K171224 · Irvine Scientific Sales Co., Inc. · MQL · Aug 31, 2017 · Obstetrics/Gynecology
Device Facts
Record ID
K171224
Device Name
Arctic Sperm Cryopreservation Medium
Applicant
Irvine Scientific Sales Co., Inc.
Product Code
MQL · Obstetrics/Gynecology
Decision Date
Aug 31, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.6180
Device Class
Class 2
Indications for Use
Arctic™ Sperm Cryopreservation Medium is intended for use in assisted reproductive procedures involving the cryopreservation and storage of human sperm.
Device Story
Arctic™ Sperm Cryopreservation Medium is a defined, single-use chemical medium used in clinical assisted reproductive technology (ART) laboratories. It facilitates the cryopreservation and storage of human sperm samples. The medium contains salts, buffering agents (including MOPS), cryoprotectants, vitamins, amino acids, minerals, carbohydrates, surfactants, energy sources, and human serum albumin. In practice, clinicians mix donor semen with the medium before cryopreservation in liquid nitrogen. The medium protects sperm cells during the freezing and thawing process, maintaining motility for subsequent use in reproductive procedures. The device is supplied in 5 mL sterilized vials.
Clinical Evidence
Bench testing only. Performance validated via pH, osmolality, and sterility testing (USP <71>). Endotoxin testing (USP <85>) confirmed levels ≤1.0 EU/ml. Sperm cryosurvival assay compared subject medium to predicate using donor semen; success defined as ≥80% of control motility post-thaw, post-gradient separation, and two hours post-wash. Accelerated shelf-life testing confirmed stability over 18 months.
Technological Characteristics
Defined medium containing salts, MOPS buffer, cryoprotectants, vitamins, amino acids, minerals, carbohydrates, surfactant, energy source, and human serum albumin. Osmolality: 2300-2600 mOsm/kg; pH: 7.25-7.54. Aseptically filled into sterilized vials (SAL 10^-3). Validated per ISO 13408-1:2008 and ISO 13408-2:2003.
Indications for Use
Indicated for use in assisted reproductive procedures involving the cryopreservation and storage of human sperm.
Regulatory Classification
Identification
Reproductive media and supplement are products that are used for assisted reproduction procedures. Media include liquid and powder versions of various substances that come in direct physical contact with human gametes or embryos (including water, acid solutions used to treat gametes or embryos, rinsing solutions, sperm separation media, supplements, or oil used to cover the media) for the purposes of preparation, maintenance, transfer or storage. Supplements are specific reagents added to media to enhance specific properties of the media (e.g., proteins, sera, antibiotics, etc.).
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing). The device, when it is phosphate-buffered saline used for washing, and short-term handling and manipulation of gametes and embryos; culture oil used as an overlay for culture media containing gametes and embryos; and water for assisted reproduction applications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym followed by the full name of the agency, "U.S. Food & Drug Administration," on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name in a sans-serif font.
January 13, 2022
Irvine Scientific Sales Co., Inc. Jayme Yamaguchi-Owens Senior Manager Regulatory Affairs 2511 Daimler Street Santa Ana, CA 92705
Re: K171224
> Trade/Device Name: ArcticTM Sperm Cryopreservation Medium Regulation Number: 21 CFR§ 884.6180 Regulation Name: Reproductive media and supplements Regulatory Class: II Product Code: MQL
Dear Jayme Yamaguchi-Owens:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated August 31, 2017. Specifically, FDA is updating this SE Letter to include the correct version of the 510(k) Summary.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Monica Garcia, Ph.D., OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices, (240) 402-2791, monica.garcia@fda.hhs.gov.
Sincerely,
## Monica D. Garcia -S
Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/2 description: The image is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the border. Inside the circle is a stylized image of three human profiles facing right, stacked on top of each other. The profiles are simple and abstract, with no facial details.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 31, 2017
Irvine Scientific Sales Co., Inc. Jayme Yamaguchi-Owens Senior Manager Regulatory Affairs 2511 Daimler Street Santa Ana, CA 92705
K171224 Re:
Trade/Device Name: Arctic™ Sperm Cryopreservation Medium Regulation Number: 21 CFR§ 884.6180 Regulation Name: Reproductive Media and Supplements Regulatory Class: II Product Code: MOL Dated: July 31, 2017 Received: August 1, 2017
Dear Jayme Yamaguchi-Owens:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{2}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
# Charles Viviano -S
For Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K171224
Device Name Arctic™ Sperm Cryopreservation Medium
Indications for Use (Describe)
Arctic™ Sperm Cryopreservation Medium is intended for use in assisted reproductive procedures involving the cryopreservation and storage of human sperm.
Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
ll.
lll.
