K171346 · Icu Medical · MRZ · Aug 25, 2017 · General Hospital
Device Facts
Record ID
K171346
Device Name
Sapphire Sets
Applicant
Icu Medical
Product Code
MRZ · General Hospital
Decision Date
Aug 25, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Sapphire sets are intended for use with the Sapphire Infusion System to deliver fluids from a container to a patient's vascular system.
Device Story
Sapphire Sets are disposable, single-patient-use administration sets designed for use with the Sapphire Infusion System. Components include male/female luer adapters, piercing pins, sapphire cassettes, tubing, flow control devices, filters, in-line adapters, injection sites, check valves, pressure-activated valves, and blood chambers. The sets facilitate fluid delivery from a container to a patient's vascular system or epidural space. Used in clinical or ambulatory environments by healthcare professionals or patients. The device acts as the fluid path interface between the infusion pump and the patient; the pump controls the flow rate and volume. Benefits include reliable, controlled fluid administration for various therapies, including blood products and epidural medications.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included ISO 10993 biocompatibility assessments and performance testing per ISO 8536-4, ISO 8536-8, and ISO 1135-4 standards, covering particulate contamination, leakage, tensile strength, coring, and flow regulator performance. Flow rate accuracy testing was conducted using the Sapphire Infusion System.
Technological Characteristics
Disposable infusion set components including luer adapters, piercing pins, cassettes, tubing, flow regulators, filters, and valves. Materials are identical to previously cleared predicate/reference sets. Sterilization achieves a Sterility Assurance Level (SAL) of 10^-6. Connectivity is physical/mechanical via the Sapphire Infusion System cassette interface.
Indications for Use
Indicated for delivery of fluids from container to patient vascular system (Primary Sets), delivery of fluids including blood/blood products to patient vascular system (Blood Sets), and delivery of fluids to patient epidural space (Epidural Sets).
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image is a circular logo for the U.S. Department of Health & Human Services. The logo features the department's symbol, which is a stylized representation of three human profiles facing right, stacked on top of each other. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper half of the circle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 25, 2017
ICU Medical Amy Giertych Vice President. Global Regulatory Affairs 600 North Field Drive Lake Forest, Illinois 60045
Re: K171346
Trade/Device Name: Sapphire Sets Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: Class II Product Code: MRZ Dated: July 27, 2017 Received: July 27, 2017
Dear Amy Giertych:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerelv.
# Tara A. Ryan -S
for Lori Wiggins, MPT, CLT Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known)
K171346
Device Name Sapphire Set
Indications for Use (Describe)
Sapphire Primary Sets are indicated for the delivery of fluids from a container to a patient's vascular.
Sapphire Blood Sets are indicated for the delivery of fluids including but not limited to blood and blood products from a container to a patient's vascular system.
Sapphire Epidural Set are indicated for the delivery of fluids to a patient's epidural space.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
A summary of 510(k) substantial equivalence information in accordance with the requirements of 21 CFR 807.92 for Sapphire Sets.
| Submitter Information | |
|---------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name | ICU Medical |
| Address | 600 North Field Drive<br>Lake Forest, IL. 60046 |
| Phone number | (224)-212-4235 |
| Mobile number | (224)-507-8812 |
| Fax number | (224) 212-5221 |
| Establishment Registration<br>Number | 3005579246 (Owner/Operator #9063339) |
| Name of contact person | Amy Giertych, Vice President Global Regulatory Affairs |
| Date prepared | 8/23/2017 |
| Name of device | |
| Trade or proprietary name | Sapphire Sets |
| Common or usual name | Accessories, Pump, Infusion |
| Classification name | Infusion Pump, 21 CFR 880.5725, Class II |
| Product Code(s) | Primary Code: MRZ |
| Legally marketed device(s) to which<br>equivalence is claimed | Sapphire Infusion System - K123049 (Primary Predicate 510(k))<br>Sapphire Sets - K160492 (Reference 510(k))<br>Plum Sets - K141789 (Reference 510(k)) |
| Reason for 510(k) submission | New IV administration sets to be used with Q Core Medical Ltd.'s Sapphire Infusion System. |
