K171360 · Contec Medical Systems Co.,Ltd · DPS · Jan 22, 2018 · Cardiovascular
Device Facts
Record ID
K171360
Device Name
CONTEC™ Electrocardiograph
Applicant
Contec Medical Systems Co.,Ltd
Product Code
DPS · Cardiovascular
Decision Date
Jan 22, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2340
Device Class
Class 2
Attributes
Pediatric
Indications for Use
CONTEC™ Electrocardiograph, ECG90A, is intended to acquire ECG signals from adult or children patients through body surface ECG electrodes. The obtained ECG records can help users to analyze and diagnose heart disease. The proposed devices also have measurements and diagnostic interpretation functions, which are offered to clinician on an advisory basis. Digital Electrocardiographs shall be used in healthcare facilities by doctors and/or trained healthcare professionals.
Device Story
Device acquires 12-lead ECG signals via body surface electrodes; processes signals using internal MCU; provides real-time waveform display on LCD and thermal printing. Features include manual/auto recording, automatic parameter measurement, and diagnostic interpretation offered on advisory basis. Used in hospitals, wards, and ECG rooms by physicians or trained professionals. Power provided by internal lithium battery or AC-DC adapter. Signal path utilizes floating and optical isolation to minimize interference. Output assists clinicians in pattern and rhythm analysis for clinical diagnosis. Benefits include automated data collection and diagnostic support for heart disease management.
Clinical Evidence
No clinical study is included in this submission. Bench testing only.
Indicated for adult and pediatric patients requiring ECG signal acquisition for heart disease analysis and diagnosis. Used in healthcare facilities by doctors or trained healthcare professionals.
Regulatory Classification
Identification
An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION".
January 22, 2018
Contec Medical Systems Co., Ltd. % Ray Wang General Manager Beijing Believe Technology Service Co., Ltd. 5-402, Building #27, YangGuangYiShang, No.56, LiangXiang East Rd. FangShan District, Beijing, 102401 China
Re: K171360
Trade/Device Name: CONTEC™ Electrocardiograph Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II Product Code: DPS Dated: December 20, 2017 Received: December 22, 2017
Dear Ray Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
k171360
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
M.A. Zillehemen
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K171360
Device Name CONTECTM Electrocardiograph
## Indications for Use (Describe)
CONTECTM Electrocardiograph, ECG90A, is intended to acquire ECG signals from adult or children patients through body surface ECG electrodes. The obtained ECG records can help users to analyze and diagnose heart disease. The proposed devices also have measurements and diagnostic interpretation functions, which are offered to clinician on an advisory basis. Digital Electrocardiographs shall be used in healthcare facilities by doctors and or trained healtheare professionals.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------------------------------------------|--|
| <span style="font-size:10pt;"> <span style="font-family:Wingdings;">✔</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | |
| <span style="font-size:10pt;"> <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> | |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Tab #7 510(k) Summary
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.
The assigned 510(k) Number: _________
- 1. Date of Preparation: 05/05/2017
- 2. Sponsor Identification
Contec Medical System Co., Ltd. No.112 Qinhuang West Street, Economic & Technical Development Zone, Qinhuangdao, Hebei, 066004, China. Establishment Registration Number: 3006979678
Contact Person: Xueyong Li Position: Quality Manager Tel: 86-355-8015490 Fax: 86-355-8015490 Email:_lxy1011@163.com
- 3. Designated Submission Correspondent
Mr. Ray Wang
## Beijing Believe Technology Service Co., Ltd.
5-402, Building #27, YangGuangYiShang, No. 56, LiangXiang East Rd., FangShan District, BeiJing, China 102401
Tel: +86-18910677558 Fax: +86-10-52214696 Email: ray.wang@believe-med.com
{4}------------------------------------------------
#### 510(k) Summary
4. Identification of Proposed Device
> Trade Name: CONTECTMElectrocardiograph Common Name: Electrocardiograph Model(s): ECG90A
Regulatory Information Classification Name: Electrocardiograph Classification:II Product Code:DPS Regulation Number: CFR 870.2340 Review Panel:Cardiovascular;
Intended Use Statement:
CONTEC™ Electrocardiograph, ECG90A, is intended to acquire ECG signals from adult or children patients through body surface ECG electrodes. The obtained ECG records can help users to analyze and diagnose heart disease. The proposed devices also have measurements and diagnostic interpretation functions, which are offered to clinician on an advisory basis. Digital Electrocardiographs shall be used in healthcare facilities by doctors and/or trained healthcare professionals.
