LUXEN Zr is indicated for the production of all-ceramic inlays, multi-units bridges, onlays, and veneers without zirconium dioxide frameworks. LUXEN Smile is indicated for the production of full ceramic crowns, onlays, 3bridges and inlay bridges (anterior and molar).
Device Story
LUXEN Zr and LUXEN Smile are zirconia-based dental restorative materials provided as blocks or disks for use with CAD/CAM systems. LUXEN Zr (3Y-TZP) and LUXEN Smile (5Y-TZP) are milled by dental technicians to create custom dental prosthetics, including crowns, bridges, inlays, and onlays. The devices are non-sterile and intended for permanent oral contact. The final restorations are placed by dental professionals to restore tooth function and aesthetics. The materials are selected based on shade and thickness requirements to match patient dentition.
Clinical Evidence
Bench testing only. Performance testing included uniformity, freedom from extraneous materials, radioactivity, chemical solubility, flexural strength, and linear thermal expansion per ISO 6872:2015. Biocompatibility testing conducted per ISO 10993-1, including cytotoxicity (ISO 10993-5), oral mucosal irritation (ISO 10993-10), sensitization (ISO 10993-10), and acute systemic toxicity (ISO 10993-11).
Technological Characteristics
Materials: 3Y-TZP (LUXEN Zr) and 5Y-TZP (LUXEN Smile) zirconia powders. Form factors: blocks, disks, Wieland, and D-95 types. Standards: ISO 6872:2015 (Type II, Class 5/4b), ISO 13356, ISO 10993-1. Non-sterile. CAD/CAM compatible.
Indications for Use
Indicated for dental patients requiring all-ceramic inlays, onlays, veneers, crowns, or multi-unit bridges (up to 3 elements) for tooth restoration.
Regulatory Classification
Identification
Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
DENTALMAX Co., Ltd Jindong Kim President 50-7, Pungsesandan 2-ro Pungse-myeon, Dongnam-gu, Cheonan-si Chungcheongnam-do, 31217 Kr
Re: K171585
Trade/Device Name: LUXEN Zr, LUXEN Smile Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain Powder For Clinical Use Regulatory Class: Class II Product Code: EIH Dated: September 25, 2018 Received: October 2, 2018
Dear Jindong Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
October 31, 2018
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, Digitally signed by Mary Mary S. Date: 2018.10.31 Runner -S3/ 08:38:16 -04'00' For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K180718
Device Name ORTHODONTIC PLASTIC BRACKETS AND ACCESSORIES
Indications for Use (Describe) ORTHODONTIC PLASTIC BRACKETS AND ACCESSORIES are indicated for orthodontic movement of natural teeth.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
# LUXEN Zr, LUXEN Smile
Date: October 29, 2018
## I. SUBMITTER
DENTALMAX Co., Ltd 50-7, Pungsesandan 2-ro, Pungse-myeon, Dongnam-gu, Cheonan-si, Chungcheongnam-do, 31217, Republic of Korea TEL : +82-41-523-3580 FAX : +82-41-523-3581 Contact Name: Kim, Jindong Email: jdkim@hedent.co.kr
### II. DEVICE NAME
| Name of Device: | LUXEN Zr, LUXEN Smile |
|----------------------------|-----------------------------------|
| Regulation Name: | Powder, Porcelain |
| Classification Name: | Porcelain powder for clinical use |
| Classification regulation: | 872.6660 |
| Regulatory Class: | 2 |
| Product Code: | EIH |
### III. PREDICATE DEVICE
K142987, DD Bio ZX2, Dental Direkt of Amerika UG K142987, DD cube X2, Dental Direkt of Amerika UG
#### IV. DEVICE DESCRIPTION
LUXEN Zr and LUXEN Smile are used for custom made dental restorations using a CAD/CAM system. The LUXEN Zr and LUXEN Smile are provided in 4 different shapes - block, disk, Wieland and D-95. The LUXEN Zr is composed of 3 mol% TZP. LUXEN Smile is composed of 5 mol% TZP. The weight of yttria is classified in accordance with ISO 6872:2015 Type2, class 5 for 3Y-TZP Zirconia, and ISO 6872:2015 Type2, class 4b for 5Y-TZP Zirconia, respectively.
