The Micromed BRAIN QUICK system is intended to acquire and store physiological signals for EEG and/or polysomnography (PSG) and to transfer data to separate PSG analysis software. The system includes: -hardware recorders intended to acquire and temporarily store physiological signals for EEG and to transfer the data to separate polysomnographic analysis software, and -software which is used for the signal acquisition, display, storage, process and measure of biologic signals. The hardware recorders (amplifiers) are BRAIN SPY PLUS, MORPHEUS, SD LTM 32 EXPRESS and SD LTM 64 EXPRESS. They are intended to acquire the same kind of signals by using the SystemPlus EVOLUTION software. The recorder models differ based on the number of available EEG channels (16 in the BRAIN SPY PLUS, 24 in the MORPHEUS, 32 in SD LTM 32 EXPRESS and 64 in SD LTM 64 EXPRESS). A system with up to 256 channel can be obtained by cascading 4 SD LTM 64 EXPRESS amplifiers. The device is intended to be used by physicians, and other medical professions that are trained in EEG and/or PSG.
Device Story
Modular physiological recording system; acquires EEG/PSG bio-potential signals via third-party electrodes. Hardware includes amplifiers (BRAIN SPY PLUS, MORPHEUS, SD LTM 32/64 EXPRESS) and communication interface to PC. SystemPlus EVOLUTION software processes, displays, and stores signals. Used in clinics, hospitals, ORs, and sleep labs by trained medical professionals. No direct patient contact; device is non-sterile and reusable. Physician interprets displayed traces; no automated diagnostic judgments or physiologic alarms provided. Benefits include standardized signal acquisition and data management for neurological and sleep assessment.
Clinical Evidence
Bench testing only. Compliance with IEC 60601-1 (safety) and IEC 60601-2-26 (EEG essential performance) verified. Electromagnetic compatibility and immunity testing performed per IEC 60601-1-2. Software verification conducted per design control procedures to ensure conformance to system requirements.
Technological Characteristics
Modular system; amplifiers (16-64 channels, up to 256 cascaded); 16-bit A/D conversion; 8192 Hz sampling rate; >100dB CMRR; >10^9 to >10^12 Ohm input impedance. Connectivity via Bluetooth or cable. Power via isolation transformer/batteries. Software-based signal processing. Non-sterile, reusable. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-26.
Indications for Use
Indicated for adult and pediatric patients requiring acquisition and storage of physiological signals for EEG and/or polysomnography (PSG). Used by trained physicians and medical professionals in research, clinical, hospital, operating room, epilepsy monitoring, and sleep laboratory environments.
Regulatory Classification
Identification
An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.
Predicate Devices
Blackrock NeuroMed Cervello Bio-Potential Signal Acquisition System-Basic (K122196)
MICROMED MORPHEUS, MICROMED BRAIN SPY PLUS (K071782)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 1, 2018
Micromed S.p.A % Joseph Ouellette Senior Consultant EAS Consulting Group 1700 Diagonal Road Suite 750 Alexandria, Virginia 22314
Re: K171384
Trade/Device Name: Micromed BRAIN OUICK System Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: Class II Product Code: GWQ, OLV Dated: January 2, 2018 Received: January 2, 2018
Dear Mr. Ouellette:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Michael J. Hoffmann -S
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K171384
Device Name Micromed BRAIN QUICK system
#### Indications for Use (Describe)
The Micromed BRAIN QUICK system is intended to acquire and store physiological signals for EEG and/or polysomnography (PSG) and to transfer data to separate PSG analysis software.
The system includes:
-hardware recorders intended to acquire and temporarily store physiological signals for EEG and to transfer the data to separate polysomnographic analysis software, and
-software which is used for the signal acquisition, display, storage, process and measure of biologic signals.
The hardware recorders (amplifiers) are BRAIN SPY PLUS, MORPHEUS, SD LTM 32 EXPRESS and SD LTM 64 EXPRESS. They are intended to acquire the same kind of signals by using the SystemPlus EVOLUTION software. The recorder models differ based on the number of available EEG channels (16 in the BRAIN SPY PLUS, 24 in the MORPHEUS, 32 in SD LTM 32 EXPRESS and 64 in SD LTM 64 EXPRESS). A system with up to 256 channel can be obtained by cascading 4 SD LTM 64 EXPRESS amplifiers.
