U-Lite is indicated for the visualization of structures and dynamic processes in the human body using and fluid flow analysis for diagnosis in the following clinical applications: Abdominal Small Parts Vascular Cardiac Obstetrics Gynecology Fetal Pediatrics
Device Story
U-Lite EXP is a notebook-sized, battery-operated, handheld diagnostic ultrasound system. It acquires and displays real-time ultrasound data via a 7-inch touch-activated LED screen. The device utilizes exchangeable probes to transmit ultrasonic energy into the body and processes received echoes to generate B-mode, M-mode, Color Doppler, Power Doppler, and Pulse Wave Doppler images. It is intended for use by clinicians for diagnostic imaging and fluid flow analysis across various clinical applications including abdominal, cardiac, and obstetric settings. The system supports DICOM connectivity and can be used with a mobile cart or table-top docking. Output is viewed directly on the tablet display to assist healthcare providers in clinical decision-making and diagnosis.
Clinical Evidence
No clinical study was conducted. Substantial equivalence is supported by bench testing, including acoustic output, biocompatibility, thermal, electrical, electromagnetic, and mechanical safety testing in accordance with recognized standards (e.g., IEC 60601-1, IEC 60601-2-37, ISO 10993).
Technological Characteristics
Handheld, battery-operated ultrasound system with 7-inch LED touch display. Supports exchangeable probes. Operates in B, M, Color Doppler, Power Doppler, and Pulse Wave Doppler modes. Complies with IEC 60601-1 (safety), IEC 60601-2-37 (ultrasonic diagnostic equipment), and ISO 10993 (biocompatibility). Connectivity includes DICOM.
Indications for Use
Indicated for visualization of human body structures and dynamic processes, including fluid flow analysis, for diagnostic purposes in abdominal, small parts, vascular, cardiac, obstetric, gynecologic, fetal, and pediatric clinical applications.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 3, 2017
Sonoscanner % E. J. Smith Consultant Smith Associates 1468 Harwell Avenue CROFTON MD 21114
Re: K171164
Trade/Device Name: U-Lite EXP Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: June 31, 2017 Received: August 1, 2017
Dear E. J. Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Michael D. O'Hara For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known)
#### K171164
Device Name U-Lite EXP
#### Indications for Use (Describe)
U-Lite is indicated for the visualization of structures and dynamic processes in the human body using and fluid flow analysis for diagnosis in the following clinical applications:
Abdominal Small Parts Vascular Cardiac Obstetrics Gynecology Fetal Pediatrics
Type of Use (Select one or both, as applicable)
[x] Prescription Use (Part 21 CFR 801 Subpart D)
] Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.
{3}------------------------------------------------
Probe: PR50: Convex probe for U-Lite EXP
Intended Use: Ultrasound imaging, measurement and analysis of the human body as follows:
| Clinical<br>Application | Mode of Operation | | | | | | | | |
|------------------------------------------|-------------------|---|-----|-----|------------------|------------------|-------------------------------|-----------------------------------|----------------------|
| | B | M | PWD | CWD | Color<br>Doppler | Power<br>Doppler | Tissue<br>Harmonic<br>Imaging | Combined*<br>(B+Color<br>Doppler) | Other**<br>(Specify) |
