K171402 · KARL STORZ Endoscopy-America, Inc. · EOB · Feb 2, 2018 · Ear, Nose, Throat
Device Facts
Record ID
K171402
Device Name
KARL STORZ Slim Nasopharyngolaryngoscope
Applicant
KARL STORZ Endoscopy-America, Inc.
Product Code
EOB · Ear, Nose, Throat
Decision Date
Feb 2, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4760
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The KARL STORZ Slim Nasopharyngolaryngoscope System is intended for visualization purposes during ENT diagnostic and therapeutic procedures.
Device Story
Reusable, flexible videoscope; utilizes CMOS chip for image capture; internal LED light source. Used for visualization/access of nasal lumens and airway anatomy during ENT diagnostic and therapeutic procedures. Operated by clinicians in clinical settings. Device provides real-time visualization to the healthcare provider to assist in clinical decision-making and procedural guidance. Benefits include improved visualization of airway anatomy for diagnostic and therapeutic interventions.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via non-clinical bench testing, including electrical safety, EMC, biocompatibility, and HLD validation studies.
Technological Characteristics
Flexible videoscope; 8.5 Fr (2.83 mm) distal tip; 350 mm working length; 1.2 mm working channel; 270° up/down deflection. CMOS imager; 90° FOV; 0° direction of view; internal LED light source. Sterilization: STERRAD NX, STERRAD 100NX, STERIS SYSTEM 1E, V-PRO maX, V-PRO 60, EtO. HLD: CIDEX. Standards: ISO 8600-1, ISO 8600-3, ISO 8600-4, ISO 10993 series, ANSI/AAMI ES 60601-1, IEC 60601-1-2.
Indications for Use
Indicated for visualization and access of nasal lumens and airway anatomy (including nasopharyngeal and trachea) during diagnostic and therapeutic procedures in adult and pediatric patients.
Regulatory Classification
Identification
A nasopharyngoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the nasopharyngoscope and is intended to examine or treat the nasal cavity and nasal pharynx. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel and flexible plastic. This generic type of device includes the antroscope, nasopharyngolaryngoscope, nasosinuscope, nasoscope, postrhinoscope, rhinoscope, salpingoscope, flexible foreign body claw, flexible biopsy forceps, rigid biopsy curette, flexible biospy brush, rigid biopsy forceps and flexible biopsy curette, but excludes the fiberoptic light source and carrier.
Predicate Devices
KARL STORZ Rhino-Laryngo-Broncho-Fiberscope (K981458)
KARL STORZ Flexible Video-Neuro-Endoscope System (K161112)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services seal on the left and the FDA acronym and name on the right. The seal is a stylized depiction of a human figure, while the FDA acronym and name are written in blue, with the acronym in a square and the name in a sans-serif font.
February 2, 2018
KARL STORZ Endoscopy-America, Inc. AnnaLisa Smullin Regulatory Engineer 2151 E. Grand Avenue El Segundo, CA 90245
Re: K171402
Trade/Device Name: KARL STORZ Slim Nasopharyngolaryngoscope Regulation Number: 21 CFR 874.4760 Regulation Name: Nasopharyngoscope (Flexible or Rigid) and Accessories Regulatory Class: Class II Product Code: EOB Dated: January 2, 2018 Received: January 3, 2018
Dear AnnaLisa Smullin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely yours,
Eric A. Mann -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K171402
Device Name
KARL STORZ Slim Nasopharyngolaryngoscope System
Indications for Use (Describe)
The KARL STORZ Slim Nasopharyngoscope System is indicated to provide visualization and access of the nasal lumens and airway anatomy (including nasopharyngeal and trachea) during diagnostic and therapeutic procedures in adult and pediatric patients.
Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for STORZ. The logo is blue and consists of the word "STORZ" in large, bold letters. Below the word "STORZ" is the phrase "KARL STORZ - ENDOSKOPE" in smaller letters. The logo is simple and professional.
