The M-CATH Microcatheter is intended to provide support to the guide wire while performing PCI (percutaneous coronary intervention) and to facilitate the exchange of guide wires while maintaining access to distal vasculature. The device is also intended to deliver radiopaque contrast media into the coronary vasculature.
Device Story
M-CATH microcatheter is a manually operated device used by physicians trained in PCI. It consists of a catheter shaft with braided mesh reinforcement, a hypotube, and a hub. The device is inserted into the coronary vasculature to provide support for a 0.014" guide wire and to facilitate guide wire exchange while maintaining distal access. It also serves as a conduit for delivering radiopaque contrast media. The outer surface features a hydrophilic coating to enhance maneuverability and patient comfort. The device is compatible with 5F guiding catheters. By providing stable access and support, it assists the physician in navigating complex coronary anatomy during interventional procedures, potentially improving procedural success and patient outcomes.
Clinical Evidence
Bench testing only. Testing included dimensional verification, compatibility, pushability, trackability, bond strength, tip pull, flexibility, kink resistance, torque strength, coating lubricity, particulate evaluation, radiopacity, pressure testing, and flow rate. Biocompatibility testing was conducted per ISO 10993, including cytotoxicity, sensitization, irritation, acute systemic toxicity, hemocompatibility, hemolysis, complement activation, and in vivo thromboresistance.
Indicated for patients undergoing percutaneous coronary intervention (PCI) requiring guide wire support, guide wire exchange, or delivery of radiopaque contrast media into the coronary vasculature.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
Public Health Service
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Acrostak (Schweiz) AG Carmen Herraez Clinical and Regulatory Affairs Manager Stegackerstrasse 14 CH-8409 Winterthur Switzerland
Re: K171176
Trade/Device Name: M-CATH Microcatheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY, KRA Dated: August 9, 2017 Received: August 17, 2017
Dear Carmen Herraez:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K171176
Device Name M-CATH Microcatheter
Indications for Use (Describe)
The M-CATH Microcatheter is intended to provide support to the guide wire while performing PCI (percutaneous coronary intervention) and to facilitate the exchange of guide wires while maintaining access to distal vasculature. The device is also intended to deliver radiopaque contrast media into the coronary vasculature.
Type of Use (*Select one or both, as applicable*)
| <div style="display:flex; align-items:center;"> <div style="margin-right:5px;">☑</div> <div>Depreciative Use (Part 21 CFR 601, Subpart D)</div> </div> |
|--------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"> <div style="margin-right:5px;">☐</div> <div>Over-The-Counter Use (21 CFR 201.66)</div> </div> |
> Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image contains the logo for Acrostak. The logo features a red plus sign on the left, followed by the word "Acrostak" in a dark gray, sans-serif font. Below "Acrostak" is the text "Devices for interventional applications" in a smaller, lighter gray font, underlined with a red line.
8409 Winterthur Switzerland Phone +41 (0)52 233 95 51 Fax+41 (0)52 233 95 50
# 510(k) Summary [as required by 21 CFR 807.92] M-CATH™ Microcatheter
| DATE PREPARED | August 30th, 2017 |
|-----------------------|-------------------------------|
| TRADE NAME | M-CATH™ microcatheter |
| COMMON NAME | Microcatheter |
| DEVICE CLASSIFICATION | Class II per 21 CFR §870.1250 |
| PRODUCT CODE | DQY, Percutaneous Catheter |
### 1. PREDICATE DEVICE
The predicate devices are Finecross® MG and Asahi Corsair, which are manufactured by Terumo Corporation and Asahi Intecc Co LTD, respectfully. The predicate devices have been cleared through the premarket notification process (Finecross® MG in K082519 and Asahi Corsair in K083127).
### 2. INTENDED USE
The M-CATH microcatheter is intended to provide support to the guide wire while performing PCI (percutaneous coronary intervention) and to facilitate the exchange of guide wires while maintaining access to distal vasculature. The device is also intended to deliver radiopaque contrast media into the coronary vasculature.
### 3. DEVICE DESCRIPTION
M-CATH™ consists of a catheter shaft, a hypotube and a hub. The catheter shaft has reinforcing braided mesh with high torquability. The outer surface of the catheter is coated with a hydrophilic polymer with a high lubricity upon moistening, hence, enhancing maneuverability and patient comfort.
The microcatheter has one distal radiopaque marker proximal at the distal tip.
