K171365 · Arthrex, Inc. · MBH · Nov 7, 2017 · Orthopedic
Device Facts
Record ID
K171365
Device Name
Arthrex Knee Systems
Applicant
Arthrex, Inc.
Product Code
MBH · Orthopedic
Decision Date
Nov 7, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3565
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Arthrex iBalance BiCompartmental Arthroplasty System is intended to be used as a multicompartmental knee arthroplasty in patients with: - Moderately disabling joint disease of the knee resulting from painful osteoarthritis or post traumatic arthritis; - Correction of functional deformities; - Revision of previous unsuccessful partial knee replacement or other procedure; The BiCompartmental Arthroplasty System is not intended to be used as a dual-condyle or tri-compartmental knee. These components are single use only and are intended for implantation with bone cement. The Arthrex iBalance PFJ components are intended for use in patellofemoral knee arthroplasty in patients with: - Degenerative arthritis in the distal femur and patella; - A history of patellar dislocation or fracture; - Failed previous surgery (arthroscopy, tibial tubercle elevation, lateral release) where pain, deformity or dysfunction persists. These components are single use only and are intended for implantation with bone cement. The Arthrex iBalance TKA System is indicated for use in individuals undergoing surgery for: - Painful, disabling joint disease of the knee resulting from degenerative arthritis, rheumatic arthritis: - Post-traumatic loss of knee joint configuration and function - Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability; - Revisions of previous unsuccessful knee replacement or other procedure. Additional indications for posteriorly stabilized components: - Ligamentous instability requiring implant bearing surfaces with increased constraint; - Absent or non-functioning posterior cruciate ligament. These devices are single use only and are intended for implantation with the exception of porous coated femoral components which can be used cemented (biological fixation). Arthrex iBalance UKA System is intended for use in unicompartmental knee arthroplasty as a result of: - Moderately disabling joint disease of the knee resulting from painful osteoarthritis or post traumatic arthritis; - Correction of functional deformities; - Revision of previous unsuccessful unicompartmental knee replacement or other procedure; - As an alternative to tibial osteotomy in patients with unicompartmental osteoarthritis. These components are single use only and are intended for implantation with bone cement.
Device Story
Arthrex Knee Systems (TKA, UKA, PFJ, BiCompartmental) are orthopedic implants consisting of femoral, tibial, and patellar components. Used by orthopedic surgeons in clinical settings to replace damaged knee joint surfaces. Implants are fixed via bone cement or biological fixation (porous-coated femoral components). This submission updates labeling to include 'MR Conditional' status per FDA guidance. Device provides structural replacement for diseased or deformed knee joints, aiming to restore function and reduce pain. No software or algorithmic processing involved.
Clinical Evidence
No clinical data provided. Evidence consists of non-clinical bench testing and in-vivo electromagnetic simulation to establish 'MR Conditional' status per FDA guidance, and bacterial endotoxin testing per EP 2.6.14/USP <85>.
Technological Characteristics
Passive orthopedic implants (femoral, tibial, patellar components). Materials include metal and polymer. Fixation via bone cement or biological fixation (porous-coated). MR Conditional status established via electromagnetic simulation. Single-use.
Indications for Use
Indicated for patients with painful, disabling knee joint disease (osteoarthritis, rheumatoid arthritis, post-traumatic arthritis), functional deformities, history of patellar dislocation/fracture, or failed prior knee surgeries requiring revision. Specific indications vary by system (TKA, UKA, PFJ, BiCompartmental) regarding ligamentous stability and compartment involvement.
Regulatory Classification
Identification
A knee joint patellofemorotibial metal/polymer porous-coated uncemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device is designed to achieve biological fixation to bone without the use of bone cement. This identification includes fixed-bearing knee prostheses where the ultra high molecular weight polyethylene tibial bearing is rigidly secured to the metal tibial base plate.
Special Controls
*Classification.* Class II (special controls). The special control is FDA's guidance: “Class II Special Controls Guidance Document: Knee Joint Patellofemorotibial and Femorotibial Metal/Polymer Porous-Coated Uncemented Prostheses; Guidance for Industry and FDA.” See § 888.1 for the availability of this guidance.
Predicate Devices
Arthrex iBalance UKA System Vitamin E Tibial Bearing (K161060)
Arthrex iBalance BiCompartmental Arthroplasty System (K160461)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration. The Department of Health & Human Services logo is on the left, and the U.S. Food & Drug Administration logo is on the right. The FDA logo is blue and white and includes the words "FDA U.S. FOOD & DRUG ADMINISTRATION".
