC current<br>through electrodes when the<br>device is on but no pulses<br>are being applied (µA) | 0 | 0 |
{6}------------------------------------------------
| Number of Output<br>Preprogrammed Modes | 4<br>4 - TENS | 1<br>1 - TENS |
|-----------------------------------------------|----------------------------------------------------------------------------|----------------------------------------------------------------------------|
| Number of Output Channels | 1 | 1 |
| Regulated Current or Regulated<br>Voltage? | Voltage | Voltage |
| Waveform | Asymmetrical biphasic<br>(0 net DC charge) | Asymmetrical biphasic<br>(0 net DC charge) |
| Amplitude | 0 – 80mA<br>(Measured @ 500 ohm load) | 0 – 80mA<br>(Measured @ 500 ohm load) |
| Pulse Width | 120 - 250μs<br>(Preprogrammed) | 120 - 240μs<br>(Preprogrammed) |
| Pulse Rate | 2 - 120 Hz | 5 - 120 Hz |
| Software/Firmware/<br>Microprocessor Control? | Yes | Yes |
| Automatic Overload Trip? | No | No |
| Automatic No-Load Trip? | No | No |
| Automatic Shut Off? | Yes | Yes |
| User Override Control? | Yes<br>Off button immediately stops<br>treatment | Yes<br>Off button immediately stops<br>treatment |
| Indicator Display | On/Off Status? Yes<br>Low Battery? Yes<br>Voltage/Current Level? No | On/Off Status? Yes<br>Low Battery? Yes<br>Voltage/Current Level? No |
| Timer Range | Nonadjustable<br>30 minutes | Nonadjustable<br>30 minutes 42 seconds |
| Compliance with 21 CFR 898 | N/A, device does not contain<br>electrode lead wires or patient<br>cables. | N/A, device does not contain<br>electrode lead wires or<br>patient cables. |
| Weight (lbs., oz.) | 4.8 oz.<br>w/ batteries included | 4.8 oz.<br>w/ batteries included |
| Dimensions (in.)<br>[W x H x D] | 7.5 (w) x 3.5 (h) x 0.7 in (d) | 7.5 (w) x 3.5 (h) x 0.7 in (d) |
{7}------------------------------------------------
| Voluntary Standards | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| • FDA Recognition<br>Number 13-79: IEC<br>62304 Edition 1.1<br>2015-06<br>• FDA Recognition<br>Number 19-4:<br>ANSI/AAMI ES60601-<br>1:2005 Ed. 3 + C1:2009<br>+ A1:2012<br>• FDA Recognition<br>Number 19-16/19-7:<br>ANSI/AAMI<br>HA60601-1-<br>11:2015/2011<br>• FDA Recognition 19-8.<br>IEC 60601-1-2:2014<br>Ed. 4<br>• FDA Recognition<br>Number 19-14: IEC<br>60601-1-11 Edition 2.0<br>2015-01<br>• FDA Recognition<br>Number 17-11: IEC<br>60601-2-10 Edition 2.0<br>2012-06<br>• FDA Recognition<br>Number 5-40: ISO 14971<br>Second edition 2007-03-01 | • AAMI ANSI ES60601-<br>1:2005/(R)2012 And<br>A1:2012<br>• IEC 60601-1-2 Edition<br>2014-02<br>• IEC 60601-1-11 Edition<br>1.0 2010-04<br>• IEC 60601-2-10 Edition<br>1.0 2012-06<br>• AAMI/ANSI/ISO<br>10993-5:2009<br>• AAMI/ANSI/ISO 10993-<br>10:2010 |
#### 5.7 Conclusion
The basis for substantial equivalence for the referenced predicate device and for the second-generation WiTouch Pro TENS device is non-clinical data and conformity with recognized standards.
Clinical testing was not required to support substantial equivalence for the secondgeneration WiTouch Pro TENS device. This is due to the intended use is well established and understood, and the hardware and software verification and validation demonstrate that the second-generation WiTouch Pro TENS device performs comparably to the predicate device that is marketed for the same intended use.
Based on the performance testing and the similarities of the indications for use and the technological characteristics, it can be concluded that the second-generation WiTouch Pro TENS device is as safe and effective as, and substantially equivalent to, the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.