The device is intended to be used by physicians, and other medical professions that are trained in EEG and/or PSG.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## K171384 510(k) Summary
#### SUBMITTER
Micromed S.p.A Via Giotto, 2-31021 Mogliano Veneto (TV) Italy
Phone: +39-041-593700 Fax: +39-041-5397011 Email: monica.camillo@micromed.eu
Contact Person Monica Camillo Date Prepared: January 31, 2018
#### SUBJECT DEVICE
| Trade Name: | Micromed BRAIN QUICK System |
|------------------------|-----------------------------------------|
| Common or usual names: | Electroencephalograph |
| Proprietary name: | BRAIN QUICK |
| Classification name: | Electroencephalograph (21 CFR 882.1400) |
| Regulatory Class: | II |
| Product Code: | GWQ, OLV |
#### PREDICATE DEVICE
The predicate devices used in this submission are:
- K122196 Blackrock NeuroMed Cervello Bio-Potential Signal Acquisition . System-Basic
- . K071782 MICROMED MORPHEUS, MICROMED BRAIN SPY PLUS
#### DEVICE DESCRIPTION
The Micromed BRAIN QUICK System is a multi-modal physiological recording system, for use in research institutions, clinics, hospitals, operating rooms, epilepsy monitoring unit environments and/or sleep laboratories. The Micromed BRAIN OUICK System is a modular system which can be configured in different versions; however, the intended use, operating principle and technology remain the same for the different configurations of the Micromed BRAIN QUICK system. The basic unit consists of one or more acquisition amplifiers, a communication interface for transferring data to a PC, a PC with monitor and user interface devices and software for the processing, storage and display of the acquired traces. The power supply is provided through an isolation transformer allowing the connection of several devices to the transformer secondary winding.
The Micromed BRAIN QUICK system acquires EEG signals (bio-potential derived from the brain activity) from third-party commercial electrodes connected to the patient. The Micromed BRAIN QUICK system processes the signal (amplification and filtering), converts it from analogical to digital representation and transmits the result to the software which controls the display of the resulting signal on a monitor, the storage of the
{4}------------------------------------------------
signal on a digital storage media (e.g. hard disk or DVD), the analysis for deriving signal measures. The clinical meaning of the displayed trace is determined by the physician only, based upon the signal morphology.
The device is not intended to directly come in contact with any part of the patient body. The device uses and is compatible with EEG commercial electrodes. The device is neither sterile nor sterilizable. It is reusable. The expected lifetime is ten years.
Intended use, users, environment and indications for use are reported in the intended use section of this 510(k) Summary.
Key Performance Specifications and Characteristics of the Device are reported in the comparison to predicate devices section of this 510(k) Summary.
## INTENDED USE
The Micromed BRAIN OUICK System is intended to record, display, store and archive bio-potential signals. It can do this wireless via Class I Bluetooth or via cables.
The Indications for Use statement for the Micromed BRAIN QUICK system is not identical to the predicate device; however, the differences do not alter the intended use of the device nor do they affect the safety and effectiveness of the device relevant to the predicate:
- The Micromed BRAIN QUICK System has overall the same intended use in the same environments as the predicate device.
- Both the Micromed BRAIN OUICK System and the predicate have the same intended use for the recording, display and storage and archiving of bio-potential signals.
- The Micromed BRAIN QUICK system IFU has been reworded in order to highlight the presence and functions of the software.