November 7, 2017
Arthrex Inc. David L. Rogers Project Manager, Regulatory Affairs 1370 Creekside Boulevard Naples, Florida 34108-1945
Re: K171365
Trade/Device Name: Arthrex Knee Systems Regulation Number: 21 CFR 888.3565 Regulation Name: Knee joint patellofemorotibial metal/polymer porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: MBH, HSX, KRR Dated: October 3, 2017 Received: October 11, 2017
Dear Mr. Rogers:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
### Page 2 - Mr. David Rogers
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known)
K171365
Device Name
Arthrex iBalance BiCompartmental Arthoplasty System
Indications for Use (Describe)
When used concurrently, the Arthrex iBalance UKA and PFJ systems create the Arthrex iBalance BiCompartmental Arthroplasty System. The Arthrex iBalance BiCompartmental Arthroplasty System is intended to be used as a multicompartmental knee arthroplasty in patients with:
- Moderately disabling joint disease of the knee resulting from painful osteoarthritis or post traumatic arthritis;
- · Correction of functional deformities;
- · Revision of previous unsuccessful partial knee replacement or other procedure;
The BiCompartmental Arthroplasty System is not intended to be used as a dual-condyle or tri-compartmental knee.
These components are single use only and are intended for implantation with bone cement.
Type of Use (Select one or both, as applicable)
| <span style="font-family: sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
|----------------------------------------------------------------------------------------------|-----------------------------------------------|
|----------------------------------------------------------------------------------------------|-----------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known)
K171365
Device Name Arthrex iBalance PFJ
### Indications for Use (Describe)
The Arthrex iBalance PFJ components are intended for use in patellofemoral knee arthroplasty in patients with:
- · Degenerative arthritis in the distal femur and patella;
- · A history of patellar dislocation or fracture;
- Failed previous surgery (arthroscopy, tibial tubercle elevation, lateral release) where pain, deformity or dysfunction persists.
These components are single use only and are intended for implantation with bone cement.
Type of Use (*Select one or both, as applicable*)
| <span></span> | Prescription Use (Part 21 CFR 801 Subpart D) |
|---------------|----------------------------------------------|
| <span></span> | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{4}------------------------------------------------
### Indications for Use
510(k) Number (if known)
#### K171365
Device Name Arthrex iBalance TKA System
Indications for Use (Describe)
The Arthrex iBalance TKA System is indicated for use in individuals undergoing surgery for:
· Painful, disabling joint disease of the knee resulting from degenerative arthritis, rheumatic arthritis:
- · Post-traumatic loss of knee joint configuration and function
- · Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability;
- · Revisions of previous unsuccessful knee replacement or other procedure.
Additional indications for posteriorly stabilized components:
- · Ligamentous instability requiring implant bearing surfaces with increased constraint;
- · Absent or non-functioning posterior cruciate ligament.
These devices are single use only and are intended for implantation with the exception of porous coated femoral components which can be used cemented (biological fixation).
Type of Use (Select one or both, as applicable)
| <span> <b>X</b> Prescription Use (Part 21 CER 801 Subpart D) </span> |
|------------------------------------------------------------------------|
| <span>Over-The-Counter Use (21 CER 801 Subpart C)</span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{5}------------------------------------------------
### Indications for Use
510(k) Number (if known) K171365
Device Name Arthrex iBalance UKA System
#### Indications for Use (Describe)
Arthrex iBalance UKA System is intended for use in unicompartmental knee arthroplasty as a result of.
- Moderately disabling joint disease of the knee resulting from painful osteoarthritis or post traumatic arthritis;
- · Correction of functional deformities;
- · Revision of previous unsuccessful unicompartmental knee replacement or other procedure;
- · As an alternative to tibial osteotomy in patients with unicompartmental osteoarthritis.
