epoc Blood Urea Nitrogen Test, epoc Total Carbon Dioxide Test
K171247 · Epocal, Inc. · CDS · Jan 17, 2018 · Clinical Chemistry
Device Facts
Record ID
K171247
Device Name
epoc Blood Urea Nitrogen Test, epoc Total Carbon Dioxide Test
Applicant
Epocal, Inc.
Product Code
CDS · Clinical Chemistry
Decision Date
Jan 17, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1770
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K171247 · Jan 17, 2018
epoc Blood Urea Nitrogen Test, epoc Total Carbon Dioxide Test
Epocal, Inc.
Patient blood samples (venous, arterial, capillary) collected at clinical sites
Patient samples were used to perform method comparison studies against laboratory reference systems and predicate devices to validate the performance of the epoc Blood Analysis System.
Method comparison; Clinical site testing; Patient samples
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
BUN Method Comparison Study; Method comparison study per CLSI EP09-A3
Venous, arterial, and capillary blood samples; Sample Size: 433; Number of Sites: 3
TCO2 Method Comparison Study; Method comparison study
Venous, arterial, and capillary patient samples; Sample Size: 574; Number of Sites: 3
i-STAT-CHEM8+ (whole blood point-of-care system)
TCO2 concentration accuracy/bias
Indications for Use
The Blood Urea Nitrogen and Total Carbon Dioxide tests, as part of the epoc Blood Analysis System, is intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of heparinized or unanticoagulated arterial, venous or capillary whole blood in the laboratory or at the point of care. Blood Urea Nitrogen measurements from the epoc Blood Analysis System are used in the diagnosis and treatment of certain renal and metabolic diseases. Total Carbon Dioxide measurements from the epoc Blood Analysis System are used in the diagnosis and treatment of disorders associated with changes in body acid-base balance.
Device Story
The epoc Blood Analysis System is a point-of-care diagnostic platform consisting of a handheld reader, host device, and single-use BGEM test cards. This submission adds BUN and TCO2 sensors to the existing card. The BUN sensor uses an enzymatic potentiometric method where urease hydrolyzes urea into ammonium ions, which are detected by an ion-selective electrode. The TCO2 test is calculated using the Henderson-Hasselbalch equation based on measured pH and pCO2, with calibration traceable to the IFCC reference method. The operator inserts a blood sample into the test card, which is then inserted into the reader for automated analysis. Results are displayed on the host device, allowing clinicians to make immediate decisions regarding renal function and acid-base status. The system provides rapid, bedside diagnostic information, reducing turnaround time compared to central laboratory testing.
Clinical Evidence
No clinical trials were performed. Evidence consists of analytical bench testing, including precision (CLSI EP05-A3), linearity (CLSI EP06-A), detection limits (CLSI EP17-A2), and interference studies (CLSI EP07-A2). Method comparison studies were conducted at 3 point-of-care sites comparing the epoc system against the Roche Cobas 8000 (BUN) and i-STAT CHEM8+ (TCO2). Results showed high correlation (R² > 0.94) across venous, arterial, and capillary samples. Matrix comparison confirmed suitability of lithium heparin, sodium heparin, and un-anticoagulated blood.
Technological Characteristics
The system uses an electrochemical multi-sensor array on a single-use test card. BUN sensing is enzymatic potentiometric (urease-based). TCO2 is calculated via the Henderson-Hasselbalch equation using measured pH and pCO2. The system is a handheld, point-of-care device. Calibration is performed using NIST-traceable standards. The device is designed for use with the epoc Reader and Host. Sterilization and specific material standards are not detailed beyond general laboratory equipment safety (IEC 61010-1).
Indications for Use
Indicated for trained medical professionals to perform quantitative testing of heparinized or un-anticoagulated arterial, venous, or capillary whole blood for BUN and TCO2 levels in laboratory or point-of-care settings. Used for diagnosis and treatment of renal, metabolic, and acid-base balance disorders.
Regulatory Classification
Identification
A urea nitrogen test system is a device intended to measure urea nitrogen (an end-product of nitrogen metabolism) in whole blood, serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of certain renal and metabolic diseases.
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k171247
B. Purpose for Submission:
Addition of blood urea nitrogen (BUN) and total carbon dioxide (TCO₂) tests to a previously cleared device
C. Measurand:
Blood urea nitrogen and total carbon dioxide
D. Type of Test:
Quantitative, electromechanical biosensor for TCO₂
Enzymatic potentiometric-based sensing for BUN
E. Applicant:
Epocal, Inc.
F. Proprietary and Established Names:
epoc Blood Urea Nitrogen Test
epoc Total Carbon Dioxide Test
G. Regulatory Information:
1. Regulation section:
21 CFR 862.1770 Urea nitrogen test system
21 CFR 862.1160 Bicarbonate/carbon dioxide test system
2. Classification:
Class II
3. Product code:
CDS - Electrode, Ion Specific, Urea Nitrogen
JFL - pH Rate Measurement, Carbon-Dioxide
{1}
4. Panel:
Clinical Chemistry (75)
H. Intended Use:
1. Intended use(s):
See Indication for use below.
