← Product Code [NKB](/productcode/NKB) · K171421

# VERTICALE® Posterior Spinal Fixation System/VERTICALE® System (K171421)

_Silony Medical GmbH · NKB · Jan 10, 2018 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K171421

## Device Facts

- **Applicant:** Silony Medical GmbH
- **Product Code:** [NKB](/productcode/NKB.md)
- **Decision Date:** Jan 10, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The VERTICALE® System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities of the thoracic, lumbar and sacral spine. The VERTICALE® System is intended for non-cervical pedicle fixation and nonpedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, and/or lordosis); tumor, pseudoarthrosis; and failed previous fusion in skeletally mature patients. When used in a posterior percutaneous approach with MIS instrumentation, the VERTICALE® MIS System is intended for non-cervical pedicle fixation and nonpedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, and/or lordosis); tumor, pseudoarthrosis; and failed previous fusion in skeletally mature patients. When used for posterior non-cervical pedicle screw fixation in pediatric patients, the VERTICALE® MIS metallic implants are indicated as an adjunct to treat adolescent idiopathic scoliosis. The subject device is intended to be used with autograft and/or allograft. Pedicle screw fixation is limited to a posterior approach.

## Device Story

Posterior spinal fixation system; provides immobilization/stabilization of thoracic, lumbar, sacral spine segments. Components: pedicle screws (solid, cannulated, fenestrated), iliac screws, revision screws, hooks, rods. Implants manufactured from Ti6Al4V ELI (ASTM F136) and CoCrMo (ASTM F1537). Used in OR by surgeons; supports open or minimally invasive (MIS) posterior surgical approaches. System includes modular, bi-functional instruments. Implants act as mechanical stabilizers to facilitate spinal fusion when used with autograft/allograft. Benefits: restores spinal stability, corrects deformities, treats instabilities. Not for use with bone cement or in patients with poor bone quality.

## Clinical Evidence

No clinical data provided. Substantial equivalence supported by non-clinical bench testing, including static/fatigue mechanical testing (ASTM F1717, F1798), biocompatibility (ISO 10993-1), and validation of sterilization, packaging, and shelf life.

## Technological Characteristics

Modular posterior spinal fixation system. Materials: Ti6Al4V ELI (ASTM F136), CoCrMo (ASTM F1537). Instruments: ISO 16061, ASTM F899. Mechanical fixation via pedicle/iliac screws, hooks, and rods. Color-coded by diameter. Non-sterile/sterile processing (implants/instruments). No software or electronic components.

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Predicate Devices

- Expedium Spine System, VIPER and VIPER 2 Systems ([K131802](/device/K131802.md))
- NuVasive SpheRx II (Armada) - Pedicle Screw System ([K092287](/device/K092287.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 10, 2018

Silony Medical GmbH % Ms. Indraj Bamrah Senior Regulatory Consultant Emergo Global Consulting, LLC 2500 Bee Cave Road, Building 1, Suite 300 Austin, Texas 78746

Re: K171421

Trade/Device Name: VERTICALE® Posterior Spinal Fixation System/VERTICALE® System Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral pedicle screw system Regulatory Class: Class II Product Code: NKB, KWP Dated: December 6, 2017 Received: December 8, 2017

Dear Ms. Bamrah:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

## Page 2 - Ms. Indraj Bamrah

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Ronald P. Jean -S for

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K171421

#### Device Name

VERTICALE® Posterior Spinal Fixation System / VERTICALE® System

#### Indications for Use (Describe)

The VERTICALE® System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities of the thoracic, lumbar and sacral spine. The VERTICALE® System is intended for non-cervical pedicle fixation and nonpedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, and/or lordosis); tumor, pseudoarthrosis; and failed previous fusion in skeletally mature patients.

When used in a posterior percutaneous approach with MIS instrumentation, the VERTICALE® MIS System is intended for non-cervical pedicle fixation and nonpedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, and/or lordosis); tumor, pseudoarthrosis; and failed previous fusion in skeletally mature patients.

When used for posterior non-cervical pedicle screw fixation in pediatric patients, the VERTICALE® MIS metallic implants are indicated as an adjunct to treat adolescent idiopathic scoliosis. The subject device is intended to be used with autograft and/or allograft. Pedicle screw fixation is limited to a posterior approach.

| Type of Use (Select one or both, as applicable)                                                 |                                                                                     |
|-------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

{3}------------------------------------------------

## 510(k) Summary

## VERTICALE® POSTERIOR SPINAL FIXATION SYSTEM K171421

## 1. Submission Sponsor

| Silony Medical GmbH     |                                                   |
|-------------------------|---------------------------------------------------|
| Leinfelder Straße 60    |                                                   |
| Leinfelden-Echterdingen |                                                   |
| 70771                   |                                                   |
| Germany                 |                                                   |
| Contact:                | Bircan Tasdelen                                   |
| Title:                  | Head of Quality Management and Regulatory Affairs |

## 2. Submission Correspondent

Emergo Global Consulting, LLC 2500 Bee Cave Road Building 1, Suite 300 Austin, TX 78746 Office Phone: (512) 327.9997 Contact: Indraj Bamrah Title: Senior Regulatory Consultant

