# Vela RX PTA Balloon Dilatation Catheter (K171290)

_Curatia Medical · LIT · Jul 19, 2017 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K171290

## Device Facts

- **Applicant:** Curatia Medical
- **Product Code:** [LIT](/productcode/LIT.md)
- **Decision Date:** Jul 19, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1250
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The Vela RX PTA Balloon Dilatation Catheter is intended for use in percutaneous transluminal angioplasty (PTA) of the femoral, popliteal, infra popliteal and renal arteries.

## Device Story

Vela RX PTA Balloon Dilatation Catheter is a 0.014" rapid exchange co-axial catheter used for percutaneous transluminal angioplasty. Device consists of proximal hub/strain relief, transition tube, and distal tube with a Pebax low-compliant balloon. Two radiopaque markers facilitate positioning under fluoroscopy. Physician operates device in clinical setting to dilate stenotic peripheral arteries. Balloon expands to known diameter/length at 6 atm. Hydrophilic coating on distal shaft aids navigation. Output is mechanical dilation of target vessel; clinical benefit is improved blood flow in peripheral arteries.

## Clinical Evidence

Bench testing only. No clinical data. Performance testing included dimension, inflation/deflation time, flexibility, kink/torque strength, balloon fatigue, compliance, RBP, shaft integrity, bond tensile strength, and radiopacity. Biocompatibility testing performed per ISO 10993 (cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, hemocompatibility, genotoxicity).

## Technological Characteristics

Materials: Nylon/Pebax. Design: 0.014" rapid exchange co-axial. Balloon: Low-compliant Pebax, 1.5-4.0mm diameter, 10-40mm length. Markers: Dual Pt/Ir. Coating: Hydrophilic distal shaft. Sterilization: Ethylene oxide per ISO 11135. Standards: ISO 10555-1, ISO 10555-4, ISO 10993 series.

## Regulatory Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

## Predicate Devices

- Ultraverse RX PTA Balloon Dilatation Catheter ([K131199](/device/K131199.md))

## Reference Devices

- Navajo PTA Catheter ([K141354](/device/K141354.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol. The logo is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 19, 2017

Curatia Medical Co. % Mr. Bill Jacqmein Regulatory Affairs Consultant JCO Consulting 11218 Zest Court Blaine, MN 55449

Re: K171290

Trade/Device Name: Vela RX PTA Balloon Dilatation Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: LIT Dated: April 28, 2017 Received: May 2, 2017

Dear Mr. Jacqmein:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR

{1}------------------------------------------------

Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

# Kenneth J. Cavanaugh -S

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K171290

Device Name

Vela RX PTA Balloon Dilatation Catheter

#### Indications for Use (Describe)

The Vela RX PTA Balloon Dilatation Catheter is intended for use in percutaneous transluminal angioplasty (PTA) of the femoral, popliteal, infra popliteal and renal arteries.

| Type of Use (Select one or both, as applicable)  |                                             |
|--------------------------------------------------|---------------------------------------------|
| X   Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) SUMMARY

A 510(k) summary in accordance with the requirements of 21 CFR 807.92.

| Submitter:                   | Curatia Medical Co.<br>3227 Kifer Road<br>Santa Clara, CA 95051<br>Establishment Number: 3008361782 |                                                                             |
|------------------------------|-----------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|
| Company Contact<br>Person:   | Jessica Chiu, CEO                                                                                   |                                                                             |
|                              | Phone:<br>Fax:<br>Email:                                                                            | (408) 414-2188<br>(408) 413-3000<br>jchiu@curatiamed.com                    |
| Submission<br>Correspondent: | Bill Jacqmein, Regulatory Affairs Consultant                                                        |                                                                             |
|                              | Address:<br>Phone:<br>Email:                                                                        | 11218 Zest Court, Blaine, MN 55449<br>(404) 216-6190<br>bjacqmein@gmail.com |
| Date Prepared:               | March 2017                                                                                          |                                                                             |
| Proprietary Name:            | Vela RX PTA Balloon Dilatation Catheter                                                             |                                                                             |
| Common Name:                 | Percutaneous catheter                                                                               |                                                                             |
| Product Code:                | LIT – Catheter, Angioplasty, Peripheral, Transluminal                                               |                                                                             |
| Device Classification:       | Class II, 21 CFR 870.1250 - Percutaneous Catheter                                                   |                                                                             |
| Predicate Devices:           | Ultraverse RX PTA Balloon Dilatation Catheter (K131199                                              |                                                                             |
| Reference Device:            | Navajo PTA Catheter (K141354)                                                                       |                                                                             |