| Ophthalmic | | | | | | | | | |
| Fetal | N | N | N | | N | N | N | N | |
| Abdominal | N | N | N | | N | N | N | N | |
| Intra-operative (Specify) | | | | | | | | | |
| Intra-operative (Neuro) | | | | | | | | | |
| Laparoscopic | | | | | | | | | |
| Pediatric | N | | N | | N | N | N | N | |
| Small Organ<br>(breast, testes, thyroid) | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | |
| Adult Cephalic | | | | | | | | | |
| Trans-rectal | | | | | | | | | |
| Trans-vaginal | | | | | | | | | |
| Trans-urethral | | | | | | | | | |
| Trans-esoph. (non-Card) | | | | | | | | | |
| Musculo-skeletal<br>(Conventional) | N | N | N | | N | N | | N | |
| Musculo-skeletal<br>(Superficial) | | | | | | | | | |
| Intravascular | | | | | | | | | |
| Cardiac Adult | | | | | | | | | |
| Cardiac Pediatric | | | | | | | | | |
| Intravascular (Cardiac) | | | | | | | | | |
| Trans-esoph. (Cardiac) | | | | | | | | | |
| Intra-cardiac | | | | | | | | | |
| Other (Gynecological) | N | | | | N | N | N | N | |
| Peripheral Vessel | | | | | | | | | |
| Urology<br>(including prostate) | N | | | | N | N | N | N | |
| Clinical Application | Mode of Operation | | | | | | | | |
| | B | M | PWD | CWD | Color<br>Doppler | Power<br>Doppler | Tissue<br>Harmonic<br>Imaging | Combined*<br>(B+Color<br>Doppler) | Other**<br>(Specify) |
| Ophthalmic | | | | | | | | | |
| Fetal | | | | | | | | | |
| Abdominal | | | | | | | | | |
| Intra-operative (Specify) | | | | | | | | | |
| Intra-operative (Neuro) | | | | | | | | | |
| Laparoscopic | | | | | | | | | |
| Pediatric | N | N | N | | N | N | N | N | |
| Small Organ<br>(breast, testes, thyroid) | N | N | N | | N | N | N | N | |
| Neonatal Cephalic | N | N | N | | N | N | N | N | |
| Adult Cephalic | | | | | | | | | |
| Trans-rectal | | | | | | | | | |
| Trans-vaginal | | | | | | | | | |
| Trans-urethral | | | | | | | | | |
| Trans-esoph. (non-Card) | | | | | | | | | |
| Musculo-skeletal<br>(Conventional) | N | N | N | | N | N | N | N | |
| Musculo-skeletal<br>(Superficial) | N | N | N | | N | N | N | N | |
| Intravascular | | | | | | | | | |
| Cardiac Adult | | | | | | | | | |
| Cardiac Pediatric | | | | | | | | | |
| Intravascular (Cardiac) | | | | | | | | | |
| Trans-esoph. (Cardiac) | | | | | | | | | |
| Intra-cardiac | | | | | | | | | |
| Other (Specify) | | | | | | | | | |
| Peripheral Vessel | N | N | N | | N | N | N | N | |
| Urology<br>(including prostate) | | | | | | | | | |
| Clinical Application | Mode of Operation | | | | | | | | |
| | B | M | PWD | CWD | Color<br>Doppler | Power<br>Doppler | Tissue<br>Harmonic<br>Imaging | Combined*<br>(B+Color<br>Doppler) | Other**<br>(Specify) |
| Ophthalmic | | | | | | | | | |
| Fetal | | | | | | | | | |
| Abdominal | | | | | | | | | |
| Intra-operative (Specify) | | | | | | | | | |
| Intra-operative (Neuro) | | | | | | | | | |
| Laparoscopic | | | | | | | | | |
| Pediatric | N | N | N | | N | N | N | N | |
| Small Organ<br>(breast, testes, thyroid) | N | N | N | | N | N | N | N | |
| Neonatal Cephalic | | | | | | | | | |
| Adult Cephalic | | | | | | | | | |
| Trans-rectal | | | | | | | | | |
| Trans-vaginal | | | | | | | | | |
| Trans-urethral | | | | | | | | | |
| Trans-esoph. (non-Card) | | | | | | | | | |
| Musculo-skeletal<br>(Conventional) | N | N | N | | N | N | N | N | |
| Musculo-skeletal<br>(Superficial) | N | N | N | | N | N | N | N | |
