← Product Code [FDX](/productcode/FDX) · K171573

# Fusion Cytology Brush, CytoMax II Double Lumen Cytology Brush (K171573)

_Wilson-Cook Medical, Inc. · FDX · Dec 29, 2017 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K171573

## Device Facts

- **Applicant:** Wilson-Cook Medical, Inc.
- **Product Code:** [FDX](/productcode/FDX.md)
- **Decision Date:** Dec 29, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The Fusion® Cytology Brush and CytoMax II® Double Lumen Cytology Brush are used for collection of cells in the biliary system.

## Device Story

The Fusion® Cytology Brush and CytoMax II® Double Lumen Cytology Brush are endoscopic accessories designed for biliary cell collection. The device consists of a double-lumen catheter, a cytology brush assembly (coil spring or bullet tip), a pin vise handle, a wire guide hub, and a detachable connecting tube for flushing. Operated by a clinician, the device is passed through an endoscope over a pre-positioned wire guide to the target biliary site. The clinician actuates the device by pushing the proximal pin vise handle to extend the distal brush and pulling it back to retract the brush for sample collection. The device facilitates tissue sampling to aid in clinical diagnosis of biliary conditions.

## Clinical Evidence

Bench testing only. Testing included shelf-life validation, packaging validation, and biocompatibility testing conducted in accordance with ISO 10993-1.

## Technological Characteristics

Double-lumen catheter assembly with pin vise handle, wire guide hub, and detachable connecting tube. Brush assembly features coil spring or bullet tip. Materials modified from predicate. Biocompatibility evaluated per ISO 10993-1. Mechanical actuation via manual handle manipulation.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Wilson-Cook Double Lumen Biliary Cytology Brush ([K040324](/device/K040324.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a sans-serif font.

December 29, 2017

Wilson-Cook Medical, Inc. Sierra Lowe Regulatory Affairs Specialist - I 4900 Bethania Station Road Winston-Salem, NC 27105

Re: K171573

Trade/Device Name: Fusion® Cytology Brush, CytoMax II® Double Lumen Cytology Brush Regulation Number: 21 CFR& 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FDX Dated: November 29, 2017 Received: December 4, 2017

Dear Sierra Lowe:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Image /page/1/Picture/6 description: The image shows the name "Charles Viviano -S" in a large, sans-serif font. The text is black against a white background. The letters are evenly spaced and the text is left-aligned. The name appears to be a title or heading.

For Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known)

### K171573

Device Name Fusion® Cytology Brush CytoMax II® Double Lumen Cytology Brush

### Indications for Use (Describe)

The Fusion® Cytology Brush and CytoMax II® Double Lumen Cytology Brush are used for collection of cells in the biliary system.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for Cook Medical. The logo is white text on a red background. The word "COOK" is in large, bold letters, and the word "MEDICAL" is in smaller letters below it.

COOK ENDOSCOPY 4900 BETHANIA STATION ROAD WINSTON-SALEM, NC 27105 U.S.A. PHONE: 336.744.0157 TOLL FREE: 800.245.4707 WWW.COOKMEDICAL.COM

# 510(k) Summary

| Name:                | Wilson-Cook Medical, Inc. /Cook Endoscopy                                                                                         |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------|
| Address:             | 4900 Bethania Station Road<br>Winston-Salem, North Carolina 27105                                                                 |
| Phone:               | (336) 744-0157 (x396506)                                                                                                          |
| Fax:                 | (336) 201-5994                                                                                                                    |
| Contact:             | Sierra Lowe, Regulatory Affairs Specialist I                                                                                      |
| Date:                | May 26, 2017                                                                                                                      |
| Trade Names:         | Fusion® Cytology Brush<br>CytoMax II® Double Lumen Cytology Brush                                                                 |
| Common Name:         | Endoscopic Cytology Brush                                                                                                         |
| Classification Name: | Endoscope and accessories 21 CFR §876.1500, FDX, Class II                                                                         |
| Predicate Device:    | Wilson-Cook Double Lumen Biliary Cytology Brush, K040324, cleared May 20,<br>2004.                                                |
| Intended Use:        | The Fusion® Cytology Brush and CytoMax II® Double Lumen Cytology Brush<br>are used for collection of cells in the biliary system. |

### Device Description:

The Fusion® Cytology Brush and CytoMax II® Double Lumen Cytology Brush (subject devices) represent modifications made to the Wilson-Cook Double Lumen Biliary Cytology Brush (predicate device) currently cleared to market via 510K K040324 by Wilson-Cook Medical, Inc.

The Fusion® Cytology Brush and CytoMax II® Double Lumen Cytology Brushes consist of a double lumen catheter with ink markings, a cytology brush assembly with either a coil spring or bullet tip, a pin vise handle, wire guide hub, and a detachable connecting tube, which can be used for flushing of the wire guide hub. The endoscopic cytology brush is used by passing the device through an endoscope, over a

{4}------------------------------------------------

prepositioned wire guide to the target location. The endoscopic cytology brush assembly is located at the distal (patient contacting) end with the pin vise handle located at the proximal (non-patient contacting) end. The handle is actuated by pushing the pin vise handle forward to extend the cytology brush and then pulling backward to retract the brush.

### Substantial Equivalence:

Changes have been made to the currently cleared Double Lumen Biliary Cytology Brushes (K040324). These changes include: catheter length labeling to reflect working length, additional catheter diameters, brush tip diameters and lengths, removal of a 6cm ink marking, removal of radiopaque markers, modified wire guide access port, removal of wire guide port with extension, updated wire guide compatibility, and modified catheter materials. Performance testing consisting of non-clinical bench testing demonstrates that the subject devices met the performance requirements to fulfill their intended uses. The results of this testing provide reasonable assurance that the subject devices will function as intended. The subject devices do not raise new questions of safety or effectiveness as compared to their respective predicate devices.

### Summary of non-clinical testing:

The following non-clinical testing was conducted to demonstrate the performance of the subject device and confirmed that the subject device performs as intended.

- 0 Shelf Life Testing
- Packaging Validation .

Biocompatibility testing was performed in accordance with the FDA Guidance, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices-Part I: Evaluation and testing within a risk management process."

### Conclusion:

We believe that the subject devices are substantially equivalent to the predicate device in terms of intended use, key operating principles, materials and technological characteristics. We consider the risks associated with the modifications to the subject devices to have been adequately addressed through our Design Control Processes and do not affect safety or effectiveness of the devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K171573](https://fda-staging.innolitics.com/device/K171573)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).