OTIS Optical Coherence Tomography System

K171560 · Perimeter Medical Imaging, Inc. · NQQ · Aug 25, 2017 · Radiology

Device Facts

Record IDK171560
Device NameOTIS Optical Coherence Tomography System
ApplicantPerimeter Medical Imaging, Inc.
Product CodeNQQ · Radiology
Decision DateAug 25, 2017
DecisionSESU
Submission TypeTraditional
Regulation21 CFR 892.1560
Device ClassClass 2

Indications for Use

The OTIS 1B Optical Coherence Tomography System is indicated for use as an imaging tool in the evaluation of excised human tissue microstructure by providing two-dimensional, cross-sectional, real-time depth visualization, with image review manipulation software for identifying and annotating regions of interest.

Device Story

The Perimeter OTIS 1B is an optical coherence tomography (OCT) system used to image excised human tissue. It utilizes a near-infrared super luminescent diode (1325 nm) and an interferometer to compare a reference beam with light backscattered from the tissue sample. The device employs an automated mechanical stage to move the probe across the specimen surface, standardizing image collection. It produces 2D cross-sectional OCT images and a white-light photograph of the specimen. Clinicians use the system to view, scroll, zoom, and pan through images to identify and annotate regions of interest. The device is intended for use in clinical settings to assist in the evaluation of tissue microstructure. It provides high-resolution, sub-surface visualization, which may aid in the assessment of excised tissue samples.

Clinical Evidence

Bench testing only. Validation studies confirmed the system meets design requirements and user needs per IEC 62366. Image quality was validated by comparing OCT images of excised tissue microstructure to histopathology, demonstrating performance comparable to other OCT systems. Safety was verified through electrical, EMC, and laser safety testing (IEC 60825-1:2014, AAMI ANSI ES 60601-1).

Technological Characteristics

OCT imaging system using a 1325 ± 20 nm super luminescent diode (Class 1 laser). Features automated mechanical stage for probe positioning. Dimensions: 1.5 mm depth in tissue. Connectivity: Standalone. Software: AAMI ANSI IEC 62304 compliant. Standards: ISO 14971:2007, AAMI ANSI ISO 10993-1:2014, IEC 60601-1-2:2007.

Indications for Use

Indicated for use as an imaging tool for the evaluation of excised human tissue microstructure, providing 2D cross-sectional, real-time depth visualization and image review software for identifying and annotating regions of interest.

