B R A H M S PCT sensitive KRYPTOR

K171338 · Brahms GmbH · PRI · May 31, 2017 · Microbiology

Device Facts

Record IDK171338
Device NameB R A H M S PCT sensitive KRYPTOR
ApplicantBrahms GmbH
Product CodePRI · Microbiology
Decision DateMay 31, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.3215
Device ClassClass 2
AttributesReal-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K171338 · May 31, 2017B R A H M S PCT sensitive KRYPTORBrahms GmbHSystematic literature reviews of randomized controlled trials; Aggregated patient-level data from clinical trial datasetsMeta-analyses of clinical trial data were used to demonstrate the clinical utility of PCT-guided antibiotic therapy for lower respiratory tract infections (LRTI) and sepsis, specifically evaluating reductions in antibiotic exposure, duration, and impact on patient outcomes like mortality and length of stay.Meta-analysis; Antibiotic stewardship; Clinical utility; Systematic review

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Study-level and Patient-level Meta-analyses (LRTI); Systematic literature review and meta-analysis of RCTs; Study Period: 2004-2016Patients with suspected or confirmed lower respiratory tract infections (CAP, acute bronchitis, AECOPD); Sample Size: 4090 (study-level); 3142 (patient-level); Number of Sites: Multiple (11-13 RCTs)Standard care therapyAntibiotic initiation, duration, exposure, length of stay, mortality, complications
Study-level and Patient-level Meta-analyses (Sepsis); Systematic literature review and meta-analysis of RCTs; Study Period: 2008-2016Patients with suspected or confirmed sepsis; Sample Size: 3489 (study-level); 598 (patient-level); Number of Sites: Multiple (5-10 RCTs)Standard care therapyAntibiotic duration, antibiotic exposure, length of ICU/hospital stay, mortality

Indications for Use

The B-R-A-H-M-S PCT sensitive KRYPTOR® is an immunofluorescent assay using Time-Resolved Amplified Cryptate Emission (TRACE) technology to determine the concentration of PCT (procalcitonin) in human serum and EDTA or heparin plasma. The B·R·A·H·M·S PCT sensitive KRYPTOR® is intended to be performed on the B·R·A·H·M·S KRYPTOR® analyzer family. Used in conjunction with other laboratory findings and clinical assessments, B·R·A·H·M·S PCT sensitive KRYPTOR® is intended for use as follows: to aid in the risk assessment of critically ill patients on their first day of ICU admission for progression to severe sepsis and septic shock, to determine the change in PCT level over time as an aid in assessing the cumulative 28-day risk of all-cause mortality for patients diagnosed with severe sepsis or septic shock in the ICU or when obtained in the emergency department or other medical wards prior to ICU admission, to aid in decision making on antibiotic therapy, for inpatients or patients in the emergency department with suspected or confirmed lower respiratory tract infections (LRTI) - defined as community-acquired pneumonia (CAP), acute bronchitis, and acute exacerbation of chronic obstructive pulmonary disease (AECOPD), to aid in decision making on antibiotic discontinuation for patients with suspected or confirmed sepsis.

Device Story

Automated immunofluorescent assay; measures procalcitonin (PCT) in human serum/plasma using TRACE technology; performed on B·R·A·H·M·S KRYPTOR analyzer family. Input: patient serum/plasma sample; Transformation: sandwich immunoassay with sheep polyclonal and mouse monoclonal anti-PCT antibodies; Output: quantitative PCT concentration. Used in clinical labs/hospitals; operated by laboratory technicians. Healthcare providers interpret PCT levels alongside clinical signs/symptoms to guide antibiotic initiation/discontinuation and assess sepsis/mortality risk. Benefits: optimized antibiotic stewardship; reduced antibiotic exposure/duration; improved risk stratification for critically ill patients.

Clinical Evidence

Evidence includes two systematic literature reviews and meta-analyses (study-level and patient-level) of randomized controlled trials (RCTs). LRTI meta-analysis (11-13 RCTs, N=3142-4090) showed 38-51% reduction in antibiotic exposure and 2.9-day reduction in duration. Sepsis discontinuation meta-analysis (5-10 RCTs, N=598-3489) showed 23% reduction in antibiotic exposure and 1.5-day reduction in duration. No negative effects on mortality or length of stay observed.