#### Device Description:

The Vela RX PTA Balloon Dilatation Catheter has a 0.014" (0.36mm) rapid exchange co-axial design, which consists of proximal tube with hub and strain relief, transition tube, and distal tube with a balloon at the distal tip. The balloon is made of a Pebax low-compliant material and will expand to a known diameter and length at 6 atmospheres pressure. Two radiopaque markers under the balloon aid in positioning the balloon working length under the fluoroscopy. The nominal inflated balloon diameters range from 1.5mm to 4.0mm with balloon working lengths of 10mm to 40mm. The effective catheter length is 140cm, and is provided with a hydrophilic coating on the distal catheter shaft. The maximum compatible guide wire to be used with the Vela RX PTA Balloon Dilatation Catheter is 0.014" (0.36mm), and the compatible introducer sheath size is 4F.

{4}------------------------------------------------

#### Figure 1: Detailed Device Description

Image /page/4/Figure/2 description: This image is a technical drawing of a medical device, possibly a catheter or similar instrument, with several dimensions labeled. The total length of the device is 140cm, with the inflatable balloon section measuring 26cm. The diameter of the shaft varies, with one section measuring 0.037"/0.035" (2.8F/2.7F) and another measuring 0.027" (2.1F), and a 10cm section is also labeled.

#### Indications for Use:

The Vela RX PTA Balloon Dilatation Catheter is intended for use in percutaneous transluminal angioplasty (PTA) of the femoral, popliteal, infra popliteal and renal arteries.

## Comparison to Predicate Devices:

Vela RX PTA Balloon Dilatation Catheter is functionally equivalent to the following predicate device: Ultraverse RX PTA Balloon Dilatation Catheter (Bard Peripheral Vascular, K131199 cleared May 30, 2013).

The following table demonstrates the functional specifications of Vela RX PTA Balloon Dilatation Catheter are substantially equivalent to the predicate devices.

| Functional<br>Specification       | Vela RX PTA<br>Balloon Dilatation<br>Catheter                                                    | Ultraverse RX PTA<br>Dilation Catheter                  | Comparison Result                                                                                                                                                                                                                                                    |
|-----------------------------------|--------------------------------------------------------------------------------------------------|---------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Balloon and<br>Catheter Materials | Nylon/Pebax                                                                                      | Nylon/Pebax                                             | Same                                                                                                                                                                                                                                                                 |
| Catheter Design                   | Distal OTW w. RX<br>wire exchange,<br>Proximal hypotube                                          | Distal OTW w. RX<br>wire exchange,<br>Proximal hypotube | Same                                                                                                                                                                                                                                                                 |
| Min. Sheath Size                  | 4F                                                                                               | 4-5F                                                    | Same                                                                                                                                                                                                                                                                 |
| Max. Guide Wire<br>Size           | 0.014"                                                                                           | 0.014"                                                  | Same                                                                                                                                                                                                                                                                 |
| Catheter Length                   | 140cm                                                                                            | 80, 150, 200 cm                                         | Within range                                                                                                                                                                                                                                                         |
| Balloon OD                        | 1.5-4.0mm                                                                                        | 1.25, 1.5-4.0mm                                         | Within range                                                                                                                                                                                                                                                         |
| Balloon Length                    | 10,20,30,40mm                                                                                    | 15,20,40,60,<br>120,200,250mm                           | 5mm shorter than the<br>predicate device. But within<br>other similar products' spec.<br>(See table below)<br>Additionally, shorter balloons<br>sizes are less risk because<br>they do not cover/dilate as<br>much healthy tissue as a<br>larger balloon size would. |
| Balloon Marker                    | Dual Pt/Ir Markers                                                                               | Dual Pt/Ir Markers                                      | Same                                                                                                                                                                                                                                                                 |
| Nominal Pressure                  | 6atm                                                                                             | 6atm                                                    | Same                                                                                                                                                                                                                                                                 |
| Balloon RBP                       | 16atm for 3.5-<br>4.0mm balloons;<br>18atm for 2.5-<br>3.0mm balloons<br>20atm for 1.5-<br>2.0mm | 15-16atm                                                | RBP of 2 atm higher than its<br>predicate devices in some<br>balloon sizes, but within other<br>similar products' spec. (See<br>table below)                                                                                                                         |

