EMG Blue Nitrile Examination Gloves Powder free with tested for use with chemotherapy Drugs
K171339 · Eco Medi Glove Sdn. Bhd. · LZA · Jul 7, 2017 · General Hospital
Device Facts
Record ID
K171339
Device Name
EMG Blue Nitrile Examination Gloves Powder free with tested for use with chemotherapy Drugs
Applicant
Eco Medi Glove Sdn. Bhd.
Product Code
LZA · General Hospital
Decision Date
Jul 7, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner and for use with chemotherapy drugs. In addition these gloves were tested for use with chemotherapy drugs in accordance ASTM D6978-05 standards- Practice for assessment of Medical Glove to Permeation by chemotherapy drugs
Device Story
Disposable, non-sterile, blue nitrile latex examination glove; worn by healthcare examiners to prevent cross-contamination. Tested for permeation resistance against specific chemotherapy drugs per ASTM D6978-05. Used in clinical settings; provides a physical barrier between patient and examiner. Benefits include protection against patient-to-examiner contamination and exposure to hazardous chemotherapy agents.
Clinical Evidence
Bench testing only. Includes physical property testing (tensile strength, elongation), pinhole testing (ASTM D5151-06), residual powder testing (ASTM D6124-06), and chemotherapy drug permeation testing (ASTM D6978-05). Biocompatibility testing performed per ISO 10993-10:2010 (primary skin irritation and dermal sensitization).
Indicated for use as a disposable patient examination glove worn on the examiner's hand or finger to prevent cross-contamination between patient and examiner, and for use during handling of chemotherapy drugs.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
EMG Blue Nitrile Examination Glove Powder Free with tested for use with Chemotherapy Drugs (K161187)
Submission Summary (Full Text)
{0}------------------------------------------------
Public Health Service
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a series of human profiles facing to the right.
July 7, 2017
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
ECO Medi Glove Sdn. Bhd. Suresh Kumar Official Correspondent Lot 23826. Jalan Tembaga Kuning Kamunting Raya Industrial Estate Kamunting Perak, 34600 Malaysia
Re: K171339
Trade/Device Name: EMG Blue Nitrile Examination Gloves Powder free with tested for use with chemotherapy Drugs Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: Class I Product Code: LZA, LZC Dated: May 15, 2017 Received: May 19, 2017
Dear Suresh Kumar:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Mark S. Fellman -S
for
Lori A. Wiggins, MPT, CLT Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
#### 510(k) Number (if known) K171339
#### Device Name
EMG Blue Nitrile Examination Glove Powder Free with tested for use with Chemotherapy Drugs
Indications for Use (Describe)
A powder-free patient examination glove is a disposable device intended for use medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and use for use with chemotherapy drugs.
In addition these gloves were tested for use with chemotherapy drugs in accordance ASTM D6978-05 standards- Practice for assessment of Medical Glove to Permeation by chemotherapy drugs
Chemotherapy Drugs and Concentration
Minimum Breakthrough Detection Time(Min),ug/cm2/minutes
| 1) Carmustine (BCNU)(3.3 mg/ml) | 24.0 Minutes |
|--------------------------------------|---------------|
| 2)Cyclophoshamide (20 mg/ml) | > 240 Minutes |
