K171389 · Microport Orthopedics, Inc. · JWH · Aug 8, 2017 · Orthopedic
Device Facts
Record ID
K171389
Device Name
EVOLUTION Revision CCK System
Applicant
Microport Orthopedics, Inc.
Product Code
JWH · Orthopedic
Decision Date
Aug 8, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The EVOLUTION® Revision CCK system is indicated for use in knee arthroplasty in skeletally mature patients with the following conditions: 1. Non-inflammatory degenerative joint disease: including osteoarthritis, traumatic arthritis, or avascular necrosis; 2. Inflammatory degenerative joint disease, including rheumatoid arthritis; 3. Correction of functional deformity; 4. Revision procedures where other treatments or devices have failed; and treatment of fractures that are unmanageable using other techniques. The EVOLUTION" Medial-Pivot Total Knee System Nonporous implants are for cemented use only.
Device Story
EVOLUTION® Revision CCK (Constrained Condylar Knee) system provides surgical options for revision or complex primary total knee replacements. System includes femoral components, augments, stem offset adapters, tibial inserts, and stem caps. Used in clinical orthopedic settings by surgeons to replace damaged knee joint surfaces. Device components are implanted to restore joint function and stability in patients with degenerative disease or failed prior arthroplasty. Output is a reconstructed knee joint; clinical decision-making relies on surgeon assessment of patient anatomy and joint pathology. Benefits include improved joint stability and function for patients with complex orthopedic needs.
Clinical Evidence
No clinical data provided. Evidence consists of bench testing: bacterial endotoxin levels, patellofemoral/tibiofemoral contact area, tibial post shear fatigue, taper disassembly, and taper fatigue. Results showed performance comparable or superior to predicate K990030 and K162026.
Technological Characteristics
Materials: Cobalt chrome alloy (ASTM F75), titanium alloy (ASTM F136), UHMWPE (ASTM F648). Semi-constrained cemented prosthesis. Available in 8 standard sizes and 2 plus sizes. Includes femoral component, augments, stem offset adapters, tibial insert, and stem cap. Cemented use only.
Indications for Use
Indicated for skeletally mature patients requiring knee arthroplasty due to non-inflammatory degenerative joint disease (osteoarthritis, traumatic arthritis, avascular necrosis), inflammatory degenerative joint disease (rheumatoid arthritis), functional deformity, revision of failed treatments/devices, or unmanageable fractures.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, one behind the other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
MicroPort Orthopedics Inc. Matt Paul Regulatory Affairs Project Manager 5677 Airline Road Arlington, Tennessee 38002
August 8, 2017
Re: K171389
Trade/Device Name: EVOLUTION Revision CCK System Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: JWH Dated: May 10, 2017 Received: May 11, 2017
Dear Matt Paul:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### K171389
Device Name
## EVOLUTION® Revision CCK System
Indications for Use (Describe)
The EVOLUTION® Revision CCK System is indicated for use in knee arthroplasty in skeletally mature patients with the following conditions:
1. Non-inflammatory degenerative joint disease: including osteoarthritis, traumatic arthritis, or avascular necrosis
- 2. Inflammatory degenerative joint disease, including rheumatoid arthritis;
- 3. Correction of functional deformity
4. Revision procedures where other treatments or devices have failed; and treatment of fractures that are unmanageable using other techniques.
The EVOLUTION® Medial-Pivot Total Knee System Nonporous implants are for cemented use only.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for MicroPort Orthopedics. The logo features a blue rounded square with a white circle inside, containing a red diamond shape. To the right of the icon, the word "MicroPort" is written in blue, with "Orthopedics" written in red below it. The tagline "Full Function. Faster." is visible below the company name.