| Intravascular | | | | | | | | | |
| Cardiac Adult | | | | | | | | | |
| Cardiac Pediatric | | | | | | | | | |
| Intravascular (Cardiac) | | | | | | | | | |
| Trans-esoph. (Cardiac) | | | | | | | | | |
| Intra-cardiac | | | | | | | | | |
| Other (Specify) | | | | | | | | | |
| Peripheral Vessel | N | N | N | | N | N | N | N | |
| Urology<br>(including prostate) | | | | | | | | | |
| Clinical Application | Mode of Operation | | | | | | | | |
| | B | M | PWD | CWD | Color<br>Doppler | Power<br>Doppler | Tissue<br>Harmonic<br>Imaging | Combined*<br>(Specify) | Other**<br>(Specify) |
| Ophthalmic | | | | | | | | | |
| Fetal | N | N | N | | N | N | N | N | |
| Abdominal | | | | | | | | | |
| Intra-operative (Specify) | | | | | | | | | |
| Intra-operative (Neuro) | | | | | | | | | |
| Laparoscopic | | | | | | | | | |
| Pediatric | | | | | | | | | |
| Small Organ<br>(breast, testes, thyroid) | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | |
| Adult Cephalic | | | | | | | | | |
| Trans-rectal | N | | N | | N | N | N | N | |
| Trans-vaginal | N | N | N | | N | N | N | N | |
| Trans-urethral | | | | | | | | | |
| Trans-esoph. (non-Card) | | | | | | | | | |
| Musculo-skeletal<br>(Conventional) | | | | | | | | | |
| Musculo-skeletal<br>(Superficial) | | | | | | | | | |
| Intravascular | | | | | | | | | |
| Cardiac Adult | | | | | | | | | |
| Cardiac Pediatric | | | | | | | | | |
| Intravascular (Cardiac) | | | | | | | | | |
| Trans-esoph. (Cardiac) | | | | | | | | | |
| Intra-cardiac | | | | | | | | | |
| Other (Gynecological) | N | | N | | N | N | N | N | |
| Peripheral Vessel | | | | | | | | | |
| Urology<br>(including prostate) | N | | N | | N | N | N | N | |
| Clinical Application | Mode of Operation | | | | | | | | |
| | B | M | PWD | CWD | Color<br>Doppler | Power<br>Doppler | Tissue<br>Harmonic<br>Imaging | Combined*<br>(B+Color<br>Doppler) | Other**<br>(Specify) |
| Ophthalmic | | | | | | | | | |
| Fetal | N | N | N | | N | N | N | N | |
| Abdominal | N | | N | | N | N | N | N | |
| Intra-operative (Specify) | | | | | | | | | |
| Intra-operative (Neuro) | | | | | | | | | |
| Laparoscopic | | | | | | | | | |
| Pediatric | N | | N | | N | N | N | N | |
| Small Organ<br>(breast, testes, thyroid) | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | |
| Adult Cephalic | | | | | | | | | |
| Trans-rectal | | | | | | | | | |
| Trans-vaginal | | | | | | | | | |
| Trans-urethral | | | | | | | | | |
| Trans-esoph. (non-Card) | | | | | | | | | |
| Musculo-skeletal<br>(Conventional) | N | | N | | N | N | N | N | |
| Musculo-skeletal<br>(Superficial) | | | | | | | | | |
| Intravascular | | | | | | | | | |
| Cardiac Adult | N | N | N | | N | | N | N | |
| Cardiac Pediatric | N | N | N | | N | | N | N | |
| Intravascular (Cardiac) | | | | | | | | | |
| Trans-esoph. (Cardiac) | | | | | | | | | |
| Intra-cardiac | | | | | | | | | |
| Other (Gynecological) | N | | | | N | N | N | N | |
| Peripheral Vessel | | | | | | | | | |
| Urology<br>(including prostate) | N | | | | N | N | N | N | |
| Clinical Application | Mode of Operation | | | | | | | | |
| | B | M | PWD | CWD | Color<br>Doppler | Power<br>Doppler | Tissue<br>Harmonic<br>Imaging | Combined*<br>(B+Color<br>Doppler) | Other**<br>(Specify) |
| Ophthalmic | | | | | | | | | |
| Fetal | N | N | N | N | N | N | N | N | |
| Abdominal | N | N | N | N | N | N | N | N | |
| Intra-operative (Specify) | | | | | | | | | |
| Intra-operative (Neuro) | | | | | | | | | |
