# Endoform Topical Matrix (K171231)

_Aroa Biosurgery · KGN · Jun 14, 2017 · SU · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K171231

## Device Facts

- **Applicant:** Aroa Biosurgery
- **Product Code:** [KGN](/productcode/KGN.md)
- **Decision Date:** Jun 14, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Device Class:** Class U
- **Review Panel:** SU
- **Attributes:** Therapeutic

## Indications for Use

Endoform Topical Matrix is a sterile, single use ovine forestomach-derived extracellular matrix intended to cover, protect, and provide a moist wound environment.

## Device Story

Endoform Topical Matrix is a sterile, lyophilized, multi-laminate collagen wound dressing derived from ovine forestomach extracellular matrix (ECM). It contains natural collagen I and III. The device is supplied in various sizes and thicknesses and can be trimmed by a clinician to fit the wound. It is applied topically to cover and protect wounds, providing a moist environment to facilitate management. The device is fabricated using a proprietary 'lug lamination' process, which physically bonds multiple collagen sheets together without sewing. It is intended for use in clinical settings for acute and chronic wounds. The device functions as a physical barrier and scaffold, supporting wound management through its ECM composition. No software or electronic components are involved.

## Clinical Evidence

No clinical performance data was generated for this submission. Equivalence is supported by literature review of in vivo models of dermal injury and established clinical use of the underlying collagen matrix technology.

## Technological Characteristics

Ovine forestomach-derived extracellular matrix (collagen I and III). Multi-laminate, lyophilized sheets. Thickness 0.15-1.50 mm. Sizes 1-400 cm². Lug lamination (physical bonding). Sterile, single-use.

## Predicate Devices

- Endoform Dermal Template ([K092096](/device/K092096.md))

## Reference Devices

- Endoform Plastics and Reconstructive Matrix ([K162461](/device/K162461.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features the department's name encircling a stylized caduceus symbol. The caduceus is composed of three human profiles facing right, with a triple helix design representing health and medicine. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 14, 2017

Aroa Biosurgery Tina O'Brien Director, Regulatory Affairs 2 Kingsford Smith Place Airport Oaks, Auckland 2022, New Zealand

Re: K171231

Trade/Device Name: Endoform Topical Matrix Regulatory Class: Unclassified Product Code: KGN Dated: April 20, 2017 Received: April 27, 2017

Dear Tina O'Brien:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

# David-Krause -S

for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K171231

Device Name

Endoform Topical Matrix

Indications for Use (Describe)

Endoform Topical Matrix is indicated for the management of the following wounds:

- partial and full-thickness wounds
- · pressure ulcers
- · venous ulcers
- · diabetic ulcers
- chronic vascular ulcers
- tunnelled / undermined wounds
- · surgical wounds (donor sites/grafts, post-Moh's surgery, post-laser surgery, podiatric, wound dehiscence)
- · trauma wounds (abrasions, lacerations, second-degree burns, and skin tears)
- · draining wounds

Type of Use (Select one or both, as applicable)

× | Prescription Use (Part 21 CFR 801 Subpart D)

] Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for AROA. The logo consists of an orange circular design made up of many smaller orange ovals. To the right of the design is the text "AROA" in a bold, sans-serif font.

### 5. 510(k) Summary

| Contact<br>person/submitter | Tina O'Brien<br>Director of Regulatory Affairs<br>Aroa Biosurgery                                                      |
|-----------------------------|------------------------------------------------------------------------------------------------------------------------|
| Date prepared               | 20 April 2017                                                                                                          |
| Contact details             | 2 Kingsford Smith Place<br>Airport Oaks, Auckland<br>2022, New Zealand<br>+64 9 369 3035                               |
| Trade name                  | Endoform Topical Matrix                                                                                                |
| Product Code                | KGN                                                                                                                    |
| Common name                 | Dressing, wound, collagen                                                                                              |
| Predicate device            | Predicate - Endoform Dermal Template (K092096)<br>Reference - Endoform Plastics and Reconstructive Matrix<br>(K162461) |

### 5.1 Device Description

Endoform Topical Matrix is an advanced collagen matrix comprised of natural >70%, non-reconstituted collagen designed to cover, protect, and provide a moist wound environment. The device is supplied in a variety of sizes and thicknesses which can be trimmed by the surgeon to meet the individual patient's needs.

