The Cotton® Graduated Dilation Catheter, Fusion® Biliary Dilation Catheter, Soehendra® Biliary Dilation Catheters are used to dilate biliary strictures. These devices are indicated for adult use only. The Geenen® Graduated Dilation Catheter is used to dilate stenotic minor papilla. These devices are indicated for adult use only.
Device Story
Catheters used for biliary or minor papilla dilation; passed through endoscope over pre-positioned wire guide; advanced under fluoroscopic monitoring. Fusion model features double lumen, radiopaque catheter, and stylet wire; allows wire guide to exit proximally or distally. Other models feature radiopaque marker bands and luer lock hub. Operated by physicians in clinical settings. Device advancement to stricture site enables mechanical dilation. Benefits include relief of biliary or papilla stenosis.
Clinical Evidence
Bench testing only. Testing included visual inspection, guidewire verification, endoscope compatibility, simulated dilation, removal force, and tensile strength. Biocompatibility testing performed per ISO 10993-1.
Technological Characteristics
Catheter-based dilation system; radiopaque marker bands (tantalum); luer lock hub with connecting cap. Fusion model includes double lumen and stylet wire. Materials vary by model. Energy source: mechanical (manual advancement). Sterilization: not specified. Connectivity: none.
Indications for Use
Indicated for adult patients requiring dilation of biliary strictures (Cotton, Fusion, Soehendra catheters) or stenotic minor papilla (Geenen catheter).
Regulatory Classification
Identification
A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.
Special Controls
*Classification.* Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two distinct elements: the Department of Health & Human Services logo on the left and the FDA acronym followed by the full name of the agency on the right. The Department of Health & Human Services logo features a stylized human figure, while the FDA part includes the acronym in a blue square and the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 12, 2017
Wilson-Cook Medical, Inc. Sierra Lowe Regulatory Affairs Specialist - 1 4900 Bethania Station Road Winston-Salem, NC 27105
Re: K171548
Trade/Device Name: Cotton® Graduated Dilation Catheter, Fusion® Biliary Dilation Catheter, Soehendra® Biliary Dilation Catheter, Geenen® Graduated Dilation Catheter Regulation Number: 21 CFR§ 876.5010 Regulation Name: Biliary Catheter and Accessories Regulatory Class: II Product Code: FGE Dated: August 18, 2017 Received: August 21, 2017
Dear Sierra Lowe:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevicesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Joyce M. Whang -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K171548
#### Device Name
Cotton® Graduated Dilation Catheter, Fusion® Biliary Dilation Catheter, Soehendra® Bilary Dilation Catheter, Geenen® Graduated Dilation Catheter
Indications for Use (Describe)
The Cotton® Graduated Dilation Catheter, Fusion® Biliary Dilation Catheter, Soehendra® Biliary Dilation Catheters are used to dilate biliary strictures. These devices are indicated for adult use only.
The Geenen® Graduated Dilation Catheter is used to dilate stenotic minor papilla. These devices are indicated for adult use only.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
COOK®
MEDICAL
COOK ENDOSCOPY 4900 BETHANIA STATION ROAD WINSTON-SALEM, NC 27105 U.S.A. PHONE: 336.744.0157 TOLL FREE: 800.245.4707 WWW.COOKMEDICAL.COM
## 510(k) Summary
| Name: | Wilson-Cook Medical, Inc. /Cook Endoscopy |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | 4900 Bethania Station Road<br>Winston-Salem, North Carolina 27105 |
| Phone: | (336) 744-0157 (x396506) |
| Fax: | (336) 201-5994 |
| Contact: | Sierra Lowe, Regulatory Affairs Specialist I |
| Date: | October 13, 2017 |
| Trade Names: | Cotton® Graduated Dilation Catheter (CGDC)<br>Fusion® Biliary Dilation Catheter (FS-BDC)<br>Soehendra® Biliary Dilation Catheter (SBDC)<br>Geenen® Graduated Dilation Catheter (GGDC) |
| Common Name: | Dilation Catheter |
| Classification Name: | Biliary Catheter and Accessories 21 CFR §876.5010; FGE Class II |
| Predicate Devices: | Soehendra® Biliary Dilation Catheter, K851960, cleared April 18, 1985<br>Vennes® Pancreatic Dilation Set, K851956, cleared June 13, 1985 |
## Intended Uses:
The Cotton® Graduated Dilation Catheter, Fusion® Biliary Dilation Catheter, Soehendra® Biliary Dilation Catheters are used to dilary strictures. These devices are indicated for adult use only.