{4}------------------------------------------------
There are various color and thickness for LUXEN Zr and LUXEN Smile; LUXEN Zr
- . Shade
- White, S1, S2, S3, E0, E1, E2, E3, E4, Multi-ivory
- . Thickness
- Block type: 10, 12, 14, 15, 16, 18, 20, 22, 25
- Disk type: 10, 12, 14, 16, 18, 20, 22, 25
- Wieland type / D95 type: 10, 12, 14, 16, 18, 20, 22, 25, 30
LUXEN Smile
- . Shade
- White, S1, S2, S3, E0, E1, E2, E3, E4, Multi-white, Multi-ivory
- . Thickness
- Block type: 16
- Disk type: 10, 12, 14, 16, 18, 20, 22, 25
- Wieland type / D95 type: 10, 12, 14, 16, 18, 20, 22, 25, 30
# V. INDICATIONS FOR USE
LUXEN Zr is indicated for the production of all-ceramic inlays, multi-units bridges, onlays, and veneers without zirconium dioxide frameworks.
LUXEN Smile is indicated for the production of full ceramic crowns, onlays, 3bridges and inlay bridges (anterior and molar).
# VI. COMPARISON TO PREDICATE DEVICE
LUXEN Zr
| Characteristics | Subject Device | Predicate Device | Summary |
|-----------------|---------------------|--------------------------------|---------|
| Device Name | LUXEN Zr | DD Bio ZX2 | - |
| Manufacturer | DENTALMAX CO., LTD. | Dental Direkt of Amerika<br>UG | - |
| 510(k) Number | K171585 | K142987 | - |
{5}------------------------------------------------
| Classification Name | Porcelain powder for clinical use | Porcelain powder for clinical use | Both products are porcelain powder for clinical use. |
|-----------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | EIH | EIH | Both products are EIH. |
| Device Class | II | II | Both products are class II. |
| Indications for Use | LUXEN Zr is indicated for the production of all-ceramic inlays, multi-units bridges, onlays, and veneers without zirconium dioxide frameworks. | DD Bio Z-dental blanks are indicated for crowns, multi-unit bridges and inlay bridges. Applications include both anterior and posterior bridges. | Both products are indicated for dental restorations. |
| Blank configuration | Block type<br>Disk type<br>Wieland type<br>D-95 type | Block type<br>Disk type<br>Wieland type<br>D-95 type | Both products are block, disk, wieland and D-95 types. |
| Thicknesses | 10 mm to 25 mm | 10 mm to 25 mm | Both products are 10-25mm of thickness. |
| Crystal Morphology | 3Y-TZP | 3Y-TZP | Both products are 3Y-TZP. |
| Color | A0, A1, A2, A3, B1, B2, B3, B4, C4, B1-B2-B3-B4 LAYER | A0, A1, A2, A3, A3.5, A4, B1, B2, B3, B4, C1, C2, C3, C4, D2, D3, D4 | Similar<br>Both product are met with VITA shade guide. Shade is an option that is selected according to patient's teeth color. Application of appropriate color that matches patient's teeth tone can be possible. |
| Sintering temperature | LUXEN Zr ST : 1580 °C<br>Others : 1500 °C | 1450 °C | Similar<br>The sintering temperature is not exactly same, but each product has provided the optimum temperature.<br>LUXEN Zr is higher than predicate device, but it is confirmed that the temperature improves translucent for nature color shade. |