These components are single use only and are intended for implantation with bone cement.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------|-----------------------------------------------|
| <span style="font-family: Arial;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{6}------------------------------------------------
## 510(k) Summary or 510(k) Statement
| Date Prepared | May 5, 2017 |
|--------------------------|-------------------------------------------------------------------------------------|
| Submitter | Arthrex Inc. |
| | 1370 Creekside Boulevard |
| | Naples, FL 34108-1945 |
| Contact Person | David L Rogers |
| | Project Manager, Regulatory Affairs |
| | 1-239-643-5553, ext. 71924 |
| | david.rogers@arthrex.com |
| Name of Device | Arthrex Knee Systems |
| Common Name | Knee Prosthesis |
| Product Code | MBH, HSX, KRR |
| Classification Name | 21 CFR 888.3565: Knee joint patellofemorotibial metal/polymer porous-coated |
| | uncemented prosthesis |
| | 21 CFR 888.3560: Prosthesis, Knee, Patellofermorotibial, Semi-Constrained, |
| | Cemented, Polymer/Metal/Polymer |
| | 21 CFR 888.3520: Prosthesis, Knee, Femorotibial, Non-Constrained, |
| | Metal/Polymer |
| Regulatory Class | II |
| Predicate Device | K161060: Arthrex iBalance UKA System Vitamin E Tibial Bearing |
| | K160461: Arthrex iBalance BiCompartmental Arthroplasty System |
| | K153586: Arthrex iBalance TKA System |
| | K152382: Arthrex iBalance TKA System |
| | K152252: Arthrex iBalance TKA System |
| | K143047: Arthrex iBalance Patella Implant, Dome |
| | K133854: Arthrex iBalance TKA System |
| | K073120: ACCIN Patellafemoral System |
| | K063782: ACCIN UNI-Knee System |
| | K060670: ACCIN UNI-Knee System |
| Purpose of<br>Submission | This Special 510(k) premarket notification is submitted to obtain clearance for the |
| | modification to the device labeling for the Arthrex Knee Systems to include the |
| | "MR Conditional" statement in accordance with the FDA Guidance, "Establishing |
| | Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) |
| | Environment". |
| Device Description | The Arthrex Knee Systems encompass the following FDA cleared devices: |
| | ● Arthrex iBalance TKA System |
| | ● Arthrex iBalance UKA System |
| | ● Arthrex iBalance PFJ |
| | ● Arthrex iBalance BiCompartmental Arthroplasty System |
| | These systems are comprised of femoral, tibia, and patellar components for use |
| | in knee prosthesis. |
| Indications for Use | The Arthrex iBalance TKA System is indicated for use in individuals undergoing |
| | surgery for: |
| | ● Painful, disabling joint disease of the knee resulting from degenerative |
| | arthritis, rheumatoid arthritis or post-traumatic arthritis; |
| | ● Post-traumatic loss of knee joint configuration and function |
| | ● Moderate varus, valgus, or flexion deformity in which the ligamentous |
| | structures can be returned to adequate function and stability; |
| | |
{7}------------------------------------------------
- Revisions of previous unsuccessful knee replacement or other procedure.
Additional indications for posteriorly stabilized components:
- Ligamentous instability requiring implant bearing surfaces with ● increased constraint;
- Absent or non-functioning posterior cruciate ligament.
These devices are single use only and are intended for implantation with bone cement, with the exception of porous coated femoral components which can be used cemented or uncemented (biological fixation).
Arthrex iBalance UKA System is intended for use in unicompartmental knee arthroplasty as a result of:
- Moderately disabling joint disease of the knee resulting from painful osteoarthritis or post traumatic arthritis;
- Correction of functional deformities;
- Revision of previous unsuccessful unicompartmental knee replacement ● or other procedure;
- . As an alternative to tibial osteotomy in patients with unicompartmental osteoarthritis.
These components are single use only and are intended for implantation with bone cement.
The Arthrex iBalance PFJ components are intended for use in patellofemoral knee arthroplasty in patients with:
- Degenerative arthritis in the distal femur and patella;
- A history of patellar dislocation or fracture;
- Failed previous surgery (arthroscopy, tibial tubercle elevation, lateral release) where pain, deformity or dysfunction persists.
These components are single use only and are intended for implantation with bone cement.
When used concurrently, the Arthrex iBalance UKA and PFJ systems create the Arthrex iBalance BiCompartmental Arthroplasty System. The Arthrex iBalance BiCompartmental Arthroplasty System is intended to be used as a multicompartmental knee arthroplasty in patients with:
- Moderately disabling joint disease of the knee resulting from painful ● osteoarthritis or post traumatic arthritis;
- Correction of functional deformities;
- Revision of previous unsuccessful partial knee replacement or other procedure;
The BiCompartmental Arthroplasty System is not intended to be used as a dualcondyle or tri-compartmental knee.
These components are single use only and are intended for implantation with bone cement.
Performance Data Non-clinical testing and in-vivo electromagnetic simulation per the FDA Guidance, "Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment", demonstrated that the Arthrex Knee Systems are "MR Conditional".
{8}------------------------------------------------
| | Bacterial endotoxin per EP 2.6.14/USP <85> was conducted to demonstrate that<br>the device meets pyrogen limit specifications. |
|------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Conclusion | The Arthrex Knee Systems are substantially equivalent to the predicate device in<br>which the basic design features and intended uses are the same. Any differences<br>between the proposed device and the predicate device are considered minor and<br>do not raise questions concerning safety or effectiveness. |
| | Based on the indications for use, technological characteristics, and the summary<br>of data submitted, Arthrex Inc. has determined that the proposed device is<br>substantially equivalent to the currently marketed predicate device. |
| | |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.