ODFS Pace XL

K171396 · Odstock Medical, Ltd. · GZI · Jan 31, 2018 · Neurology

Device Facts

Record IDK171396
Device NameODFS Pace XL
ApplicantOdstock Medical, Ltd.
Product CodeGZI · Neurology
Decision DateJan 31, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.5810
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The ODFS® Pace XL is intended to provide ankle dorsiflexion in individuals who have a dropped foot as a consequence of upper motor neuron injury. By detecting the swing phase of gait through a foot switch signal, appropriate electrical stimulation of the leg and ankle muscles may improve gait by flexing the foot of persons who have lost or impaired function. There may be additional benefits from FES such as muscle re-education. prevention/retardation of disuse atrophy, increased or maintained range of joint motion and increase in local blood flow. The ODFS® Pace XL is a medical device and should only be used under medical supervision for the treatment of dropped foot following an upper motor neuron injury.

Device Story

ODFS Pace XL is a body-worn, battery-powered, single-channel external functional neuromuscular stimulator (FES). It receives input from a footswitch (wired or wireless) to detect the swing phase of gait. The device transforms this input into electrical pulses delivered via skin-surface electrodes, typically over the common peroneal nerve. It produces muscle contractions causing ankle dorsiflexion. Used in clinics or by patients at home under medical supervision. The clinician programs the stimulation parameters; the device operates automatically during gait. It can also function as a powered muscle stimulator for exercise programs. Benefits include improved gait, muscle re-education, atrophy prevention, and increased range of motion.

Clinical Evidence

Bench testing only. No clinical data presented. Performance verified through functional testing of stimulator controls, output specifications, and timing. Wireless footswitch performance verified against wired predicate. Compliance with ISO 60601-1, ISO 60601-1-2, and ISO 60601-2-10 confirmed.

Technological Characteristics

Battery-powered (9V) single-channel FES. Housing: ABS plastic. Output: Biphasic (symmetrical/balanced asymmetrical) pulses, 0-360µs width, 20-60Hz frequency. Connectivity: Wireless transceiver for footswitch accessories. Electrodes: Self-adhesive skin-surface. Standards: ISO 60601-1, ISO 60601-1-2, ISO 60601-2-10. Software: Firmware V1.3.

Indications for Use

Indicated for individuals with dropped foot due to upper motor neuron injury. Used for gait improvement via ankle dorsiflexion and as a powered muscle stimulator for muscle re-education, prevention/retardation of disuse atrophy, maintenance/increase of range of motion, and increased local blood flow.

