← Product Code [KGO](/productcode/KGO) · K171375

# Gammex PI Breach Detect Powder Free Surgical Glove (K171375)

_Ansell Healthcare Products, LLC · KGO · Feb 6, 2018 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K171375

## Device Facts

- **Applicant:** Ansell Healthcare Products, LLC
- **Product Code:** [KGO](/productcode/KGO.md)
- **Decision Date:** Feb 6, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4460
- **Device Class:** Class 1
- **Review Panel:** General, Plastic Surgery

## Indications for Use

Gammex PI Breach Detect Powder Free Surgical Gloves are intended to be worn by operating room personnel to protect a surgical wound from contamination.

## Device Story

Sterile, green, single-use, powder-free synthetic polyisoprene surgical glove; worn by operating room personnel. Features polyurethane inner coating for donning; outer surface treated with proprietary surface-active agent. Used as under-glove; surface-active agent and green color promote wetting and spreading of bodily fluids upon breach, improving visibility of glove failure to the wearer. Enhances detection of breaches during surgery; protects surgical wound from contamination.

## Clinical Evidence

No clinical data; bench testing only. Device performance validated against ASTM D3577-09 (dimensions, physical properties), ASTM D5151-06 (freedom from holes), and ASTM D6124-06 (powder residual). Biocompatibility testing performed per ISO 10993-5 (cytotoxicity), ISO 10993-10 (irritation/sensitization), and ISO 10993-11 (acute systemic toxicity).

## Technological Characteristics

Material: Synthetic polyisoprene rubber. Inner coating: Polyurethane polymer. Outer coating: Proprietary surface-active agent. Dimensions/Physical: Meets ASTM D3577-09. Sterilization: Radiation (ANSI/AAMI/ISO 11137-1:2006, 10^-6 SAL). Form factor: Hand-specific, beaded cuff, green color. Connectivity: None.

## Regulatory Identification

A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- DermaPrene® PI or IsoTouch Green Sterile Powder-Free Polyisoprene Surgical Gloves ([K071746](/device/K071746.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text. The word "Administration" is on the second line.

February 6, 2018

Ansell Healthcare Products LLC Don Cronk Sr Manager, QARA / Technical Services 111 Wood Avenue South, Suite 210 Iselin, New Jersey 08830

Re: K171375

Trade/Device Name: Gammex PI Breach Detect Powder Free Surgical Glove Regulation Number: 21 CFR 878.4460 Regulation Name: Surgeon's Glove Regulatory Class: Class I Product Code: KGO Dated: December 20, 2017 Received: December 27, 2017

Dear Don Cronk:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Michael J. Ryan -S

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K171375

Device Name:

Gammex PI Breach Detect Powder Free Surgical Glove

Indications for Use (Describe)

Gammex PI Breach Detect Powder Free Surgical Gloves are intended to be worn by operating room personnel to protect a surgical wound from contamination.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

x | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510k Summary

#### 510(k): K171375

#### Submitter

Ansell Healthcare Products LLC. 111 Wood Avenue South, Suite 210 Iselin, NJ 08830 USA

#### Contact Person:

Donald J. Cronk Senior Manager QA/RA/Technical Services Phone: (775) 470-7106 Email: don.cronk@ansell.com

#### Date Prepared

January 22, 2018

#### Name of Device

| Trade Names:               | Gammex PI Breach Detect Powder Free Surgical Glove |  |
|----------------------------|----------------------------------------------------|--|
| Common Name:               | Surgeon's Gloves                                   |  |
| Classification Name:       | Surgeon's Gloves                                   |  |
| Classification Regulation: | 21 CFR 878.4460                                    |  |
| Device Class:              | I                                                  |  |
| Product Code:              | KGO                                                |  |
| Classification Panel:      | General and Plastic Surgery                        |  |

#### Legally Marketed Predicate Device

K071746 – DermaPrene® Pl or IsoTouch Green Sterile Powder-Free Polyisoprene Surgical Gloves Surgeon's Gloves Common Name: Classification Name: Surgeon's Gloves Classification Regulation: 21 CFR 878.4460 Device Class: -Product Code: KGO Classification Panel: General and Plastic Surgery

#### Device Description

Gammex Pl Breach Detect Powder Free Surgical Glove is a sterile, green in color, single use disposable powder free surgical glove. The glove is made of synthetic polyisoprene rubber. A polymer is applied to the inner surface of the glove to make donning easy. The glove's outer surface is coated with a surface-active agent and when used as an under-glove, this coating and the glove's green color can improve visibility of a glove breach by promoting wetting of the glove surface and spreading of bodily fluid penetrating the breach.

#### Indication for Use Statement

Gammex PI Breach Detect Powder Free Surgical Gloves are intended to be worn by operating room personnel to protect a surgical wound from contamination.