Regulatory Classification

Identification

An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Special Controls

*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Public Health Service Image /page/0/Picture/2 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. In the center of the seal is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - W O66-G609 Silver Spring, MD 20993-0002 August 25, 2017 Perimeter Medical Imaging, Inc. Elizabeth Munro Engineering Operations and Regulatory Lead 47 Colborne Street Suite 202 Toronto, M5E 1P8 Ca Re: K171560 Trade/Device Name: OTIS Optical Coherence Tomography System Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic Pulsed Echo Imaging System Regulatory Class: Class II Product Code: NOO Dated: May 25, 2017 Received: Mav 30, 2017 Dear Elizabeth Munro: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act and the limitations described below. The general controls of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading. The Office of Device Evaluation has determined that there is a reasonable likelihood that this device will be used for an intended use not identified in the proposed labeling and that such use could cause harm. Therefore, in accordance with Section 513(i)(1)(E) of the Act, the following limitation must appear in the warnings section device's labeling: {1}------------------------------------------------ "The safety and effectiveness of this device for diagnostic analysis (i.e. differentiating normal versus specific abnormalities) in any tissue microstructure or specified disease has not been evaluated." Furthermore, the indication for use "the OTIS 1B Optical Coherence Tomography System is indicated for use as an imaging tool in the evaluation of excised human tissue microstructure by providing two-dimensional, cross-sectional, real-time depth visualization, with image review manipulation software for identifying and annotating regions of interest" must be prominently displayed in all labeling, including pouch box, and carton labels, instructions for use, and other promotional materials, in close proximity to the trade name, of a similar point size, and in bold orint. Please note that the above labeling limitations are required by Section 513(i)(1)(E) of the Act. Therefore, a new 510(k) is required before these limitations are modified in any way or removed from the device's labeling. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification if the limitation statement described above is added to your labeling. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm_for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. {2}------------------------------------------------ You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, William H. Maisel -S William H. Maisel, MD, MPH Acting Director, Office of Device Evaluation Deputy Center Director for Science Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # SECTION 5 INDICATIONS FOR USE STATEMENT {4}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K171560 ### Device Name OTIS 1B Optical Coherence Tomography System ### Indications for Use (Describe) The OTIS 1B Optical Coherence Tomography System is indicated for use as an imaging tool in the evaluation of excised human tissue microstructure by providing two-dimensional, real-time depth visualization, with image review manipulation software for identifying and annotating regions of interest. | Type of Use (Select one or both, as applicable) | | |----------------------------------------------------------------------------------|---------------------------------------------------------------------------------| | <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {5}------------------------------------------------ # SECTION 6 510(K) SUMMARY {6}------------------------------------------------ # 510(k) Summary | 1. Basic Information - Submitter | | |-----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------| | 510(k) Owner: | Perimeter Medical Imaging, Inc. | | Address: | 47 Colborne Street, Suite 202<br>Toronto, Ontario, Canada, M5E 1P8 | | Official Contact: | Elizabeth Munro<br>Engineering Operations and Regulatory Lead<br>Phone: 877-278-4614<br>Fax: 647-360-0302<br>emunro@perimetermed.com | Date Summary Prepared: August 23, 2017 ## 2. Device Name Trade Name: Perimeter OTIS™ 1B Optical Coherence Tomography (OCT) System Common Name: Optical Coherence Tomography (OCT) System Classification Name: Ultrasonic pulsed echo imaging system Regulatory Classification: 21 CFR 892.1560 Product Code: NQQ Classification: Class II ## 3. Legally Marketed Predicate Devices Perimeter OTIS™ 1.0 Optical Coherence Tomography (OCT) System, K160240 # 4. Device Description The Perimeter OTIS 1B is based on optical coherence tomography (OCT) imaging which is similar to ultrasound, but uses non-ionizing, low-power optical radiation to produce high resolution, sub-surface images of a tissue sample. Due to the extremely high velocity of light, optical echoes (reflected and backscattered light from the sample) cannot be measured directly using a photodetector. Instead, OCT devices use an interferometer to compare a reference beam of light to the backscattered light returning from the tissue sample. The features in an OCT image are created by changes in the optical properties (namely scattering, absorption, and index of refraction) of the sample. The Perimeter OTIS 1B collects and displays OCT images of human tissue with comparable image quality to other previously 510(k)-cleared OCT imaging systems and, specifically, the Perimeter OTIS 1.0 System (K160240). Like its predicate, the Perimeter OTIS 1B has automated the OCT scanning of the specimen surface, standardizing the image collection process. {7}------------------------------------------------ The Perimeter OTIS 1B includes a white light image (e.g., photograph) of the specimen, to assist the clinician in viewing the OCT images with a reference to the white light image and the OCT image scan position. The Perimeter OTIS 1B also includes user interface options, such as the ability to "scroll" through the OCT images, across a specimen's surface, as well as