AR N Type Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. AR N Type Implant System is for two stage surgical procedures. It is intended for delayed load.
Device Story
AR N Type Implant System consists of endosseous dental implants and associated abutments; used to replace missing teeth in mandible or maxilla. Implants are surgically placed in jaw bone; osseointegrate to provide prosthetic anchorage. System includes fixtures (CP Ti Gr 4) with RBM surface treatment and various abutments (cover screws, healing, angled, cemented, solid). Used by dental professionals in clinical settings. Output is a stable foundation for dental restorations; restores chewing function. Benefits include tooth replacement and improved oral function.
Clinical Evidence
Bench testing only. Includes sterilization validation (ISO 11137, ISO 11737), steam sterilization validation (ANSI/AAMI ST79), surface chemical/SEM analysis, comparative fatigue testing (ISO 14801:2007), and biocompatibility (cytotoxicity ISO 10993-5, sensitization ISO 10993-10).
Indicated for partially or fully edentulous adult patients requiring single or multiple-unit dental restorations, including cemented, screw-retained, or overdenture restorations, and fixed bridgework support. For two-stage surgical procedures with delayed loading.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
January 31, 2018
Biotem Co., Ltd. c/o Joyce Bang Consultant Provision Consulting Group Inc. 14071 Peyton Dr. #967 Chino Hills, California 91709
Re: K171297
Trade/Device Name: AR N Type Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: November 2, 2017 Received: January 4, 2018
Dear Joyce Bang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Mary S. Runner -S
For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K171297
Device Name AR N Type Implant System
#### Indications for Use (Describe)
AR N Type Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. AR N Type Implant System is for two stage surgical procedures. It is intended for delayed load.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| <span style="font-size:100%">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:100%">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 16. 510(K) SUMMARY
#### Submitter:
Hong Koo Yeo Biotem Co., Ltd. 57, Sasang-ro , 401 beon-gil , Sasang-gu, Busan , Korea Tel : +82-70-8220-4934 Fax : +82-51-908-8257
#### Official Correspondent/US Agent:
Jovce Bang Provision Consulting Group, Inc. 1370 Valley Vista Suite 200 Phone: +1-909-550-0131 EXT.1131 Date Prepared: January 26, 2018
#### Device Information:
Device Name: AR_N Type Implant System Classification Name: Endosseous Dental Implant Common Name: Endosseous Dental Implant Classification: Class II Product Code: DZE Subsequent Product Code: NHA Regulation number: 21 CFR 872.3640
#### Predicate devices
• CMI IS Implant System (K113554) manufactured by Neobiotech Implant Co., Ltd.
#### Indication for use
AR N Type Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, screw retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. AR_N Type Implant System is for two stage surgical procedures. It is intended for delayed loading.
#### Device Description
The AR_N Type Implant System is a dental implant system made of CP TiGr 4 per ASTM F67, intended to be surgically placed in the bone of the upper or lower jaw arches for loading after a conventional healing period.
The implants may be used to replace one or more missing teeth. The surface of the implants has been treated with RBM (Resorbable Blast Media). The AR_N Type Implant System is offered in the following sizes.
| Platform | Body Diameter | Total Length (mm) |
|----------|---------------|----------------------------------|
| Ø 3.7 | Ø 3.5 | 7.5, 8.5, 10.0, 11.5, 13.0, 15.0 |
| Ø 4.2 | Ø 4.0 | 7.5, 8.5, 10.0, 11.5, 13.0, 15.0 |
| Ø 4.6 | Ø 4.5 | 7.5, 8.5, 10.0, 11.5, 13.0, 15.0 |
| Ø 5.1 | Ø 5.0 | 7.5, 8.5, 10.0, 11.5, 13.0, 15.0 |
{4}------------------------------------------------
## Abutment
Each product line includes corresponding abutments in multiple designs (Cover Screws, Healing Abutments, Angled Abutments, Cemented Abutments (Hex, Non-hex), and Solid Abutments) for restorations. Small diameter implants and angled abutments are not recommended for the posterior region. They are made of Titanium Grade 4 / ASTM F67, and provided as non-sterile.