| Types,<br>Class(ISO6872:2015) | Type II Class 5 | Type II Class 5 | Both products are type II Class 5. |
| Chemical<br>composition(wt.%) | Zirconia Powder Zpex<br>ZrO2+HfO2+Y2O3<br>>99.8 | ZrO2+HfO2+ Y2O3 : ≥99<br>Y2O3 : <6 | Similar<br>The raw material of |
| | | | |
| | Other oxides : <0.2 | Al2O3 : <<0.15<br>Other oxides : <0.15 | LUXEN Zr (Zpex, Zpex Yellow, Zpex Pink / Manufacturer: Tosoh) is consisted of less amount of aluminium oxide than predicate device due to translucency. However, both products meet the requirement of ISO 13356 (Al2O3 ≤ 0.5%).<br><br>Also, Zpex Yellow or Zpex Pink contain Fe2O3 or Er2O3, respectively, while predicate device contains Fe2O3, Er2O3 and MnO2. However, both products meet the requirement of ISO 13356 (other oxide < 0.5%). |
| | ZrO2 : 83-92 | | |
| | Y2O3 : 4-6 | | |
| | HfO2 : <5 | | |
| | Al2O3 : 0.03-0.07 | | |
| | SiO2 : Max 0.02 | | |
| | Fe2O3 : 0.12-0.18 | | |
| | Zirconia Powder Zpex Yellow<br>ZrO2+HfO2+Y2O3 : >99.8<br>Other oxides : <0.2<br>ZrO2 : 83-92<br>Y2O3 : 4-6<br>HfO2 : <5<br>Al2O3 : 0.03-0.07<br>SiO2 : Max 0.02<br>Fe2O3 : 0.12-0.18 | | |
| | Zirconia Powder Zpex Pink<br>ZrO2+HfO2+Y2O3 : >99.8<br>Other oxides : <0.2<br>ZrO2 : 79-92<br>HfO2 : <5<br>Er2O3 : 8-11<br>Al2O3 : 0.03-0.07<br>SiO2 : Max 0.02<br>Fe2O3 : Max 0.01 | | |
| Flexural strength(MPa) | 1038 ± 135 | 1250 | Similar Flexural strength is slightly lower, but higher than required by ISO 6872 for Class 5 dental ceramics(>800 MPa). |
| Thermal expansion coefficient(20-500°C)<br>Coefficient of thermal expansion | 10.7 X 10-6K-1 | 10 X 10-6K-1 | Both products are 10 X 10-6K-1. |
| Chemical solubility(µg/cm²) | 0 | Not listed | - |
| Biocompatibility | Device is biocompatible when used as directed by dental professionals per ISO 10993-1. | ISO 10993 and ISO 7405 | Both products are biocompatible. |
| Sterile | Non-sterile | Non-sterile | Both products are non-sterile. |
| Characteristics | Subject Device | Predicate Device | Summary |
| Device Name | LUXEN Smile | DD cube X2 | - |
| Manufacturer | DENTALMAX CO., Ltd. | Dental Direkt GmbH | - |
| 510(k) Number | K171585 | K150196 | - |
| Classification Name | Porcelain powder<br>for<br>clinical use | Porcelain powder<br>for<br>clinical use | Both<br>products<br>are<br>porcelain powder<br>for<br>clinical use. |
| Product Code | EIH | EIH | Both products are EIH. |
| Device Class | II | II | Both products are class<br>II. |
| Indications for Use | LUXEN Smile is indicated<br>for the production of<br>of full ceramic crowns,<br>onlays, 3- bridges and<br>inlay bridges(anterior<br>and molar). | Dental blanks made<br>from DD cubeX2 are<br>indicated for crowns,<br>multi-unit bridges (up to<br>a maximum of 3<br>elements) and inlay<br>bridges. Applications<br>include both, anterior<br>and posterior bridges. | Both<br>products<br>are<br>indicated for<br>dental<br>restorations. |
| Blank configuration | Block type<br>Disk type<br>Wieland type<br>D-95 type | Disk type<br>Wieland type | Both products are Disk<br>and Wieland type. |