| | Micromed BRAIN QUICK<br>(Subject Device) | Blackrock NeuroMed Cervello<br>Basic Bio-Potential Signal<br>Acquisition System<br>K122196 | Micromed<br>Morpheus, Brain<br>Spy Plus<br>K071782 |
|---------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended<br>Use | The Micromed BRAIN QUICK<br>system is intended to record,<br>display, store and archive bio-<br>potential signals. | Acquire, display, store, and archive<br>electroencephalographic signals from<br>the brain using a full montage array<br>and user-specified locations. | Acquire, display, store,<br>and archive<br>electroencephalographic<br>signals from the brain<br>using a full montage<br>array (i.e., 16 or more<br>electrodes) and user<br>specified locations. |
| Intended<br>Use<br>Environment | Research institutions, clinics,<br>hospitals, operating rooms,<br>epilepsy evaluation unit<br>environments or sleep laboratories. | Not stated | Medical facility,<br>physician's office,<br>laboratory, clinic or<br>nursing home or outside<br>of a medical facility<br>under supervision of a<br>medical professional |
| Target<br>Patient<br>Population | Adult and pediatrics | Not stated | The device will be<br>available on all patient<br>populations as<br>determined by a trained<br>professional |
| | Micromed BRAIN QUICK<br>(Subject Device) | Blackrock NeuroMed Cervello<br>Basic Bio-Potential Signal<br>Acquisition System<br>K122196 | Micromed<br>Morpheus, Brain<br>Spy Plus<br>K071782 |
| Indications<br>for Use | The Micromed BRAIN QUICK<br>system is intended to acquire and<br>store physiological signals for<br>EEG and/or polysomnography<br>(PSG) and to transfer data to<br>separate PSG analysis software.<br><br>The system includes:<br>-hardware recorders intended to<br>acquire and temporarily store<br>physiological signals for EEG<br>and/or PSG and to transfer the data<br>to separate polysomnographic<br>analysis software, and<br>-software which is used for the<br>signal acquisition, display, storage,<br>process and measure of biologic<br>signals.<br><br>The hardware recorders<br>(amplifiers) are BRAIN SPY<br>PLUS, MORPHEUS, SD LTM 32<br>EXPRESS and SD LTM 64<br>EXPRESS. They are intended to<br>acquire the same kind of signals<br>by using the SystemPlus<br>EVOLUTION software. The<br>recorder models differ based on<br>the number of available EEG<br>channels (16 in the BRAIN SPY<br>PLUS, 24 in the MORPHEUS, 32<br>in SD LTM 32 EXPRESS and 64<br>in SD LTM 64 EXPRESS). A<br>system with up to 256 channel can<br>be obtained by cascading 4 SD<br>LTM 64 EXPRESS amplifiers.<br><br>The device is intended to be used<br>by physicians, technicians, and<br>other medical professions that are<br>trained in EEG and/or PSG. | Up to 64 channels with one Cervello<br>hardware device (Amplifier) and up<br>to 128 channels by cascading 2<br>Cervello devices using the Cervello<br>software. The device is intended to<br>acquire and store physiological<br>signals for EEG and/or PSG, and to<br>transfer the data to separate<br>polysomnographic analysis software.<br><br>The devices are intended to be used<br>by physicians, technicians and other<br>medical professions that are trained<br>in EEG and/or PSG.<br><br>The Cervello Basic System does not<br>make any judgment of normality or<br>abnormality of the displayed signals<br>or the results of an analysis. In no<br>way are any of the functions<br>represented as being in and of<br>themselves diagnostic. | Acquire and store<br>physiological signals for<br>EEG and Sleep Studies,<br>and to transfer the data<br>to separate<br>polysomnographic<br>analysis software. The<br>devices are intended for<br>use by physicians,<br>technicians, or other<br>medical professionals<br>that are trained in EEG<br>and/ or PSG. |
| Use<br>limitations | The BRAIN QUICK system does<br>not make any judgment of<br>normality or abnormality of the<br>displayed signals or the result of<br>an analysis. In no way are any of<br>the functions represented as being<br>in themselves diagnostic.<br><br>Moreover, the system is not a<br>monitoring system: no physiologic<br>alarms are provided. | The System is not a monitoring<br>system. No physiologic alarms are<br>provided. The acquisition and display<br>of bio-potential signals is for the<br>interpretation and use of the<br>clinician. The devices do not make<br>any judgment of normality or<br>abnormality of the displayed signals. | These devices do not<br>provide alarms and is<br>not intended for use as<br>an automated apnea<br>monitor. The devices do<br>not make any judgment<br>of normality or<br>abnormality of the<br>displayed signals. |
| Bio-<br>Potential<br>Signals | Electroencephalography (EEG),<br>Video EEG, Respiration, Heart | Electroencephalography (EEG)<br>VideoEEG, Respiration, Heart Rate, | Electroencephalography<br>(EEG) |
| | Micromed BRAIN QUICK<br>(Subject Device) | Blackrock NeuroMed Cervello<br>Basic Bio-Potential Signal<br>Acquisition System<br>K122196 | Micromed<br>Morpheus, Brain<br>Spy Plus<br>K071782 |
| Recorded | Rate, SPO2 | SPO2 | Electrocardiography<br>(ECG)<br>Respiration, Heart Rate,<br>SPO2 |
| Potential<br>Clinical<br>Applications | Bio-potential signal acquisition:<br>amplification, recording,<br>digitization, display and storage. | Bio-potential signal amplification,<br>recording, display, digitization,<br>retrieval and display. | The device function is<br>the acquisition of<br>bioelectric signals, as is<br>typical for EEG<br>amplifiers and Holter<br>recorders. |
{5}------------------------------------------------
{6}------------------------------------------------
### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES
As shown in the following table, the BRAIN QUICK system shares similar technological characteristics with the predicates the main technological characteristics.
