The sponsor conducted a method comparison study using 867 clinical urine samples to demonstrate substantial equivalence by comparing the performance of the DUS R-50S system against the predicate device (Clinitek Status+).
Concordance/agreement between DUS R-50S and predicate device
Indications for Use
The DUS R-50S System provides a qualitative measurements for specific gravity, pH, leukocytes, nitrite, protein, glucose, ketone, urobilinogen, bilirubin, blood, microalbumin and creatinine in urine specimens. These measurements are used to aid in the diagnosis of metabolic disorders, kidney function anomalies, urinary tract infections and liver function. The system is intended for prescription, in vitro diagnostic use only.
Device Story
Portable automated urinalysis system; analyzes urine specimens using DUS 10 and DUS 2AC reagent strips. Analyzer reads color changes on test pads; automatically displays semi-quantitative results for 12 analytes (leukocytes, nitrite, urobilinogen, protein, pH, blood, specific gravity, ketone, bilirubin, glucose, microalbumin, creatinine). Used in clinical settings by healthcare professionals. Provides rapid screening for metabolic and renal conditions; aids clinical decision-making by identifying potential anomalies requiring further diagnostic investigation. Benefits patients through non-invasive, efficient monitoring of urinary health markers.
Clinical Evidence
Bench testing only. Precision evaluated per CLSI EP05-A3; linearity per CLSI EP06-A; interference per CLSI EP07-A2; method comparison per CLSI EP09-A2. Method comparison study used 867 clinical urine samples compared against the predicate device, showing high exact agreement and agreement within one color block across all analytes.
Technological Characteristics
Reflectance photometer; white light source; RGB sensor. Plastic reagent strips with chemical pads. Dimensions/form factor: semi-automated benchtop analyzer. Connectivity: standalone (no wireless/network transmission). Calibration: automatic via internal white plastic calibration bar. Software: embedded firmware for image processing and color interpretation.
Indications for Use
Indicated for qualitative and semi-quantitative measurement of urine analytes (glucose, blood, creatinine, microalbumin, leukocytes, nitrites, urobilinogen, protein, pH, specific gravity, ketones, bilirubin) to aid in diagnosis of metabolic disorders, kidney function anomalies, urinary tract infections, and liver function. For prescription, in vitro diagnostic use.
Regulatory Classification
Identification
A urinary glucose (nonquantitative) test system is a device intended to measure glucosuria (glucose in urine). Urinary glucose (nonquantitative) measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, hypoglycemia, and hyperglycemia.
Predicate Devices
Siemens Clinitek Status + Urine Chemistry Analyzer (Multistix 10SG) (k091216)
Submission Summary (Full Text)
{0}
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number:
k171521
B. Purpose for Submission:
New Device
C. Measurand:
Measurement of the following analytes in urine: glucose, blood, creatinine, microalbumin, leukocytes, nitrites, urobilinogen, protein, pH, specific gravity, ketones, bilirubin
D. Type of Test:
Qualitative and semi-quantitative urinalysis
E. Applicant:
DFI Co., Ltd.
F. Proprietary and Established Names:
DUS R-50S (Urine Chemistry System)
G. Regulatory Information:
1. Regulation section:
| Name | Regulation | Product code | Device class |
| --- | --- | --- | --- |
| Urinary Glucose (non-quantitative) test system | 21 CFR § 862.1340 | JIL | II |
| Occult Blood test | 21 CFR 864.6550 | JIO | II |
| Creatinine test system | 21 CFR 862.1225 | JFY | II |
| Urinary Protein or Albumin (nonquantitative) test system | 21 CFR 862.1645 | JIR | I |
| Leukocyte peroxidase | 21 CFR 864.7675 | LJX | I |
| Nitrite (nonquantitative) test system | 21 CFR 862.1510 | JMT | I |
{1}
| Urinary Urobilinogen (nonquantitative) test system | 21 CFR 862.1785 | CDM | I |
| --- | --- | --- | --- |
| Urinary pH (nonquantitative) test | 21 CFR 862.1550 | CEN | I |
| Specific Gravity | 21 CFR 862.2800 | JRE | I |
| Ketones (nonquantitative) test system | 21 CFR 862.1435 | JIN | I |
| Urinary Bilirubin and its conjugates (nonquantitative) test system | 21 CFR 862.1115 | JJB | I |
| Automated Urinalysis System | 21 CFR 862.2900 | KQO | I |
2. Panel:
Chemistry (75)
Hematology (81)
H. Intended Use:
1. Intended use(s):
See indications for use statement below.
2. Indication(s) for use:
The DUS R-50S System provides a qualitative and semi-quantitative measurements for specific gravity, pH, leukocytes, nitrite, protein, glucose, ketone, urobilinogen, bilirubin, blood, microalbumin and creatinine in urine specimens. These measurements are used to aid in the diagnosis of metabolic disorders, kidney function anomalies, urinary tract infections and liver function. The system is intended for prescription, in vitro diagnostic use only.
The DUS R-50S System consists of the following:
DUS R-50S Analyzer
DUS 10 Reagent Strips for urinalysis include test pads for leukocytes, nitrite, urobilinogen, protein, pH, blood, specific gravity, ketone (acetoacetic acid), bilirubin and glucose.
DUS 2AC Reagent Strips for urinalysis include test pads for microalbumin and creatinine.
3. Special conditions for use statement(s):
For prescription use only.
For in vitro diagnostic use only.
{2}
4. Special instrument requirements:
DUS R-50S Analyzer
Device Description:
The DUS R-50S (Urine Chemistry System) consists of the DUS R-50S Analyzer and the DUS Series Urine reagent strips (DUS 10 reagent strips and the DUS 2AC reagent strips). The DUS R-50S Analyzer is a semi-automated urine chemistry analyzer. The analyzer uses the principle of light-reflection. The strip is illuminated by white light and the reflected light from the strip is detected by the sensor. The RGB signals are digitized and this digitized image is interpreted by the processor. The parameter values are determined based on the changes in the pad color. The DUS R-50S Urine Chemistry system reports qualitative and semi-quantitative results including the date and time of the measurement, sequence number and ID stored in the analyzer.