| Laparoscopic | | | | | | | | | |
| Pediatric | N | N | N | N | N | N | N | N | |
| Small Organ<br>(breast, testes, thyroid) | N | N | N | N | N | N | N | N | |
| Neonatal Cephalic | N | N | N | N | N | N | N | N | |
| Adult Cephalic | | | | | | | | | |
| Trans-rectal | | | | | | | | | |
| Trans-vaginal | | | | | | | | | |
| Trans-urethral | | | | | | | | | |
| Trans-esoph. (non-Card) | | | | | | | | | |
| Musculo-skeletal<br>(Conventional) | N | N | N | N | N | N | N | N | |
| Musculo-skeletal<br>(Superficial) | N | N | N | N | N | N | N | N | |
| Intravascular | | | | | | | | | |
| Cardiac Adult | | | | | | | | | |
| Cardiac Pediatric | | | | | | | | | |
| Intravascular (Cardiac) | | | | | | | | | |
| Trans-esoph. (Cardiac) | | | | | | | | | |
| Intra-cardiac | | | | | | | | | |
| Other (Specify) | | | | | | | | | |
| Peripheral Vessel | N | N | N | N | N | N | N | N | |
| Urology<br>(including prostate) | N | N | N | N | N | N | N | N | |
{4}------------------------------------------------
Probe: PR51: Linear probe 40mm for U-Lite EXP
Intended Use: Ultrasound imaging, measurement and analysis of the human body as follows:
{5}------------------------------------------------
Probe: PR52: Linear probe 50mm for U-Lite EXP
Intended Use: Ultrasound imaging, measurement and analysis of the human body as follows:
{6}------------------------------------------------
Probe: PR53: Endocavitary probe for U-Lite EXP
Intended Use: Ultrasound imaging, measurement and analysis of the human body as follows:
{7}------------------------------------------------
Probe: PR54: Phased Array probe for U-Lite EXP
Intended Use: Ultrasound imaging, measurement and analysis of the human body as follows:
{8}------------------------------------------------
Probe: PR55: MicroConvex probe for U-Lite EXP
Intended Use: Ultrasound imaging, measurement and analysis of the human body as follows:
{9}------------------------------------------------
# 510(k) Summary
| Sponsor: | Sonoscanner |
|-------------------|-------------------------------------------------------|
| Address: | 6 rue Andre Voguet<br>Ivry Sur Seine, France 94200 |
| Telephone Number: | +33 954-971-557 |
| Email Address: | richard@sonoscanner.com |
| Contact Person: | Bruno Richard<br>Scientific Director, Quality Manager |
Summary Preparation Date: July 31, 2017
DEVICE NAME:
| Trade Name: | U-Lite EXP |
|----------------------|------------------------------------------------------|
| Common/Usual Name: | Ultrasonic Pulsed Doppler Imaging System |
| Classification Name: | Ultrasonic Pulsed Doppler Imaging System |
| Regulation Number: | 21 CFR 892.1550, 21 CFR 892.1560 and 21 CFR 892.1570 |
| Product Code: | IYN, IYO and ITX |
| Device Class: | Class II |
| Panel: | Radiology |
DEVICE DESCRIPTION:
The U-Lite EXP are notebook-size, battery operated, general purpose track 3 diagnostic ultrasound system. It is used to acquire and display high-resolution, real-time ultrasound data and display the data as B Mode, Color Mode, Power Doppler Imaging, and Pulse Wave Doppler spectrum.
The handheld U-Lite EXP tablet's display is a high-resolution 7-inch color LED screen and the controls are intuitive and easy to use. Controls are touch-activated – there are no knobs or switches. The U-Lite EXP is equipped with a lithium-ion battery. When fully charged, the battery can give the system a total autonomy of up to 1h30 in the scanning mode.