| | Micromed BRAIN QUICK<br>(Subject Device) | Blackrock NeuroMed<br>Cervello Basic Bio-Potential<br>Signal Acquisition System | Micromed Morpheus,<br>Brain Spy Plus |
|---------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| | | K122196 | K071782 |
| System<br>Components | Ambulatory and Stationary<br>Amplifiers,<br>Adapter Cables, Signal Processor,<br>Power Supply, (Interface to the<br>user's Monitor and PC) | Ambulatory and Stationary<br>Amplifiers,<br>Adapter Cables, Signal Processor,<br>Power Supply, (Interface to the<br>user's Monitor and Personal<br>Computer) | Ambulatory: Amplifier<br>Basic: Amplifier, Adapter<br>Cable,<br>Signal Processor, Power<br>Supply |
| Accessories | Ambulatory holster, data<br>acquisition unit mounting pole,<br>video camera mounting pole,<br>equipment cart. | Ambulatory holster, data<br>acquisition unit mounting pole,<br>equipment cart. | None stated |
| Number of<br>EEG Signal<br>Recording<br>Channels | System with SD LTM 64/32<br>EXPRESS: Up to 64 channels<br>with one device. Up to 256<br>channels by cascading four<br>devices<br>System with MORPHEUS/ | 64 channels with one device. 128<br>channels by cascading two<br>devices. | 24 channels |
| | BRAIN SPY PLUS: 24 | | |
| Amplifier<br>Input<br>Impedance | System with SD LTM 64/32<br>EXPRESS: >1012 Ω<br>System with MORPHEUS/<br>BRAIN SPY PLUS: >109 Ω (diff) | >109 Ω (diff) | >109 Ω (diff) |
| Amplifier<br>DC Signal<br>Range | System with SD LTM 64/32<br>EXPRESS:<br>1.48 /4.8V (0.46/4.69<br>mV/digit) System with<br>MORPHEUS/<br>BRAIN SPY PLUS: 1V<br>(15μV/digit) | 1V (15μV/digit) | 1V (15μV/digit) |
| Amplifier<br>Frequency | System with SD LTM 64/32 | 0.15 - 220 Hz | 0.15 - 220 Hz |
| | Micromed BRAIN QUICK<br>(Subject Device) | Blackrock NeuroMed<br>Cervello Basic Bio-Potential<br>Signal Acquisition System<br>K122196 | Micromed Morpheus,<br>Brain Spy Plus<br>K071782 |
| Response | System with MORPHEUS/<br>BRAIN SPY PLUS: 0.15 – 220<br>Hz | | |
| A/D<br>Conversion | System with SD LTM 64/32<br>EXPRESS: 16 bit for EEG<br>channels, 10 bit for DC channels | 16 bit | 16 bit |
| | System with MORPHEUS/<br>BRAIN SPY PLUS: 16 bit | | |
| Sampling<br>Rate | 8192 Hz per channel | 8192 Hz per channel | 8192 Hz per channel |
| CMRR | >100dB@50Hz between G1 and<br>all other inputs | >100dB@20Hz between G1 and<br>all other inputs | >100dB@20Hz between G1<br>and all other inputs |
| Noise | System with SD LTM 64/32<br>EXPRESS:<br><0.5µV r.m.s.@128Hz sampling<br>rate<br>System with MORPHEUS:<br><0.5µV r.m.s. @256Hz sampling<br>rate | <0.5µV r.m.s. @128Hz sampling<br>rate | <0.5µV r.m.s.@256Hz<br>sampling rate |
| Power<br>Source | Acquisition unit, 2x 1,5V DC AA,<br>alkaline batteries, computer (for<br>software), 120 VAC | Acquisition unit, 2x 1,5V DC AA,<br>alkaline batteries, computer (for<br>software), 120 VAC | 2x 1,5V DC AA, alkaline<br>batteries |
| Digital and<br>Serial Input<br>Channels | Yes | Yes | Yes |
| Measures | Impedance check<br>Amplitude | Impedance check<br>Amplitude | Impedance check |
| | Frequency spectrum | Frequency spectrum | N/A |
| | Time interval | Time interval | N/A |
| Extraction<br>of Spike and<br>Field<br>Potentials | No | No | N/A |
| Alarms | No | No | No |
| Stimulation | None | None | None |
| Video<br>camera | Available | Available | Available |
| Patient<br>Contact | None - The device interfaces to<br>user supplied bio-potential<br>electrodes. | None - The device interfaces to<br>user supplied bio-potential<br>electrodes. | None - The device interfaces<br>to user supplied bio-<br>potential electrodes. |
| Operating<br>System | Windows XP, 7, 8.1 | Windows XP, 7, Server 2008 R2,<br>Dual Core 2 GHz Processor | N/A |
| Device<br>Input | Keyboard and Mouse | Keyboard and Mouse | N/A |
| Display | Computer Screen &<br>Built-in LCD | Computer Screen | Built-in LCD |