### 5.2 Intended Use and Indications for Use

## 5.1.1 Intended Use

Endoform Topical Matrix is a sterile, single use ovine forestomach-derived extracellular matrix intended to cover, protect, and provide a moist wound environment.

### 5.1.2 Indications for Use

Endoform Topical Matrix is indicated for use in the management of the following wounds:

- . partial and full-thickness wounds
- . pressure ulcers
- . venous ulcers
- . diabetic ulcers

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for AROA. The logo consists of an orange circular design made up of smaller orange circles, resembling a sun or a burst of energy. To the right of the circular design is the word "AROA" in bold, black, sans-serif font.

### Endoform Topical Matrix - Traditional 510(k)

- . chronic vascular ulcers
- . tunnelled / undermined wounds
- surgical wounds (donor sites/grafts, post-Moh's surgery, post-laser . surgery, podiatric, wound dehiscence)
- . trauma wounds (abrasions, lacerations, second-degree burns, and skin tears)
- . draining wounds

### 5.3 Technological Characteristics Comparison

The subject Endoform Topical Matrix device is identical to the predicate Endoform Dermal Template with respect to raw material, tissue processing, intended use, and indications for use. The only difference between the subject and predicate devices is the number of layers and use of a different lamination process during manufacturing.

The subject device is a multi-layer device that is laminated via a proprietary process termed 'lug lamination', which is a physical means of fabricating the devices, such that the individual sheets of the device are physically bonded to one another as opposed to using embroidery to sew the laminate of collagen matrix sheets. The following table details the similarities and differences of the subject, predicate, and reference devices:

| Device Status          | Subject Device<br>Endoform Topical Matrix                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Predicate Device<br>Endoform Dermal Template<br>(K092096)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Reference Device<br>Endoform Plastics &<br>Reconstruction Matrix<br>(K162461)                                                                                                                                                                                                                                  |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer           | Aroa                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Aroa                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Aroa                                                                                                                                                                                                                                                                                                           |
| 510K Number            | K171231                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | K092096                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | K162461                                                                                                                                                                                                                                                                                                        |
| Device name            | Endoform Topical Matrix                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Endoform Dermal Template                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Endoform Plastics &<br>Reconstruction Matrix                                                                                                                                                                                                                                                                   |
| Classification<br>Name | dressing, wound, collagen<br>(KGN)<br>Unclassified                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | dressing, wound, collagen<br>(KGN)<br>Unclassified                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | surgical mesh (FTL)<br>21 CFR 878.3300<br>Class II                                                                                                                                                                                                                                                             |
| Intended use           | Endoform Topical Matrix is<br>a sterile, single use ovine<br>forestomach-derived<br>extracellular matrix<br>intended to cover, protect,<br>and provide a moist<br>wound environment                                                                                                                                                                                                                                                                                                                                  | Endoform Dermal Template<br>is supplied sterile and is<br>intended for single use in<br>the<br>treatment of the following<br>wounds:<br>* partial and full-thickness<br>wounds<br>* pressure ulcers<br>* venous ulcers<br>* diabetic ulcers<br>* chronic vascular ulcers<br>* tunnelled / undermined<br>wounds<br>* surgical wounds (donor<br>sites/grafts, post-Moh's                                                                                                                                                                                | Endoform Plastics and<br>Reconstructive Matrix is for<br>implantation to reinforce<br>soft tissue where weakness<br>exists in patients requiring<br>soft tissue repair or<br>reinforcement in plastic and<br>reconstructive surgery. The<br>device is supplied sterile<br>and is intended for one-time<br>use. |
| Device Status          | Subject Device<br>Endoform Topical Matrix                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Predicate Device<br>Endoform Dermal Template<br>(K092096)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Reference Device<br>Endoform Plastics &<br>Reconstruction Matrix<br>(K162461)                                                                                                                                                                                                                                  |
|                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | surgery, post-laser surgery,<br>podiatric,<br>wound dehiscence)<br>* trauma wounds<br>(abrasions, lacerations,<br>second-degree burns, and<br>skin tears)<br>* draining wounds                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                |
| Indications for Use    | Endoform Topical Matrix is<br>indicated for use in the<br>management of the<br>following wounds:<br>* partial and full-thickness<br>wounds<br>* pressure ulcers<br>* venous ulcers<br>* diabetic ulcers<br>* chronic vascular ulcers<br>* tunnelled / undermined<br>wounds<br>* surgical wounds (donor<br>sites/grafts, post-Moh's<br>surgery, post-laser<br>surgery, podiatric,<br>wound dehiscence)<br>* trauma wounds<br>(abrasions, lacerations,<br>second-degree burns,<br>and skin tears)<br>* draining wounds | Endoform Dermal Template<br>is supplied sterile and is<br>intended for single use in<br>the treatment of the<br>following wounds:<br>* partial and full-thickness<br>wounds<br>* pressure ulcers<br>* venous ulcers<br>* diabetic ulcers<br>* chronic vascular ulcers<br>* tunnelled / undermined<br>wounds<br>* surgical wounds (donor<br>sites/grafts, post-Moh's<br>surgery, post-laser surgery,<br>podiatric,<br>wound dehiscence)<br>* trauma wounds<br>(abrasions, lacerations,<br>second-degree burns, and<br>skin tears)<br>* draining wounds | Endoform Plastics and<br>Reconstructive Matrix is for<br>implantation to reinforce<br>soft tissue where weakness<br>exists in patients requiring<br>soft tissue repair or<br>reinforcement in plastic and<br>reconstructive surgery. The<br>device is supplied sterile<br>and is intended for one-time<br>use. |
| Animal origin          | Ovine                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Ovine                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Ovine                                                                                                                                                                                                                                                                                                          |
| Tissue type            | Forestomach                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Forestomach                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Forestomach                                                                                                                                                                                                                                                                                                    |
| Nominal sizes          | Sizes ranging from 1 cm²<br>to 400 cm², lugged                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Fenestrated or non-<br>fenestrated sheets<br>ranging in size up to 400<br>cm²                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Sizes ranging from 1 cm² to<br>400 cm², lugged                                                                                                                                                                                                                                                                 |
| Thickness              | 0.15 - 1.50 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | 0.25 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | 0.15 – 1.50 mm                                                                                                                                                                                                                                                                                                 |
| Presentation           | Sterile, lyophilized, multi-<br>laminate lugged sheets in<br>peel pouch                                                                                                                                                                                                                                                                                                                                                                                                                                              | Sterile, lyophilized single<br>sheets in peel pouch                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Sterile, lyophilized, multi-<br>laminate lugged sheets in<br>peel pouch                                                                                                                                                                                                                                        |
| Components             | Ovine derived collagen<br>and associated ECM<br>components<br>-collagen I<br>-collagen III                                                                                                                                                                                                                                                                                                                                                                                                                           | Ovine derived collagen and<br>associated ECM<br>components<br>-collagen I<br>-collagen III                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Ovine derived collagen and<br>associated ECM<br>components<br>-collagen I<br>-collagen III                                                                                                                                                                                                                     |