DUS Urine reagent strips are plastic strips with different reagent pads affixed for the determination of leukocyte, nitrite, urobilinogen, protein, pH, blood, specific gravity, ketone, bilirubin, glucose, microalbumin and creatinine in urine. The DUS Series Urine reagent strips contain the following multi-parameter strip:
- DUS 10 reagent strips for urinalysis include test pads for leukocytes, nitrite, urobilinogen, protein, pH, blood, specific gravity, ketone (acetoacetic acid), bilirubin and glucose.
- DUS 2AC reagent strips for urinalysis include test pads for microalbumin and creatinine.
J. Substantial Equivalence Information:
1. Predicate device name(s):
Siemens Clinitek Status + Urine Chemistry Analyzer (Multistix 10SG)
2. Predicate 510(k) number(s):
k091216
{3}
3. Comparison with predicate:
| Similarities and Differences | | |
| --- | --- | --- |
| Item | Subject Device k171521 | Predicate Device k091216 Clinitek Status |
| Intended Use | Qualitative and semi-quantitative measurement of glucose, bilirubin, ketone, specific gravity, blood, pH, protein, urobilinogen, nitrite, leukocyte esterase, creatinine and microalbumin | Same |
| Measurands | Glucose, bilirubin, ketone, specific gravity, ketones, blood, pH, protein, urobilinogen, nitrite, leukocyte, creatinine and microalbumin | Albumin, bilirubin, blood (occult), creatinine, glucose, ketone, leukocytes, nitrite, pH, protein, protein-to-creatinine ratio, albumin-to-creatinine ratio, specific gravity, urobilinogen and human chorionic gonadotropin (hCG) |
| Sample type | Human urine | Same |
| Data type | Qualitative and semi-quantitative | Same |
| Instrument Optical System | Reflectance photometer | Same |
| Display | Touch screen | Same |
| Test strip | DUS 10 and DUS 2AC | Multistix 10 SG |
| Assay reaction time | 60 seconds | Same |
K. Standard/Guidance Document Referenced (if applicable):
CLSI-Evaluation of Precision Performance of Clinical Chemistry Devices-EP05-A3
CLSI-Evaluation of Linearity of Quantitative Analytical Methods-EP06-A
CLSI-Interference Testing in Clinical Chemistry-EP07-A2
CLSI-Method Comparison and Bias Estimation Using Patients Samples-EP09-A2
L. Test Principle:
Urobilinogen: The test is based on the Ehrlich's reaction. Color changes from light orange-pink to dark pink.
Glucose: Glucose oxidase catalyzes the oxidation of glucose to form hydrogen peroxide. The hydrogen peroxide thus formed then oxidizes a chromogen on the reaction pad by the action of peroxidase.
{4}
5
Bilirubin: Azo-coupling reaction of bilirubin with a diazonium salt in an acid medium to form an azo dye. Color changes from light tan to beige or light pink.
Ketones: The test is based on the Legal’s test-nitroprusside reaction. Acetoacetic acid in an alkaline medium reacts with nitroferricyanide to produce a color change from beige to purple.
pH: The test is based on the double indicator system. Indicator’s methyl red and bromothymol blue are used to give distinct color changes from orange to green to blue.
Blood: The test is based on the pseudo-peroxidase activity of the heme moiety of hemoglobin and myoglobin. The chromogen is oxidized by a hydroperoxide in the presence of heme and changes color from yellow to blue.
Specific Gravity (SG): Ionic solutes present in the urine cause protons to be released from a polyelectrolyte. As the protons are released the pH decreases and produces a color change of bromothymol blue from blue-green to yellow-green.
Protein: The test is based on the protein-error of-indicators reaction. When pH is held constant by a buffer, indicator dyes release H+ ions because of the protein present and change color from yellow to blue green.
Nitrite: The test is based on the diazotization reaction of nitrite with an aromatic amine to produce a diazonium salt. It is followed by an azo-coupling reaction of this diazonium salt with an aromatic compound on the reaction pad. The azo dye produced causes a color change from white to pink.
Leukocyte: This test pad contains an indoxyl ester and diazonium salt. It is followed by an azo-coupling reaction of the aromatic amine formed by leukocytes esterase with a diazonium salt on the reaction pad. The azo dye produced causes a color change from beige to violet.
Microalbumin: This test is based on dye binding using sulfonephthalein dye. At a constant pH, albumin binds with sulfonephthalein dye to develop a blue color. The resulting color ranges from pale green to aqua blue.
Creatinine: This test is based on the reaction of creatinine with a dye-metal complex. At an alkaline condition, creatinine reacts with a dye-metal complex to form a purplish brown color complex.
{5}
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Precision testing was done in accordance with CLSI EP5-A3. The within-run and within-day precisions were performed by trained laboratory professionals at three clinical sites, using two levels of commercially available urine based quality control material. The studies utilized three DUS R-50S analyzers and three different lot numbers of DUS 10 and DUS 2AC reagent strips.
The within-run precision study was conducted by three laboratory professionals testing 2 levels of commercially available urine based quality control material. Ten test strips from three different lots at 3 sites were used in this study (30 tests at each site = 90 replicates per level). The within-day precision study was conducted by three laboratory professionals each testing one test strip a day from three different lots at 3 sites for ten days (30 tests at each site = 90 replicates per level).
