K171355 · Surefire Medical, Inc. · DQO · May 17, 2017 · Cardiovascular
Device Facts
Record ID
K171355
Device Name
Surefire Precision Infusion System
Applicant
Surefire Medical, Inc.
Product Code
DQO · Cardiovascular
Decision Date
May 17, 2017
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1200
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The Surefire Precision Infusion System is intended for use in angiographic procedures. It delivers radiopaque media and therapeutic agents to selected sites in the peripheral vascular system.
Device Story
Coaxial microcatheter system with expandable distal tip; used in angiographic procedures to deliver radiopaque media, flush solutions, and embolic agents (beads/microspheres) to peripheral vascular sites. Physician operates device by advancing/retracting inner microcatheter relative to outer sheath to expand/collapse tip; expansion improves infusion efficiency while maintaining antegrade flow. Teflon inner liner provides lubricity; hydrophilic coating facilitates navigation. Radiopaque markers assist in positioning. Used in clinical settings by interventionalists. Benefits include precise delivery of therapeutic agents to target vessels.
Clinical Evidence
Bench testing and animal study performed. Bench testing verified tensile strength, kink radius, torque resistance, infusion efficiency, antegrade flow, burst pressure, and coating durability. Animal study assessed acute performance in a simulated clinical environment; results confirmed the device met user needs and performed comparably to the predicate.
Technological Characteristics
Coaxial microcatheter; Teflon inner liner; hydrophilic coating; distal expandable tip; radiopaque markers. Sizes: 0.021-0.025 inch ID; 120-150 cm length; 1.5-6 mm vessel compatibility. Sterilization: Ethylene Oxide (EtO). Materials biocompatible per ISO 10993-1. No software or electronic components.
Indications for Use
Indicated for use in angiographic procedures to deliver radiopaque media and therapeutic agents to selected sites in the peripheral vascular system in patients requiring peripheral vascular intervention.
Regulatory Classification
Identification
An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.
Surefire Infusion System (marketed as Surefire Precision Infusion System) (K160662)
Submission Summary (Full Text)
{0}------------------------------------------------
Public Health Service
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, overlapping each other.
May 17, 2017
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Surefire Medical, Inc. % Mark Job Official Correspondent Regulatory Technology Services. LLC 1394 25th Street, NW Buffalo, Minnesota 55313
Re: K171355
Trade/Device Name: Surefire Precision Infusion System Regulation Number: 21 CFR 870.1200 Regulation Name: Diagnostic intravascular catheter Regulatory Class: Class II Product Code: DOO Dated: March 6, 2017 Received: May 9, 2017
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Fernando
Aguel
Fernando
Aguel-S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K171355
Device Name Surefire Precision Infusion System
Indications for Use (Describe)
The Surefire Precision Infusion System is intended for use in angiographic procedures.
It delivers radiopaque media and therapeutic agents to selected sites in the peripheral vascular system.
Type of Use (Select one or both, as applicable)
| <span style="font-size: 10pt;"> <span style="font-family: Symbol;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|---------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-size: 10pt;"> <span style="font-family: Symbol;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
### Device Name
Surefire Precision Infusion System
## Manufacturer Name and Address
Surefire Medical, Inc. 6272 W. 91st Avenue Westminster, CO 80031 Owner Operator Number: 10038066
## Submitter Contact Information
Surefire Medical, Inc. 6272 W. 91st Avenue Westminster, CO 80031 Lynne Aronson, VP Regulatory Affairs and Quality Assurance Contact: Phone: 303-426-1222 Fax: 303-426-1223
## Common, Classification & Proprietary Names
| Common Name: | diagnostic intravascular catheter |
|----------------------------|------------------------------------|
| Classification Name: | diagnostic intravascular catheter |
| Proprietary Name: | Surefire Precision Infusion System |
| Classification: | Class II |
| Classification Panel: | Cardiovascular Devices |
| Classification Regulation: | 21 CFR 870.1200 |
| Product Code: | DQO |
## Predicate Device
| | ■ Surefire Infusion System 021 | K143588 |
|--|--------------------------------------------------------------------------------|---------|
| | ■ Surefire Infusion System<br>(marketed as Surefire Precision Infusion System) | K160662 |
# Device Description
The Surefire Precision Infusion Systems are coaxial microcatheters with a female luer lock hub at the proximal end, and the Surefire Expandable Tip at the distal end that is sized for use in various sized vessels. The Surefire Precision Infusion Systems are currently available in the following sizes:
| Catalog Number | Inner Diameter | Length | Tip / Vessel Size |
|----------------|----------------|--------|-------------------|
| SPS-21120-35 | 0.021 inch | 120 cm | 1.5 - 3.5 mm |
| SPS-21150-35 | 0.021 inch | 150 cm | 1.5 - 3.5 mm |
| SPS-25120-40 | 0.025 inch | 120 cm | 2 - 4 mm |
| SPS-25150-40 | 0.025 inch | 150 cm | 2 - 4 mm |
| SPS-25120-60 | 0.025 inch | 120 cm | 4 - 6 mm |
| SPS-25150-60 | 0.025 inch | 150 cm | 4 - 6 mm |
The Surefire Precision Infusion System, serves as the conduit for physician-specified agents such as contrast agents, flush solutions, and embolic beads. It is compatible with standard guide wires up to 0.018", and embolic hydrogel particles ≤500μm, and glass microspheres ≤110μm (0.021" lumen) or <150um (0.025" lumen). The Surefire Precision Infusion System has a Teflon inner liner to provide a lubricious surface for passage of physician-specified agents and other accessory devices. The device is hydrophilically coated.