The shaft has a central guidewire lumen which permits the use of a guide wire with maximum diameter of 0.014" (0.36 mm). The guidewire lumen begins at the proximal luer port and ends at the distal tip. At the proximal end the catheter has a hub with one port. The catheter is compatible with any 5F guiding catheter.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for Acrostak. The logo features a red plus sign on the left, followed by the word "Acrostak" in gray, bold font. Underneath the word "Acrostak" is a thin red line, and below that is the text "Devices for interventional applications" in a smaller, lighter gray font.
The main materials are stainless steel, PTFE and Pebax (braided shaft), stainless steel and Teflon (hypotube), Makrolon (luer), and Pt/Ir alloy (radiopaque marker). The coating is hydrophilic, based on hyaluronic acid.
#### 4. PRINCIPLE OF OPERATION / TECHNOLOGY
M-CATH™ microcatheter is operated manually or by a manual process. The catheter should only be used by physicians trained in the performance of PCI.
#### 5. SPECIFICATIONS
| Usable length of catheter: | 135 cm |
|--------------------------------------------|------------------|
| Outer diameter of catheter (distal end): | 2.25 Fr (0.75mm) |
| Outer diameter of catheter (middle): | 3.30 Fr (1.10mm) |
| Outer diameter of catheter (proximal end): | 2.10 Fr (0.70mm) |
| Inner diameter of catheter (distal end): | 0.016" (0.40mm) |
| Inner diameter of catheter (proximal end): | 0.018" (0.45mm) |
| Radiopaque markers: | 1 |
#### 6. PERFORMANCE DATA
Bench testing was performed on the M-CATH™ microcatheter to determine substantial equivalence. The following testing/assessments were performed:
- 1) Dimensional Verification
- 2) Ancillary Device Compatibility Test
- Pushability and Trackability Testing 3)
- 4) Catheter Bond Strength
- Tip Pull Test 5)
- ର) Flexibility and Kink Test
- 7) Torque Strength Test
- 8) Coating Lubricity
- a) Particulate Evaluation – Coating Integrity
- 10) Particulate Evaluation - Coating Integrity (Coating Inspection)
- Radiopacity 11)
- Maximum Labeled Pressure 12)
- 13) Media Flow Rate
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Acrostak. The logo consists of a red plus sign on the left, followed by the word "Acrostak" in gray. A red line is underneath the word "Acrostak", and below the line is the text "Devices for interventional applications" in gray.
The in vitro bench tests demonstrated that M-CATH™ microcatheter met all acceptance criteria and performed similarly to the predicate devices. Performance data demonstrates that the device functions as intended, and is substantially equivalent to the predicate devices. The M-CATH™ microcatheter shelf life will be 3 years.
#### 7. BIOCOMPATIBILITY TESTING
Blood contacting materials were tested in accordance with the test recommendations in International Standard ISO 10993, "Biological Evaluation of Medical Devices - Part I: Evaluation and Testing." The blood contacting materials were found to be biocompatible. The following biocompatibility and chemical characterization tests were performed:
- Cytotoxicity .
- 트 Sensitization
- 트 Intracutaneous Reactivity/Irritation
- Acute Systemic Toxicity
- . Hemocompatibility
- . Hematology: ASTM Hemolysis
- Complement Activation Assay (Sc5b9)
- I In vivo thromboresistance
- 트 Chemical Characterization
#### 8. SUBSTANTIAL EQUIVALENCE
M-CATH™ microcatheter is substantially equivalent in intended use, principle of operation, technological characteristics, materials and performance to the Finecross® MG and Asahi Corsair, which are manufactured by Terumo Corporation and Asahi Intecc Co LTD, respectfully.
The submitted data demonstrates that the device is substantially equivalent as the legally marketed devices, and that the minor differences in materials and design do not raise different questions of substantial equivalence to the predicate devices.
| Submitter Name and Address: | Acrostak (Schweiz) AG<br>Stegackerstrasse 14<br>CH-8409 Winterthur,<br>Switzerland<br>T.+41 22 347 68 70<br>F.+41 52 233 95 50 |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Carmen Herraez<br>Clinical and Regulatory Affairs Manager<br>T.+41 (0)22 533 23 03<br>F.+41 52 233 95 50<br>carmen.herraez@acrostak.com |
### 9. SUBMITTER INFORMATION
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.