The within-run and within-day precision studies for all 3 sites are summarized below:
| Level 1 control | | | | | |
| --- | --- | --- | --- | --- | --- |
| | | Within-run (90) | | Within-day (90) | |
| Item | Expected test results | Exact agreement (%) | Agreement within +/-one block (%) | Exact agreement (%) | Agreement within +/-one block (%) |
| Urobilinogen | Normal | 90/90 (100%) | 90/90 (100%) | 90/90 (100%) | 90/90 (100%) |
| Glucose | Negative | 90/90 (100%) | 90/90 (100%) | 90/90 (100%) | 90/90 (100%) |
| Bilirubin | Negative | 90/90 (100%) | 90/90 (100%) | 90/90 (100%) | 90/90 (100%) |
| Ketones | Negative | 90/90 (100%) | 90/90 (100%) | 90/90 (100%) | 90/90 (100%) |
| SG | 1.020 | 90/90 (100%) | 90/90 (100%) | 90/90 (100%) | 90/90 (100%) |
| Blood | Negative | 90/90 (100%) | 90/90 (100%) | 90/90 (100%) | 90/90 (100%) |
| PH | 6 | 90/90 (100%) | 90/90 (100%) | 90/90 (100%) | 90/90 (100%) |
| Protein | Negative | 90/90 (100%) | 90/90 (100%) | 90/90 (100%) | 90/90 (100%) |
| Nitrite | Negative | 90/90 (100%) | 90/90 (100%) | 90/90 (100%) | 90/90 (100%) |
| Leukocytes | Negative | 90/90 (100%) | 90/90 (100%) | 90/90 (100%) | 90/90 (100%) |
| Creatinine | 10mg/dL | 90/90 (100%) | 90/90 (100%) | 90/90 (100%) | 90/90 (100%) |
| Microalbumin | 10mg/L | 90/90 (100%) | 90/90 (100%) | 90/90 (100%) | 90/90 (100%) |
{6}
| Level 2 control | | | | | |
| --- | --- | --- | --- | --- | --- |
| | | Within-run (90) | | Within-day (90) | |
| Item | Expected test results | Exact agreement (%) | Agreement within +/-one block (%) | Exact agreement (%) | Agreement within +/-one block (%) |
| Urobilinogen | 4 mg/dL | 90/90 (100%) | 90/90 (100%) | 90/90 (100%) | 90/90 (100%) |
| Glucose | 1000 mg/dL | 90/90 (100%) | 90/90 (100%) | 90/90 (100%) | 90/90 (100%) |
| Bilirubin | 4 mg/dL | 90/90 (100%) | 90/90 (100%) | 90/90 (100%) | 90/90 (100%) |
| Ketones | 40 mg/dL | 90/90 (100%) | 90/90 (100%) | 90/90 (100%) | 90/90 (100%) |
| SG | 1.020 | 89/90 (98.9%) | 90/90 (100%) | 90/90 (100%) | 90/90 (100%) |
| Blood | 120 RBC/μL | 90/90 (100%) | 90/90 (100%) | 90/90 (100%) | 90/90 (100%) |
| PH | 7 | 90/90 (100%) | 90/90 (100%) | 89/90 (98.9%) | 90/90 (100%) |
| Protein | 100 mg/dL | 90/90 (100%) | 90/90 (100%) | 90/90 (100%) | 90/90 (100%) |
| Nitrite | Positive | 90/90 (100%) | 90/90 (100%) | 90/90 (100%) | 90/90 (100%) |
| Leukocytes | 70 WBC/μL | 90/90 (100%) | 90/90 (100%) | 90/90 (100%) | 90/90 (100%) |
| Creatinine | 200mg/dL | 90/90 (100%) | 90/90 (100%) | 90/90 (100%) | 90/90 (100%) |
| Microalbumin | 150mg/L | 90/90 (100%) | 90/90 (100%) | 90/90 (100%) | 90/90 (100%) |
# b. Linearity/assay reportable range:
The linearity study was conducted following the recommendations in CLSI EP6-A. Samples were created by spiking known concentrations of each standard material into a negative urine pool or by serial dilution of a high concentration urine sample with negative urine.
{7}
Testing was performed on three instruments using three different lot numbers of test strips (DUS 2AC and DUS 10) by three laboratory professionals. Each sample was tested in 10 replicates with each lot of reagent strips.
| Analyte | Urine sample concentration tested | Color block output | %Exact match |
| --- | --- | --- | --- |
| Urobilinogen | Negative/Norm | 0.1 | 100% (90/90) |
| | 1 mg/dL | 1 mg/dL | 100% (90/90) |
| | 2 mg/dL | 2 mg/dL | 98.8% (89/90) |
| | 4 mg/dL | 4 mg/dL | 98.8% (89/90) |
| | 8 mg/dL | 8 mg/dL | 100% (90/90) |
| Glucose | Negative | Neg. | 100% (90/90) |
| | 100 mg/dL | ± (100 mg/dL) | 100% (90/90) |
| | 250 mg/dL | +250 mg/dL | 98.8% (89/90) |
| | 500 mg/dL | ++500 mg/dL | 96.6% (87/90) |
| | 1000 mg/dL | +++ 1000 mg/dL | 98.8% (89/90) |
| | 2000 mg/dL | ++++2000 mg/dL | 100% (90/90) |
| Bilirubin | Negative | Neg. | 100% (90/90) |
| | 1 mg/dL | + | 98.8% (89/90) |
| | 2 mg/dL | ++ | 98.8% (89/90) |
| | 4 mg/dL | +++ | 100% (90/90) |
| Ketones | Negative | Neg. | 100% (90/90) |
| | 5 mg/dL | ± 5 mg/dL | 100% (90/90) |
| | 15 mg/dL | + 15 mg/dL | 96.6% (87/90) |
| | 40 mg/dL | ++ 40 mg/dL | 98.8% (89/90) |
| | 80 mg/dL | +++ 80 mg/dL | 97.7% (88/90) |
| | 160 mg/dL | ++++ 160 mg/dL | 100% (90/90) |
| Specific gravity | 1.000 | 1.000 | 100% (90/90) |
| | 1.005 | 1.005 | 98.8% (89/90) |
| | 1.010 | 1.010 | 97.7% (88/90) |
| | 1.015 | 1.015 | 96.6% (87/90) |
| | 1.020 | 1.020 | 98.8% (89/90) |
| | 1.025 | 1.025 | 96.6% (87/90) |
| | 1.030 | 1.030 | 100% (90/90) |
| Nitrite | Negative | Neg. | 100% (90/90) |
| | 0.05 mg/dL | Trace | 100% (90/90) |
| | 10 mg/dL | Pos | 100% (90/90) |
| Blood | Negative | Neg | 100% (90/90) |
| | 10 RBC/μL | Trace | 100% (90/90) |
| | 25 RBC/μL | + 25 RBC/μL | 100% (90/90) |
| | 80 RBC/μL | ++ 80 RBC/μL | 98.8% (89/90) |
| | 200 RBC/μL | +++ 200 RBC/μL | 97.7% (88/90) |
| pH | 5 | 5 | 100% (90/90) |
| | 6 | 6 | 98.8% (89/90) |
| | 6.5 | 6.5 | 98.8% (89/90) |
| | 7 | 7 | 98.8% (89/90) |
{8}
9
| | 7.5 | 7.5 | 97.7% (88/90) |
| --- | --- | --- | --- |
| | 8 | 8 | 100% (90/90) |
| | 8.5 | 8.5 | 100% (90/90) |
| Protein | Negative | Neg. | 100% (90/90) |
| | 15 mg/dL | Trace | 98.8% (89/90) |
| | 30 mg/dL | + 30 mg/dL | 97.7% (88/90) |
| | 100 mg/dL | ++ 100 mg/dL | 97.7% (88/90) |
| | 300 mg/dL | +++ 300 mg/dL | 97.7% (88/90) |
| | 1000 mg/dL | ++++ 1000 mg/dL | 100% (90/90) |