2. Indication(s) for use:
The Blood Urea Nitrogen and Total Carbon Dioxide tests, as part of the epoc Blood Analysis System, is intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous or capillary whole blood in the laboratory or at the point of care.
Blood Urea Nitrogen measurements from the epoc Blood Analysis System are used in the diagnosis and treatment of certain renal and metabolic diseases.
Total Carbon Dioxide measurements from the epoc Blood Analysis System are used in the diagnosis and treatment of disorders associated with changes in body acid-base balance.
3. Special conditions for use statement(s):
For prescription use and point-of-care use.
For in vitro diagnostic use only.
4. Special instrument requirements:
epoc Blood Analysis System
I. Device Description:
The epoc BUN test is being added as an additional sensor to the existing single use test card that is used with the epoc Blood Analysis System. The epoc TCO₂ test is being added based on the standard Henderson-Hasselbalch equation (using the measured pCO₂ and pH values); this test is metrologically traceable to the IFCC TCO₂ reference method. The test card is inserted into the epoc Reader and all analytical steps are performed automatically. The epoc Blood Analysis System is an in vitro diagnostic device system for the quantitative testing of blood gases, electrolytes, and metabolites in venous, arterial, and capillary whole blood samples. The epoc System is comprised of 3 major subsystems: epoc Host, epoc Reader and epoc BGEM Test Card. The main accessory used with the epoc System includes the epoc Care-Fill Capillary Tubes used to collect and introduce capillary blood samples into the epoc Test Card.
2
{2}
The epoc test card panel configuration currently includes sensors for the determination of pH, pCO₂, pO₂, Na, K, iCa, Cl, Glu, lactate, creatinine, and hematocrit in arterial, venous, and capillary blood samples cleared previously in k061597, k090109, k092849, k093297, and k113726. This premarket notification adds blood urea nitrogen (BUN) and total carbon dioxide (TCO₂) quantitation to the epoc BGEM Test Card and Blood Analysis System.
## J. Substantial Equivalence Information:
1. Predicate device name(s):
i-STAT Chem8+ Cartridge (with i-STAT Portable Clinical Analyzer)
2. Predicate 510(k) number(s):
k053110
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Candidate Device: epoc Blood Urea Nitrogen test and epoc Total Carbon Dioxide test (k171247) | Predicate Device: BUN and TCO₂ Tests using the i-STAT Chem8+ Cartridge (with i-STAT Portable Clinical Analyzer) (k053110) |
| Intended use | Intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous or capillary whole blood in the laboratory or at the point of care.
BUN are used in the diagnosis and treatment of certain renal and metabolic diseases.
TCO₂ measurements are used in the diagnosis and treatment of disorders associated with changes in body acid-base balance. | Same |
| Sample type | Venous, arterial and capillary whole blood | Same |
{3}
| Similarities | | |
| --- | --- | --- |
| Item | Candidate Device: epoc Blood Urea Nitrogen test and epoc Total Carbon Dioxide test (k171247) | Predicate Device: BUN and TCO2 Tests using the i-STAT Chem8+ Cartridge (with i-STAT Portable Clinical Analyzer) (k053110) |
| Technology | An electrochemical multi-sensor array integrated into a single-use test that is interpreted by a handheld reader and associated software | Same |
| Reportable ranges (TCO2) | 5-50 mmol/L | Same |
| Differences | | |
| --- | --- | --- |
| Item | Candidate Device: epoc Blood Urea Nitrogen test and epoc Total Carbon Dioxide test (k171247) | Predicate Device: BUN and TCO2 Tests using the i-STAT Chem8+ Cartridge (with i-STAT Portable Clinical Analyzer) (k053110) |
| Measured Parameter | pH; Carbon Dioxide, Partial Pressure (pCO2); Oxygen, Partial Pressure (pO2); Sodium (Na); Potassium (K); Ionized Calcium (iCa); Chloride (Cl); Glucose (Glu); Lactate (Lac); Creatinine (Crea); Hematocrit (Hct); Blood Urea Nitrogen (BUN); Total CO2 (TCO2) | Sodium (Na); Potassium (K); Ionized Calcium (iCa); Chloride (Cl); Glucose (Glu); Creatinine (Crea); Hematocrit (Hct); Urea Nitrogen (BUN); Total CO2 (TCO2) |
| Calculated Parameter | Bicarbonate (cHCO3-); Calculated Total Carbon Dioxide (cTCO2); (only available when measured TCO2 is not available) Base Excess (BE); Oxygen Saturation (cSO2); Alveolar Oxygen (A); Arterial Alveolar Oxygen Tension Gradient (A-a); Arterial Alveolar Oxygen Tension Ratio (a/A); Anion Gap (AGap, AGapK); Estimated Glomerular Filtration Rate (eGFR, eGRF-a); Hemoglobin (cHgb) BUN/Creatinine ratio (BUN/Crea) | Anion Gap (AnGap); Hemoglobin (Hgb) |
| Reportable ranges BUN | 3-120 mg/dL | 3-140 mg/dL |
{4}
| Differences | | |
| --- | --- | --- |
| Item | Candidate Device: epoc Blood Urea Nitrogen test and epoc Total Carbon Dioxide test (k171247) | Predicate Device: BUN and TCO2 Tests using the i-STAT Chem8+ Cartridge (with i-STAT Portable Clinical Analyzer) (k053110) |
| Sample volume | At least 92 μL | 95 μL |
## K. Standard/Guidance Document Referenced (if applicable):
CLSI EP05-A3, Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline—Third Edition; 2014.