## 510(k) Summary
This 510(k) Summary is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CFR 807.92. All data included in this document is accurate and complete to the best of KSEA's knowledge.
| Submitter: | KARL STORZ Endoscopy-America, Inc.<br>2151 E. Grand Avenue<br>EI Segundo, CA 90245 |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | AnnaLisa Smullin<br>Regulatory Engineer<br>Phone: (424) 218-8376<br>Fax: (424) 218-8519 |
| Date of<br>Preparation: | January 30, 2018 |
| Type of 510(k)<br>Submission: | Traditional |
| Device<br>Identification: | Trade Name: KARL STORZ Slim Nasopharyngolaryngoscope<br>System<br>Classification Name: Nasopharyngoscope (21 CFR Part 874.4760) |
| Regulatory<br>Class: | II |
| Product Code: | EOB |
| Guidance<br>Document: | Not Applicable for EOB product code |
| Recognized<br>Consensus<br>Standards: | Not Applicable for EOB product code |
| Predicate<br>Device(s): | Primary Predicate Device: KARL STORZ Endoscopy-America's<br>Karl Storz Rhino-Laryngo-Broncho-Fiberscope (K981458)<br>Secondary Predicate Device: MAHE INTL., INC's MAHE Pediatric<br>Naso Pharyngoscope (K030857)<br>Reference: KARL STORZ Endoscopy-America's KARL STORZ<br>Flexible Video-Neuro-Endoscope System (K161112) |
| Device<br>Description: | The KARL STORZ Slim Nasopharyngolaryngoscope (Model<br>Number: 11161V) is a reusable, flexible videoscope used for<br>examination of the nasal lumens and airway anatomy. The KARL<br>STORZ Slim Nasopharyngolaryngoscope System is identical to the |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image is a logo for Karl Storz Endoskope. The word "STORZ" is written in large, bold, blue letters. Below that, the words "KARL STORZ - ENDOSKOPE" are written in smaller, blue letters.
| | KARL STORZ Flexible Video-Neuro-Endoscope System, which is<br>currently distributed under K161112. | | | | | |
|-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|------------------------------------------------------------|----------------------------------------------|---------------------------------------------------------------------------------------------------|--|
| Intended Use: | The KARL STORZ Slim Nasopharyngolaryngoscope System is<br>intended for visualization purposes during ENT diagnostic and<br>therapeutic procedures. | | | | | |
| Indications For<br>Use: | The KARL STORZ Slim Nasopharyngolaryngoscope System is<br>indicated to provide visualization and access of the nasal lumens and<br>airway anatomy (including nasopharyngeal and trachea) during<br>diagnostic and therapeutic procedures in adult and pediatric patients. | | | | | |
| Technological<br>Characteristics: | | Subject Device | Primary<br>Predicate<br>Device,<br>K981458 | Secondary<br>Predicate<br>Device,<br>K030857 | Reference<br>Device,<br>K161112 | |
| | | KARL STORZ<br>Slim<br>Nasopharyngo-<br>laryngoscope<br>System | KARL STORZ<br>Rhino-<br>Laryngo-<br>Broncho-<br>Fiberscope | MAHE<br>Pediatric<br>Nasopharyngo-<br>scope | KARL STORZ<br>Flexible Video-<br>Neuro-<br>Endoscope<br>System | |
| | | Physical Characteristics | | | | |
| | Type of<br>Scope | Flexible | Flexible | Flexible | Flexible | |
| | Distal Tip<br>Diameter | 8.5 Fr, 2.83 mm<br>(elliptical:<br>3.2 mm x 2.4 mm) | 3.5 mm | 2.8 mm | 8.5 Fr, 2.83 mm<br>(elliptical:<br>3.2 mm x 2.4 mm) | |
| | Working<br>Shaft<br>Length | 350 mm | 340 mm | 300 mm | 350 mm | |
| | Working<br>Channel<br>Diameter | 1.2 mm | 1.2 mm | No working<br>channel | 1.2 mm | |
| | Deflection<br>(°) | Up: 270°<br>Down: 270° | Up: 180°<br>Down: 100° | Up: 120°<br>Down: 100° | Up: 270°<br>Down: 270° | |
| | Optical Characteristics | | | | | |
| | Type of<br>Imager | CMOS chip | Fiberscope | Fiberscope | CMOS chip | |
| | Field of<br>View | 90° | 85° | 95° | 90° | |
| | Direction of<br>View | 0° | 0° | 0° | 0° | |
| | Light<br>Source | Internal LED | External Xenon | External LED | Internal LED | |
| | Cleaning and Sterilization Methods | | | | | |
| | Cleaning | Manual | Manual | Manual | Manual | |
| | Sterilization | - STERRAD NX<br>- STERRAD<br>100NX<br>- STERIS<br>SYSTEM 1E<br>- V-PRO maX<br>- V-PRO 60<br>- EtO | - EtO | Unknown | - STERRAD NX<br>- STERRAD<br>100NX<br>- STERIS<br>SYSTEM 1E<br>- V-PRO maX<br>- V-PRO 60<br>- EtO | |
| | HLD | - CIDEX | - CIDEX | Unknown | No HLD | |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Karl Storz Endoskope. The logo is in blue and features the word "STORZ" in large, bold letters. Below the word "STORZ" is the phrase "KARL STORZ - ENDOSKOPE" in a smaller font.