| Leukocytes | Negative | Negative | 100% (90/90) |
| | 15 WBC/μL | 15 WBC/μL | 100% (90/90) |
| | 70 WBC/μL | 70 WBC/μL | 97.7% (88/90) |
| | 125 WBC/μL | 125 WBC/μL | 97.7% (88/90) |
| | 500 WBC/μL | 500 WBC/μL | 98.8% (89/90) |
| Microalbumin | 10 mg/L | 10 mg/L | 100% (90/90) |
| | 30 mg/L | 30 mg/L | 98.8% (89/90) |
| | 80 mg/L | 80 mg/L | 97.7% (88/90) |
| | 150 mg/L | 150 mg/L | 98.8% (89/90) |
| Creatinine | 10 mg/L | 10 mg/L | 100% (90/90) |
| | 50 mg/L | 50 mg/L | 100% (90/90) |
| | 100 mg/L | 100 mg/L | 98.8% (89/90) |
| | 200 mg/L | 200 mg/L | 96.6% (87/90) |
| | 300 mg/L | 300 mg/L | 97.7% (88/90) |
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability: No traceability is claimed
Calibration: The DUS R-50S Analyzer performs a "self-test" and calibration each time it is turned on. Each time a test is run, the analyzer re-calibrates using a white plastic calibration bar located at the upper end of the loading plate. Reflectance measurements from the bar must match the factory set calibration.
Self-life stability: The stability studies was performed to assess the shelf life (closed bottle) stability for the DUS Series Urine reagent strips on the DUS R-50S Analyzer. The protocol and acceptance criteria were reviewed and found sufficient. The stability studies support the sponsors claim that the DUS Series Urine reagent strips, unopened, are stable for 26 months at 2-30 °C at <50% humidity.
Open vial stability: Open bottle stability studies were performed on the DUS Series Urine reagent strips used on the DUS R-50S Analyzer. The protocol and acceptance criteria were reviewed and found sufficient. Based on the studies, once opened the test strips are stable for 6 months when stored from 2-30 °C at <50% humidity.
{9}
Quality Control: No urine controls are provided with this device. In the labeling, the sponsor recommends the use of commercially available quality control material.
# d. Detection limit:
A detection limit study was performed to determine the analytical sensitivity or the lowest concentration of each analyte that can be distinguished from negative. Urine samples were prepared by spiking a negative human urine pool with standard materials to create 3 levels across the measuring range for each analyte concentration.
Each samples concentration was analyzed 30 times using three reagent strip lots, on three DUS R-50S Analyzer. Sensitivity was defined as the cutoff which $95\%$ of the contrived pool measurements were trace or positive. The detection limit study and the linearity study support the following reportable ranges for the DUS 10 reagent strips and the DUS 2AC reagent strips:
DUS 10 reagent strips:
| Test | Reportable Range |
| --- | --- |
| Urobilinogen | Arbitrary: normal to 3+Semi-quantitative: 0.1 – 8 mg/dL |
| Glucose | Arbitrary: Negative to 4+Semi-quantitative: Neg – 2000 mg/dL |
| Bilirubin | Arbitrary: normal to 3+Semi-quantitative: Neg – 4 mg/dL |
| Ketones | Arbitrary: Negative to 4+Semi-quantitative: Neg – 160 mg/dL |
| Specific Gravity | 1.000 – 1.030 |
| Blood | Arbitrary: normal to 3+Semi-quantitative: Neg – 200 RBC/ μL |
| pH | 5 – 8.5 |
| Protein | Arbitrary: Negative to 4+Semi-quantitative: Neg – 1000 mg/dL |
| Nitrite | Semi-quantitative: Neg – 10 mg/dL |
| Leukocytes | Arbitrary: normal to 3+Semi-quantitative: Neg - 500 WBC/μL |
{10}
DUS 2AC reagent strips
| Microalbumin | Semi-quantitative: 10 – 150 mg/L |
| --- | --- |
| Creatinine | Semi-quantitative: 10 – 300 mg/dL |
e. Analytical specificity:
Potential interferents and drugs were evaluated to assess the interfering effect on the performance of the DUS R-50S (Urine Chemistry System). A urine sample pools was prepared at 3 concentrations (negative, low positive and high positive) for each urine chemistry analyte. A negative urine pool was spiked with potential interfering substances at various interference concentrations and analyzed in three replicates using three lots of DUS strips (DUS10, DUS2AC) on three DUS R-50S analyzer. Interference was defined as change in output of $\pm 1$ color block between the spiked and unspiked control sample. The results of the interference studies are summarized below:
| Potential Interfering Substance | Highest concentration at which no interference was observed |
| --- | --- |
| Albumin (protein) | 750 mg/dL |
| Ascorbic acid | 15 mg/dL |
| Acetaminophen | 25 mg/dL |
| Azo gantrinsin (sulfamethoxazole) | 140 mg/dL |
| Bilirubin | 3 mg/dL |
| Captopril | 100 mg/dL |
| Sodium hypochlorite | 0.1% |
| Creatinine | 800 mg/dL |
| Nitrofurantoin | 10 mg/dL |
| Oxalic acid | 30 mg/dL |
| p-amino salicylic acid | 375 mg/dL |
| β-D-Glucose (glucose) | 1000 mg/dL |
| Phenazopyridine | 25 mg/dL |
| Riboflavin | 12 mg/dL |
| Selenium | 150 mg/dL |
| Formaldehyde | 100 mg/dL |
| Hemoglobin | 3 mg/dL |
| Lithium acetoacetate (ketones) | 20 mg/dL |
| Tetracycline | 20 mg/dL |
| Urobilinogen | 10 mg/dL |
| pH | 7 |
| Specific gravity | 1.030 |
The following table shows the substances which did interfere with one or more of the
{11}
DUS Series Urine Reagent Strip analytes. Results are expressed as the lowest concentration of interfering substances that exhibited interference and the resulting change in output of color block.