CLSI EP06-A, Evaluation of the linearity of Quantitative Measurement Procedures: a Statistical Approach; Approved Guideline—Second Edition; 2003.
CLSI EP07-A2, Interference Testing in Clinical Chemistry; Approved Guideline - Second Edition; 2002.
CLSI EP09-A3, Measurement Procedure Comparison and Bias Estimation Using Patient Samples; Approved Guideline - Third Edition; 2013.
CLSI EP17-A2, evaluation of detection capability for clinical laboratory measurement procedures; approved guideline - Second Edition.
IEC 61010-1, Safety requirements for electrical equipment for measurement, control and laboratory use – Part 1: General requirements; IECEE, 2010.
CLSI H11-A4, Procedures for the Collection of Arterial Blood Specimens; Approved Standard—Forth Edition; 2004.
## L. Test Principle:
### Urea Nitrogen (BUN) Sensor
The sensor module consists of an epoxy foil supporting array of foil electrode contacts on the outer side (or contact surface) and an array of sensor membranes on the inner side (or sensor surface). The epoc BUN electrode design is an enzymatic potentiometric-based sensing device. The BUN electrode uses the enzyme urease to hydrolyze urea to ammonium ions, as follows:
$$
\mathrm{Urea} + \mathrm{H_2O} + 2\mathrm{H^+} + \mathrm{Urease} \rightarrow 2\mathrm{NH_4^+} + \mathrm{CO_2}
$$
A potentiometric ion-selective electrode then detects the enzymatically-produced ammonium ion. The sensor is designed as a two-layer device, each layer providing the functions described above. The concentration of ammonium ions is obtained from the measured potential using the Nernst equation.
{5}
# Total Carbon Dioxide
The epoc System calculates total carbon dioxide using the measured and reported values of pH and $p\mathrm{CO}2$ according to the standard Henderson-Hasselbalch equation:
Calculated $\mathrm{TCO}_2$ ( $\mathrm{cTCO}_2$ ) = $\mathrm{cHCO}_3^-$ + 0.0307 x $p\mathrm{CO}_2$
where: $\mathrm{LOG~cHCO_3^- = pH + LOG~}p\mathrm{CO}_2 - 7.608$
This calculated $\mathrm{TCO}_2$ ( $\mathrm{cTCO}_2$ ) value is metrologically traceable to the epoc pH and $p\mathrm{CO}_2$ measurements, which are in turn traceable to primary standard reference materials for pH and $p\mathrm{CO}_2$ .
The measured $\mathrm{TCO}_2$ is achieved by calibrating a function of pH and $p\mathrm{CO}_2$ against the accepted IFCC Reference Measurement Procedure for Total Carbon Dioxide by mathematically modifying the calculated $\mathrm{TCO}_2$ equation in order to match the IFCC reference values. Hence, the measured $\mathrm{TCO}_2$ will be metrologically traceable to the IFCC $\mathrm{TCO}_2$ reference method.
# M. Performance Characteristics (if/when applicable):
# 1. Analytical performance:
# a. Precision/Reproducibility:
# Internal Precision Study
The precision study was conducted following the CLSI EP5-A3 guideline. Two levels of aqueous controls were tested in duplicate on 2 separate runs per day over 20 days using 4 lots of test cards and 11 epoc Blood Analysis Systems. The total precision results are shown in the table below:
| Total Precision: BUN | | | | |
| --- | --- | --- | --- | --- |
| Level | Mean (mg/dL) | N | SD | %CV |
| Low | 7.1 | 320 | 0.32 | 4.5 |
| High | 51.7 | 320 | 1.16 | 2.3 |
| Total Precision: TCO2 | | | | |
| --- | --- | --- | --- | --- |
| Level | Mean (mmol/L) | N | SD | %CV |
| L1 | 16.2 | 320 | 1.02 | 6.3 |
| L3 | 30.7 | 320 | 0.92 | 3.0 |
External precision studies at POC sites:
Additional precision studies were performed at 3 point-of-care (POC) sites with multiple POC operators.