{4}------------------------------------------------

#### Technological Characteristics

Gammex Pl Breach Detect Powder Free Surgical Gloves have the following technological characteristics as compared to ASTM or equivalent standards:

| Characteristics                         | Standard/Test/FDA Guidance     | Result Summary                                                                                                                                                                           |
|-----------------------------------------|--------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Physical Characteristics:               |                                |                                                                                                                                                                                          |
| Dimensions                              | ASTM D3577-09                  | Meets ASTM D3577-09 requirements for length, width and thickness                                                                                                                         |
| Length                                  | Minimum 265mm                  | Average value:<br>Minimum 297 mm                                                                                                                                                         |
| Palm Width (mm)                         |                                | Average value in mm:                                                                                                                                                                     |
| Size 5.5                                | $70 \pm 6$                     | 76                                                                                                                                                                                       |
| Size 6.0                                | $76 \pm 6$                     | 82                                                                                                                                                                                       |
| Size 6.5                                | $83 \pm 6$                     | 86                                                                                                                                                                                       |
| Size 7.0                                | $89 \pm 6$                     | 91                                                                                                                                                                                       |
| Size 7.5                                | $95 \pm 6$                     | 98                                                                                                                                                                                       |
| Size 8.0                                | $102 \pm 6$                    | 100                                                                                                                                                                                      |
| Size 8.5                                | $108 \pm 6$                    | 112                                                                                                                                                                                      |
| Size 9.0                                | $114 \pm 6$                    | 120                                                                                                                                                                                      |
| Thickness (mm) – single wall            |                                | Average value in mm                                                                                                                                                                      |
| Finger                                  | Minimum 0.10                   | 0.25                                                                                                                                                                                     |
| Palm                                    | Minimum 0.10                   | 0.21                                                                                                                                                                                     |
| Cuff                                    | Minimum 0.10                   | 0.17                                                                                                                                                                                     |
| Physical Properties:                    | ASTM D3577-09                  | Meets ASTM D3577-09 requirements<br>for tensile strength, ultimate<br>elongation and stress at 500%<br>elongation before and after<br>accelerated aging for synthetic<br>surgical gloves |
| Freedom from holes                      | ASTM D3577-09<br>ASTM D5151-06 | Meets ASTM D3577-09 and ASTM<br>D5151-06 requirements of AQL 1.5                                                                                                                         |
| Powder Residual                         | ASTM D3577-09<br>ASTM D6124-06 | Meets applicable requirement for<br>powder free; ≤ 2 mg per glove                                                                                                                        |
| Sterility                               | ANSI/AAMI/ISO 11137-1:2006     | Meets ANSI/AAMI/ISO 11137-1:2006<br>requirement of $10^{-6}$ SAL.                                                                                                                        |
| Biocompatibility                        |                                |                                                                                                                                                                                          |
| ISO In Vitro Cytotoxicity               | ISO 10993-5:2009               | Under the conditions of the study,<br>undiluted and 1:2 dilution were<br>cytotoxic. 1:4, 1:8, 1:16, 1:32 and 1:64<br>are not cytotoxic                                                   |
| ISO Skin Irritation Study               | ISO10993-10:2010               | Under the conditions of the study,<br>not an irritant                                                                                                                                    |
| ISO Maximization<br>Sensitization Study | ISO 10993-10:2010              | Under the conditions of the study,<br>not a sensitizer                                                                                                                                   |
| ISO acute systemic<br>toxicity          | ISO 10993-11: 2006             | Under the conditions of the study, no<br>evidence of systemic toxicity                                                                                                                   |

