The Rhythm View Workstation is a computerized system that assists in the diagnosis of complex cardiac arrhythmias. The Rhythm View Workstation is used to analyze electrogram and electrocardiogram signals and display them in a visual format.
Device Story
RhythmView Workstation 6.1 is a computerized diagnostic system used in clinical electrophysiology (EP) studies. It ingests electrical signals from multi-polar EP catheters placed within the heart. Proprietary algorithms compute spatial organization of electrical activity during arrhythmias, identifying rotors or focal beat sources. The system transforms these signals into dynamic, visual representations, including electrical activity maps, stability maps, and rotational activity profiles. Physicians use these interactive displays to identify arrhythmia patterns and guide clinical decision-making during procedures. The workstation includes tools for signal quality indication, epoch timeline management, and procedure documentation. It provides a simplified, dynamic view of wave propagation to aid in the diagnosis of complex arrhythmias like AF, VT, and VF.
Clinical Evidence
Bench testing only. Validation included software and user verification/validation protocols. Testing confirmed the accuracy of Stability Maps and Summation Maps, validated that Rotational Activity Profile (RAP) detection of rotational targets is at least as effective as prior versions, and utilized simulated user testing to evaluate UI updates. Clinical data sets were used to validate the Stability Map and default filter settings.
Technological Characteristics
Programmable diagnostic computer. Operates on 120V, 60Hz power with a 40 Ah, 12V DC Lithium Iron Phosphate battery. Connectivity includes USB data transfer from EP systems and two-port switch data transfer. Software is C++ based. Complies with ISO 14971:2012, IEC 60601-1 (3rd Ed), and IEC 60601-1-2:2007. Features include signal quality indicator (SQI), RAP intensity scale, spotlight feature, stability map, and epoch timeline.
Indications for Use
Indicated for patients with complex cardiac arrhythmias, including focal AT, AFL, other SVT, AF, VT, and VF, to assist in diagnosis during electrophysiological studies.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract symbol that resembles three stylized human profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 1, 2017
Abbott Electrophysiology Dennis Pozzo Senior Regulatory Affairs Specialist 3668 S. Gever Road. Suite 365 St. Louis, Missouri 63127
Re: K171583
Trade/Device Name: RhythmView Workstation 6.1 Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II Product Code: DOK Dated: August 4, 2017 Received: August 8, 2017
Dear Dennis Pozzo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
M.A. Wilhelm
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K171583
Device Name
RhythmView Workstation 6.1
Indications for Use (Describe)
The Rhythm View Workstation is a computerized system that assists in the diagnosis of complex cardiac arrhythmias. The Rhythm View Workstation is used to analyze electrogram and electrocardiogram signals and display them in a visual format.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
Page 1 of 1
PSC Publishing Services (101) 443-6740 EF
{3}------------------------------------------------
# Section 5: 510(k) Summary per 21CFR §807.92
| Submitter's<br>information | Abbott Electrophysiology<br>3668 S. Geyer Road, Suite 365<br>St. Louis, MO 63127<br>Contact: Dennis Pozzo<br>Phone 314-300-6580 | |
|-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Device/<br>classification<br>name | Device Name: RhythmView Workstation 6.1<br>●<br>Classification/Common name: Programmable diagnostic computer<br>●<br>Product Code/Classification No .: DQK/21 CFR 870.1425<br>●<br>The marketed device to which substantial equivalence is claimed:<br>●<br>RhythmView Workstation, K161240, cleared August 10, 2016<br>RhythmView Workstation, K151245, cleared September 15, 2015<br>(ventricular function) | |
| Device<br>description | RhythmView uses electrical signals collected from the electrodes of one or<br>more multi-polar electrophysiology (EP) catheters. It provides a dynamic,<br>simplified representation of wave propagation.<br>The RhythmView computes and displays electrical rotors or focal beat<br>sources that may sustain human heart rhythm disorders including focal AT,<br>AFL, other SVT, AF, VT and VF in a given patient. The product takes as<br>input electrical signals recorded during the heart rhythm disorder under<br>consideration, typically from multiple specified locations within the heart<br>during an electrophysiological study. The RhythmView then uses proprietary<br>patented algorithms and methods to compute spatial organization during the<br>heart rhythm disorder. These computed elements are displayed graphically in<br>interactive form for review to aid diagnosis by the physician during an<br>electrophysiology study. | |
| Indications for<br>use | The RhythmView Workstation is a computerized system that assists in the<br>diagnosis of complex cardiac arrhythmias. The RhythmView Workstation is<br>used to analyze electrogram and electrocardiogram signals and display them<br>in a visual format. | |