- 3. Date Prepared
December 06, 2017

## 4. Device Identification

Trade/Proprietary Name: VERTICALE® Posterior Spinal Fixation System / VERTICALE® System

|                       | Common/Usual Name: | Pedicle Screw Spinal System                             |
|-----------------------|--------------------|---------------------------------------------------------|
| Classification Name:  |                    | Pedicle Screw Spinal System                             |
| Regulation Number:    |                    | 21 CFR 888.3070 Thoracolumbosacral Pedicle Screw System |
| Product Code:         |                    | NKB, KWP                                                |
| Device Class:         |                    | Class II                                                |
| Classification Panel: |                    | Orthopedic                                              |

{4}------------------------------------------------

## 5. Legally Marketed Predicate Device(s)

Primary Predicate

K131802. Expedium Spine System,VIPER and VIPER 2 Systems Medos International Sárl.

Additional Predicate

K092287, NuVasive SpheRx II (Armada) - Pedicle Screw System, NuVasive Inc.

## 6. Indication for Use Statement

The VERTICALE® System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine The VERTICALE® System is intended for non-cervical pedicle fixation and nonpedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with deqeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor, pseudoarthrosis; and failed previous fusion in skeletally mature patients.

When used in a posterior percutaneous approach with MIS instrumentation, the VERTICALE® MIS System is intended for non-cervical pedicle fixation and nonpedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, and/or lordosis); tumor, pseudoarthrosis; and failed previous fusion in skeletally mature patients.

When used for posterior non-cervical pedicle screw fixation in pediatric patients, the VERTICALE® and VERTICALE® MIS metallic implants are indicated as an adjunct to fusion to treat adolescent idiopathic sooliosis. The subject device is intended to be used with autograff and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

## 7. Device Description

The VERTICALE® System is a posterior fixation system for stabilizing the thoracic, lumbar and sacral spine. The VERTICALE® System is a modular, versatile fixation system consisting of pedicle screws, iliac screws, revision screws, hooks and rods for the fusion of treated spinal segments.

The pedicle screws are available in a range of sizes, and in solid, cannulated and fenestrated designs, which when used in combination with the rods, can be used for a range of indications. The pedicle screws and hooks are manufactured from titanium alloy and the rods, from titanium alloy or cobalt chrome alloy.

The VERTICALE® System comprises modular ergonomically designed, bi-functional two-in-one (2 in 1) instruments with modular handle options. The VERTICALE® System also offers a set of 

{5}------------------------------------------------

minimal invasive instrumentation (MIS) to facilitate minimally invasive implantation of the components.

The VERTICALE® System is color coded by diameter for ease of identification.

The VERTICALE® implants are manufactured from Ti6Al4V ELI conforming to ASTM F136 and CoCrMo conforming to ASTM F1537. The VERTICALE® System specific instruments are manufactured from materials conforming to ISO 16061 and ASTM F899.

The safety and effectiveness of this device has not been established when used in conjunction with bone cement or for use in patients with poor bone quality (e.q., osteoporosis, osteopenia). This device is intended only to be used with saline or radiopaque dye.

#### 8. Substantial Equivalence Discussion

The VERTICALE® System has been shown to be substantially equivalent to the predicate device with respect to indications for use, principles of operation, technological characteristics, materials, and performance testing.

#### 9. Non-Clinical Performance Data

As part of demonstrating safety and effectiveness of VERTICALE® System and in showing substantial equivalence to the predicate devices, Silony Medical GmbH completed a number of non-clinical performance tests. The VERTICALE® System meets all the requirements for overall design, sterilization, biocompatibility, and performance testing results confirming that the design output meets the design inputs and specifications for the device.

The VERTICALE® System passed all the testing in accordance with internal requirements, national standards, and international standards shown below to support substantial equivalence of the subject device:

- Mechanical testing Tested in accordance with ASTM F1717-15 for static compression . bending, static tension bending, static torsion and fatigue with acceptable results.
- . Mechanical testing - Tested in accordance with ASTM F1798-13 for static axial slipping, polyaxial slipping, head splay and fatigue with acceptable results.
- . Biocompatibility - Biological evaluation was performed according to 10993-1 on known materials supporting that no toxic substances or materials are used; materials used for the subject device are materials all used in marketed implant devices.
- o Validation activities have been successfully conducted for cleaning and sterilization, packaging, shelf life and transport.

#### 10. Clinical Performance Data

There was no human clinical testing required to support the medical device as the indications for use is equivalent to the predicate device. These types of devices, including the predicate devices, have been on the market for many years with proven safety and efficacy for the use of 

{6}------------------------------------------------

the device. The non-clinical testing detailed in this submission supports the substantial equivalence of the device.

## 11. Statement of Substantial Equivalence

By definition, a device is substantially equivalent to a predicate device when the device has the same intended use and the same technological characteristics as the previously cleared predicate device; or the device has the same intended use and different technological characteristics, but it can be demonstrated that the device is substantially equivalent to the predicate device and that the new device does not raise additional questions regarding its safety and effectiveness as compared to the predicate devices.

The VERTICALE® System, as designed and manufactured, is determined to be substantially equivalent to the referenced predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K171421](https://fda-staging.innolitics.com/device/K171421)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