Regulatory Classification

Identification

An external functional neuromuscular stimulator is an electrical stimulator that uses external electrodes for stimulating muscles in the leg and ankle of partially paralyzed patients (e.g., after stroke) to provide flexion of the foot and thus improve the patient's gait.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text. January 31, 2018 Odstock Medical Ltd Steven Crook Quality and Regulatory Manager Salisbury District Hospital Salisbury, Wiltshire, SP2 8BJ UK Re: K171396 Trade/Device Name: ODFS Pace XL Regulation Number: 21 CFR 882.5810 Regulation Name: External Functional Neuromuscular Stimulator Regulatory Class: Class II Product Code: GZI, IPF Dated: October 27, 2017 Received: November 13, 2017 Dear Steven Crook: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good {1}------------------------------------------------ manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. # Vivek J. Pinto -S for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K171396 Device Name ODFS® Pace XL #### Indications for Use (Describe) The ODFS® Pace XL is intended to provide ankle dorsiflexion in individuals who have a dropped foot as a consequence of upper motor neuron injury. By detecting the swing phase of gait through a foot switch signal, appropriate electrical stimulation of the leg and ankle muscles may improve gait by flexing the foot of persons who have lost or impaired function. There may be additional benefits from FES such as muscle re-education, prevention/retardation of disuse atrophy, increased or maintained range of joint motion and increase in local blood flow. The ODFS® Pace XL is a medical device and should only be used under medical supervision for the treatment of dropped foot following an upper motor neuron injury. | Type of Use (Select <i>one</i> or <i>both</i> , as applicable) | |----------------------------------------------------------------| |----------------------------------------------------------------| X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows a logo with the letters "OML" in white on a blue background. Below the logo, the words "LEADING REHABILITATION" and "THROUGH TECHNOLOGY" are written in blue. The text is centered below the logo. The logo and text are likely for a company or organization that specializes in rehabilitation through technology. # Odstock Medical Limited National Clinical FES Centre Salisbury District Hospital Salisbury Wilitshire SP2 8BJ UK Tel +44 (0) 1722 439541 Fax +44 (0) 1722 425263 steven.crook@odstockmedical.com www.odstockmedical.com ### 510(k) Summary (Traditional format) For the ODFS® Pace XL #### Sponsor/Owner/Manufacturer Odstock Medical Limited Laing Building Salisbury District Hospital Salisbury Wiltshire SP2 8BJ United Kingdom Odstock Medical Ltd is a Registered Establishment, number: 3005344585 | Contact person: | Steven Crook, Quality and Regulatory Manager | |-----------------|----------------------------------------------| | Telephone: | 0044 1722 439541 | | Fax: | 0044 1722 425263 | | Email: | steven.crook@odstockmedical.com | 30th January 2018 Date prepared lan Downie (Chairman). Lisa Thomas (Director) · Philip Casson (CEO) · Paul Taylor (Director) Odstock Medical Limited (Company No: 5532620) is part of Salis NHS Foundation Trus {4}------------------------------------------------ #### DEVICE NAME | Trade/Proprietary Name: | ODFS® Pace XL | |-------------------------|----------------------------------------------------------------------------------------| | Common/Usual Name: | Dropped foot stimulator | | Classification Name: | External Functional Neuromuscular Stimulator (FES)<br>Powered Muscle Stimulator (NMES) | | Classification No.: | 21 CFR 882.5810 | | Classification Code: | Class II | | Product Codes: | GZI and IPF | #### Predicate Devices | Predicate Device | K number | Comment | |------------------|----------|-----------------------------------------------------------------------| | ODFS® Pace | K102115 | Odstock Medical Ltd | | ODFS® III V6.2 | K050991 | Odstock Medical Ltd (transferred from<br>Salisbury District Hospital) | #### Device Description The ODFS® Pace XL is an external functional neuromuscular stimulator. It consists of a small, body worn, battery powered, single channel stimulator which is triggered by a footswitch worn in the shoe. The footswitch options are: - 1. Shoe insole incorporating wireless transceiver which provides wireless triggering - 2. 