| 3)Doxorubicin Hydrochloride (2mg/ml) | > 240 Minutes |
| 4)Etoposide (20mg/ml) | > 240 Minutes |
| 5)Fluorouracil (50mg/ml) | > 240 Minutes |
| 6)Methorexate (25mg/ml) | > 240 Minutes |
| 7)Paclitaxel (6mg/ml) | > 240 Minutes |
| 8)Thiotepa (10mg/ml) | 56.9 Minutes |
| 9)Carboplatin (10mg/ml) | > 240 Minutes |
| 10)Cisplatin (1 mg/ml) | > 240 Minutes |
| 11)Dacarabazine (1 mg/ml) | > 240 Minutes |
| 12)Ifosfamide (5 mg/ml) | > 240 Minutes |
| 13)Mitomycin (0.5mg/ml) | > 240 Minutes |
| 14)Mitoxantrone (2 mg/ml) | > 240 Minutes |
| 15)Vincristine Sulfate (1 mg/ml) | > 240 Minutes |
The Maximum testing is 240 minutes. Please note that the following drugs have extremely low Permeation time, Carmustine (BCNU)(3.3 mg/ml) - Minimum breakthrough detection time : 24.0 minutes Thiotepa (10mg/ml) - Minimum breakthrough detection time : 56.9 minutes
Type of Use (Select one or both, as applicable)
] Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
*(formerly known as Sinetimed Consumables Sdn. Bhd.)*
Jalan Tembaga Kuning, Kamunting Raya Industrial Estate, 34600 Taiping, Perak Darul Ridzuan. MALAYSIA. +60-5-891 2777 FAX +60-5-891 2000
## 510(K) Summary EMG Blue Nitrile Examination Gloves Powder Free with Tested for use with Chemotherapy Drugs
### 1.0 Submitter :
| Company Name | : ECO MEDI GLOVE SDN. BHD. |
|-------------------|---------------------------------------------------------------------------------------------------------------|
| Company Address : | : Lot 23826, Jalan Tembaga Kuning<br>Kamunting Raya Industrial Estate,<br>34600, Kamunting Perak<br>Malaysia. |
| Contact Person | : Mr Suresh Kumar |
| Telephone No | : 603-60283033 |
| Email | : suresh@ecomediglove.com.my |
- 2.0 Preparation Date : 4th July 2017
## 3.0 Name of the Device
Trade Name / Proprietary Name : EMG Blue Nitrile Examination Gloves Powder Free with tested for use with Chemotherapy drugs.
Device Name : Nitrile Patient Examination gloves.
Device Classification Name : Patient Examination gloves (21 CFR 880.6250).
Device Class : Class I.
Product Code : Nitrile-LZA and LZC.
{5}------------------------------------------------
*(formerly known as Sinetimed Consumables Sdn. Bhd.)*
amunting Raya Industrial Estate, 34600 Taiping. Perak Darul Ridzuan. MALAYSIA. +60-5-891 2777
#### 4.0) Identification of The Legally Marketed Device :
EMG Blue Nitrile Examination Gloves Powder Free with tested for use with Chemotherapy drugs.LZC, which meets all the requirement of ASTM D 6319-10 and FDA 21 CFR 880.6250.It is equivalent to K161187, EMG Blue Nitrile Examination Glove Powder Free with tested for use with Chemotherapy Drugs.
#### 5.0 Device Description
The subject device in this 510(k) Notification is EMG Blue Nitrile Examination Gloves Powder Free with tested for use with Chemotherapy drugs. The subject device is a patient examination glove made from nitrile latex compound, Blue color, powder free and non-sterile (Per 21 CFR 880.6250, class I). The device meets all the specifications in ASTM D6319-10, Standard specification for Nitrile Examination Gloves. Additionally, the gloves have been tested for biocompatibility and permeability to chemotherapy drugs.
The Blue Nitrile Medical Examination Gloves ,Powder Free, is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner (product Code LZA) and is used with chemotherapy drugs (Product code LZC). The subject device is substantially equivalent to the legally marketed Nitrile Medical Examination Gloves (product Code LZA and LZC).