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for AROA. The logo consists of an orange circular design on the left and the word "AROA" in black on the right. The circular design is made up of several smaller orange shapes arranged in a circular pattern.

Endoform Topical Matrix – Traditional 510(k)

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for AROA. The logo consists of an orange circular design made up of smaller circles, followed by the text "AROA" in black, sans-serif font. The circular design is positioned to the left of the text.

### 5.4 Non-Clinical Performance Data

As the subject and predicate devices are made of identical material (ovine forestomach tissue) and the multi-layer design and manufacture has been cleared via the reference device, no additional bench performance testing was warranted.

The use of collagen wound dressings is well-established in clinical use and studies in various animal models of dermal injury. Additionally, the collagen matrix used in all Aroa Biosurgery Endoform devices has been previously studied in in vivo models of dermal injury and is currently in wide-spread clinical use for the management of acute and chronic wounds.

In vivo use of the subject Endoform Topical Matrix was evaluated via a literature review to demonstrate that the subject device is appropriate for the proposed indications based on equivalence to commercially available devices with equivalent composition and indications.

### 5.5 Clinical Performance Data

Substantial equivalence was not based on an assessment of clinical performance data.

### 5.6 Conclusions

Based on the subject, predicate, and reference devices being completely identical in raw material and tissue processing steps, as well as the published data supporting use of the collagen matrix for the proposed indications, the introduction of Endoform Topical Matrix does not raise any new questions on safety and effectiveness and the results support a determination of substantial equivalence.

---

**Source:** [https://fda-staging.innolitics.com/device/K171231](https://fda-staging.innolitics.com/device/K171231)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).