| Abutments | Connection | Angulation (°) | Sizes | |
|-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------|-------------------|-----------------------|
| | | | Cuff Heights (mm) | Platform diameter (Ø) |
| Angled<br>abutment | Assembled with Fixture and<br>Angled Abutment (Hex) by<br>inserting Abutment body with set<br>screw. The connection type is<br>internal Hexagon shape. | 15, 25 | 2, 4 | 4.5~5.8 |
| Cemented<br>Abutment<br>(Hex) | Assembled with fixture and<br>cemented abutment (Hex) by<br>inserting abutment body with set<br>screw. The connection type is<br>Internal Hexagon shape | 0 | 1~6 | 4.5~6.5 |
| Cemented<br>Abutment<br>(Non-Hex) | Assembled with Fixture and<br>Cemented Abutment (Non-Hex)<br>by inserting Abutment body with<br>set screw. The connection type is<br>internal Non-Hexagon shape. | 0 | 1~6 | 4.5~6.5 |
| Healing<br>Abutment | Assembled with Fixture and<br>Healing Abutment by inserting the<br>screw part. And the connection is<br>internal Hexagon shape | 0 | 1~5 | 4.5~6.5 |
| Solid<br>Abutment | Assembled with fixture by<br>inserting set screws. The<br>connection type is internal<br>Hexagon shape. | 0 | 1~6 | 4.5~6.5 |
| Cover Screw | Assembled with Fixture by<br>inserting screw part.<br>And the connection is internal<br>Hexagon shape. | 0 | 2.7 | 2.8 -3.3 |
| Set screw | Connection with abutment with<br>screw | 0 | 10 (length) | 2.1, 2.5 |
## Substantial Equivalence Comparison Chart
| | Subject Device | Predicate Device |
|---------------|--------------------------|-----------------------|
| 510(k) Number | K171297 | K113554 |
| Device Name | AR_N Type Implant System | CMI IS Implant System |
| Manufacturer | Biotem Co., Ltd. | Neobiotech Co., Ltd. |
{5}------------------------------------------------
| | | Principles of Operation | Subject device is an endosseous<br>implants and are integrated in the<br>bone (osseointegration). They are<br>intended to be used for surgical<br>placement in the upper or lower jaw<br>to provide a means for prosthetic<br>attachment for anchoring or<br>supporting tooth replacements to<br>restore chewing function. | Predicate device is an endosseous<br>implants and are integrated in the<br>bone (osseointegration). They are<br>intended to be used for surgical<br>placement in the upper or lower jaw<br>to provide a means for prosthetic<br>attachment for anchoring or<br>supporting tooth replacements to<br>restore chewing function. |
|---------------------|--------|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | | | AR_N Type Implant System is<br>indicated for use in partially or fully<br>edentulous mandibles and maxillae,<br>in support of single or multiple-unit<br>restorations including; cemented<br>retained, screw retained, or<br>overdenture restorations, and<br>terminal or intermediate abutment<br>support for fixed bridgework. AR_N<br>Type Implant System is for two stage<br>surgical procedures. It is intended for<br>delayed load. | The CMI Implant IS System is<br>indicated for use in partially or fully<br>edentulous mandibles and maxillae, in<br>support of single or multiple-unit<br>restorations including; cemented<br>retained, screw retained, or<br>overdenture restorations, and<br>terminal or intermediate abutment<br>support for fixed bridgework. IS<br>System is dedicated for two stage<br>surgical procedures and for<br>immediate loading when there is<br>good primary stability and an<br>appropriate occlusal load. Also,<br>implants with diameters larger than 5<br>min are indicated for molar regions. |
| | | | | |
| | | | | |
| | | | | |
| | | | | |
| | | | | |
| | | Design | Internal Hex Submerged<br>Macro thread | Internal Hex Submerged<br>Macro thread |
| | | Material | CP Ti Gr 4<br>ASTM F67 | CP Ti Gr 4<br>ASTM F67 |
| | | Fixture Sterilization | Gamma sterilization | Gamma sterilization |
| | | Fixture Diameter (mm) | Ø3.7 - 5.1 | Ø3.5 - 8.0 |
| | | Fixture Length (mm) | 7.5 - 15.0 | 7.0 - 15.0 |