| Dimension | Size S | Size M | Size L | Size XL |
|------------------------------------------|--------|--------|--------|---------|
| Overall Length (mm) (Minimum) | 230 | 230 | 230 | 230 |
| Width (± 5mm) | 85 | 95 | 105 | 115 |
| Thickness at Palm (mm)<br>(Minimum) | 0.05 | 0.05 | 0.05 | 0.05 |
| Thickness at Finger Tip<br>(mm)(Minimum) | 0.05 | 0.05 | 0.05 | 0.05 |
### 6.1 Specification for Nitrile gloves: 6.2 Dimension and Thickness of Gloves
{6}------------------------------------------------
#### ECO Medi 5262-D)
(for
Lot 23826, Jalan Tembaga Kuning, Kamunting Raya Industrial Estate, 34600 Taiping. Perak Darul Ridzuan. MALAYSIA. TEL +60-5-891 2777 FAX +60-5-891 2999
| Measurement | Before Ageing | After Aging at 70 °C for<br>168 hrs @ 100°C for 22 hrs |
|-------------------------|----------------------|--------------------------------------------------------|
| Tensile Strength (MPa) | 14min | 14 Min |
| Ultimate Elongation (%) | 500min | 400min |
| Pin-hole Level | AQL 2.5 | AQL 2.5 |
| | Inspection Level G-1 | Inspection Level G-1 |
# 6.1.2 Gloves Physical Properties and Holes
Gloves meet all the specification listed in ASTM D 6319-10
| Characteristics | Acceptance Criteria | EMG Blue Nitrile<br>Examination Gloves<br>Powder Free with<br>tested for use with<br>chemotherapy drugs.<br>K171339 | EMG Blue Nitrile<br>Examination Gloves,<br>Powder Free with<br>tested for use with<br>chemotherapy drugs,<br>K161187 | Comparison | | | | |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|-------------------|----------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| Product Code | LZA and LZC | LZA and LZC | LZA and LZC | Same | | | | |
| Intended use | A powder free patient<br>examination glove is a<br>disposable device<br>intended for medical<br>purposes that is worn<br>on the examiner's hand<br>or finger to prevent<br>contamination between<br>patient and examiner.<br>The device is for over-<br>the-counter use.<br>Intended use. | A powder free patient<br>examination glove is a<br>disposable device<br>intended for medical<br>purposes that is worn on<br>the examiner's hand or<br>finger to prevent<br>contamination between<br>patient and examiner.<br>The device is for over-<br>the-counter use. | A powder free patient<br>examination glove is a<br>disposable device<br>intended for medical<br>purposes that is worn on<br>the examiner's hand or<br>finger to prevent<br>contamination between<br>patient and examiner.<br>The device is for over-<br>the-counter use. | Same | | | | |
| Material use | Nitrile latex compound | Nitrile latex compound | Nitrile latex compound | Same | | | | |
| Color | Blue | Blue | Blue | Same | | | | |
| Sterility | Non sterile | Non sterile | Non sterile | Same | | | | |
| Single used | Single used | Single used | Single used | Same | | | | |
| | | | | | | | | |
| Dimensions | Overall Length (mm)<br>Min 230 mm<br>Width (± 5mm)<br>Size S = <b>85mm</b><br>Size M = <b>95mm</b><br>Size L = <b>105mm</b><br>Size XL = <b>115mm</b><br>Thickness at Palm (mm)<br>Min; 0.05 mm<br>Thickness at Finger Tip (mm)<br>Min 0.05 mm | Size : Medium<br>Palm Width: 98 mm<br>Length : 247 mm<br>Finger Thickness : 0.08mm<br>Palm Thickness : 0.05mm<br>Meets ASTM D6319-10 | Size : Medium<br>Palm Width: 99 mm<br>Length : 247 mm<br>Finger Thickness : 0.11mm<br>Palm Thickness : 0.07mm<br>Meets ASTM D6319-10 | Same | | | | |