The Geenen® Graduated Dilation Catheter is used to dilate stenotic minor papilla. These devices are indicated for adult use only.
{4}------------------------------------------------
#### Description of the Devices:
## Biliary Dilation Catheters
The subject devices consist of a catheter with a luer lock hub with a connecting cap fitting. The Fusion® Biliary Dilation Catheter is constructed with a double lumen, radiopaque catheter with a stylet wire that is removed prior to contrast or fluid injection. The subject devices are used by passing the device through an endoscope, over a pre-positioned wire guide. The device is then advanced under fluoroscopic monitoring; during advancement, the pre-positioned wire guide will exit the proximal (non-patient contacting) end of the catheter through the connecting cap and luer lock hub. The Fusion® Biliary Dilation Catheter includes an additional option which allows the wire guide to exit at the distal end of the device (patient contacting). The dilation catheter is advanced to the appropriate position where dilation of the stricture is completed.
#### Dilation Catheter
The subject device consists of a catheter with a radiopaque marker band and a luer lock hub with a connecting cap fitting. The subject device is used by passing the device through an endoscope, over a prepositioned wire guide. The device is then advanced under fluoroscopic monitoring; during advancement, the pre-positioned wire guide will exit the proximal (non-patient contacting) end of the catheter through the connecting cap and luer lock adapter. The dilation catheter is advanced to the appropriate position where dilation of the stricture is performed.
#### Substantial Equivalence:
### Biliary Dilation Catheter
Minor changes were made to the predicate dilation catheters cleared to market via K851960. Changes to the subject devices include: additional catheter lengths and diameters, additional tapered tip lengths and diameters, and the addition of tantalum radiopaque marker bands. For the Fusion® Biliary Dilation Catheter (FS-BDC), there was a change to the catheter material and the addition of a stylet wire.
#### Dilation Catheter
Minor changes have been made to the predicate dilation catheter cleared to market via (K851956). Changes to the subject device include: additional catheter lengths and diameters, additional tapered tip lengths and diameters, addition of tantalum radiopaque marker bands, exclusion of the wire guide that was previously supplied along with the device, and updated wire guide compatibility.
{5}------------------------------------------------
## Performance Data:
Performance testing consisting of non-clinical bench testing demonstrates that the subject devices met the performance requirements to fulfill their intended uses. The results of this testing provides reasonable assurance that the subject devices will function as intended. The subject devices do not raise new questions of safety or effectiveness as compared to their respective predicate devices.
#### Summary of non-clinical testing:
The following non-clinical testing was conducted to demonstrate the performance of the subject device and confirmed that the subject device performs as intended.
- . Shelf Life Testing
- Packaging Validation .
- Non-clinical, bench testing of the subject devices methods included: visual inspection, pre-. position guidewire verification, endoscope channel compatibility, dilatation simulated removal, and removal force and tensile strength tests
Biocompatibility testing was performed in accordance with the FDA Guidance, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices-Part I: Evaluation and testing within a risk management process."
#### Conclusion:
We believe that the subject devices are substantially equivalent to the predicate devices in terms of intended use, key operating principles, materials and technological characteristics. We consider the risks associated with the modifications to the subject devices to have been adequately addressed through our Design Control Processes and do not affect safety or effectiveness of the devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.