{6}
i. An external precision study was performed using three levels of commercially available controls. For each level of control, each operator ran a maximum 12 test cards from 3 lots. The results for within-run precision and total precision results are shown in the tables below:
| BUN | | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| Level | Mean (mg/dL) | N | Within Run | | Total Precision | |
| | | | SD | %CV | SD | %CV |
| L1 | 7.1 | 168 | 0.24 | 3.4 | 0.26 | 3.7 |
| L2 | 17.7 | 171 | 0.45 | 2.5 | 1.11 | 6.3 |
| L3 | 52.1 | 170 | 1.06 | 2.0 | 1.54 | 3.0 |
| TCO2 | | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| Level | Mean (mmol/L) | N | Within Run | | Total Precision | |
| | | | SD | %CV | SD | %CV |
| L1 | 15.9 | 172 | 0.44 | 2.8 | 0.50 | 3.1 |
| L2 | 19.7 | 172 | 1.00 | 5.1 | 1.12 | 5.7 |
| L3 | 30.4 | 169 | 0.58 | 1.9 | 1.05 | 3.4 |
ii. A whole blood precision study was performed using freshly collected lithium heparin venous whole blood samples. The 2 types of sample delivery methods were evaluated (syringe and capillary tubes) at all 3 POC sites. A total of 12 operators (4 operators per site) participated in the study. Within-run precision for each individual operator was calculated. The results of all sites and all operators are shown in the tables below:
| BUN Whole Blood Precision: Syringe Delivery Mode | | | | | |
| --- | --- | --- | --- | --- | --- |
| Level | N | Range (mg/dL) | Mean (mg/dL) | SD | %CV |
| Lo | 136 | 3.7 - 10.6 | 7.0 | 0.6 | 7.2% |
| Normal | 136 | 12.5 – 35.5 | 17.3 | 0.7 | 4.1% |
| Hi | 134 | 51.8 - 72.4 | 57.4 | 1.3 | 2.3% |
| BUN Whole Blood Precision: Capillary tube Delivery Mode | | | | | |
| --- | --- | --- | --- | --- | --- |
| Level | N | Range (mg/dL) | Mean (mg/dL) | SD | %CV |
| Lo | 135 | 5.9 - 9.7 | 7.6 | 0.5 | 7.0% |
| Normal | 135 | 11.9 – 20.8 | 15.6 | 0.6 | 3.9% |
| Hi | 136 | 51.3 - 60.3 | 55.5 | 1.6 | 2.9% |
{7}
8
| TCO₂ Whole Blood Precision: Syringe Delivery Mode | | | | | |
| --- | --- | --- | --- | --- | --- |
| Level | n | Range (mmol/L) | Mean (mmol/L) | SD | %CV |
| Lo | 136 | 5.0 - 15.9 | 10.5 | 0.4 | 3.7% |
| Normal | 136 | 22.3 - 30.9 | 27.5 | 0.4 | 1.4% |
| Hi | 134 | 33.8 - 40.0 | 36.5 | 0.6 | 1.5% |
| TCO₂ Whole Blood Precision: Capillary tube Delivery Mode | | | | | |
| --- | --- | --- | --- | --- | --- |
| Level | n | Range (mmol/L) | Mean (mmol/L) | SD | %CV |
| Lo | 134 | 11.2 - 15.0 | 34.1 | 1.4 | 3.5% |
| Normal | 137 | 22.4 - 28.3 | 25.6 | 0.7 | 2.9% |
| Hi | 139 | 31.7 - 36.2 | 34.1 | 0.7 | 2.1% |
b. Linearity/assay reportable range:
The linearity study was performed based on the CLSI EP6-A guideline. Lithium heparin whole blood samples with nine levels of analyte concentrations spanning the entire measuring range of each assay were prepared and tested. Regression analysis was performed as per CLSI EP6-A guideline; results are shown below.
BUN
| Claimed Range | Test range | Slope | Intercept | R² |
| --- | --- | --- | --- | --- |
| 3-120 mg/dL | 4-120 mg/dL | 1.0204 | 0.3996 | 0.9979 |
TCO₂
| Claimed Range | Test range | Slope | Intercept | R² |
| --- | --- | --- | --- | --- |
| 5-50 mmol/L | 4-50 mmol/L | 0.9032 | 3.3176 | 0.9994 |
The results of the linearity studies support the claimed measuring ranges described in the tables above.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability:
Calibration of the epoc system is performed using both primary and secondary NIST traceable standards.
The calibration fluid is prepared gravimetrically from pure materials and is handled anaerobically during the test card manufacturing process until it is sealed between metallized foils in each epoc test.
