SurgicalPreview enables visualization and measurement of cerebral blood vessels for preoperational planning and sizing for neurovascular interventions and surgery. SurgicalPreview also allows for the ability to computationally model the placement and deployment of neurointerventional devices. General functionalities are provided such as: - Segmentation of neurovascular structures - Automatic centerline detection - Visualization of CT scan images for 2D review and 3D reconstruction - Measurement and annotation tools - Reporting tools Information provided by the software is not intended in any way to eliminate, replace or substitute for, in whole or in part, the healthcare provider's judgment and analysis of the patient's condition.
Device Story
SurgicalPreview is a standalone software application for Windows/Mac OS; used by physicians for preoperative planning of endovascular surgery. Input: DICOM CT scan studies. Processing: software segments neurovascular structures; performs automatic centerline detection; reconstructs 2D slices into 3D models; enables computational modeling of neurointerventional device placement/deployment. Output: 2D/3D visualizations, measurements, annotations, and reports via web browser (WebGL). Clinical use: supports physician decision-making; does not replace clinical judgment. Benefits: assists in sizing and planning for neurovascular interventions.
Clinical Evidence
Bench testing only. Validation included DICOM import, patient manager functionality, image display/processing, anatomic reconstruction, computational modeling of specific vascular devices (Codman Enterprise, Pipeline Embolization Device), measurement accuracy, and report generation. No clinical data provided.
Technological Characteristics
Software-only device; runs on standard Windows/Mac OS computers. DICOM 3.1 compliant. Connectivity via secure internet file server. WebGL-based output. Software lifecycle per IEC 62304; risk management per ISO 14971.
Indications for Use
Indicated for visualization and measurement of cerebral blood vessels for preoperational planning and sizing for neurovascular interventions and surgery, including computational modeling of neurointerventional device placement and deployment.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 8, 2017
EndoVantage, LLC % H. Semih Oktay, Ph.D. Regulatory Consultant CardioMed Device Consultants, LLC 1783 Forest Drive, Suite 254 Annapolis, Maryland 21401
Re: K171534
Trade/Device Name: SurgicalPreview Regulation Number: 21 CFR 892.2050 Regulation Name: Picture Archiving and Communications System Regulatory Class: Class II Product Code: PZO Dated: October 6, 2017 Received: October 10, 2017
Dear Dr. Oktay:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Nicole G. Ibrahim -S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K171534
Device Name EndoVantage SurgicalPreview
#### Indications for Use (Describe)
SurgicalPreview enables visualization and measurement of cerebral blood vessels for preoperational planning and sizing for neurovascular interventions and surgery. SurgicalPreview also allows for the ability to computationally model the placement and deployment of neurointerventional devices.
General functionalities are provided such as:
- · Segmentation of neurovascular structures
- · Automatic centerline detection
- · Visualization of CT scan images for 2D review and 3D reconstruction
- · Measurement and annotation tools
- · Reporting tools
Information provided by the software is not intended in any way to eliminate, replace or substitute for, in whole or in part, the healthcare provider's judgment and analysis of the patient's condition.
Type of Use (Select one or both, as applicable)
| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> |
|-------------------------------------------------------------------------------------------------|
| <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Endovantage. The logo is red and features a stylized globe-like graphic on the left side. The text "endovantage" is written in a bold, sans-serif font and is also red.
SurgicalPreview™
| 510(k) SUMMARY | K171534 |
|-----------------------------|------------------------------------------------------------------------------------------|
| Submitter: | EndoVantage, LLC<br>8444 N. 90th Street<br>Suite 125<br>Scottsdale, AZ 85258 |
| Contact Person: | Robert S. Green<br>President<br>Phone: (480) 291-5031<br>Email: simulate@endovantage.com |
| Summary Date: | October 6, 2017 |
| Device Trade Name: | SurgicalPreviewTM |
| Device Common Name: | Radiological Image Processing Software |
| Device Classification Name: | Software for visualization of vascular anatomy and<br>intravascular devices |
| Regulatory Description: | Picture Archiving and Communication System |
| Regulation Number: | 21 CFR 892.2050 |
| Product Code: | PZO |
| Regulatory Classification: | Class II |
| Predicate Device: | EndoSize (K160376) |
### Device Description
SurgicalPreview" is a stand-alone software application that runs on any standard Windows or Mac OSX based computer. It enables physicians to upload patient CT scan studies from various data sources, view them, and process the images. SurgicalPreview™ provides a clinical decision support system during the preoperative planning of endovascular surgery.
SurgicalPreview™ enables visualization and measurement of different vascular structures such as vessels, aneurysms, and other anomalies. SurgicalPreview™ also allows for the ability to computationally model the placement and deployment of neurointerventional devices. SurgicalPreview™ can reconstruct 2D scan slices into 3D models of the patient, and can display supporting DICOM CT scan data. It works with DICOM CT scan images and can access multiple DICOM data files.
The device does not contact the patient, nor does it control any life sustaining devices. Information provided by the software is not intended in any way to eliminate, replace or
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Endovantage. The logo features a stylized globe-like graphic on the left, with red lines forming a grid pattern. To the right of the graphic is the word "endovantage" in red, with a slightly blurred effect. The font is sans-serif and appears to be bolded.
substitute for, in whole or in part, the healthcare provider's judgement and analysis of the patient's condition.