'Out of shoe' wireless transceiver with 'in shoe' heel switch. Provides wireless triggering. - Wired 'in shoe' heel switch for where 1 & 2 are not appropriate or to cover an 3. emergency, e.g. loss of battery power in wireless accessory. The triggering element and detection is the same across all options and option 3 is common to the ODFS® Pace (predicate device). Electrical pulses are generated by the device in order to stimulate muscle contractions in the lower limb. (Neuromuscular stimulation) Triggering with a footswitch permits the stimulation to be varied according to the gait of the user. Stimulation produces contraction of the appropriate muscles to cause dorsiflexion of the ankle in individuals who have impaired or absent ability to pick up their foot during walking as a result of a neurological injury. This paralysis affecting the foot is commonly known as 'drop foot' or 'dropped foot'. The electrical stimulation is delivered via self adhesive skin surface electrodes. The electrodes may be flexibly placed according to the optimum response of the patient but typically are located over the common peroneal nerve as it passes near the head of the fibula. The user may be offered the below knee ODFS® Leg Cuff accessory as aid to locating electrodes repeatedly day to day. An additional option is to mount the ODFS® Pace XL upon the ODFS® Leg Cuff. {5}------------------------------------------------ For individuals presenting with weakness and/or spasticity a muscle stimulation exercise program can be delivered by the ODFS® Pace XL. In this case the footswitch trigger is not used and the ODFS® Pace functions as a powered muscle stimulator. A clinician would determine the need for this application which can be enabled from a setup option in the ODFS® Pace XL. (Default is to have the exercise disabled) #### Intended Use The ODFS® Pace XL is intended to provide ankle dorsiflexion in individuals who have a dropped foot as a consequence of upper motor neuron injury. By detecting the swing phase of gait through a foot switch signal, appropriate electrical stimulation of the leg and ankle muscles may improve gait by flexing the foot of persons who have lost or impaired function. There may be additional benefits from FES such as muscle re-education. prevention/retardation of disuse atrophy, increased or maintained range of joint motion and increase in local blood flow. The ODFS® Pace XL is a medical device and should only be used under medical supervision for the treatment of dropped foot following an upper motor neuron injury. #### Technological Characteristics and Substantial Equivalence The technological characteristics of the ODFS® Pace XL and predicate devices are tabulated below. All the devices are battery powered and deliver electrical stimulation pulses of similar waveform, pulse width and frequency. The ODFS® Pace is a development of the ODFS® III v6.2 and therefore has identical indications for use as an FES device, as does the ODFS® Pace XL. The ODFS® Pace XL has internal software (firmware) version 1.3 whereas the cleared ODFS® Pace has firmware version 1.2. Internal circuitry is unchanged apart from addition of the wireless transceiver expansion circuit. The footswitch contains a force sensitive resistor within a thin proprietary package. As a wired footswitch this may be located flexibly within the shoe. Generally the position is under the heel. Two wireless footswitch versions are available for the ODFS® Pace XL. An insole, which has a fixed position for the switch element, fits inside the full length of the shoe. The thickness is of the order of 6mm and a non-functioning equivalent insole is supplied for the other shoe. The thickness of the insole is a limitation which prevents the use in certain shoes. The second wireless version (OML LINQ™ ) makes use of the thin wired footswitch with a shortened cable. This is connected to a small out of shoe transceiver that can be flexibly located with a clip or retained in a sock. Extensive performance testing was part of the wireless footswitch development to ensure equivalence to the wired version. {6}------------------------------------------------ | Feature | ODFS® Pace XL | ODFS® Pace | ODFS® III v6.2 | |---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Firmware | V1.3 | V1.2 | Not applicable | | Waveform | Biphasic (symmetrical<br>and balanced<br>asymmetrical) | Biphasic (symmetrical<br>and balanced<br>asymmetrical) | Biphasic (symmetrical<br>and balanced<br>asymmetrical) | | Pulsewidth | $0\mu s$ - $360\mu s$ (±10%)<br>according to patient<br>selection (3.6 $\mu s$ steps) | $0\mu s$ - $360\mu s$ (±10%)<br>according to patient<br>selection (3.6 $\mu s$ steps) | 7 $\mu s$ - $365\mu s$ (±10%)<br>according to patient<br>selection | | Frequency | Default is 40Hz.