{8}
For $TCO_{2}$
The pCO2-pH (and hence the $\mathrm{HCO^{-}}$ ) the stability in the calibration fluid as part of the epoc test card was previously cleared under k061597.
# d. Detection limit:
The study was performed as per the CLSI EP17-A2 guideline. Test samples were prepared from whole blood using lithium heparin anticoagulant. 4 blank and 4 low concentration samples were each prepared separately for BUN and $\mathrm{TCO_2}$ . Each sample was tested using two epoc BGEM Test Cards lots over 3 days.
A minimum of 64 replicates of measurements (16 replicates x 4 samples) per lot were used to calculate the LoB, LoD, and LoQ of BUN. For $\mathrm{TCO_2}$ , a minimum of 60 replicates of measurements per lot were used (15 replicates x 4 samples). The LoB was calculated as the $95^{\text{th}}$ percentile using the nonparametric method and a risk level, $\beta$ , of 0.05. The LoD was evaluated from the lowest analyte level that had at least $95\%$ of the replicates greater than the LoB. The limit of quantitation was determined according to accuracy goal based on total error (TE); for BUN the TE accuracy goal was $2\mathrm{mg / dL}$ and for $\mathrm{TCO_2}$ of $4\mathrm{mmol / L}$ . The results of the detection limits study are summarized below:
| Analyte | LoB | LoD | LoQ | Claimed Range |
| --- | --- | --- | --- | --- |
| BUN (mg/dL) | 2 | 3 | 3 | 3 - 120 |
| TCO2(mmol/L) | 4 | 4.3 | 4.3 | 5 - 50 |
# e. Analytical specificity:
An interferent testing of the BUN and $\mathrm{TCO_2}$ measurements on the epoc Blood Analysis System was performed per the CLSI EP07-A2 guideline. Potential interferences were evaluated at BUN concentrations of 8 and $20\mathrm{mg / dL}$ and $\mathrm{TCO_2}$ concentrations of 20 and $35\mathrm{mmol / L}$ . In each of these tests, human specimens were aliquoted into 2 samples. The test samples were spiked with interferent, while the control samples were spiked with the solvent of the interferent. The bias between the control sample and the test sample with added interferent was calculated. The concentration of interfering substance considered as causing no clinically significant interference is defined as a bias (difference between the test and the control sample) of: $1.77\mathrm{mg / dL}$ for BUN; $3.32\mathrm{mmol / L}$ for $\mathrm{TCO_2}$ concentrations $\leq 20\mathrm{mmol / L}$ and $2.95\mathrm{mmol / L}$ for $\mathrm{TCO_2}$ concentrations $>20\mathrm{mmol / L}$ .
# Summary of interference studies for BUN:
The interferences tested and found to be clinically non-significant are summarized below:
| Substance tested | Highest concentration tested showing no significant interference |
| --- | --- |
| Acetaminophen | 20 mg/dL |
{9}
| Substance tested | Highest concentration tested showing no significant interference |
| --- | --- |
| Acetoacetic Acid (Li salt) | 21.6 mg/dL |
| Acetyl Salicylic Acid | 65.2 mg/dL |
| Ammonium ions | 5.35 mg/dL |
| Ascorbate (Na) | 6.8 mg/dL |
| Benzalkonium Chloride | 0.002% |
| Bilirubin Conjugated | 28.8 mg/dL |
| Bilirubin Unconjugated | 25 mg/dL |
| Bromide | 386 mg/dL |
| Cefazolin (Na) | 125.9 mg/dL |
| Ceftriaxone (Na) | 96.6 mg/dL |
| CO2 High (NaHCO3) | 630.6 mg/dL |
| Citrate (Na tribasic dihydrate) | 588 mg/dL |
| Dopamine HCl | 0.1 mg/dL |
| EDTA | 744 mg/dL |
| Ethanol | 400 mg/dL |
| Fluoride (Na) | 399 mg/dL |
| Gallamine Triethiodide | 4.46 mg/dL |
| Glucose | 504 mg/dL |
| Glutathione oxidized | 156 mg/dL |
| Glutathione reduced | 156 mg/dL |
| Glycolic Acid | 38 mg/dL |
| Hematocrit (High) | 60% vs 45% (Spun) |
| Hematocrit (Low) | 20% vs 45% (Spun) |
| Heparin (Na) | 20 U/mL |
| β-Hydroxybutyrate (Na) | 252 mg/dL |
| Hydroxyurea | 15.2 mg/dL |
| Ibuprofen | 50 mg/dL |
| Intralipid | 500 mg/dL |
| Iodide (Na) | 19.5 mg/dL |
| Lactate (Na) | 74 mg/dL |
| L-Cysteine | 12 mg/dL |
| L-dopa | 0.5 mg/dL |
| Lithium (Cl) | 13.5 mg/dL |
| Metamizole (Na) | 210.8 mg/dL |
| Methotrexate | 90 mg/dL |
| N-Acetyl Cysteine | 163 mg/dL |
| Nithiodote (Na Thiosulfate) | 264 mg/dL |
| Oxalate (K) Monohydrate | 4 mg/dL |
| Pentothal (Na) | 6.5 mg/dL |
| Perchlorate (Na) | 12.2 mg/dL |
{10}
| Substance tested | Highest concentration tested showing no significant interference |
| --- | --- |
| pH – Low | pH <6.8 |
| pH – High | pH>8 |
| Protein – Low | 3.5 g/dL |
| Protein – High | 10 g/dL |
| Salicylic Acid (Na salicylate) | 69.5 mg/dL |
| Thiocyanate (K) | 16.7 mg/dL |
| Uric acid | 23.5 mg/dL |
Clinically significant interfering substances for BUN measurements are itemized below:
- Samples contaminated with benzalkonium salts used as coatings for in-dwelling lines may cause elevated BUN results. For proper line-flushing procedures refer to CLSI H11-A4.