| Thicknesses | 10 mm to 25 mm | 10 mm to 25 mm | Both products are 10-<br>25mm of thickness. |
| Crystal Morphology | 5Y-TZP | 5Y-TZP | Both products are 5Y-<br>TZP. |
| Color | A0, A1, A2, A3, B1, B2,<br>B3, B4, C4, A1-A2-B3-B4<br>LAYER,<br>B1-B2-B3-B4<br>LAYER | A0, A1, A2, A3, A3.5,<br>B2, C2, D3 | Similar<br>Both product are met<br>with VITA shade guide.<br>Shade is an option that is<br>selected according to<br>patient's teeth color.<br>Application of<br>appropriate color that<br>matches patient's teeth<br>tone can be possible. |
| Sintering temperature | 1450 °C | 1450 °C | Both products are<br>1450 °C. |
| Types,<br>Class(ISO6872:2015) | Type II Class 4b | Type II Class 4 | The proposed device<br>LUXEN Smile is a Type II<br>class 4b device |
| Chemical composition | Zirconia Powder Zpex Smile<br>ZrO2+HfO2+Y2O3 : >99.8<br>Other oxides : <0.2<br>ZrO2 : 80-92<br>Y2O3 : 8-10<br>HfO2 : <5<br>Al2O3 : 0.04-0.06<br>SiO2 : Max 0.02<br>Fe2O3 : <0.1<br><br>Zirconia Powder Zpex Yellow<br>ZrO2+HfO2+Y2O3 : >99.8<br>Other oxides : <0.2<br>ZrO2 : 83-92<br>Y2O3 : 4-6<br>HfO2 : <5<br>Al2O3 : 0.03-0.07<br>SiO2 : Max 0.02<br>Fe2O3 : 0.12-0.18<br><br>Zirconia Powder Zpex Pink<br>ZrO2+HfO2+Y2O3 : >99.8<br>Other oxides : <0.2<br>ZrO2 : 79-92<br>HfO2 : <5<br>Er2O3 : 8-11<br>Al2O3 : 0.03-0.07<br>SiO2 : Max 0.02<br>Fe2O3 : Max 0.01 | ZrO2+HfO2+ Y2O3 : ≥99<br>Y2O3 : <10<br>HfO2 : ca.0.2<br>Al2O3 : <0.1<br>Other oxides : ≤0.05 | 6872:2015 standard<br>while the predicate device DD cube X² is a<br>Type II class 4 (Catalog, Dental Direkt GmbH) in accordance with ISO 6872:2008 standard.<br>Similar<br>The raw material of LUXEN Smile (Zpex Smile, Zpex Yellow, Zpex Pink / Manufacturer: Tosoh) is consisted of less amount of aluminium oxide than predicate device due to translucency. However, both products meet the requirement of ISO 13356 (Al2O3 ≤ 0.5%).<br>Also, Zpex Yellow or Zpex Pink contain Fe2O3 or Er2O3, respectively, while predicate device contains Fe2O3, Er2O3 and MnO2. However, both products meet the requirement of ISO 13356 (other oxide < 0.5%). |
| Flexural strength | 770 ± 66 MPa | > 750-800 MPa | Similar<br>Flexural strength is slightly lower, but higher than required by ISO 6872 for Class 4b dental ceramics(>500 MPa). |
| Thermal expansion coefficient(20-500°C)<br>Coefficient of thermal expansion | $10.3 X 10^{-6}K^{-1}$ | $10 X 10^{-6}K^{-1}$ | Both products are $10 X 10^{-6}K^{-1}$ . |
| Chemical solubility | 0 μg/cm² | 15 μg/cm² | Similar<br>The lower value of |
| | | | chemical solubility has<br>more stable, and both<br>results of meet the<br>requirement of ISO 6872<br>for Class 4b dental<br>ceramics(>2000 $\mu$ g/cm2).<br>Therefore, it is<br>equivalent. |
| Biocompatibility | Device is biocompatible<br>when used as directed<br>by dental professionals<br>per ISO 10993-1. | EN ISO 10993-1, -5 | Both products are<br>biocompatible. |
| Sterile | Non-sterile | Non-sterile | Both products are non-<br>sterile.…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.