<br>20 – 60 Hz in 5Hz steps<br>available | Default is 40Hz.<br>20 – 60 Hz in 5Hz steps<br>available | 40Hz ± 10% | | Indications for use | The ODFS® Pace XL is<br>intended to provide ankle<br>dorsiflexion in individuals<br>who have a dropped foot<br>as a consequence of upper<br>motor neuron injury. By<br>detecting the swing phase<br>of gait through a foot<br>switch signal, appropriate<br>electrical stimulation of<br>the leg and ankle muscles<br>may improve gait by<br>flexing the foot of persons<br>who have lost or impaired<br>function. There may be<br>additional benefits from<br>FES such as muscle re-<br>education,<br>prevention/retardation of<br>disuse atrophy, increased<br>or maintained range of<br>joint motion and increase<br>in local blood flow.<br>As a NMS/NMES device:<br>relaxation of muscle<br>spasm; prevention or<br>retardation of disuse<br>atrophy; increasing local<br>blood circulation; muscle<br>re-education; maintaining<br>or increasing range of<br>motion. | The ODFS® Pace is<br>intended to provide ankle<br>dorsiflexion in individuals<br>who have a dropped foot<br>as a consequence of upper<br>motor neuron injury. By<br>detecting the swing phase<br>of gait through a foot<br>switch signal, appropriate<br>electrical stimulation of<br>the leg and ankle muscles<br>may improve gait by<br>flexing the foot of persons<br>who have lost or impaired<br>function. There may be<br>additional benefits from<br>FES such as muscle re-<br>education,<br>prevention/retardation of<br>disuse atrophy, increased<br>or maintained range of<br>joint motion and increase<br>in local blood flow.<br>As a NMS/NMES device:<br>relaxation of muscle<br>spasm; prevention or<br>retardation of disuse<br>atrophy; increasing local<br>blood circulation; muscle<br>re-education; maintaining<br>or increasing range of<br>motion. | To provide ankle<br>dorsiflexion in individuals<br>who have a dropped foot<br>as a consequence of upper<br>motor neuron injury. By<br>detecting the swing phase<br>of gait through a foot<br>switch signal, appropriate<br>electrical stimulation of<br>the leg and ankle muscles<br>may improve gait by<br>flexing the foot of persons<br>who have lost or impaired<br>function. There may be<br>additional benefits from<br>FES such as muscle re-<br>education,<br>prevention/retardation of<br>disuse atrophy, increased<br>or maintained range of<br>joint motion and increase<br>in local blood flow. | | Accessories | Wireless heel switch (2<br>versions), wired heel<br>switch. Electrode locator<br>cuff (Leg Cuff) | Heel switch (wired) | Heel switch (wired) | | Electrode size &<br>shape | 5cm x 5cm | 5cm x 5cm | 5cm x 5cm | # 5.6 Technological Characteristics and Substantial Equivalence - Table {7}------------------------------------------------ #### 5.7 Performance testing The ODFS® Pace XL system has been subjected to testing to verify that the device meets its functional and output specifications. Production units pass functional testing of stimulator controls, settings and indicators, stimulus output and timing, footswitch operation, and power consumption. Parameters measured under the conditions specified in FDA Guidance Document for Powered Muscle Stimulator 510(k)s are included in the ODFS® Pace 510(k) submission. These are identical for the ODFS® Pace XL and reproduced below. (Appendix 1) Wireless footswitch production models have been tested to functionally perform as reliably and effectively as the wired version. In addition to functional and operational performance testing the ODFS® Pace XL passed the applicable requirements of the following standards: - ISO 60601-1-1:2006 Medical Electrical Equipment - General requirements for basic safety and essential performance. - O ODFS Pace XL is electrically identical to the ODFS Pace in respect of the pulse delivery to the patient. A wireless transceiver circuit is additional and activated when a wireless footswitch accessory is in use. - o Performance is verified with wired and wireless footswitches. - . ISO 60601-1-2:2007 Medical Electrical Equipment - General requirements for basic safety and essential performance. Collateral standard: Electromagnetic compatibility - Requirements and tests - o