EVOLUTION® Revision CCK System 510(k) Summary
# 510(k) Summary of Safety and Effectiveness
In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CFR 807, this information serves as a Summary of Safety and Effectiveness for the use of EVOLUTION® Revision CCK System.
| Submitted by: | MicroPort Orthopedics Inc.<br>5677 Airline Rd, Arlington TN, 38002<br>Phone: 866-872-0211<br>Fax: 855-446-2247 |
|-----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date: | August 2, 2017 |
| Contact Person: | Matt Paul |
| Proprietary Name: | EVOLUTION® Revision CCK System |
| Common Name: | Total knee implant |
| Classification Name<br>and Reference: | 21 CFR 888.3560 - Knee joint patellofemorotibial metal/polymer/metal semi-<br>constrained cemented prosthesis.<br>Class II |
| Subject Product Code<br>and Panel Code: | Orthopedics/87/JWH |
| Predicate Device: | K093552 EVOLUTION® MP Total Knee System<br>K142550 EVOLUTION® MP Revision Femur<br>K162026 EVOLUTION® Revision Tibial System<br>K093552 EVOLUTION® MP Total Knee System |
{4}------------------------------------------------
K171389 p.2/5
Image /page/4/Picture/1 description: The image is the logo for MicroPort Orthopedics. The logo features a blue square with a white circle inside, and a red diamond in the center of the circle. To the right of the square is the text "MicroPort" in blue, with "Orthopedics" in red below it. Below that is the text "Full Function. Faster" in blue.
EVOLUTION® Revision CCK System 510(k) Summary
| Reference Devices: | Femoral Component: |
|--------------------|---------------------------------------------------------------------------|
| | K990030 ADVANCE® Revision (CCK System) |
| | K142550 EVOLUTION® MP Revision Femur |
| | K063731 ADVANCE® STATURE Femoral Component |
| | K901992 Whiteside Ortholoc Modular Post. Stab. Knee System |
| | K932677 AXIOM Total Knee System |
| | K863668/K871769 Omnifit Series I Total Knee System |
| | K892800/K895639/K911572/K930831 MG II Total Knee System – Flat, AP Lipped |
| | K862837 Omnifit Series I Total Knee System |
| | K882322/K896542/K901457/K914072 Continuum (CKS) Primary Total Knee System |
| | K953439 Ultrack (ADVANCE®) Total Knee |
| | K972626 ADVANCE® Knee System |
Femoral Augment: K990030 ADVANCE® Revision CCK System
#### Tibial Insert:
K093552 EVOLUTION® MP CS/CR and PS Tibial Inserts K990030 ADVANCE® Revision CCK Tibial Insert K162026 EVOLUTION® Revision Tibial Base and Screw K973524 ADVANCE® Modular Tibial Component, Screw K063731 ADVANCE® STATURE Femoral Component K953439 Ultrack (ADVANCE®) Total Knee System K932677 AXIOM Total Knee System K972770 ADVANCE® Ultra-Congruent Tibial Insert K102380 EVOLUTION® MP Total Knee System K152298 EVOLUTION® BIOFOAM Tibial Base K063128 ADVANCE® Spiked Porous Tibial Base
MicroPort Orthopedics BV Hoogoorddreef 5, 1101 BA Amsterdam, The Netherlands +31 20 545 01 00
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for MicroPort Orthopedics. The logo consists of a blue square with rounded corners, inside of which is a white circle containing a red star-like shape. To the right of the square is the text "MicroPort" in blue, with "Orthopedics" underneath in red. Below that is the phrase "Full Function. Faster."
#### Device Information
#### A. Intended Use
The EVOLUTION® Revision CCK system is indicated for use in knee arthroplasty in skeletally mature patients with the following conditions:
- 1. Non-inflammatory degenerative joint disease: including osteoarthritis, traumatic arthritis, or avascular necrosis;
- 2. Inflammatory degenerative joint disease, including rheumatoid arthritis;
- 3. Correction of functional deformity;
- 4. Revision procedures where other treatments or devices have failed; and treatment of fractures that are unmanageable using other techniques.
The EVOLUTION" Medial-Pivot Total Knee System Nonporous implants are for cemented use only.