- Citrate will have no significant effect up to 6.0 mmol/L (176.5 mg/dL) after which it will decrease the BUN concentration by up to 0.26 mg/dL BUN per mmol/L citrate.
- EDTA will have no significant effect up to 4.5 mmol/L (167 mg/dL) after which it will decrease the BUN concentration by up to 0.43 mg/dL BUN per mmol/L EDTA.
- Glutathione reduced will have no significant effect up to 1.7 mmol/L (52.2 mg/dL), after which it will increase the BUN concentration by up to 1.91 mg/dL BUN per mmol/L glutathione reduced. Blood glutathione (GSH) in human subjects is ~0.79-1.05 mmol/L. Long term oral glutathione reduced supplementation (250-1,000 mg/day administered for 6 months) increases glutathione plasma levels by ~0.2-8 μmol/L (~0.01-0.25 mg/dL). Short-term, oral intake of glutathione reduced does not affect plasma glutathione levels.
- β-Hydroxybutyrate will have no significant effect up to 17.2 mmol/L (216.9 mg/dL), after which it will decrease the BUN concentration by up to 0.11 mg/dL BUN per mmol/L hydroxybutyrate. The reference range for β-hydroxybutyrate in plasma is <0.4 to 0.5 mmol/L. β-hydroxybutyrate concentration over 3 mmol/L are indicative of ketoacidosis; in very severe diabetic ketoacidosis the concentration may exceed 25 mmol/L.
- Hydroxyurea will have no significant effect up to 1.3 mmol/L (9.9 mg/dL), after which it will increase the BUN concentration by up to 1.61 mg/dL BUN per mmol/L hydroxyurea. The recommended dose of hydroxyurea for patients range from 15 mg/kg/day to 30 mg/kg/day. A treatment dose of 2,000 mg/day (~30mg/kg) results in maximum plasma concentration of ~800μmol/L with oral administration and ~1 mmol/L with intravenous method.
{11}
- N-acetylcysteine will have no significant effect up to $9.2\mathrm{mmol / L}$ (150.1 mg/dL), after which it will increase the BUN concentration by up to 0.11 mg/dL BUN per mmol/L N-acetylcysteine. It has been reported that 1 mmol/L N-acetyl cysteine is therapeutically unattainable in plasma. The therapeutic level for N-acetyl cysteine is 0.3 mmol/L.
- Nithiodote will have no significant effect up to $4.1\mathrm{mmol / L}$ (64.8 mg/dL) after which it will decrease the BUN concentration by up to $0.41\mathrm{mg / dL}$ BUN per mmol/L Nithiodote. The expected peak sodium thiosulfate plasma concentration following a $12.5\mathrm{g}$ of Nithiodote is $16.7\mathrm{mmol / L}$ .