| Physical<br>properties | <b>Before Ageing</b><br>Tensile Strength (MPa)<br>= Min 14 Mpa<br>Ultimate Elongation (%)<br>= Min 500 %<br><b>After Aging at 70°C for<br/>168 hrs @ 100°C for<br/>22 hrs</b><br>Tensile Strength (MPa)<br>= Min 14 Mpa<br>Ultimate Elongation (%)<br>= Min 400 % | Tensile strength (Mpa) :<br>Before aging : 25.45Mpa<br>After Accelerated aging:<br>26.33Mpa<br>Ultimate Elongation(%):<br>Before aging:600%<br>After aging : 550%<br>Meets ASTM D6319-10 | Tensile strength (Mpa) :<br>Before aging : 25.45Mpa<br>After Accelerated aging:<br>26.33Mpa<br>Ultimate Elongation(%):<br>Before aging:600%<br>After aging : 550%<br> | Same | | | | |
| | Freedom from<br>pinholes | AQL 2.5<br>Inspection Level G-1 | AQL 2.5<br>Meets ASTM D5151-06 | AQL 2.5<br>Meets ASTM D5151-06 | | | | |
| | Residual<br>Powder | < 2.0 mg/glove | Average 0.17 mg/glove<br>Meets ASTM D6124-06 | Average 0.19 mg/glove<br>Meets ASTM D6124-06 | | | | |
| | Biological<br>Evaluation on<br>Medical Device<br>-Primary Skin<br>Irritation Test | ISO 10993-10:2010 | Under the conditions of<br>this study, the test article<br>was a non-irritant. | Under the conditions of<br>this study, the test article<br>was a non-irritant. | | | | |
| | | | | | Biological<br>Evaluation on<br>Medical<br>Device-<br>Dermal<br>Sensitization<br>Assay | ISO 10993-10:2010 | Under the conditions of<br>this study, the test article<br>was a non-sensitizer. | Under the conditions of<br>this study, the test article<br>was a non- sensitizer. |
| | | | | | | | | |
{7}------------------------------------------------
# ECO Medi Glove Sdn. Bhd. (815262-D)
(formerly known as Sinetimed Consumables Sdn. Bhd.)
(formerly known as Sinetimed Consumables Sdn. Bhd.)
Lot 23826, Jalan Tembaga Kuning, Kamunting Raya Industrial Estate, 34600 Taiping. Perak Darul Ridzuan. MALAYSIA. TEL +60-5-891 2777 FAX +60-5-891 2999
{8}------------------------------------------------
## 510K No : K171339
# ECO Medi Glove Sdn. Bhd. (815262-D)
(formerly known as Sinetimed Consumables Sdn. Bhd.)
*(formerly known as Sinetimed Consumables Sdn. Bhd.)*
Lot 23826, Jalan Tembaga Kuning, Kamunting Raya Industrial Estate, 34600 Taiping. Perak Darul Ridzuan. MALAYSIA. TEL +60-5-891 2777 FAX +60-5-891 2999
{9}------------------------------------------------
# ECO Medi Glove Sdn. Bhd. (815262-D)
(formerly known as Sinetimed Consumables Sdn. Bhd.)
*(formerly known as Sinetimed Consumables Sdn. Bhd.)*
Lot 23826, Jalan Tembaga Kuning, Kamunting Raya Industrial Estate, 34600 Taiping. Perak Darul Ridzuan. MALAYSIA. TEL +60-5-891 2777 FAX +60-5-891 2999
| 7) Methorexate<br>(25mg/ml or<br>25,000ppm),<br>Minimum Breakthrough<br>detection time: > 240<br>minutes | 9) Thiotepa (10mg/ml or<br>10,000ppm),<br>Breakthrough time: 67.8<br>minutes. | |
|----------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|--|
| 8) Paclitaxel (6mg/ml or<br>6,000ppm),<br>Minimum Breakthrough<br>detection time:<br>>240minutes | | |
| 9) Thiotepa (10mg/ml or<br>10,000ppm),<br>Minimum Breakthrough<br>detection time:<br>56.9 minutes | | |
| 10) Ifosfamide<br>(50mg/ml)<br>Minimum Breakthrough<br>detection time: >240<br>minutes | | |
| 11) Mitoxantrone<br>(2mg/ml) ,<br>Minimum Breakthrough<br>detection time: >240<br>minutes | | |
| 12) Vincristine Sulfate<br>(1mg/ml) ,<br>Minimum Breakthrough<br>detection time: >240<br>minutes | | |
| 13) Carboplatin<br>(10.0mg/ml), Minimum<br>Breakthrough detection<br>time: >240 minutes | | |
| | 14) Dacarbazine (DTIC)<br>(10.0mg/ml) Minimum<br>Breakthrough detection<br>time: >240 minutes | |
| | 15) Mitomicin C<br>(0.5mg/ml) Minimum<br>Breakthrough detection<br>time: >240 minutes | |
{10}------------------------------------------------
(formerly known as Sinetimed Consumables Sdn. Bhd.)