ODFS Pace XL tested by third party report supplied for inclusion in submission. - o Wireless accessories are also tested and included in the submitted report. - I ISO 60601-2-10:2001 Medical electrical equipment. Particular requirements for the safety of nerve and muscle stimulators - o ODFS Pace XL outputs etc. unchanged from ODFS Pace Based on the information provided above the ODF5® Pace XL is substantially equivalent to legally marketed predicate devices. Performance data is included on the next page. {8}------------------------------------------------ | Performance data | | | | ODFS® Pace XL<br>Modified device | ODFS® Pace<br>Marketed device | ODFS III v6.2<br>Marketed device | |----------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------|--------------------------------------------------------------------| | Feature | ODFS® Pace (cleared device)<br>ODFS® Pace XL (modification) | ODFS III (cleared device) | 510(k) number | Not cleared | K102115 | K050991 | | Waveform | Biphasic (symmetrical and balanced<br>asymmetrical) | Biphasic (symmetrical and balanced<br>asymmetrical) | Device Name | ODFS® Pace XL dropped<br>foot stimulator | ODFS® Pace dropped foot<br>stimulator | ODFS Dropped foot<br>stimulator (ODFS III V6.2) | | Shape | (Quasi)-rectangular | (Quasi)-rectangular | Manufacturer | Odstock Medical Ltd | Odstock Medical Ltd | Odstock Medical Ltd | | Maximum Output<br>Voltage<br>(+/-10%) | 86V peak; 8Vrms @ 500R<br>104V peak; 11.5Vrms @1k<br>125V peak; 14Vrms @ 2K<br>140V peak; 18Vrms @10K | 85V peak; 9Vrms @500R<br>115V peak; 11Vrms @1K<br>130Vpeak;13Vrms@2K<br>150V peak; 14Vrms@10K | Power sources | 9V Alkaline battery<br>(ANSI 1604A)<br>9V LiPo rechargeable<br>(same physical form –<br>removed for charging) | 9V Alkaline battery (ANSI<br>1604A) | 9V Alkaline battery (ANSI<br>1604A) | | Maximum output<br>current<br>(+/-10%) | 172mA @ 500R<br>62mA @ 2K<br>14mA @ 10K | 170mA peak:18mArms@500R<br>155mA peak:11mArms@1K<br>65mApeak:6.5mArms@2K<br>15mA peak:1.4mArms@10K | Method of line<br>current isolation | No mains power<br>connection. Internal 9V<br>battery only | No mains power<br>connection. Internal 9V<br>battery only | No mains power<br>connection. Internal 9V<br>battery only | | Pulse width | 0 - 360 microseconds (+/- 10%) in 3.6<br>microsecond steps | 7 to 360 microseconds, clinician<br>selectable | Patient leakage<br>current | Normal condition: <10μA<br>Single fault condition:<br><25μA | Normal condition: <10μA<br>Single fault condition:<br><25μA | Normal condition: <10μA<br>Single fault condition:<br><100μA | | Frequency | 20Hz to 60Hz in 5Hz steps (+/-10%) | 40Hz (+/- 4Hz) | No. of output modes | 2, Symmetrical and<br>Balanced Asymmetrical | 2, Symmetrical and<br>Balanced Asymmetrical | 2, Symmetrical and<br>Balanced Asymmetrical | | Interferential<br>modes:<br>Beat frequency<br>For multiphasic<br>waveforms only | Not Applicable | Not Applicable | Number of output<br>channels | 1 | 1 | 1 | | Symmetrical<br>phases | Yes - symmetrical mode | Yes - symmetrical mode | Synchronous or<br>alternating? | Not applicable - single<br>channel device | Not applicable - single<br>channel device | Not applicable - single<br>channel device | | Phase duration | 360 (+/- 20) microseconds | 365 (+/- 20) microseconds | Channel isolation<br>method | Not applicable - single<br>channel device | Not applicable - single<br>channel device | Not applicable - single<br>channel device | | Net charge (µC per<br>pulse) | 0µC | 0µC | Regulated current or<br>regulated voltage? | Regulated voltage, source<br>resistance effectively =<br>360Ω | Regulated voltage, source<br>resistance effectively =<br>360Ω | Regulated voltage, source<br>resistance effectively =<br>360Ω | | Max phase charge | 50μC @ 500R | 50μC @ 500R | Software/<br>Firmware/<br>Microprocessor<br>control? | Yes | Yes | No | | Maximum current<br>density 1 | 0.9mArms/cm² @ 500R<br>0.6mArms/cm² @ 1K | 0.9mArms/cm² @ 500R<br>0.6mArms/cm² @ 1K | Automatic overload<br>trip | No | No | No | | Maximum power<br>density (mW/cm2) | 3.9mW/cm2 @ 500R<br>3.2mW/cm2 @ 1K | 3.9mW/cm2 @ 500R<br>3.2mW/cm2 @ 1K | Automatic No-Load<br>trip | No | No | No | | Burst Mode (i.e.