## Device Description
This product is an electrocardiograph collecting 12-lead ECG signal simultaneously and printing ECG waveform with thermal printing system, which features in, recording and displaying ECG waveform with manual or auto mode, measuring and diagnosing ECG waveform parameters automatically, prompting for "Lead off" and "Lack of paper", optional interface languages(Chinese/ English), case database management.
This product is used to assist doctors to analyze and print the ECG waveform of children and adults, which can be applied to ECG room, ward, hospital, etc.
The ECG machine provides medical department human ECG wave group for pattern and rhythm analysis that applied in clinical diagnosis and research.
Essential Performance:Sync collection and display of 12-lead ECG.Display of HR.Print ECG wave with different format,Filter settings,and Auto-measurement of ECG data.
Fig 7-1 Working Frame of ECG90A
{5}------------------------------------------------
Image /page/5/Figure/2 description: The image shows a block diagram of a system. The diagram includes the following components: button system, printing system, control system, power supply system, signal collection system, and LCD display system. The control system is in the center of the diagram, and the other components are connected to it via arrows.
ECG90A is divided into button system, printing system, power supply system, LCD display system, signal collection system, control system
Button system: to collect the user's key operation input, send the control system to process,
Printing system: print the print data sent by the control system, heat the printer head to record the waveform on the thermal printing paper, and feedback whether the lack of paper;
Power system: provide power supply for each system module;
The proposed device adopts two kinds of power supply modes: build-in lithium battery DC power supply and AC-DC adapter power supply.
AC-DC adapter power supply: after the external power adapter connecting with the proposed device, the AC voltage is reduced to 12V via the AC-DC power adapter. The 12V voltage is used to charge the lithium battery via the charging management chip;meanwhile the 12V voltage is reduced to 8.6V via a voltage converter, which is used to power the following circuit and printer. Then the 8.6V voltage is reduced again to 5.0V voltage via the voltage converter, and the 5.0V voltage pass through power isolation and is used to power the signal collection and processing module; meanwhile the 5V voltage will be reduces to 3.3 Vvoltage, which is used to power the control module. Build-in lithium battery DC power supply: when the proposed device is connected with AC-DC adapter power supply, the build-in lithium battery will be charged automatically. The charged battery will power the whole system as DC power supply when the proposed device is not connected with the external AC-DC adapter power supply
LCD display system: display various of settings of waveform information to the user for viewing :
Signal collection system: collect the waveform data,submit the final data to the control system to process; the signal acquisition and process module adopts floating and optical isolation to reduce the external disturbance to signal. After filtering and amplification, the A/C conversion of acquired ECG signal is completed on MCU (Micro Controller Unit); and then the data is transmitted to control system.
{6}------------------------------------------------
# K171360
#### 510(k) Summary
after processing the data will be displayed on the LCD screen and the waveform will be printed on the thermal paper. MCU can identify and response the key operation through connecting line, and MCU can connect and communicate with the SD card and USB through the signal line.
Control system: the MCU is used to control the whole system operation. The device parameters setting, the processing of acquired signal, the printing, screen display and charging are all controlled by this module.
- న్. Identification of Predicate Device(s)
Predicate Device :
510(k) Number: K131900 Product Name: CONTECTM Electrocardiograph Model Name: ECG300G Manufacturer: Contec Medical System Co., Ltd.
- Non-Clinical Test Conclusion 6.