Table 1: Functional Specification Comparison

{5}------------------------------------------------

| Indication for Use | For use in<br>percutaneous<br>transluminal<br>angiogplasty (PTA)<br>of the femoral,<br>popliteal, infra<br>popliteal and renal<br>arteries. | For use in<br>Percutaneous<br>Transluminal<br>Angioplasty of the<br>renal, femoral,<br>popliteal, infra-<br>popliteal, tibial, and<br>peroneal arteries.<br>This catheter is not<br>for use in coronary<br>arteries. | Similar |
|--------------------|---------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
|--------------------|---------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|

In addition to the Ultraverse, the Curatia Navajo PTA catheter is also provided as a reference.
Please see the comparison table below.

| Functional Specification | Vela RX (CML)                                                                                                                      | Navajo PTA<br>(CML)                                                                                                                                                                                                                                                       |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| FDA registration No.     | K171290                                                                                                                            | K141354                                                                                                                                                                                                                                                                   |
| Min. Sheath Size         | 4F                                                                                                                                 | 5F                                                                                                                                                                                                                                                                        |
| Max. Guide Wire Size     | 0.014"                                                                                                                             | 0.035"                                                                                                                                                                                                                                                                    |
| Catheter Length          | 140 cm                                                                                                                             | 80 to 130 cm                                                                                                                                                                                                                                                              |
| Balloon OD               | 1.5-4.0 mm                                                                                                                         | 5 to 10 mm                                                                                                                                                                                                                                                                |
| Balloon Length           | 10,20,30,40 mm                                                                                                                     | 40 to 80 mm                                                                                                                                                                                                                                                               |
| Nominal Pressure         | 6 atm                                                                                                                              | 6 to 8 atm                                                                                                                                                                                                                                                                |
| Balloon RBP              | 20atm: 1.5-2.0mm<br>18 atm: 2.5-3.0 mm<br>16 atm: 3.5-4.0 mm                                                                       | 14 to 16 atm                                                                                                                                                                                                                                                              |
| Indication for Use       | For use in percutaneous<br>transluminal angiogplasty<br>(PTA) of the femoral,<br>popliteal, infra popliteal<br>and renal arteries. | The catheter is intended to dilate<br>stenoses in the peripheral arteries (iliac,<br>femoral, ilio-femoral, popliteal, infra<br>popliteal, renal arteries); and for<br>the treatment of obstructive lesions of<br>native or synthetic arteriovenous dialysis<br>fistulae. |

{6}------------------------------------------------

#### Comparison of Indications for Use to Predicate Devices:

Based on the above comparison, the indications for use of the Vela RX PTA Balloon Dilatation Catheter is similar to that of the Ultraverse. Therefore, the Vela RX PTA Balloon Dilatation Catheter can be considered substantially equivalent to its predicate device.