# Summary of interference studies for $\mathrm{TCO_2}$
The interferences tested and found to be clinically non-significant are summarized below:
| Substance tested | Highest concentration tested showing no significant interference |
| --- | --- |
| Acetaminophen | 20 mg/dL |
| Acetoacetic Acid (Li salt) | 21.6 mg/dL |
| Acetylsalicylic Acid | 65.2 mg/dL |
| Ammonium ions | 5.35 mg/dL |
| Ascorbate (Na) | 6.8 mg/dL |
| Benzalkonium Chloride | 0.001% |
| Bilirubin Conjugated | 26.8 mg/dL |
| Bilirubin Unconjugated | 25 mg/dL |
| Bromide | 386 mg/dL |
| Cefazolin (Na) | 125.9 mg/dL |
| Ceftriaxone (Na) | 96.6 mg/dL |
| Citrate (Na tribasic dihydrate) | 588 mg/dL |
| Dopamine HCl | 0.1 mg/dL |
| EDTA | 186mg/dL |
| Ethanol | 400 mg/dL |
| Fluoride (Na) | 399 mg/dL |
| Gallamine Triethiodide | 4.46 mg/dL |
| Glucose | 504mg/dL |
| Glutathione oxidized | 156 mg/dL |
| Glutathione reduced | 156 mg/dL |
| Glycolic Acid | 38 mg/dL |
| Hematocrit (High) | 60% vs 45% (Spun) |
| Hematocrit (Low) | 20% vs 45% (Spun) |
| Heparin (Na) | 20 U/mL |
| β-Hydroxybutyrate (Na) | 252 mg/dL |
| Hydroxyurea | 15.2 mg/dL |
{12}
| Substance tested | Highest concentration tested showing no significant interference |
| --- | --- |
| Ibuprofen | 50 mg/dL |
| Intralipid | 500 mg/dL |
| Iodide (Na) | 19.5 mg/dL |
| Lactate (Na) | 74 mg/dL |
| L-Cysteine | 12 mg/dL |
| L-dopa | 0.5 mg/dL |
| Lithium (Cl) | 13.5 mg/dL |
| Metamizole (Na) | 210.8 mg/dL |
| Methotrexate | 90 mg/dL |
| N-Acetyl Cysteine | 163 mg/dL |
| Nithiodote (Na Thiosulfate) | 264 mg/dL |
| Oxalate (K) Monohydrate | 4 mg/dL |
| Pentothal (Na) | 6.5 mg/dL |
| Perchlorate (Na) | 12.2 mg/dL |
| pH – Low | pH <6.8 |
| Protein – Low | 3.5 g/dL |
| Protein – High | 10 g/dL |
| Salicylic Acid (Na salicylate) | 69.5 mg/dL |
| Thiocyanate (K) | 16.7 mg/dL |
| Uric acid | 23.5 mg/dL |
Clinically significant interfering substances for $\mathrm{TCO}_2$ measurements are itemized below:
- Samples contaminated with benzalkonium salts used as coatings for in-dwelling lines may cause significant decrease in $\mathrm{TCO}_2$ results. For proper line-flushing procedures refer to CLSI H11-A4.
- Citrate will have no significant effect up to $11.8\ \mathrm{mmol/L}$ ($347.0\ \mathrm{mg/dL}$) after which it will increase the $\mathrm{TCO}_2$ concentration by up to $0.24\ \mathrm{mmol/L}\ \mathrm{TCO}_2$ per mmol/L citrate.
- EDTA will have no significant effect up to $4.8\ \mathrm{mmol/L}$ ($178.7\ \mathrm{mg/dL}$) after which it will increase the $\mathrm{TCO}_2$ concentration by up to $0.57\ \mathrm{mmol/L}\ \mathrm{TCO}_2$ per mmol/L EDTA.
- N-acetyl cysteine will have no significant effect up to $9.6\ \mathrm{mmol/L}$ ($156.7\ \mathrm{mg/dL}$) after which it will increase the $\mathrm{TCO}_2$ concentration by up to $0.54\ \mathrm{mmol/L}\ \mathrm{TCO}_2$ per mmol/L N-acetyl cysteine. It has been reported that $1\ \mathrm{mmol/L}$ N-acetyl cysteine is therapeutically unattainable in plasma. The therapeutic level for N-acetyl cysteine is $0.3\ \mathrm{mmol/L}$.
{13}
f. Assay cut-off:
Not applicable.
# 2. Comparison studies:
a. Method comparison with predicate device:
Method comparison studies were performed at 3 POC sites by POC operators. Lithium heparin venous and arterial whole blood samples and fresh capillary whole blood samples were analyzed for BUN using the epoc Blood Analysis System (candidate device) and compared to matched lithium heparin plasma samples analyzed using BUN assay on the Roche Cobas 8000 Modular Analyzer. Lithium heparin venous and arterial whole blood samples and fresh capillary whole blood samples were analyzed for $\mathrm{TCO_2}$ using the epoc Blood Analysis System (candidate device) and the i-STAT CHEM8+/i-STAT system (predicate device). The results of the overall performance of the device at all the sites are summarized in the tables below.