| Analyte | Potential interfering substance | Concentration at which interference was observed | Change in color block output |
| --- | --- | --- | --- |
| Urobilinogen | Azo gantrinsin (sulfamethoxazol) | ≥105mg/dL | +1 |
| | Bilirubin | ≥4 mg/dL | +1 |
| | Formaldehyde | ≥150 mg/dL | -1 |
| | Nitrofurantion | ≥40 mg/dL | +1 |
| | p-Amino salicylic acid | ≥750 mg/dL | +1 |
| | Phenazopridine | ≥37 mg/dL | +1 |
| | Riboflavin | ≥25 mg/dL | +1 |
| | Selenium | ≥220 mg/dL | +1 |
| Glucose | Sodium hypochlorite | ≥1% | +1 |
| | Formaldehyde | ≥150 mg/dL | +1 |
| | Lithium acetoacetate | ≥40mg/L | -1 |
| | Ascorbic acid | ≥40 mg/dL | -1 |
| | Bilirubin | ≥6 mg/dL | +1 |
| | Specific gravity | ≥1.040 | +1 |
| Bilirubin | Selenium | ≥220 mg/L | +1 |
| | p-amino salicylic acid | ≥12500 mg/L | +1 |
| | Phenazopridine | ≥37 mg/L | +1 |
| | Ascorbic acid | ≥40 mg/dL | -1 |
| Ketones | Bilirubin | ≥4 mg/dL | +1 |
| | Captopril | ≥100 mg/L | +1 |
| | Specific gravity | ≥1.040 | +1 |
| Specific gravity | pH | ≥9 | +1 |
| Protein | Acetaminophen | ≥50 mg/dL | +1 |
| | Bilirubin | ≥6 mg/dL | +1 |
| | Hemoglobin | ≥5 mg/dL | +1 |
| | Nitrofurantoin | ≥20 mg/dL | +1 |
| | Riboflavin | ≥20 mg/dL | -1 |
{12}
| Analyte | Potential interfering substance | Concentration at which interference was observed | Change in color block output |
| --- | --- | --- | --- |
| | Urobilinogen | >20 mg/dL | -1 |
| | pH | ≥9 | +1 |
| Blood (hemolyzed and non-hemolyzed) | Albumin | ≥1000 mg/L | -1 |
| | Ascorbic acid | ≥30 mg/dL | -1 |
| | Bilirubin | ≥6 mg/dL | +1 |
| | Captopril | ≥150 mg/dL | -1 |
| | Sodium hypochlorite | ≥0.5 % | +1 |
| | Formaldehyde | ≥150 mg/dL | +1 |
| | Specific gravity | ≥1.040 | -1 |
| | Nitrofurantoin | ≥30 mg/dL | +1 |
| | Riboflavin | ≥50 mg/dL | +1 |
| | Urobilinogen | ≥20 mg/dL | -1 |
| Nitrite | Ascorbic acid | ≥40 mg/dL | -1 |
| | Bilirubin | ≥4 mg/dL | +1 |
| | Nitrofurantoin | ≥20 mg/dL | +1 |
| | Phenazopyridine | ≥37 mg/dL | +1 |
| | Riboflavin | ≥50 mg/dL | +1 |
| | Selenium | ≥300 mg/dL | +1 |
| | Urobilinogen | ≥15 mg/dL | +1 |
| Leukocytes | Albumin | ≥1000 mg/dL | -1 |
| | Bilirubin | ≥4 mg/dL | +1 |
| | Captopril | ≥150 mg/dL | -1 |
| | Sodium hypochlorite | ≥0.5 % | +1 |
| | Formaldehyde | ≥150 mg/dL | +1 |
| | Specific gravity | ≥1.040 | -1 |
| | Nitrofurantoin | ≥30 mg/dL | +1 |
| | Oxalic acid | ≥45 mg/dL | -1 |
| | Glucose | ≥1500 mg/dL | -1 |
| | Phenazopyridine | ≥50 mg/dL | +1 |
| | Riboflavin | ≥37 mg/dL | +1 |
| | Selenium | ≥300 mg/dL | +1 |
| | Tetracycline | ≥40 mg/dL | -1 |
{13}
| Analyte | Potential interfering substance | Concentration at which interference was observed | Change in color block output |
| --- | --- | --- | --- |
| | Urobilinogen | ≥ 15 mg/dL | +1 |
| Microalbumin | Acetaminophen | ≥ 50 mg/dL | +1 |
| | Bilirubin | ≥ 4 mg/dL | +1 |
| | Hemoglobin | ≥ 5 mg/dL | +1 |
| | Nitrofurantoin | ≥ 20 mg/dL | +1 |
| | Phenazopyridine | ≥50 mg/dL | +1 |
| | Riboflavin | ≥ 25 mg/dL | -1 |
| | Urobilinogen | ≥ 15 mg/dL | -1 |
| | pH | ≥ 9 | +1 |
| Creatinine | Sodium hypochlorite | ≥ 1% | +1 |
| | Hemoglobin | ≥ 5 mg/dL | +1 |
| | Nitrofurantoin | ≥ 30 mg/dL | +1 |
| | Riboflavin | ≥37 mg/dL | +1 |
| | Urobilinogen | ≥ 20 mg/dL | -1 |
The labeling includes the following limitations to address the interferences observed:
As with all laboratory tests, definitive diagnostic or therapeutic decisions should not be based on any single result of method. Substances that cause abnormal urine color may affect the readability of test pads in urinalysis reagent strips.