#### B. Device Description
EVOLUTION® Revision CCK (Constrained Condylar Knee) is being introduced to supplement MicroPort Orthopedics' knee product lines to provide options in the case of revision or complex primary total knee replacements. The design features are summarized below:
- Components manufactured from cobalt chrome alloy, titanium alloy, and UHMWPE conforming to ASTM F75, F136, and F648, respectively
- . Available in 8 standard sizes, left and right
- Available in 2 plus sizes, left and right
- . System includes femoral component, augments, stem offset adapters, tibial insert and stem cap
## C. Substantial Equivalence Information
The design features and materials of the subject devices are substantially equivalent to those of the predicate devices. The indications for use are identical to the predicate devices. The fundamental scientific technology of the modified devices has not changed relative to the predicate device. The safety and effectiveness of the subject devices are adequately supported by the substantial equivalence information, materials information, and analysis data provided within this Premarket Notification.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for MicroPort Orthopedics. The logo features a blue square with rounded corners, inside of which is a white circle containing a red star-like shape. To the right of the square is the text "MicroPort" in blue, with "Orthopedics" in red below it. Below that is the text "Full Function, Faster" in a smaller font.
K171389 p.4/5 EVOLUTION® Revision CCK System
510(k) Summary
# D. Nonclinical Testing
The subject EVOLUTION® Revision CCK system was evaluated for bacterial endotoxins and found to be less than the USP endotoxin limit of 20 EU/device.
The subject EVOLUTION® Revision CCK system was evaluated for patellofemoral contact area, tibiofemoral contact area, tibial post shear fatigue, taper disassembly, and taper fatigue.
Results for patellofemoral contact area testing concluded the subject EVOLUTION® Revision CCK system to perform as well as the predicate K990030 implants in respect to measured contact area. Results for tibiofemoral contact area testing concluded the subject EVOLUTION® Revision CCK System to possess greater maximum contact area than the predicate K990030 implants. Results of tibial post shear fatigue testing concluded the subject EVOLUTION® Revision CCK system to demonstrate greater fatigue resistance than the predicate K990030. Results of taper disassembly testing indicate the subject EVOLUTION® Revision CCK system will withstand typical in vivo disassociation forces as reported in peer reviewed literature. Results of taper fatigue testing demonstrated the subject EVOLUTION® Revision CCK Femoral constructs to achieve taper disassembly values greater than those of the predicate K162026.
# E. Clinical Testing
Clinical data was not provided for the subject devices.
{7}------------------------------------------------
K171389 p.5/5
Image /page/7/Picture/1 description: The image is the logo for MicroPort Orthopedics. The logo features a blue square with rounded corners on the left side, inside of which is a white circle with a red diamond shape in the center. To the right of the square is the text "MicroPort" in blue, with "Orthopedics" in red below it. Below that is the text "Full Function. Faster."
EVOLUTION® Revision CCK System 510(k) Summary
## F. Component and Accessory Compatibility
The subject devices are compatible with the following previously cleared MicroPort Orthopedics products.
#### Femoral
| K162026 | EVOLUTION® Revision Stem Extensions |
|---------|-----------------------------------------------------------|
| K162026 | EVOLUTION® Revision Straight Stem Extension Adapters |
| K142550 | EVOLUTION® Revision Distal and Posterior Femoral Augments |
#### Patella
| K122218 | ADVANCE® Patella |
|---------|------------------|
| K953439 | ADVANCE® Patella |
#### Tibial
| K162026 | EVOLUTION® Revision Tibial Base |
|---------|------------------------------------------------------|
| K162026 | EVOLUTION® Revision Tibial Augments |
| K162026 | EVOLUTION® Revision Modular Keel |
| K162026 | EVOLUTION® Revision Tibial Stem Offset Adapters |
| K162026 | EVOLUTION® Revision Straight Stem Extension Adapters |
| K162026 | EVOLUTION® Revision Stem Extensions |
The EVOLUTION® Revision CCK System instrumentation includes reamers, trials, guides, broaches, bushings, spacers, drivers, and clamps.
#### G. Conclusion
The design features, materials information, predicate testing and analysis data provided in this premarket notification adequately support the substantial equivalence of EVOLUTION® Revision CCK.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.