to zoom and pan in more closely to examine areas of interest, as desired. #### Indications for Use 5. The OTIS 1B Optical Coherence Tomography System is indicated for use as an imaging tool in the evaluation of excised human tissue microstructure by providing twodimensional, cross-sectional, real-time depth visualization with image review manipulation software for identifying and annotating regions of interest. | PRODUCT COMPARISON TABLE | | | |-----------------------------------------------------------------|----------------------------------------------------------------|----------------------------------------------------------------| | | Perimeter OTIS™ 1B<br>[New Device] | Perimeter OTIS™ 1.0<br>K160240<br>[Predicate] | | Measurement Technique | Optical Coherence Tomography | Optical Coherence Tomography | | Center Wavelength<br>[NIR: 700 – 1400 nm] | 1325 ± 20 nm | 1325 ± 20 nm | | Optical Source | Super Luminescent Diode | Super Luminescent Diode | | Optical Radiation Safety | Safe for Indicated Use,<br>Class 1 Laser | Safe for Indicated Use,<br>Class 1 Laser | | Lateral Resolution<br>(Nominal) | ≤ 50 μm<br>(20 μm) | ≤ 50 μm | | Lateral Range (x-axis):<br>Single B-scan<br>WF-OCT Scan | (1.7 mm) 2 mm<br>500 mm [250 2 mm-strips] | (1.7 mm) 2 mm<br>500 mm [250 2 mm-strips] | | Axial Resolution<br>(Nominal)<br>[free space units] | ≤ 15 μm<br>(11.7 ± 2.0 μm) | ≤ 15 μm | | In-depth tissue range<br>[free space units] | 1.5 mm in tissue<br>[minimum of 2.2 mm in free space<br>(7mm)] | 1.5 mm in tissue<br>[minimum of 2.2 mm in free space<br>(7mm)] | | Patient Applied Part | No Patient Applied Part | No Patient Applied Part | | Input Devices:<br>Keyboard<br>Pointing<br>Foot pedal (optional) | Yes<br>Yes<br>No | Yes<br>Yes<br>No | | Electrical Voltage Frequency | 108-132 V,<br>60 Hz [North America Use] | 108-132 V,<br>60 Hz [North America Use] | #### 6. Substantial Equivalence Summary {8}------------------------------------------------ As the OTIS 1B device is identical in construction to OTIS 1.0, the Systems may be collectively refer to as "OTIS Systems." #### Product and Quality Management Standards 7. The Perimeter OTIS Systems were designed and developed under design controls per 21 CFR 820.30, in addition to risk management, ISO 14971:2007 Application of Risk Management to Medical Devices. In addition, testing was performed to ensure that the Perimeter OTIS System complies with the following recognized standards: - IEC 60825-1:2014, Safety of Laser Products Part 1: Equipment Classification and . Requirements - . AAMI ANSI IEC 62304 Medical Device Software - Software Life Cycle Management - AAMI ANSI ISO 15223-1:2012 Medical Devices Symbols to be used with Medical . Devices Labels, Labeling and Information to be Supplied - Part 1: General Requirements - . AAMI ANSI ISO 10993-1:2014 Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing within a Risk Management Process - AAMI ANSI ES 60601-1: 2005/(R)2012, A1:2012, C1:2009/(R)2012 and . A2:2010/(R)2012 Medical Electrical Equipment - Part 1 - General Requirements for Basic Safety and Essential Performance - . IEC 60601-1-2:2007 Medical Electrical Equipment Part 1-2: General Requirements for the Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements and Tests #### Technological Characteristics 8. The Perimeter OTIS 1B is substantially equivalent to the Perimeter OTIS 1.0 [K160240] as the devices are identical and have the same intended use. Both Perimeter OTIS Systems provide a standardized "stepping" of the OCT image acquisition probe across a pre-set scan area (e.g., 3 cm - 5 cm). The Perimeter OTIS Systems perform this surface image acquisition by using an automated mechanical stage and precise probe positioning. The Perimeter OTIS 1B and the predicate device, OTIS 1.0, use near infrared (NIR) light to produce Optical Coherence Tomography images. The Perimeter OTIS Systems have a center wavelength of 1325 ± 20 nm. The Perimeter OTIS Systems position the probe automatically during imaging - reducing the operator workload and ensuring consistent {9}------------------------------------------------ sample coverage. The intended use of the Perimeter OTIS 1B is the same as OTIS 1.0; the indications for use for OTIS 1B are encompassed within the intended use of the OTIS Systems; and the device constructions are identical. Therefore, Perimeter OTIS 1B is as safe and effective as the predicate device. ### 9. Performance and Safety Testing Perimeter completed verification and validation activities under Perimeter Design Control procedures to ensure that verification studies demonstrated that outputs met input requirements, and that validation studies demonstrated that the Perimeter OTIS fulfilled the intended use and user needs. Perimeter validated the usability of the OTIS Systems per IEC 62366:2007 - Application of usability engineering to medical devices. Testing was performed with representative users in actual and simulated use environments. Perimeter validated that the OTIS 1B can obtain OCT images with sufficient image quality to identify excised tissue microstructure features as compared to histopathology, at a level comparable to other OCT imaging systems. External laser, electrical, and EMC testing successfully demonstrated the safety of the Perimeter OTIS Systems in its intended environment. Perimeter OTIS Systems employ a near infrared super luminescent diode for optical image acquisition, as well as a sufface sensing laser system. Both laser systems are classified as Class 1 laser, per IEC 60825-1:2014, requiring no special optical safety precautions. ### 10. Conclusions Perimeter Medical Imaging, Inc. has demonstrated that the OTIS 1B Optical Coherence Tomography System has the same intended use and is substantially equivalent to its predicate devices, the OTIS™ 1.0 Optical Coherence Tomography System. The Perimeter OTIS 1B, as it is identical in construction to the OTIS 1.0, complies with recognized standards and guidelines for electrical safety, and biocompatibility. Testing has demonstrated that the devices is as safe and effective as the predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...