Urinary ascorbic acid concentrations as low as $40\mathrm{mg / dl}$ can cause interference in specimens with low concentrations of glucose, blood, nitrite and bilirubin.
Urobilinogen: The absence of urobilinogen in the specimen can't be determined. The test area will react with interfering substances known to react with Ehrlich's reagent, such as p- amino salicylic acid. Drugs containing azo gantrisin (sulfamethoxazole) or high bilirubin may give a masking golden color. Preservative formaldehyde may cause false negative. The test is not reliable method for the detection of porphobilinogen.
Glucose: Ascorbic acid (more than $40\mathrm{mg / dl}$ ) may cause false negative result at the low level of glucose. Ketones reduce the sensitivity of the test. Moderately high ketone level ( $>40\mathrm{mg / dl}$ ) may cause a false negative for specimen containing small amount of glucose ( $100\mathrm{mg / dl}$ ). Chlorine Bleach ( $\geq 1\%$ ), low SG and formaldehyde urine may cause false positive result at the low level.
{14}
15
**Bilirubin**: Metabolites of drugs, such as selenium (≥220 mg/dL), phenazopyridine (≥37 mg/dL) may cause false positives. p-Amino salicylic acid (≥1500 mg/dL) can produce a yellow-orange to red color response, which may result in false positive bilirubin readings. Ascorbic acid (≥40 mg/dL) may cause false negative.
**Ketones**: Low level false positive reactions may be seen in highly concentrated urine specimens (high specific gravity) or in specimens containing large amounts of levodopa metabolites drug such as captopril. The uroprotective drug mesna (sodium 2-mercaptoethane sulfonate) and other free-sulfhydryl compounds produce false-positive results in ketone methods that are based on the Legal reaction (alkaline sodium nitroprusside).
**pH**: If excessive urine remains on the strip because of improper test procedure, it is possible that the acidic buffer in the protein pad comes out and affect the pH pas, then pH results may be erroneously decreased. This phenomenon is called “carry-over effect.”
**Blood**: Elevated specific gravity or protein in urine may reduce the reactivity of the blood test pad. Microbial peroxidase associated with urinary tract infection may cause false positive results. Ascorbic acid concentrations (>30 mg/dl) may cause false negatives at the low blood concentrations. Substances that cause abnormal urine color, such as drug containing azo dyes, nitrofurantoin and riboflavin may cause false positive results. Strong oxidizing cleaning agents such as chlorine bleach cause false positive results.
**Specific Gravity (SG)**: High-buffered alkaline urine may cause diminished result, whereas high buffered acidic urine may cause slightly elevated result.
**Protein**: False positive results may be found in strongly basic urine (pH 9). The interpretation of results is also difficult in turbid urine specimens. Metabolites of drugs, such as acetaminophen, hemoglobin may cause false positives. Pigments such as bilirubin and azo-containing compounds cause false positive results.
**Nitrite**: Ascorbic acid (>40 mg/dL) may cause false negative result with urine containing low levels of nitrite (<0.03 mg). The negative result does not always mean that the patient is free from bacteriuria. Medication such as phenazopyridine or other azo-containing compounds or other dyes cause false positive results.
**Microalbumin**: The following substances may cause false positive results; a large amount of hemoglobin (≥5 mg/dl), visibly bloody urine, highly alkaline urine (pH ≥9), disinfectant including quaternary ammonium compound. Substances that cause abnormal urine color, such as drug containing nitrofurantoin, riboflavin may affect the results (false positives).
**Creatinine**: Nitrofurantoin (≥200 mg/L), Riboflavin (≥50 mg/L) and a large amount of
{15}
hemoglobin(≥5mg/dl) cause false positive results. Visibly dark brown color urine may affect the results (false positives). Substances that cause abnormal urine color, such as drug containing nitrofurantoin, riboflavin may affect the results (false positives).
Microalbumin to Creatinine Ratio: A low microalbumin result (10mg/L) in combination with strongly diluted urine (creatinine result of 10mg/dl) could indicate a microalbumin concentration below the sensitivity limit. In that case, consider testing a new specimen, preferably a first morning collection, for greater confidence in the result.
f. Assay cut-off:
Not applicable
2. Comparison studies:
a. Method comparison with predicate device:
The method comparison study was conducted at three clinical sites with a total of 867 samples (Site A 255 samples, Site B 280 samples and Site C 332). Three trained laboratory professionals performed the testing at each site. Fresh urine samples were obtained at each medical facility. The samples were processed within 4 hours. The results from the DUS R-50S urine chemistry system (DUS R-50S instrument, DUS10 and DUS2AC reagent strips) was compared to the predicate device (Siemens Clinitek Status + urine chemistry instrument using Multistix 10SG and CLINITEK Microalbumin 2 test strips). No altered samples were used in this study. The results of the method comparison study for the combined sites are shown in the table below. The data for the combined sites is representative of the data collected at each site.