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- IEC 60601-1:2005+A1:2012, Medical Electrical Equipment- Part 1: General requirements for basic a. safety and essential performance
- b. IEC 60601-1-2:2007, Medical Electrical Equipment-Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility-Requirements And Tests
- IEC 60601-2-25:2011 Medical Electrical Equipment Part 2-25: Particular Requirements For The C. Basic Safety And Essential Performance Of Electrocardiographs. (Cardiovascular)
- ISO 10993-5:2009. Biological Evaluation of Medical Device. Part 5-Tests for Vitro cytotoxicity. d.
- ISO 10993-10:2010 Standard, Biological Evaluation of Medical Device, Part 10-Test for irritation e. and delay-type hypersensitivity
- 7. Clinical Test Conclusion
No clinical study is included in this submission.
{7}------------------------------------------------
## 8. Substantially Equivalent (SE) Comparison
| Item | Proposed Device(s) | Predicate Device(s) | |
|------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------|
| Device name | ECG90A CONTEC™ Electrocardiograph | ECG300G CONTEC™ Electrocardiograph | |
| Classification Name | Electrocardiograph | Electrocardiograph | |
| Product Code | DPS | DPS | |
| Regulation Number | CFR 870.2340 | CFR 870.2340 | |
| Comparison Statement | The proposed device has same classification information as the predicate device. | | |
| Intended Use | CONTEC™ Electrocardiographs, ECG90A, are intended to acquire<br>ECG signals from adult or children patients through body surface ECG<br>electrodes. The obtained ECG records can help users to analyze and<br>diagnose heart disease. The proposed devices also have measurements<br>and diagnostic interpretation functions, which are<br>offered to clinician on an advisory basis. Digital Electrocardiographs<br>shall be used in healthcare facilities by doctors and/or trained<br>healthcare professionals. | CONTEC™ Electrocardiographs, ECG300G, are intended to acquire<br>ECG signals from adult patients through body surface ECG<br>electrodes. The obtained ECG records can help users to analyze and<br>diagnose heart disease. The proposed devices also have measurements<br>and diagnostic interpretation functions, which are offered to clinician<br>on an advisory basis. Digital Electrocardiographs shall be used in<br>healthcare facilities by doctors and/or trained healthcare<br>professionals. | |
| Comparison Statement | The proposed device has similar intended use as the predicate device. | | |
| Main Unit Technical Specifications | | | |
| Channel | 3 channel | 3 channel | |
| collection mode | Simultaneous 12-lead collection | Simultaneous 12-lead collection | |
| Recording mode | Automatic /Manual/rhythm | Automatic /Manual/rhythm | |
| Patient leak current | <10 µA | <10 µA | |
| Frequency response | 0.5 Hz~150 Hz (-3 dB~0.4 dB) | 0.5 Hz~150 Hz (-3 dB~0.4 dB) | |
| Noise level | <15 µVp-p | <15 µVp-p | |
| CMRR | | >60 dB, >100 dB(filter) | >60 dB, >100 dB(filter) |
| Input way | | Floating and defibrillation protection | Floating and defibrillation protection |
| Measurement Function | | Have measurement function | Have measurement function |
| Input CIR current | | ≤50nA | ≤50nA |
| Input impedance | | ≥50 ΜΩ | ≥50 ΜΩ |
| Paper | Auto record | 6.25 mm/s、12.5 mm/s、25 mm/s、50 mm/s, error: ±5 % | 25mm/s, 50mm/s, error: ±5% |
| | Manual record | 6.25 mm/s、12.5 mm/s、25 mm/s、50 mm/s, error: ±5 % | 5mm/s,6.25mm/s,10mm/s,12.5mm/s,25mm/s,50mm/s,error:±5% |
| | Rhythm record | 6.25 mm/s、12.5 mm/s、25 mm/s、50 mm/s, error: ±5 % | 25mm/s, 50mm/s, error: ±5% |
| EMG interference filter | | 25 Hz or 35 Hz (-3 dB) | 25Hz/35Hz (-3dB) |
| | Recording way | Thermal printing system | Thermal printing system |