BUN Regression Analysis Summary for epoc Blood Analysis System vs. Roche Cobas 8000 Modular Analyzer
| Venous Samples - BUN | | | | | |
| --- | --- | --- | --- | --- | --- |
| | N | Intercept | Slope | R² | Range (mg/dL) |
| Site 1 | 50 | 0.8 | 0.953 | 0.996 | 4-109 |
| Site 2 | 49 | 0.8 | 0.962 | 0.996 | 4-118 |
| Site 3 | 50 | -0.6 | 1.035 | 0.997 | 4-118 |
| All Sites | 149 | 0.2 | 0.990 | 0.995 | 4-118 |
| Arterial Samples - BUN | | | | | |
| --- | --- | --- | --- | --- | --- |
| | N | Intercept | Slope | R² | Range (mg/dL) |
| Site 1 | 42 | 0.9 | 0.973 | 0.995 | 7-106 |
| Site 2 | 49 | 0.9 | 0.943 | 0.996 | 8-118 |
| Site 3 | 50 | 1.0 | 1.028 | 0.997 | 3-107 |
| All Sites | 141 | 0.9 | 0.977 | 0.994 | 3-118 |
| Capillary Samples - BUN | | | | | |
| --- | --- | --- | --- | --- | --- |
| | N | Intercept | Slope | R² | Range (mg/dL) |
| Site 1 | 48 | 0.4 | 1.001 | 0.995 | 5-117 |
| Site 2 | 45 | 0.5 | 0.945 | 0.997 | 6-112 |
| Site 3 | 50 | 0.0 | 0.995 | 0.998 | 6-118 |
| All Sites | 143 | 0.2 | 0.986 | 0.996 | 5-118 |
{14}
$\mathsf{TCO}_2$ Regression Analysis Summary for epoc Blood Analysis System vs i-STAT-CHEM8+ Cartridges (Venous, Arterial, and Capillary Samples)
| Venous Samples - TCO2 | | | | | |
| --- | --- | --- | --- | --- | --- |
| | N | Intercept | Slope | R² | Range (mmol/L) |
| Site 1 | 54 | -0.80 | 1.041 | 0.978 | 10-49 |
| Site 2 | 60 | -5.60 | 1.146 | 0.969 | 11-45 |
| Site 3 | 50 | -1.30 | 1.057 | 0.957 | 13-45 |
| All Sites | 164 | -2.80 | 1.086 | 0.953 | 10-49 |
| Arterial Samples - TCO2 | | | | | |
| --- | --- | --- | --- | --- | --- |
| | N | Intercept | Slope | R² | Range (mmol/L) |
| Site 1 | 53 | -2.5 | 1.125 | 0.965 | 11-41 |
| Site 2 | 53 | -2.3 | 1.098 | 0.972 | 7-44 |
| Site 3 | 54 | -0.8 | 1.024 | 0.974 | 19-41 |
| All Sites | 160 | -1.8 | 1.079 | 0.966 | 7-44 |
| Capillary Samples - TCO2 | | | | | |
| --- | --- | --- | --- | --- | --- |
| | N | Intercept | Slope | R² | Range (mmol/L) |
| Site 1 | 103 | 1.7 | 0.947 | 0.949 | 8-49 |
| Site 2 | 76 | -0.12 | 1.029 | 0.946 | 9-49 |
| Site 3 | 71 | -0.36 | 1.053 | 0.954 | 9-47 |
| All Sites | 250 | 0.65 | 0.999 | 0.946 | 8-49 |
# b. Matrix comparison:
A comparison was performed to demonstrate the equivalence between lithium heparin whole blood and un-anticoagulated venous whole blood and sodium heparin venous whole blood for testing BUN and $\mathrm{TCO_2}$ levels. The results of linear regression for heparinized (Na heparin) vs. lithium heparin, and non-anticoagulated venous samples vs lithium heparin are shown below:
| BUN | | | | |
| --- | --- | --- | --- | --- |
| Matrix | N | Intercept | Slope | R² |
| No Additive | 62 | 0.1 | 1.000 | 0.989 |
| Sodium heparin | 62 | 0.3 | 0.981 | 0.992 |
| TCO2 | | | | |
| --- | --- | --- | --- | --- |
| Matrix | N | Intercept | Slope | R2 |
| No Additive | 62 | 0.1 | 0.980 | 0.980 |
| Sodium heparin | 62 | 0.6 | 0.970 | 0.989 |
The results of the matrix comparison study support that lithium heparin, sodium heparin and non-anticoagulated whole blood specimens are suitable for use with the epoc Blood Urea Nitrogen Test and epoc Total Carbon Dioxide Test.
{15}
16
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable.
b. Clinical specificity:
Not applicable.
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable.
4. Clinical cut-off:
Not applicable.
5. Expected values/Reference range:
The following expected values are provided in the product insert based on the literature:
| Analyte | Reference Range |
| --- | --- |
| TCO_{2} | 22-29 mmol/L (arterial) |
| | 23-30 mmol/L (venous) |
| BUN | 8-26 mg/dL |
- Tietz Textbook of Clinical Chemistry and Molecular Diagnostics, fourth edition, C.A. Burtis, E.R. Ashwood, and D.E. Bruns eds., Elsevier Saunders, St. Louis, 2006.
- E. Statland, Clinical Decision Levels for Lab Tests, Medical Economic Books, Oradell, Ni, 1987.
- Pruden E.L., Siggaard-Andersen 0., and Tietz N.W., Chapter 30 (Blood Gases and pH), of Tietz Textbook of Clinical Chemistry, Second Edition, ed. C.A. Burtis and E.R. Ashwood. W.B. Saunders Company, Philadelphia, 1994
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.