Urobilinogen
| New device (mg/dL) | Predicate device (mg/dL) | | | | |
| --- | --- | --- | --- | --- | --- |
| | Norm | 1 | 2 | 4 | 8 |
| 8 | | | | 3 | 41 |
| 4 | | | 2 | 86 | |
| 2 | | | 115 | 1 | |
| 1 | | 84 | | | |
| Norm | 535 | | | | |
| | | | | | |
| Total | 535 | 84 | 117 | 90 | 41 |
| Exact agreement | 100% | 100% | 98% | 96% | 100% |
| Within 1 block | 100% | 100% | 100% | 100% | 100% |
{16}
Glucose
| New device (mg/dL) | Predicate device (mg/dL) | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | Neg | 100 | 250 | 500 | 1000 | 2000 |
| 2000 | | | | | | 24 |
| 1000 | | | | 1 | 41 | |
| 500 | | | 2 | 59 | | |
| 250 | | 1 | 71 | 3 | | |
| 100 | | 71 | | | | |
| Neg | 594 | | | | | |
| | | | | | | |
| Total | 594 | 72 | 73 | 63 | 41 | 24 |
| Exact agreement | 100% | 99% | 97% | 94% | 100% | 100% |
| Within 1 block | 100% | 100% | 100% | 100% | 100% | 100% |
Bilirubin
| New device (mg/dL) | Predicate device (mg/dL) | | | |
| --- | --- | --- | --- | --- |
| | Neg | 1 | 2 | 4 |
| 4 | | | | 46 |
| 2 | | 4 | 87 | 1 |
| 1 | | 82 | 3 | |
| Neg | 644 | | | |
| | | | | |
| Total | 644 | 86 | 90 | 47 |
| Exact agreement | 100% | 95% | 97% | 98% |
| Within 1 block | 100% | 100% | 100% | 100% |
Ketones
| New device (mg/dL) | Predicate device (mg/dL) | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | Neg | 5 | 15 | 40 | 80 | 160 |
| 160 | | | | | | 9 |
| 80 | | | | 3 | 39 | |
| 40 | | | | 60 | | |
| 15 | | | 86 | 1 | | |
| 5 | | 60 | 7 | | | |
| Neg | 602 | | | | | |
| | | | | | | |
| Total | 602 | 60 | 93 | 64 | 39 | 9 |
| Exact agreement | 100% | 100% | 92% | 94% | 100% | 100% |
| Within 1 block | 100% | 100% | 100% | 100% | 100% | 100% |
{17}
Blood
| New device (RBC/μL) | Predicate device (RBC/μL) | | | | |
| --- | --- | --- | --- | --- | --- |
| | Neg | 10 | 25 | 80 | 200 |
| 200 | | | | 4 | 159 |
| 80 | | | 8 | 150 | 1 |
| 25 | | 7 | 145 | 5 | |
| 10 | | 159 | 1 | | |
| Neg | 228 | | | | |
| | | | | | |
| Total | 228 | 166 | 154 | 159 | 160 |
| Exact agreement | 100% | 96% | 94% | 94% | 99% |
| Within 1 block | 100% | 100% | 100% | 100% | 100% |
Protein
| New device (mg/dL) | Predicate device (mg/dL) | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | Neg | 15 | 30 | 100 | 300 | 1000 |
| 1000 | | | | | 1 | 17 |
| 300 | | | | 4 | 51 | |
| 100 | | | 5 | 104 | | |
| 30 | | 4 | 126 | 1 | | |
| 15 | | 133 | 2 | | | |
| Neg | 419 | | | | | |
| | | | | | | |
| Total | 419 | 137 | 133 | 109 | 52 | 17 |
| Exact agreement | 100% | 97% | 95% | 95% | 98% | 100% |
| Within 1 block | 100% | 100% | 100% | 100% | 100% | 100% |
Nitrite
| New device | Predicate device | | |
| --- | --- | --- | --- |
| | Neg | Trace | Pos |
| Pos | | | 311 |
| Trace | | 125 | 4 |
| Neg | 427 | | |
| | | | |
| Total | 427 | 125 | 315 |
| Exact agreement | 100% | 100% | 99% |
| Within 1 block | 100% | 100% | 100% |
{18}
Leukocytes
| New device (mg/dL) | Predicate device (mg/dL) | | | | |
| --- | --- | --- | --- | --- | --- |
| | Neg | 15 | 70 | 125 | 500 |
| 500 | | | | 2 | 108 |
| 125 | | | 3 | 136 | 6 |
| 70 | | | 122 | 4 | |
| 15 | | 105 | 2 | | |
| Neg | 379 | | | | |
| | | | | | |
| Total | 379 | 105 | 127 | 142 | 114 |
| Exact agreement | 100% | 100% | 96% | 96% | 95% |
| Within 1 block | 100% | 100% | 100% | 100% | 100% |
pH
| New device | Predicate device | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | 5 | 6 | 6.5 | 7 | 7.5 | 8 | 8.5 |
| 8.5 | | | | | | 5 | 72 |
| 8 | | | | | 5 | 103 | |
| 7.5 | | | | 7 | 106 | 3 | |
| 7 | | | 3 | 151 | 5 | | |
| 6.5 | | 5 | 182 | 2 | | | |
| 6 | 1 | 143 | 7 | | | | |
| 5 | 91 | 6 | | | | | |
| | | | | | | | |
| Total | 92 | 154 | 192 | 160 | 116 | 111 | 42 |
| Exact agreement | 99% | 93% | 95% | 94% | 91% | 93% | 100% |
| Within 1 block | 100% | 100% | 100% | 100% | 100% | 100% | 100% |
Specific gravity
| New device | Predicate device | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | 1.000 | 1.005 | 1.010 | 1.015 | 1.020 | 1.025 | 1.030 |
| 1.030 | | | | | | 5 | 82 |
| 1.025 | | | | | 5 | 107 | |
| 1.020 | | | | 6 | 167 | 2 | |
| 1.015 | | | 5 | 166 | 7 | | |
| 1.010 | | 5 | 147 | 5 | | | |
| 1.005 | | 104 | 4 | | | | |
| 1.000 | 50 | | | | | | |
| | | | | | | | |
| Total | 50 | 109 | 156 | 177 | 179 | 114 | 82 |
| Exact agreement | 100% | 95% | 94% | 94% | 93% | 94% | 100% |
| Within 1 block | 100% | 100% | 100% | 100% | 100% | 100% | 100% |
{19}
20
Creatinine
| New device (mg/dL) | Predicate device (mg/dL) | | | | |
| --- | --- | --- | --- | --- | --- |
| | 10 | 50 | 100 | 200 | 300 |
| 300 | | | | 7 | 188 |
| 200 | | | 7 | 191 | 11 |
| 100 | | 9 | 198 | 9 | |
| 50 | | 151 | 6 | | |
| 10 | 90 | | | | |
| | | | | | |
| Total | 90 | 160 | 211 | 207 | 199 |
| Exact agreement | 100% | 94% | 94% | 92% | 94% |
| Within 1 block | 100% | 100% | 100% | 100% | 100% |
Microalbumin
| New device (mg/dL) | Predicate device (mg/dL) | | | |
| --- | --- | --- | --- | --- |
| | 1 | 3 | 8 | 15 |
| 15 | | | | 445 |
| 8 | | 3 | 98 | 3 |
| 3 | | 72 | 8 | |