{5}------------------------------------------------

#### Substantial Equivalence

|                                            | Predicate Device                                                                                                         | Subject Device                                                                                                                                                                                                                                                                              | Substantial<br>Equivalence |
|--------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|
| Trade Name                                 | DermaPrene® PI or IsoTouch<br>Green Sterile Powder-Free<br>Polyisoprene Surgical Gloves                                  | Gammex® PI Breach Detect<br>Powder Free Surgical Glove                                                                                                                                                                                                                                      | Not applicable             |
| 510(k) Number                              | K071746                                                                                                                  | K171375                                                                                                                                                                                                                                                                                     | Not applicable             |
| Submitter                                  | Ansell Healthcare Products LLC                                                                                           | Ansell Healthcare Products LLC                                                                                                                                                                                                                                                              | Same                       |
| Product Code                               | KGO                                                                                                                      | KGO                                                                                                                                                                                                                                                                                         | Same                       |
| Regulation<br>Number                       | 21 CFR 878.4460                                                                                                          | 21 CFR 878.4460                                                                                                                                                                                                                                                                             | Same                       |
| Regulation<br>Name                         | Surgeon's Glove                                                                                                          | Surgeon's Glove                                                                                                                                                                                                                                                                             | Same                       |
| Regulatory Class                           | Class I                                                                                                                  | Class I                                                                                                                                                                                                                                                                                     | Same                       |
| Device Type                                | Type 2<br>Gloves compounded primarily<br>from rubber cement or synthetic<br>rubber latex                                 | Type 2<br>Gloves compounded primarily<br>from rubber cement or<br>synthetic rubber latex                                                                                                                                                                                                    | Same                       |
| Indications for<br>use                     | These gloves are intended to be<br>worn by operating room<br>personnel to protect a surgical<br>wound from contamination | These gloves are intended to<br>be worn by operating room<br>personnel to protect a surgical<br>wound from contamination.<br>When used as an underglove,<br>the glove is intended to<br>promote visibility of a fluid that<br>breaches/enters the glove via<br>an increased spreading rate. | Different                  |
| Prescription or<br>Over-The<br>Counter-Use | Over-The-Counter-Use                                                                                                     | Over-The-Counter-Use                                                                                                                                                                                                                                                                        | Same                       |
| Material                                   | Synthetic Polyisoprene Rubber                                                                                            | Synthetic Polyisoprene Rubber                                                                                                                                                                                                                                                               | Same                       |
| Coating (inner)                            | Polyurethane polymer inner<br>coating to aid donning                                                                     | Polyurethane polymer inner<br>coating to aid donning                                                                                                                                                                                                                                        | Same                       |
| Coating (outer)                            | None                                                                                                                     | Proprietary solution to aid in<br>wetting/spreading of fluid<br>when used as an underglove                                                                                                                                                                                                  | Different                  |
|                                            |                                                                                                                          | and the outer surgeon's glove<br>is breached.                                                                                                                                                                                                                                               |                            |
| Design                                     | Single use                                                                                                               | Single use                                                                                                                                                                                                                                                                                  | Same                       |
|                                            | Sterile                                                                                                                  | Sterile                                                                                                                                                                                                                                                                                     | Same                       |
|                                            | Powder-Free                                                                                                              | Powder-Free                                                                                                                                                                                                                                                                                 | Same                       |
|                                            | Hand-specific                                                                                                            | Hand-specific                                                                                                                                                                                                                                                                               | Same                       |
|                                            | Beaded cuff                                                                                                              | Beaded cuff                                                                                                                                                                                                                                                                                 | Same                       |
| Color                                      | Green                                                                                                                    | Green                                                                                                                                                                                                                                                                                       | Same                       |
| Dimension &<br>physical<br>properties      | Meets ASTM D3577-09<br>requirements                                                                                      | Meets ASTM D3577-09<br>requirements                                                                                                                                                                                                                                                         | Same                       |
| Freedom from<br>holes                      | Meets ASTM D3577-09<br>requirements of AQL 1.5                                                                           | Meets ASTM D3577-09<br>requirements of AQL 1.5                                                                                                                                                                                                                                              | Same                       |
| Powder-Free                                | Meets applicable definition for<br>powder free; ≤ 2mg per glove                                                          | Meets applicable definition for<br>powder free; ≤ 2mg per glove                                                                                                                                                                                                                             | Same                       |
| Sterilization<br>method                    | Radiation                                                                                                                | Radiation                                                                                                                                                                                                                                                                                   | Same                       |
| Sterility<br>Assurance Level<br>(SAL)      | 10-6 SAL                                                                                                                 | 10-6 SAL                                                                                                                                                                                                                                                                                    | Same                       |
| Shelf life                                 | 3 years                                                                                                                  | 3 years                                                                                                                                                                                                                                                                                     | Same                       |
| Biocompatibility                           | "Under the conditions of the<br>study, not an irritant" and<br>"Under the conditions of the<br>study, not a sensitizer"  | "Under the conditions of the<br>study, not an irritant" and<br>"Under the conditions of the<br>study, not a sensitizer"                                                                                                                                                                     | Same                       |

#### Substantial Equivalence Comparison Table

{6}------------------------------------------------

The subject device is manufactured from polyisoprene with polyurethane polymer inner coating to aid donning. The glove's outer surface is coated with a surface-active agent and when used as an under-glove, this coating and the glove's green color can improve visibility of a glove breach by promoting wetting of the glove surface and spreading of bodily fluid penetrating the breach. Aside from the outer surface treatment, the subject device's materials are functionally equivalent to those of the cited predicate.

The subject device meets the applicable requirements for surgeon's gloves regarding dimensions and sizes, physical properties, freedom from holes, and powder residues as found in the following standards: ASTM D3577, ASTM D5151 and ASTM D6124. The subject device passes biological reactivity testing for dermal sensitization and irritation, in accord with the ISO 10993 series of standards.

#### Performance Data

A clinical study was not conducted on the subject or predicate devices.

#### Substantial Equivalence Statement

{7}------------------------------------------------

The Gammex PI Breach Detect Powder Free Surgeon's Glove are as safe and effective as the predicate device with respect to design, technological characteristics, intended use and conformance to standard requirements.

#### Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the subject device is as safe, as effective, and performs as well as the legally marketed device, K071746

---

**Source:** [https://fda-staging.innolitics.com/device/K171375](https://fda-staging.innolitics.com/device/K171375)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).