| Device description | Sapphire Sets are intended for use with the Sapphire Infusion System. Sapphire sets are<br>comprised of various components including the following: male luer adapter with cap, female<br>luer with cap, piercing pin connector, sapphire cassette, tubing, flow control device, filter, in-<br>line adapter, injection site assembly, check valve, pressure activated valve and blood chamber.<br>Sapphire sets are configured to ensure the intended use of the device is met. The sets are<br>disposable devices for single patient use. |
| Intended Use of Device | Sapphire sets are intended for use with the Sapphire Infusion System to deliver fluids from a<br>container to a patient's vascular system. |
| Indication for use | Sapphire Primary Sets are indicated for the delivery of fluids from a container to a patient's<br>vascular system.<br>Sapphire Blood Sets are indicated for the delivery of fluids including but not limited to blood<br>and blood products to a patient's vascular system.<br>Sapphire Epidural Sets are indicated for the delivery of fluids from a container to a patient's<br>epidural space. |
{4}------------------------------------------------
| Summary of the technological characteristics of the device compared to the predicate device | | |
|---------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Characteristic | Predicate 510(k) | Proposed Device |
| Indications for Use | The Q Core Sapphire infusion pump is intended for the controlled delivery through intravascular, subcutaneous, intra-arterial and epidural routes. The pump is designed to deliver saline, Total Parenteral Nutrition (TPN), lipids, IV medication, epidural medication, blood and blood products.<br>The Sapphire pump includes the following infusion modes for all intended uses: Continuous, Intermittent, TPN, PCA, Multi-step, and Epidural. The pump is intended to be used by both licensed health care professionals in a clinical environment, and home users in an ambulatory environment. The Sapphire pump is designed to follow the patient through the various care areas, and is suitable for use in the different settings. The dedicated Q Core administration sets for the Sapphire pump are intended for single-patient use and single-use only. | Sapphire Sets are indicated for the delivery of fluids from a container to a patient's vascular system.<br>Sapphire Blood Sets are indicated for the delivery of fluids including but not limited to blood and blood products to a patient's vascular system.<br>Sapphire Epidural Sets are indicated for the delivery of fluids from a container to a patient's epidural space. |
| Characteristic | Predicate/Reference 510(k) | Proposed Device |
| Design and Materials of Construction | The design and materials of construction of all components for the sets, are as cleared under predicate/reference 510(k) noted below for each set component. | The design and materials of construction for all set components listed below have been previously cleared under the associated predicate/reference 510(k) |
| | Piercing Pin | Piercing Pin |
| | new | 75-0704 |
| | K160492 (reference 510(k)) | 85-0143 |
| | K160492 (reference 510(k)) | 85-0142 |
| | K160492 (reference 510(k)) | 85-0288 |
| | K160492 (reference 510(k)) | 67-1760 |
| | K160492 (reference 510(k)) | 90-5710 |
| | Female Adapter | Female Adapter |
| | K160492 (reference 510(k)) | 90-9674 |
| | Male Adapter | Male Adapter |
| | K160492 (reference 510(k)) | 85-0250 |
| | Cap/Hood | Cap/Hood |
| | K160492 (reference 510(k)) | 75-0042 |
| | K160492 (reference 510(k)) | 75-0735 |
| | K160492 (reference 510(k)) | 75-2322 |
| | K160492 (reference 510(k)) | 90-9664 |
| Characteristic | Predicate/Reference 510(k) | Proposed Device |
| | K160492 (reference 510(k)) | 90-9665 |
| Cassette | | Cassette |
| | K160492 (reference 510(k)) | 85-0239 |
| | K160492 (reference 510(k)) | 67-2115 |
| Tubing | | Tubing |
| | K160492 (reference 510(k)) | 75-0322 |
| | K160492 (reference 510(k)) | 75-0324 |
| | K160492 (reference 510(k)) | 75-0364 |
| | K141789 (reference 510(k)) | 75-0941 |
| | K160492 (reference 510(k)) | 75-5236 |
| Flow Control | | Flow Control |
| | K141789 (reference 510(k)) | 67-1942 |
| | K160492 (reference 510(k)) | 75-1004 |
| | K160492 (reference 510(k)) | 75-1099 |
| | K160492 (reference 510(k)) | 96-4509 |
| | K141789 (reference 510(k)) | 96-4882 |
| In-Line Adapter | | In-Line Adapter |
| | K141789 (reference 510(k)) | 75-4967 |
| | K160492 (reference 510(k)) | 90-6979 |
| Injection Site | | Injection Site |
| | K160492 (reference 510(k)) | 67-2060 |
| Check Valve | | Check Valve |
| | K160492 (reference 510(k)) | 67-1042 |
| | K160492 (reference 510(k)) | 67-1589 |
| Filter |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.