K171293 · Genesis Fracture Care, Inc. · HRS · Oct 23, 2017 · Orthopedic
Device Facts
Record ID
K171293
Device Name
G3™ Active Plate® Miniature System
Applicant
Genesis Fracture Care, Inc.
Product Code
HRS · Orthopedic
Decision Date
Oct 23, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The G3™ Active Plate® Miniature System is intended for fracture fixation in patients 18 years of age and older requiring stabilization of mal-unions, osteotomies and repair of closed and open fractures in small bones. The system is indicated for the fixation of small bone fractures including, but not limited to, fractures of the hand, foot, wrist and ankle.
Device Story
G3 Active Plate Miniature System is a bone fixation device for small bones (hand, foot, wrist, ankle). System consists of 8-hole straight plates and 1.5mm diameter locking screws. Plates feature internal link elements embedded in a silicone elastomer sheath; once screws are inserted, these elements allow independent, controlled axial translation of the screws. Used by orthopedic surgeons in clinical settings for fracture stabilization. Output is mechanical fixation of bone fragments. Benefits include controlled axial translation to facilitate fracture healing. Device is purely mechanical; no software or electronic components.
Clinical Evidence
Bench testing only. Mechanical testing included axial, torsional, and shear stiffness, and four-point bend testing per ASTM F382-08. Engineering analysis compared torsional strength, axial pullout strength, and insertion/removal torque against predicate devices.
Technological Characteristics
Materials: Titanium alloy (Ti6Al4V-ELI), cobalt chrome alloy (Co-28Cr-6Mo), and silicone elastomer. Components: 8-hole straight plates and 1.5mm diameter locking screws (8-24mm lengths). Features: Internal link elements in elastomer sheath for controlled axial translation. Mechanical testing per ASTM F382-08.
Indications for Use
Indicated for patients 18+ years old requiring stabilization of mal-unions, osteotomies, and repair of closed/open fractures in small bones, including hand, foot, wrist, and ankle.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Hand Innovations (Zimmer Biomet Inc.) Mini Fragment Plates (K061748)
DePuy Synthes Variable Angle Locking Hand System (K141527)
DePuy Synthes - Small Bone Locking Plating System (K083364)
Wright Medical - WMT Implantable K-wires (K132895)
Reference Devices
Genesis Fracture Care (Zimmer Biomet Inc.) G3™ Active Plate® Large Fragment System (K152242)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue.
Genesis Fracture Care, Inc. % Christine Scifert Managing Partner MRC-X, LLC 6075 Poplar Ave Suite 500 Memphis, Tennessee 38119 October 23, 2017
## Re: K171293
Trade/Device Name: G3TM Active Plate® Miniature System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS. HWC Dated: September 18, 2017 Received: September 21, 2017
Dear Christine Scifert:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.qov
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
## 510(k) Number (if known)
K171293
#### Device Name
G3TM Active Plate® Miniature System
#### Indications for Use (Describe)
The G3™ Active Plate® Miniature System is intended for fracture fixation in patients 18 years of age and older requiring stabilization of mal-unions, osteotomies and repair of closed and open fractures in small bones. The system is indicated for the fixation of small bone fractures including, but not limited to, fractures of the hand, foot, wrist and ankle.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
K171293 Page 1/3
# 510(k) Summary G3™ Active Plate® Miniature System September 18, 2017
| Company: | Genesis Fracture Care, Inc.<br>13568 SE 97th Ave<br>Suite 202<br>Clackamas, OR 97015<br>503-528-4048<br>503-413-5216 (fax) |
|--------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Primary Contact: | Christine Scifert |
| Company Contact: | Michael Bottlang |
| Trade Name: | G3™ Active Plate® Miniature System |
| Common Name: | Plate, Fixation, Bone<br>Screw, Fixation, Bone |
| Classification: | II |
| Regulation Number: | 888.3030 - Single/multiple component metallic bone fixation appliances<br>and accessories<br>888.3040 - Smooth or threaded metallic bone fixation fastener |
| Panel: | 87-Orthopedic |
| Product Code(s): | HRS, HWC |
The G3™ Active Plate® Miniature System is a straight plate and locking screw Device Description: system intended to be used for fracture fixation of the hand and foot. All implantable components are manufactured from medical grade titanium alloy (Ti6Al4V-ELI), medical grade cobalt chrome alloy (Co-28Cr-6Mo) and silicone elastomer in the subject device.
The plate has 8 holes. The plates incorporate link elements, which are constrained within the plate and embedded in an elastomer sheath that is bonded to both the link element. Once locking screws are inserted, the active elements allow for independent controlled axial translation of the screws. The screws are 1.5 mm diameter and come in lengths ranging from 8 mm to 24 mm.
{4}------------------------------------------------
Indications for Use: The G3™ Active Plate® Miniature System is intended for fracture fixation in patients 18 years of age and older requiring stabilization of mal-unions, osteotomies and repair of closed and open fractures in small bones. The system is indicated for the fixation of small bone fractures including, but not limited to, fractures of the hand, foot, wrist and ankle.
Substantial Equivalence: The subject components were demonstrated to be substantially equivalent to the following systems previously cleared by the FDA:
Primary Predicate
- Hand Innovations (Zimmer Biomet Inc.) Mini Fragment Plates (K061748) ●
Additional Predicates
- DePuy Synthes Variable Angle Locking Hand System (K141527) ●
- DePuy Synthes - Small Bone Locking Plating System (K083364)
- Small Bone Innovations SBI K-Wires (K051605) ●
- . Wright Medical - WMT Implantable K-wires (K132895)
Reference Devices
- Genesis Fracture Care (Zimmer Biomet Inc.) G3™ Active Plate® Large Fragment . System (K152242)
The subject G3™ Active Plate® Miniature System has demonstrated to be substantially equivalent to the predicates as the products are similar in indications, materials, geometry, and active feature.
| | Subject of Present 510(k):<br>G3 TM Active Plate® Miniature Plate | Predicate Devices |
|--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|
| Intended Use/ Indications<br>for Use | The G3 TM Active Plate® Miniature<br>System is intended for fracture fixation<br>in patients 18 years of age and older<br>requiring stabilization of mal-unions,<br>non-unions, osteotomies and repair of<br>closed and open fractures in small<br>bones. The system is indicated for the<br>fixation of small bone fractures | Inclusive |
{5}------------------------------------------------
| | including, but not limited to, fractures of the hand, foot, wrist and ankle. | |
|-------------------------|-------------------------------------------------------------------------------------------------|-----------|
| Materials | Titanium Alloy (Plate)<br>Silicone – NuSil (Plate)<br>Cobalt Chrome (Screws) | Identical |
| Geometry and Dimensions | Plates: 8 holes<br>Length: 25 mm<br>Locking Screws:<br>Diameter: 1.5mm<br>Lengths: 8 mm – 24 mm | Inclusive |
Performance Testing: Mechanical testing including axial, torsional and shear stiffness and fourpoint bend per ASTM F382-08 was performed on the subject G3™ Active Plate® Miniature System and the results show the subject plates to be substantially equivalent to the predicate devices. Engineering analysis was used to compare the torsional strength, axial pullout strength and insertion/removal torque of the subject screws to the predicates.
## Conclusion
Based on the test results and the comparison to the predicate devices, the subject device is determined to be substantially equivalent to the predict devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.