| | | Surface treatment | RBM | RBM |
| | | Diameters | Ø4.5 - 6.5mm | Ø3.5 - 6.5 |
| | | Lengths | 4.0 - 9.0mm | 4.0-9.0mm |
| Abutment | Angled | | 15°, 25° | 15°,25° |
| Abutment material | | | CP Ti Gr 4<br>ASTM F67 | CP Ti Gr 4<br>ASTM F67 |
| | | | | |
| | | Abutment sterilization | Non-sterile | Non-sterile |
| Product Code | | | DZE, NHA | DZE, NHA |
{6}------------------------------------------------
| | Subject abutments<br>Biotem Product (AR N type) | | | Predicate abutments<br>NEO BIOTECH Product (IS-II type) | | |
|-----------------------------------|-------------------------------------------------|----------------------|----------------|---------------------------------------------------------|----------------------|----------------|
| | Platform<br>diameter (Ø) | Cuff heights<br>(mm) | Angulation (°) | Platform<br>diameter (Ø) | Cuff heights<br>(mm) | Angulation (°) |
| Healing<br>abutment | 4.5~6.5 | 1~5 | 0 | 4.0~6.0 | 2~7 | 0 |
| Cemented<br>abutment<br>(hex) | 4.5~6.5 | 1~6 | 0 | 4.5~6.5 | 1~4 | 0 |
| Cemented<br>abutment<br>(non-hex) | 4.5~6.5 | 1~6 | 0 | 4.5~6.5 | 1~4 | 0 |
| Solid<br>abutment | 4.5~6.5 | 1~6 | 00 | 4.5~6.5 | 1~7 | 0 |
| Angled<br>abutment | 4.5~5.8 | 2,4 | 15, 25 | 4.5~5.7 | 2,3,4 | 15, 25 |
## Basis for Substantial Equivalence
The AR N Type Implant System has a substantially equivalent intended use as the identified predicate. Both are used for mandible and maxilla endosseous dental implant and accessories. The AR N Type Implant System is similar in fundamental scientific technology to the predicate device in that they all have been designed, manufactured and tested in compliance with FDA's Class II special controls guidance document root-form endosseous dental implants and endosseous dental implant abutments. They all share same internal hexagon submerged connection system. The subject and predicate devices are both bone-level implants that share same tapered body design and cutting edge. The subject and predicate devices are similar in size, same surface treatment and materials.
## Non-Clinical Test data
The subject device was tested to evaluate its performance as below.
- Sterilization Validation testing for sterile devices (fixtures) has been performed in accordance with ISO 11137, SIO 11737-1& ISO 11737-2 for gamma sterilization
- Steam Sterilization validation for non-sterile devices (abutments) has been performed in accordance with ANSI/AAMI ST79.
- Surface Characteristics Test Report Chemical and SEM image analyses have been performed to verify that there is no residual after RBM treatment on the fixtures.
- Comparative Fatigue testing of the subject device and predicate device has been performed in accordance with ISO 14801:2007.
- Cytotoxicity Test performed according to ISO 10993-5:2009
- · Sensitization test performed according to ISO 10993-10:2010
Those tests have been performed to evaluate the substantial equivalence in the surface characteristics compared to the predicate device. The result of the above tests have met the standard, and proved the substantial equivalence with the predicate device. Non-clinical testing consisted of performance of testing in accordance with the FDA guidance "Class II Special Controls Guidance Document Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments." The results of the non-clinical testing demonstrate that the subject device is substantially equivalent to the predicate device.
## Conclusions
Overall, the AR N Type Implant System has the following similarities to the predicate devices:
{7}------------------------------------------------
*has the same intended use,
*uses the same operating principle,
*incorporates the same basic design,
*incorporates the same material and the surface treatment.
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification, we conclude that the AR_N Type Implant System is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.