Technological Characteristics

Homogeneous sandwich immunoassay; TRACE (Time-Resolved Amplified Cryptate Emission) technology; 19-minute assay duration. Reagents: lyophilized sheep polyclonal and mouse monoclonal anti-PCT antibodies. Sample volume: 50 µL. Connectivity: B·R·A·H·M·S KRYPTOR analyzer family. Standards: CLSI EP05-A3, EP06-A, EP07-A2, EP17-A2, EP25-A, EP21-Ed2; ISO 14971, ISO 15223-1; IEC 62304, IEC 61010-1.

Indications for Use

Indicated for critically ill patients on first day of ICU admission for risk assessment of progression to severe sepsis/septic shock; patients with severe sepsis/septic shock in ICU, ED, or medical wards for 28-day all-cause mortality risk assessment; inpatients or ED patients with suspected/confirmed LRTI (CAP, acute bronchitis, AECOPD) for antibiotic therapy decision-making; and patients with suspected/confirmed sepsis for antibiotic discontinuation decision-making. Not for use in patients <18 years, pregnant women, or immunocompromised individuals.

Regulatory Classification

Identification

A device to detect and measure non-microbial analyte(s) in human clinical specimens to aid in assessment of patients with suspected sepsis is identified as an in vitro device intended for the detection and qualitative and/or quantitative measurement of one or more non-microbial analytes in human clinical specimens to aid in the assessment of patients with suspected sepsis when used in conjunction with clinical signs and symptoms and other clinical and laboratory findings.

Special Controls

A device to detect and measure non-microbial analyte(s) in human clinical specimens to aid in assessment of patients with suspected sepsis must comply with the following special controls:

*Classification.* Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include the device's detailed Indications for Use statement describing what the device detects and measures, the results provided to the user, whether the measure is qualitative and/or quantitative, the clinical indications for which the test is to be used, and the specific population(s) for which the device use is intended. (2) Premarket notification submissions must include detailed documentation of the device description, including (as applicable), all device components, software, ancillary reagents required but not provided, explanation of the device principle and methodology, and for molecular devices include detailed documentation of the primer/probe sequence, design, and rationale for sequence selection. (3) Premarket notification submissions must include detailed documentation of applicable analytical studies, such as, analytical sensitivity (Limit of Detection, Limit of Blank, and Limit of Quantitation), precision, reproducibility, analytical measuring range, interference, cross-reactivity, and specimen stability. (4) Premarket notification submissions must include detailed documentation of a prospective clinical study or, if appropriate, results from an equivalent sample set. This detailed documentation must include the following information: (i) Results must demonstrate adequate device performance relative to a well-accepted comparator. (ii) Clinical sample results must demonstrate consistency of device output throughout the device measuring range likely to be encountered in the Intended Use population. (iii) Clinical study documentation must include the original study protocol (including predefined statistical analysis plan), study report documenting support for the Indications for Use(s), and results of all statistical analyses. (5) Premarket notification submissions must include evaluation of the level of the non-microbial analyte in asymptomatic patients with demographic characteristics ( *e.g.,* age, racial, ethnic, and gender distribution) similar to the Intended Use population.(6) As part of the risk management activities performed under 21 CFR 820.10(c) design and development, you must document an appropriate end user device training program that will be offered as part of your efforts to mitigate the risk of failure to correctly operate the instrument. (7) A detailed explanation of the interpretation of results and acceptance criteria must be included in the device's 21 CFR 809.10(b)(9) compliant labeling, and a detailed explanation of the interpretation of the limitations of the samples ( *e.g.,* collected on day of diagnosis) must be included in the device's 21 CFR 809.10(b)(10) compliant labeling.

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION ## TRIAGE-QUICK REVIEW DECISION SUMMARY 510(k) #: K171338 This 510(k) was reviewed under OIR’s Triage-Quick Review Program. This program represents an internal workload management tool intended to reduce internal FDA review resources for 510(k) applications that are of good quality upon receipt by FDA. The information in the 510(k) is complete and supports a substantial equivalence (SE) determination. Please refer to the applicant’s 510(k) summary for a summary of the information that supports this SE determination.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...