{4}------------------------------------------------
The Surefire Precision Infusion System has an outer sheath to facilitate expanding and collapsing the Surefire Expandable Tip. There are three radiopaque markers located just proximal and distal to the Surefire Expandable Tip. The Tip can be expanded or collapsed for re-positioning during an interventional procedure by simply retracting or advancing the inner microcatheter while holding the outer sheath stationary. When expanded, the Expandable Tip is designed to improve infusion efficiency of compatible embolic agents while maintaining antegrade flow in the vessel.
The Surefire Precision Infusion System is provided sterile (EtO) for single patient use.
## Indications for Use
The Surefire Precision Infusion System is intended for use in angiographic procedures. It delivers radiopaque media and therapeutic agents to selected sites in the peripheral vascular system.
## Biocompatability Testing
Biocompatibility testing (listed below) of the patient-contact materials used in the construction of the modified Surefire Precision Infusion System were performed in accordance with ISO 10993-1 for an external communicating device in contact with circulating blood with a limited duration of less than 24 hours. The testing was conducted in accordance with GLP by NAMSA (Northwood, OH).
Additionally, testing for thrombogenicity was performed on the modified Surefire Precision Infusion System Catheter as a part of an Animal Study.
The results of the biocompatibility testing did not indicate any significant biological reaction that would affect the patient due to contact with the materials used in the device construction.
| Category | Standard | Test Method |
|--------------------------------------------|--------------|----------------------------------------------------------------------------------------|
| Cytotoxicity | ISO 10993-5 | Cytotoxicity Study Using the ISO Elution Method – 1x Minimal Essential Media Extract |
| Sensitization | ISO 10993-10 | ISO Maximization Sensitization Study – Extract – 0.9% Sodium Chloride Solution Extract |
| | | ISO Maximization Sensitization Study – Extract – Sesame Oil, NF Extract |
| Irritation or<br>Intracutaneous Reactivity | ISO 10993-10 | ISO Intracutaneous Study – Extract – 0.9% Sodium Chloride Solution Extract |
| | | ISO Intracutaneous Study – Extract – Sesame Oil, NF Extract |
| Systemic Toxicity | ISO 10993-11 | ISO Systemic Toxicity Study – Extract – 0.9% Sodium Chloride Solution Extract |
| | | ISO Systemic Toxicity Study – Extract – Sesame Oil, NF Extract |
| | | Pyrogen – Material Mediated – 0.9% Sodium Chloride Solution Extract |
| Hemocompatability | ISO 10993-4 | ASTM Hemolysis – CMF-PBS Extract |
| | | C3a Complement Assay – Normal Human Serum Extract |
| | | SC5b-9 Complement Assay – Normal Human Serum Extract |
| | | Coagulation – ASTM Partial Thromboplastin Time |
# Performance Testing
The following design verification / validation tests were performed as a result of the risk analysis assessment of the material modifications. The test results demonstrate that the modified Surefire Precision Infusion System meets the same performance specifications and acceptance criteria as the predicate device.
{5}------------------------------------------------
- Tensile (Pull) Strengths (distal tip) ■
- Kink Radius
- 트 Torque Resistance
- . Infusion Efficiency
- . Antegrade Flow
- . Particulates
- 트 Embolic Agent Compatibility
- Burst Pressure
- Coating Durability and Uniformity
- Frictional Force
- Base Catheter Insertion/Retraction Force
- I Visual and Dimensional
- . EtO Residuals
The following testing was leveraged from previous testing of the predicate device:
- I Tensile (Pull) Strengths (proximal hub/handle)
- Corrosion
- I Diagnostic Agent Compatibility
- . Packaging & Labeling Visual Inspection
- . Package Seal Strength
- . Package Integrity
## Animal Testing
An animal study was performed to assess the comparative acute performance of the modified Surefire Precision Infusion System to the predicate device, as defined by physicians in a simulated clinical environment. The modified Surefire Precision Infusion System was found to be acceptable in all evaluated categories, met the defined user needs, and performed comparably to the predicate device.
## Substantial Equivalence
The modified Surefire Precision Infusion System is substantially equivalent in intended use, design, and technology/principles of operation to the predicate.
Comparative Summary: Design / Technological Characteristics
The modified Surefire Precision Infusion System and predicate device are coaxial microcatheters with the Surefire Expandable Tip at the distal end. Both infusion systems have an outer catheter to facilitate expanding and collapsing the Surefire Expandable Tip.
The modified Surefire Precision Infusion System and the predicate device are constructed of similar materials utilizing similar construction and manufacturing processes.
The predicate and modified are available in the same range of sizes: inner diameters 021" to 025". lengths 120 cm to 150 cm, and tip/vessel size 1.5 mm to 6 mm.
The modified Surefire Precision Infusion System and predicate device are provided in identical packaging, sterilized by ethylene oxide, and labeled for single use only.
## Comparative Summary: Indications for Use
The modified Surefire Precision Infusion System has the same indications for use as the predicate device. Both devices are intended for use in angiographic procedures to deliver radiopaque media and therapeutic agents to selected sites in the peripheral vascular system.
## Comparative Summary: Performance
Animal and bench performance test data demonstrate that the modified Surefire Precision Infusion System performance is comparable to the predicate device.
In summary, the modified Surefire Precision System is substantially equivalent in intended use, design, and technology/principles of operation to the predicate device. Animal and bench performance test data demonstrate that the modified Surefire Precision Infusion System performance is comparable to the predicate device. Differences between the devices do not raise any issues of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.