| | Micromed BRAIN QUICK<br>(Subject Device) | Blackrock NeuroMed<br>Cervello Basic Bio-Potential<br>Signal Acquisition System<br>K122196 | Micromed Morpheus,<br>Brain Spy Plus<br>K071782 |
| Safety<br>Standards<br>Compliance | IEC 60601-1:2005 + /A1:2011 | IEC 60601-1:1998 | IEC 60601-1 (1988)+<br>A1:1991 +<br>A2:1995 (2nd edition) |
| | IEC 60601-1-2:2007 | IEC 60601-1-2:2007 | IEC 60601-1-1 2000-12 |
| | IEC 60601-2-26:2012 | IEC 60601-2-26: 2002 | IEC 60601-1-2:2001-09 |
| | | | IEC 60601-1-4:1996<br>+A1:1999 |
| | | | IEC 60601-2-26:2002-11 |
| | | | UL 2601-1 |
| | | | ETSI EN 301 489-1 |
| | | | ETSI EN 301 489-17 |
| | | | ETSI EN 300 328-1 |
EXPRESS: 0.15-500 Hz
{7}------------------------------------------------
{8}------------------------------------------------
#### Discussion
#### Comparison with the predicate devices
The Micromed BRAIN QUICK system is a composition system based on devices already on the market in the U.S.A. The BRAIN QUICK system is comprised of one or more amplifiers, software and some accessories needed for supporting (PC, monitor, keyboard, separation transformer) or complementing (cart, video camera) the functions of the devices.
No patient contacting parts are included.
The integration of the amplifier, the software and the supporting accessories devices has already been performed in a similar system sold by Blackrock Neuromed, that is, the predicate device Cervello Basic system.
#### Similarities
- . The amplifier used in the predicate system is identical to the SD LTM 64 EXPRESS amplifier used in the Micromed BRAIN QUICK system. It is a relabeling of the same device).
- . The MORPHEUS amplifier used in the Micromed BRAIN OUICK system is identical to the amplifier submitted by Micromed under K071782.
- The software used in the Micromed BRAIN QUICK system provides the user with the same functionalities as the software included in the predicate Cervello Basic system.
#### Differences
- . The Micromed BRAIN OUICK system differs from the Blackrock NeuroMed system in that the two systems use a different cart and a different isolation transformer.
{9}------------------------------------------------
- . The software used in the Micromed BRAIN QUICK system is an updated version of the one used in the predicate system. The updates to the software version do not impact the safety or effectiveness of the subject device.
### PERFORMANCE TESTING
Safety Tests have been performed to verify compliance with IEC 60601-1 and IEC 60601-2-26 to ensure that there are no potential hazards on patients, other persons, or the surroundings, and that basic safety and essential performance for EEG devices have been attained.
Electromagnetic Compatibility tests according to IEC 60601-1-2 have been performed to ensure no intolerable magnetic disturbances are introduced into its electromagnetic environment.
Immunity tests to IEC 60601-1-2 have been performed to ensure that the EEG equipment has the ability to operate satisfactorily in its electromagnetic environment
No specific guidance document on performance is required for EEG/PSG devices.
The device capability is equivalent to the features specified for predicate devices.
Verification of software was conducted according to Micromed design control procedures: it ensures that the system conforms to all the system design requirements
## CONCLUSION
The Micromed BRAIN QUICK System has the same manner of use with basic functions as the predicate device and the Micromed BRAIN OUICK system has a similar requirement for training and expectations of user. The safety requirements and expectations are the same. The systems have comparable performance in terms of data sampling and accuracy. They use identical patient connectors and methodology.
Based on the information provided in this submission, we conclude that the Micromed BRAIN OUICK system is substantially equivalent to the predicate and is safe and effective for its intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.