| Non-Clinical<br>Performance<br>Data: | Electrical Safety and Electromagnetic Compatibility Summary<br>The device design, including the electrical safety and<br>electromagnetic compatibility (EMC), has not changed since<br>clearance of the reference device, the KARL STORZ Flexible Video-<br>Neuro-Endoscope System in K161112. The electrical safety and<br>EMC data submitted in K161112 was in compliance with the<br>following FDA recognized standards:<br>• ANSI/AAMI ES 60601-1:2005 (Recognition Number: 19-4)<br>• IEC 60601-1-2:2007 (Recognition Number: 19-8)<br>As the subject device remains identical to the reference device, no<br>additional electrical safety or EMC data is needed for the KARL<br>STORZ Slim Nasopharyngolaryngoscope System.<br><br>Bench Testing Summary<br>The device design, including the performance has not changed since<br>clearance of the reference device, the KARL STORZ Flexible Video-<br>Neuro-Endoscope System in K161112. The performance data<br>submitted in K161112 was in compliance with the following FDA<br>recognized standards:<br>• ISO 8600-1: Fourth edition: 2015-10-15 (Recognition<br>Number: 9-110)<br>• ISO 8600-3: First edition: 1997-07-01 (Recognition Number:<br>9-84)<br>• ISO 8600-4: Second edition: 2014-03-15 (Recognition<br>Number: 9-94)<br>As the subject device remains identical to the reference device, no<br>additional performance data is needed for the KARL STORZ Slim<br>Nasopharyngolaryngoscope System.<br><br>Biocompatibility Summary<br>The device design, including the materials, has not changed since<br>clearance of the reference device, the KARL STORZ Flexible Video-<br>Neuro-Endoscope System in K161112.The Biological Evaluation<br>data submitted in K161112 was in compliance with the following<br>FDA recognized standards:<br>• ISO 10993-1:2009/(R) 2013 (Recognition Number: 2-156)<br>• ISO 10993-5:2009/ (R) 2014 (Recognition Number: 2-153)<br>• ISO 10993-10:2010 (Recognition Number: 2-173)<br>• ISO 10993-11:2006/ (R) 2010 (Recognition Number: 2-118)<br>As the subject device remains identical to the reference device, no<br>additional biological evaluation data is needed for the KARL<br>STORZ Slim Nasopharyngolaryngoscope System. |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | High Level Disinfection (HLD) Summary |
| | The reference device, the KARL STORZ Flexible Video-Neuro-<br>Endoscope System (K161112), is a critical device per Spaulding<br>Classification, whereas the subject device is a semi-critical device,<br>requiring a thorough cleaning followed by sterilization or high level<br>disinfection (HLD). In order to support an additional HLD method<br>(CIDEX), a new HLD validation study was performed using a<br>representative device. The HLD data submitted was in compliance<br>with the following FDA standards. Where applicable, FDA standard<br>recognition numbers are noted:<br>• AAMI TIR12: 2010<br>• ISO 15883-5:2005<br>• AAMI TIR30:2011<br>• AAMI/ANSI/ISO 11737-1:2006/ (R)2011 (Recognition<br>Number: 14-227)<br>• ASTM E1837-96:2014 |
| Clinical<br>Performance<br>Data: | Clinical testing was not required to demonstrate substantial<br>equivalence to the predicate devices. Non-clinical bench testing was<br>sufficient to establish the substantial equivalence of the<br>modifications. |
| Conclusion: | The conclusions drawn from the nonclinical tests demonstrate that<br>the subject device, KARL STORZ Slim Nasopharyngolaryngoscope<br>System performs as well as or better than the legally marketed<br>predicate devices. |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image is a logo for "STORZ KARL STORZ - ENDOSKOPE". The word "STORZ" is in large, bold, blue letters. Below that, the words "KARL STORZ - ENDOSKOPE" are in smaller, blue letters.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.