| Technological<br>characteristics | The RhythmView Workstation currently allows the user to:<br>Review and select a time sequence of electrical signals from various electrodes; Analyze the signals; View a graphic display (Electrical Activity) of the signal potentials showing progressive depolarization and repolarization in greyscale for the particular arrhythmia; Play/Replay the animated graphic representation of electrical signals. Review various display options to assist the provider with the identification of arrhythmia patterns. Evaluate the quality of the electrograms exported from the EP system to RhythmView Workstation ("Signal quality indicator" (SQI)) Enable an option to improvements usability of RAP ("RAP intensity scale" and "Spotlight" feature) Review the composite RAP profile (Stability Map) over the entire epoch under analysis (vs. just a single 4-second time segment) A method of thresholding the Stability Map image analysis tool for delineating the persistence of RAP over the course of the entire epoch A mechanism for displaying and selecting all time segments within an epoch ("epoch timeline") Create a procedure history within RhythmView ("Procedure Notebook") and allow the user to add freeform text notes. The RhythmView software is being modified to improve the feedback and usability during a RhythmView procedure by updating the following functions: Caliper tool; ECG lead; SQI; RAP calculation; and Stability map | |
{4}------------------------------------------------
# Section 5: 510(k) Summary per 21CFR §807.92, Continued
{5}------------------------------------------------
# Section 5: 510(k) Summary per 21CFR §807.92, Continued
| Device Characteristic | Predicate<br>RhythmView<br>Workstation | Proposed<br>RhvthmView<br>Workstation |
|------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Signal processing | Yes | Yes |
| Post-processing display | Yes | Yes |
| Grid display of electrode signals | Yes | Yes |
| Programming Language | C++ | C++ |
| Export of processed file into video<br>format | Yes | Yes |
| OTS Software requirements | Same | Same |
| Display options for review of<br>processed signals<br>RAP display (optional) | · Electrical Activity<br>· Contours Only<br>· DContours<br>• Rotational Activity<br>Profile<br>• Composite RAP<br>image ("Stability<br>Map")<br>• Epoch timeline<br>· Spotlight<br>Multi-color with<br>monochromatic<br>option available | · Electrical Activity<br>• Contours Only<br>· DContours<br>· Rotational Activity<br>Profile<br>· Composite RAP<br>image ("Stability<br>Map")<br>• Epoch timeline<br>· Spotlight<br>Multi-color with<br>monochromatic option<br>available |
| RFID Reader/Writer Function | Yes | Yes |
| Data transfer via Two Port Switch | Yes | Yes |
| Direct data transfer via USB cable to<br>RV Workstation from EP system | Option available | Option available |
| Atrial Function | Yes | Yes |
| *Ventricular Function | Option not available | Option available |
| Signal Quality Indicator | Option available | Option available |
| Procedure Notebook | Option available | Option available |
| Device Characteristic | Predicate RhythmView<br>Workstation | Proposed RhythmView<br>Workstation |
| <b>**Spotlight Feature</b> | Option available | Option available |
| <b>**Stability Map</b> | Option available | Option available |
| <b>**Epoch Timeline</b> | Option available | Option available |
| Electrical Rating (Technological Comparison) | | |
| Electrical rating – Typical | 120V, 60Hz | 120V, 60Hz |
| Battery | 40 Ah, 12 V DC,<br>Lithium Iron Phosphate | 40 Ah, 12 V DC, Lithium<br>Iron Phosphate |
| RF Class per CISPR11 | Group1 | Group 1 |
{6}------------------------------------------------
#### Section 5: Summary per 21CFR §807.92, Continued
*Ventricular Function: predicate 510(k) K151245 **Atrial function only
Applicable standards
- ISO 14971: 2012, Medical Devices Application of risk management ● to medical devices
Since there are no hardware updates, the EMC and Electrical Safety reports referenced K161240 still apply.
- . IEC 60601-1, 3rd Edition, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2: 2007, Medical electrical equipment Part 1-2: ● General requirements for basic safety and essential performance -Collateral standard: Electromagnetic compatibility - Requirements and tests
{7}------------------------------------------------
# Section 5: 510(k) Summary per 21CFR §807.92, Continued
| Performance<br>data | Bench testing has been conducted through software and user verification/<br>validation protocols to ensure the RhythmView Workstation meets its<br>intended use and user needs. |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The testing encompassed:<br>Validation of the Stability Map and the default Stability Filter setting<br>using clinical data sets Verification that the Summation Maps generated are correct<br>compilations of the RAP profiles from individual segments Validate that RAP detects known rotational targets at least as well as<br>prior versions of RAP. Validate the RAP functionality in RhythmView through Physician<br>reads Simulated User testing to evaluate new features of UI |
| Conclusion | The testing has demonstrated that the SW updates for RhythmView V6.1<br>provide reasonable assurance that the proposed device conforms to the |
appropriate requirements for its intended use. Therefore, it is substantially equivalent to the predicate device, safe and effective for its intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.