| | Specification of<br>recording paper | 50 mm (W)×20 m(L) high-speed thermal paper | 80mm (W)*20m (L) High-speed thermal paper |
| Sensitivity selections | | 5/10/20 mm/mV, error: ±5 %, Standard sensitivity is<br>10 mm/mV±0.2 mm/mV | 2.5/5/10/20/40mm/mV,error:±5%.Standard sensitivity is<br>10mm/mV±0.2mm/mV |
| Net weight | | 0.5 Kg | 1.7kg |
| Power supply | | 12 V adapter/7.4 V, 2000 mAh rechargeable lithium battery | AC: 100-240V, 50/60Hz/DC: 7.4V, 3700 mAh lithium rechargeable<br>battery |
| | Dimensions(mm) | 207 mm(L)×96 mm(W)×62 mm(H) | 315mm (L)*215mm (W)*77mm (H) |
| Skin Contacted Material | | Chest Suction ball ( silicone, nickel plated copper) | Chest Suction Electrode (Tin Alloy) |
| | | Limb Clamp Electrode (ABS, PMO +10%fiberglass, nickel plated<br>brass) | Limb Clamp Electrode(Tin Alloy and ABS) |
| | | ECG lead cable (TPU) | ECG lead cable (TPU) |
| Operation Environment | | Temperature: +5℃~ +40℃ | Temperature: +5°C~ +40°C |
| | | Relative humidity: ≤80 % | Relative humidity: ≤80 % |
| | | Atmosphere pressure: 700 hPa ~ 1060 hPa | Atmosphere pressure: 700 hPa ~ 1060 hPa |
| | Temperature: -40°C~+55°C, | Temperature: -40°C~+55°C, | |
| Storage environment | Relative Humidity: ≤95%, no condensation | Relative Humidity: ≤95%, no condensation | |
| | Atmosphere pressure: 500 hPa ~ 1060 hPa, | Atmosphere pressure: 500 hPa ~ 1060 hPa, | |
| Sterile | No | No | |
| Single Use | No | No | |
| Comparison Statement: | The proposed device has the similar main unit specifications with the predicate device. | | |
| Applied Standards: | | | |
| Biocompatibility | ISO10993-5&ISO10993-10 | ISO10993-5&ISO10993-10 | |
| Electrical Safety | IEC60601-1 | IEC60601-1 | |
| EMC | IEC60601-1-2 | IEC60601-1-2 | |
| Performance | IEC 60601-2-25 | IEC 60601-2-25 | |
| Comparison Statement | The proposed probe has same applied Standards with the predicate device. | | |
# Table 7-1 Comparison of Technology Characteristics
{8}------------------------------------------------
## 510(k) Summary
{9}------------------------------------------------
{10}------------------------------------------------
9. Substantially Equivalent (SE) Conclusion SE Analysis :
The subject device has same classification information, same intended use, same indication for use, similar product design, similar specification, same safety elements, similar applied Standards as predicate device.
The differences are included as followings:
Analyse 1: The device and predicate device have difference in Paper speed, paper speed is a manifestation of ECG reporting on paper, per IEC 60601-2-25:2011 Section 201.12.4.108.3 ECG reporting on paper, it stated that the device shall provide at least two recording speeds. 25mm/s and 50mm/s. the proposed device has passed the IEC60601-2-25 test , therefore the paper speed of the proposed device meet the requirements of the standard. we believe these differences will not affect the effectiveness and safety compared with the predicate device.
Analyse 2 : Although the Power supply specifications of ECG90A is different from the predicate device.but both the predicate device and the proposed device has passed the IEC60601-1 safety test. we believe these differences will not affectiveness and safety compared with the predicate device.
Analyse 3: Although the skin Contacted Material of ECG90A is different from the predicate device, but both the predicate device and the proposed device has passed the ISO10993 series test, we believe these differences will not affect the effectiveness and safety compared with the predicate device.
Conclusion: The proposed device is Substantially Equivalent (SE) to the predicate device which is US legally market device. Therefore, the subject device is determined as safe and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.