#### Comparison of Technological Characteristics to Predicate Devices:

Based on the above comparison, the design, construction, and performance characteristics of the Vela RX PTA Balloon Dilatation Catheter is similar to that of Ultraverse. Therefore, the Vela RX PTA Balloon Dilatation Catheter can be considered substantially equivalent to its predicate devices.

## Summary of Performance Data and Substantial Equivalence:

The Vela RX PTA Balloon Dilatation Catheters were designed and verified in accordance with the risk analysis and product requirements. All tests confirmed the products met the pre-defined acceptance criteria. The Vela RX PTA Balloon Dilatation Catheters have been tested and shown to be compliant with the following standards documents:

- . ISO 10555-1:2013- Intravascular catheters - Sterile, single-use intravascular catheters -Part 1: General requirements
- . ISO 10555-4:2013- Intravascular catheters - Sterile, single-use intravascular catheters -Part 4: Balloon dilatation catheter
- . ISO 10993-1:2009- Biological evaluation of medical devices - Part 1: Evaluation and testina
- ISO 10993-3:2014- Biological evaluation of medical devices Part 3: Tests for . genotoxicity, carcinogenicity and reproductive toxicity
- . ISO 10993-4:2002+AMD1:2006-Biological testing of medical and dental materials and devices - Part 4: Selection of tests for interactions with blood
- ISO 10993-5: 2009- Biological evaluation of medical devices Part 5: Tests for in vitro . cytotoxicity
- ISO 10993-10:2010 Biological evaluation of medical devices Part 10: Tests for irritation ● and skin sensitization
- ISO 10993-11:2006- Biological testing of medical and dental materials and devices Part . 11: Tests for systemic toxicity
- . ISO 10993-12:2012- Biological testing of medical and dental materials and devices – Part 12: Sample Preparation And Reference Materials
- . ISO 11135: 2014 Sterilization of health-care products-Etylene oxide-Requirements for the development, validation and routine control of a sterilization process for medical devices
- . EN ISO 11607-1: 2009(R2014)- Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems

{7}------------------------------------------------

## Performance Testing:

To demonstrate substantial equivalence of Vela RX PTA Balloon Dilatation Catheter to the predicate devices, the technological characteristics and performance criteria were evaluated using the bench testing recommendations outlined in the FDA Guidance Document "Class II Special Controls Guidance Document for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters" dated September 8, 2010. The following performance tests were completed:

- Dimension Catheter Preparation, In/deflation Time Catheter Flexibility and Kink Torque Strength Balloon Fatigue (w/ & w/o stent) Balloon Compliance Balloon Rated Rupture Pressure (w/ & w/o stent ) Catheter Shaft Pressure Integrity GW Lumen Collapse Pressure Catheter Bond Tensile Strength Catheter Shaft Coating Integrity Radiopacity
The results of these tests demonstrate the technological characteristics and performance criteria of the Vela RX PTA Balloon Dilatation Catheter are adequate for its intended use, and is substantially equivalent to the predicate devices.

## Biocompatibility:

To demonstrate the biological safety of the body-contacting materials and substantial equivalence of the Vela RX PTA Balloon Dilatation Catheter to its predicate devices, the following biocompatibility testing was performed in accordance with ISO 10993 and the FDA Guidance Document "Use of International Standard ISO 10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing"" dated June, 2016:

Cytotoxicity Sensitization Irritation (Intracutaneous Reactivity) Systemic Toxicity (acute) Pvrogenicitv Hemocompatibility (Hemolysis, Thrombogenicity, and Immunology) Genotoxicity (Bacterial Reverse Mutation Assay, Mouse Lymphoma Mutagenesis Assay)

The results from these tests demonstrate that Vela RX PTA Balloon Dilatation Catheter is biocompatible for its intended use.

#### Conclusion:

Based on comparison of indications for use, technological characteristics, safety and performance testing, biocompatibility testing, the Vela RX PTA Balloon Dilatation Catheter has been shown to be appropriate for its indications for use and is substantially equivalent to the legally marketed predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K171290](https://fda-staging.innolitics.com/device/K171290)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).