| 1 | 238 | | | |
| | | | | |
| Total | 238 | 75 | 106 | 448 |
| Exact agreement | 100% | 96% | 92% | 99% |
| Within 1 block | 100% | 100% | 100% | 100% |
b. Matrix comparison:
Not applicable
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
{20}
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
Literature references were provided to support the stated reference ranges:
Urobilinogen: The normal urobilinogen range is 0.1 to 1.0 Ehrlich unit /dl. If results exceed the concentration of 2.0 mg/dl, the patient and the urine specimen should be evaluated further.
Glucose: The kidney normally excretes small amounts of glucose. Concentrations of 100mg/dl may be considered as abnormal if found consistently.
Bilirubin: Normally no bilirubin is detectable in urine by even the most sensitive methods. Even trace amounts of bilirubin are sufficiently abnormal to require further investigation.
Ketones: Ketone bodies should not be detected in normal urine specimens with this reagent.
pH: Urine values generally range from pH 4.5 to 8.
Blood: Normally, no hemoglobin is detectable in urine (0.010mg/dl; 3 RBC/μl). When hemoglobin appears in urine it indicates kidney disease or a urinary tract disorder. Blood may often be found in the urine of menstruating females.
Specific Gravity (SG): The normal SG of urine ranges from 1.001 to 1.035.
Protein: Normal urine specimens ordinarily contain some protein (<20mg/dL) therefore only persistent elevated levels of urine protein indicate kidney or urinary tract disease. The persistent results of trace level or over indicate significance proteinuria and thus further clinical testing is needed to evaluate the significant of results.
Nitrite: Normally no nitrite is detectable in urine.
Leukocyte: Normally no leukocytes are detectable in urine.
Microalbumin: Normal albumin levels in urine are under 2mg/dl. Microalbuminuria is indicated with results of 3~30mg/dl.
Creatinine: The urine of healthy individuals contains 10~300mg/dl of creatinine. Very low creatinine results can be caused by adulteration of the urine specimen or by severe renal failure.
21
{21}
Microalbumin to Creatinine Ratio: Microalbumin is normally present in urine at concentrations of less than 30 mg albumin/g creatinine. Microalbuminuria is indicated at a ratio result of 30-300 mg/g (abnormal) and clinical albuminuria at a ratio result of >300 mg/g (high abnormal).
Kaplan LA and Pesce AJ. Clinical Chemistry Theory Analysis and Correlation. CV Mosby Co., St. Louis, pp. 1004-1007 (1984)
Levey AS, Coresh J, Balk E, et al. National Kidney Foundation practice guidelines for chronic kidney disease: evaluation, classification, stratification. Ann Intern Med. 139:137-147; 2003
N. Instrument Name:
DUS R-50S Analyzer
O. System Descriptions:
1. Modes of Operation:
Does the applicant's device contain the ability to transmit data to a computer, webserver, or mobile device?
Yes ☐ or No ☑
Does the applicant's device transmit data to a computer, webserver, or mobile device using wireless transmission?
Yes ☐ or No ☑
2. Software:
FDA has reviewed applicant's Hazard Analysis and software development processes for this line of product types:
Yes ☐ or No ☑
3. Specimen Identification:
The DUS R-50S user manual instructs the user to input the patient and sample ID number into the DUS R-50S analyzer by pressing the number keypad of the display (Maximum 10 digits).
22
{22}
4. Specimen Sampling and Handling:
The urine sample should be collected in a clean, dry container with no added preservatives. The specimen should be tested by the user by dipping a DUS test strip into a patient’s urine specimen for no more than two seconds. The user removes the strip and excess urine and places the test strip onto the instrument. The DUS R-50S will perform the reading and evaluation of the urine strip.
5. Calibration:
The DUS R-50S analyzer calibrates using a white plastic calibration pad located on the edge of the strip tray. Calibration is performed automatically before each reagent strip is read. Reflectance measurements from the pad must match the factory set calibration. When strong variations are detected by contamination of the calibration pad or low light intensity an error message (system check failed) will be displayed.
6. Quality Control:
The sponsors states the following in their labeling: For best results, performance of reagent strips should be confirmed by commercially available quality control or assayed urine controls whenever a new bottle of DUS reagent strips are first opened. Water should NOT be used as a negative control. Controls should be tested after performing maintenance or service on the reader. Quality Control materials should be used in accordance with local, state, and/or federal requirements for QC testing.
P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
Not applicable
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
23
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.