## Indications for Use
SurgicalPreview™ enables visualization and measurement of cerebral blood vessels for preoperational planning and sizing for neurovascular interventions and surgery. SurgicalPreview™ also allows for the ability to computationally model the placement and deployment of neurointerventional devices.
General functionalities are provided such as:
- Segmentation of neurovascular structures ●
- Automatic centerline detection
- Visualization of CT scan images for 2D review and 3D reconstruction ●
- Measurement and annotation tools ●
- Reporting tools ●
Information provided by the software is not intended in any way to eliminate, replace or substitute for, in whole or in part, the healthcare provider's judgement and analysis of the patient's condition.
### Technological Characteristics
SurgicalPreview™ is a software-only device that runs on a standard computer that meets the minimum requirements. SurgicalPreview™ is not intended for use on mobile devices.
SurgicalPreview™ can use local DICOM files. The device does not contact the patient, nor does it control any life sustaining devices. Information provided by the software is not intended in any way to eliminate, replace or substitute for, in whole or in part, the healthcare provider's judgement and analysis of the patient's condition.
### Performance Data
The subject device is designed in conformance with:
- ACR/NEMA Digital Imaging Communications in Medicine (DICOM) Version 3.1 ●
- ISO 14971:2007 Medical Devices Application of Risk Management to Medical ● Devices
- IEC 62304:2006 Medical Device Software Software Life-Cycle Processes ●
All specifications of the SurgicalPreview™ software are validated by a bench test before release. Bench testing includes:
- Tests of importation of DICOM images ●
- Patient manager tests
- Tests of image display and processing
- Functioning tests for visualization of anatomic reconstruction
- Functioning tests for computational modeling of Codman Enterprise Vascular Reconstruction Device and Pipeline Embolization Device
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "endovantage" in red font. The word is written in a stylized font, with the letters connected to each other. There is a graphic to the left of the word that looks like a red dome with a gray grid pattern on it. The word "endovantage" is the main focus of the image.
- Measurement tests ●
- Reports creation and visualization tests ●
### Predicate Device Comparison
The predicate device for the SurgicalPreview™ software is EndoSize (K160376). Both SurgicalPreview™ and its predicate device have similar intended use for preoperational planning of endovascular procedures using image data from a medical scanner. A summary comparison of technological characteristics is provided in Table 1.
| Characteristic | SurgicalPreviewTM | Predicate: EndoSize |
|----------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K171534 | K160376 |
| Indications for Use | SurgicalPreviewTM enables visualization<br>and measurement of vessels for<br>preoperational planning and sizing for<br>neurovascular interventions and surgery.<br>SurgicalPreviewTM also allows for the<br>ability to computationally model the<br>placement and deployment of<br>neurointerventional devices. | EndoSize enables visualization and<br>measurement of structures of the<br>heart and vessels for preoperational<br>planning and sizing for<br>cardiovascular interventions and<br>surgery, and for postoperative<br>evaluation |
| Interface to Image<br>Sources | DICOM image data | Same |
| Import of Patient Data | Manual through keyboard/mouse,<br>automatic import with image file, study<br>creation list | Same |
| List Image Functionality | Deleting, anonymizing, search | Same |
| Image Processing | Segmentation by EndoVantage operator<br>with end-user clinician review and<br>comment | By clinician |
| 3D Assessment | Linear (length and diameter)<br>measurements, volume measurements | Same |
| Image and 3D Display | Orthogonal, color volume rendering, 2D<br>slice review, active presets, 3D view of<br>assemblies of devices | Same |
| DICOM Support | Compatible with all scanner vendor<br>DICOM datasets, storage SCP, import<br>DICOM files, DICOM compliance for<br>CT and enhanced CT, import from<br>DICOMDIR, storage SCU,<br>query/retrieve SCU | Same |
| Computer OS<br>Compatibility | MS Windows and Mac OS | Same |
| Data Interchange /<br>Transfer Method | Secure Internet File Server | Personal computer |
| Output File Format | Web browser via WebGL | Personal computer |
| Characteristic | SurgicalPreviewTM | Predicate: EndoSize |
| Preoperational Planning | Yes | Same |
| Patient Contact | No | Same |
| Human Intervention for<br>Interpretation of Images | Yes | Same |
## Table 1 - SurgicalPreviewTM and Predicate Comparative Analysis
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "endovantage" in red font. The word is slightly blurred, and there is a red graphic to the left of the word. The graphic is a curved shape with a grid pattern inside.
### Substantial Equivalence Conclusion
Based on the information provided in this 510(k) submission, SurgicalPreview™™ has the same intended use as the predicate device. SurgicalPreview™ underwent bench testing to simulate clinical use. The performance data and software verification results demonstrated that the differences in technological characteristics between SurgicalPreview™ and the predicate device do not raise different questions of safety and effectiveness and that SurgicalPreview™™ should perform as intended in the specified use conditions. Endo Vantage concludes that SurgicalPreview™ is substantially equivalent to the predicate device identified in this submission.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.