<br>pulse trains)<br>a. pulses per burst<br>b. Bursts per<br>second<br>c. Burst duration<br>(seconds)<br>d. Duty Cycle | NA - Continuous pulse train during<br>swing phase | NA - Continuous pulse train during<br>swing phase | Automatic shut off | Yes(after 4hr in 'sleep'<br>state) | Yes (after 4hr in 'sleep'<br>state) | No | | ON Time (seconds) | 0.2 (+/- 0.02) seconds to 6(+/-)<br>seconds as adjusted by clinician | 0.2 (+/- 0.02) seconds to 6(+/-)<br>seconds as adjusted by clinician | Patient override<br>control | Yes (Pause button) | Yes (Pause button) | Yes (Pause button) | | OFF Time<br>(seconds) | Varies with gait cadence of patient | Varies with gait cadence of patient | Indicator display | | | | | Additional Features<br>(if applicable) | Pulse train is initiated by heel switch<br>(wired or wireless(XL)) in patient's<br>shoe. Pulse train is programmed by<br>clinician for fixed duration or<br>terminated after footswitch contact. | Pulse train is initiated by heel<br>switch(wired) in patient's shoe. Pulse<br>train is programmed by clinician for<br>fixed duration or terminated after<br>footswitch contact. | On/Off status | Yes – LCD displays<br>function | Yes – LCD displays<br>function | No, audio at on/off,<br>pause/un-pause. Stim On<br>indicator light | | Low battery | Yes - via LCD | Yes - via LCD | Yes | | | | {9}------------------------------------------------ Substantial equivalence data – basic characteristics. {10}------------------------------------------------ | | ODFS® Pace XL<br>Modified device | ODFS® Pace<br>Marketed device | ODFS III v6.2<br>Marketed device | |----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------| | Timer range<br>(exercise) | 5 to 100 minutes in 5 min<br>steps plus unlimited | 5 to 100 minutes in 5 min<br>steps plus unlimited | Not applicable | | Compliance with<br>voluntary standards | Yes<br>EN60601-1:2006<br>EN60601-1-2:2007<br>EN60601-2-10:2001 | Yes<br>EN60601-1:2006<br>EN60601-1-2:2007<br>EN60601-2-10:2001 | Yes<br>EN60601-1<br>EN60601-1-2<br>EN60601-2-10 | | Compliance with 21<br>CFR 898 | Yes | Yes | Yes | | Weight | 112g with battery | 112g with battery | 146g with battery | | Dimensions | 7.2cm x 6.2 cm x 2.6cm | 7.2cm x 6.2 cm x 2.6cm | 6cm (w) x 3cm (h) x<br>11cm(d) | | Housing material and<br>construction | Custom moulded ABS<br>plastic enclosure with clear<br>window for LCD | Custom moulded ABS<br>plastic enclosure with clear<br>window for LCD | Moulded thermoplastic<br>case with plastic front<br>panel | | Accessories | In shoe wired footswitch.<br><br>In shoe wireless footswitch<br>(insole)<br>Or<br>In shoe footswitch with<br>external wireless<br>transmitter (OML LinqTM)<br><br>Electrode cable.<br>Self adhesive skin-surface<br>electrodes.<br>[Option] ODFS® Leg cuff<br>for electrode location and<br>device mounting. (also<br>applicable to ODFS® Pace) | In shoe wired footswitch.<br>Electrode cable.<br>Self adhesive skin-surface<br>electrodes. | In shoe wired footswitch.<br>Electrode cable.<br>Self adhesive skin-surface<br>electrodes. | Output calculations: FDA Guidance suggests measurements to be made at a load resistance of 500R. Measurements made with 10x probe and digital oscilloscope calibrated against reference 'scope with traceable calibration. | Under these conditions: | | |--------------------------------------------------------------|----------------| | R = 500R | | | Single pulse measurement: | | | peak V | 86V, | | peak instantaneous current, | 172mA | | therefore RMS voltage (scope measurement function) | 8.0V to 8.5V | | (depending on how many pulses incorporated into measurement) | | | RMS current | 16mA to 17mA | | Max phase charge (taken as peak forward charge) | 48µCoulombs | | | @ approx 320µs | ("integrate" by summing V/500 at each 80us sampling interval in a spreadsheet) {11}------------------------------------------------ Image /page/11/Figure/1 description: The figure shows voltage and charge for ODFS Pace at max settings of pulsewidth and current (PP3 battery, 8.5V). The y-axis shows Volts (V) & Charge (µC), and the x-axis shows Time (µs). The figure shows two lines, one for Int µC and one for Voltage. The Int µC line starts at around 40 µC and decreases over time, while the Voltage line starts at around 80 V and decreases to around -40 V before increasing to around 0 V. Charge balances at approx 6ms from pulse leading edge. Current density and power density calculations: Smallest recommended electrodes: 5cm x 5cm square PALS, area = 25cm2 Current density: 172/25 = 6.9mA.cm²2 peak 17/25 ==0.68mA.cm²2 RMS Power density using P=V2/R: | peak | $ (86^2/500)/25 = 0.6W\cdot cm^{-2} $ | |------|-----------------------------------------| | RMS | $ (8.5^2/500)/25 = 5.8mW\cdot cm^{-2} $ | -- X --
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...