## 510(k) Summary
#### l. General Information on Submitter
| Submitter/Address: | Irvine Scientific Sales Co., Inc.<br>2511 Daimler Street<br>Santa Ana, CA 92705<br>Telephone: 800-437-5706<br>Facsimile: 949-261-6522 |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Jayme Yamaguchi-Owens<br>Senior Manager Regulatory Affairs<br>Irvine Scientific<br>2511 Daimler Street<br>Santa Ana, CA 92705<br>Tel: 949-261-7800 ext. 254<br>Fax: 949-261-6522<br>Email: jfy@irvinesci.com |
| Date Prepared: | August 29, 2017 |
| General Information | |
| Device Name: | Arctic™ Sperm Cryopreservation Medium |
| Common Name: | Sperm Cryopreservation Medium |
| Classification Name: | Reproductive Media and Supplements (21 CFR884.6180) |
| Product Code: | MQL (Media, Reproductive) |
| Regulatory Class: | II |
#### IV. Predicate Device:
Sperm Maintenance Medium with Glycerol (K991337) manufactured by Irvine Scientific Sales, Co., Inc. This predicate device has not been subject to any design related recalls.
#### V. Description of the Device:
The Arctic™ Sperm Cryopreservation Medium is a modified version of the predicate device (K991337). It is a defined medium consisting of salts, buffering materials, cryoprotectants, vitamins, amino acids, minerals, carbohydrates, surfactant, energy source, and human serum albumin.
{5}------------------------------------------------
The Arctic™ Sperm Cryopreservation Medium is a single-use device that is aseptically filled into sterilized vials and has a sterility assurance level (SAL) of 10-3. This medium is supplied in a fill volume of 5 mL. and twelve (12) 5mL vials are included in a package, The product is tested for pH, osmolality, sperm cyrosurvival, endotoxin and sterility before lot release, and has a shelf-life of 18 months.
#### VI. Indications for Use
Arctic™ Sperm Cryopreservation Medium is intended for use in assisted reproductive procedures involving the cryopreservation and storage of human sperm.
#### VII. Comparison of Intended Use and Technological Characteristics with the Predicate Device
| Device/Predicate Device | K171224 (subject device) | K991337 (predicate device) |
|-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | Arctic™ Sperm Cryopreservation<br>Medium is intended for use in<br>assisted reproductive procedures<br>involving the cryopreservation and<br>storage of human sperm. | Sperm Maintenance Medium with<br>Glycerol is intended for use in<br>assisted reproductive procedures<br>involving the cryopreservation and<br>storage of human sperm. |
| Appearance | Clear, free of particulate | Clear, light yellow, free of particulate |
| pH | Same as predicate | 7.25-7.54 |
| Osmolality | 2300-2600 mOsm/kg | 2362-2532 mOsm/kg |
| Formulation | 21 Ingredients | 20 Ingredients |
The subject and predicate devices have the same indication - cryopreservation and storage of human sperm.
The subject and predicate devices have the same or comparable appearance, pH, and osmolality. The only differences in formulation are the replacement of L-glutamine with L-alanyl glutamine, addition of MOPS buffer, and minor modifications in concentrations of certain ingredients in the subject device. The differences noted are minor and do not raise different questions of safety andeffectiveness.
In addition to pH and osmolality, the subject and predicate devices have the same acceptance criteria for the sperm cyrosurvival assay and sterility testing, and the subject device has a more stringent acceptance criterion for endotoxin testing (≤1.0 EU/ml vs. ≤3.0 EU/ml).
#### VIII. Summary of Non-Clinical Performance Testing:
The following studies have been performed to support substantial equivalence to the predicate device:
- pH (7.25-7.54) ●
- Osmolality (2300-2600 mOsm/kg) ●
- Aseptic Processing Validation testing per ISO 13408-1:2008 and ISO 13408-● 2:2003
- Sterility testing (no microbiological growth) per USP <71> ●
{6}------------------------------------------------
- Endotoxin testing (≤1.0 EU/ml) per USP <85> o
- Sperm Cryosurvival Assay
Donor semen was mixed with the test medium and control (predicate device) and cryopreserved in liquid nitrogen. After thawing, sperm motility was determined at three time points: post thaw, post gradient separation, and two (2) hours post wash. Success of this test was based on ≥80% of control motility.
- Shelf-life testing (accelerated) was conducted to ensure that the following . acceptance criteria for product characteristics are met at time zero and the end of shelf-life:
- √ pH 7.25-7.54
- > Osmolality - 2300-2600 mOsm/kg
- > Sperm Cryosurvival Assay - ≥80% of the control motility
- Endotoxin ≤1.00 EU/ml (LAL) V
- V Sterility - No microbiological growth
## XI. Conclusion:
The subject and predicate devices have the same intended use. Although there are differences in technological characteristics between the subject and predicate devices, these differences do not raise different questions of safety or effectiveness. The performance data demonstrate that the subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.