Lot 23826, Jalan Tembaga Kuning, Kamunting Raya Industrial Estate, 34600 Taiping. Perak Darul Ridzuan. MALAYSIA. TEL +60-5-891 2777 FAX +60-5-891 2999
### 7.0 Indications for Use
A powder-free patient examination glove is a disposable device intended for medical Purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner and for use with chemotherapy drugs .
In addition these gloves were tested for use with Chemotherapy drugs in accordance With ASTM D6978-05 standards Practice for assessment of Medical Glove to Permeation by chemotherapy drugs.
| Chemotherapy Drug and concentration | Minimum Breakthrough detection time in Minutes,µg/cm²/minute |
|---------------------------------------|--------------------------------------------------------------|
| 1)Carmustine (BCNU) (3.3mg/ml) | 24 minutes |
| 2)Cyclophosphamide (20mg/ml) | > 240 minutes |
| 3)Cytarabine (10mg/ml) | > 240 minutes |
| 4)Doxorubicin Hydrochloride (2 mg/ml) | > 240 minutes |
| 5)Etoposide (20mg/ml) | > 240 minutes |
| 6)Fluorouracil (50mg/ml) | > 240 minutes |
| 7)Methorexate (25mg/ml) | > 240 minutes |
| 8) Paclitaxel (6mg/ml) | > 240 minutes |
| 9) Thiotepa (10mg/ml) | 56.9 minutes |
| 10) Cisplatin (1.0mg/ml) | > 240 minutes |
{11}------------------------------------------------
*(formerly known as Sinetimed Consumables Sdn. Bhd.)*
Raya Industrial Estate, 34600 Taiping. Perak Darul Ridzuan. MALAYSIA.
| 11) Dacarbazine (DTIC) (10.0mg/ml) | > 240 minutes |
|------------------------------------|---------------|
| 12) Ifosfamide (50.0mg/ml) | > 240 minutes |
| 13) Mitomycin C (0.5mg/ml) | > 240 minutes |
| 14) Mitoxantrone (2.0mg/ml) | > 240 minutes |
| 15) Vincristine Sulfate (1.0mg/ml) | > 240 minutes |
The maximum testing time is 240 minutes. Please note that the following drugs have extremely low permeation time,
Carmustine (BCNU) (3.3mg/ml) - Minimum Breakthrough detection time 24 Minutes
Thiotepa (10mg/ml) - Minimum Breakthrough detection time 56.9 minutes.
## 8.0 Summary of the Technological Characteristics of the Device compared to the Predicate Device for substantial equivalent discussion
There are no differences in technological characteristics of the subject device compare with the predicate device.
The gloves are made from nitrile latex compound, Blue color, Powder free and nonsterile. The gloves met all the specifications in ASTM D6319-10 Standard specification for Nitrile Examination Gloves as well Biological Evaluation on medical device. Additionally, the gloves have been tested for permeability to Chemotherapy drugs.
### 9.0 Conclusion
Based on intended uses, technological characteristics and non - clinical performance Data, the EMG Blue Nitrile Examination Gloves Powder Free with tested for use with Chemotherapy Drugs is as safe, as effective, and performs as well as the legally marketed predicate device, K161187.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.