{10}------------------------------------------------
#### INDICATIONS FOR USE:
U-Lite is indicated for the visualization of structures and dynamic processes in the human body using ultrasound imaging and fluid flow analysis for diagnosis in the following clinical applications:
Abdominal
Small Parts
Vascular
Cardiac
Obstetrics
Gynecology
Fetal
Pediatrics
PREDICATE PRODUCT COMPARISON:
#### Predicate Product Comparison Table
| DEVICE COMPARISON TABLE | | | |
|------------------------------------|---------------------------------------------------------------------------|---------------------------------------------------------------------------|-----------|
| Parameters | U-Lite K143601 | U-Lite EXP | Comment |
| Intended Use | Intended for diagnostic<br>ultrasound analysis<br>and fluid flow analysis | Intended for diagnostic<br>ultrasound analysis and<br>fluid flow analysis | Identical |
| Track | Track 3 | Track 3 | Identical |
| Dimensions | 7.5 x 5.3 x 0.8<br>inches | 7.5 x 5.3 x 0.8<br>inches | Identical |
| Weight | 1.8 lbs | 1.8 lbs | Identical |
| Configuration/Design | Notebook,<br>handheld | Notebook,<br>handheld | Identical |
| Battery Life | 1hr 30mins | 1hr 30mins | Identical |
| Display Size | 7in | 7in | Identical |
| Table Top Docking | Yes | Yes | Identical |
| Mobile Cart | No | Yes | Added |
| Scanning Modes | | | |
| B Mode | Y | Y | Identical |
| Color Mode | Y | Y | Identical |
| Power Doppler | Y | Y | Identical |
| M Mode | Y | Y | Identical |
| Pulse Wave Doppler | Y | Y | Identical |
| Indications | | | |
| Ophthalmic | N | N | Identical |
| Fetal | Y | Y | Identical |
| Abdominal | Y | Y | Identical |
| Intra-operative (Specify) | N | N | Identical |
| Intra-operative (Neuro) | N | N | Identical |
| Laparoscopic | N | N | Identical |
| Pediatric | Y | Y | Identical |
| Small Organ | Y | Y | Identical |
| Neonatal Cephalic | N | Y | No |
| Adult Cephalic | N | N | Identical |
| Trans-rectal | Y | Y | Identical |
| Trans-vaginal | Y | Y | Identical |
| Trans-urethral | N | N | Identical |
| Trans-esoph. (Non-Card.) | N | N | Identical |
| Musculo-skeletal<br>(Conventional) | Y | Y | Identical |
| Musculo-skeletal<br>(Superficial) | Y | Y | Identical |
| Intravascular | N | N | Identical |
| Cardiac Adult | Y | Y | Identical |
| Cardiac Pediatric | Y | Y | Identical |
| Intravascular (Cardiac) | N | N | Identical |
| Trans-esoph. (Cardiac) | N | N | Identical |
| Intra-cardiac | N | N | Identical |
| Gynecological | Y | Y | Identical |
| Peripheral Vessel | Y | Y | Identical |
| Urology (Including prostate) | Y | Y | Identical |
| Integrated Speaker | Y | Y | Identical |
| DICOM | N | Y | Identical |
{11}------------------------------------------------
{12}------------------------------------------------
### Summary of the Technological Characteristics of the Device Compared to the Predicate Device
The U-Lite EXP is comparable to the U-Lite (K143601) in technological characteristics and operating principle. The U-Lite EXP offers the advantage of exchangeable probes versus fixed probes of the predicate. Both devices transmit ultrasonic energy into patients, then perform post processing of received echoes to generate on-screen display of anatomic structures and fluid flow within the body, and have similar intended use and basic operating modes. Both systems allow for specialized measurements of structures and flow and calculations.
### Performance Data:
The device has been evaluated for acoustic output, biocompatibility, as well as for thermal, electrical, electromagnetic and mechanical safety, and has been found to conform to applicable medical device safety standards. The U-Lite complies with the voluntary standards:
{13}------------------------------------------------
- . EN ISO 13485:2012, EN ISO 13485:2012/AC:2012 Medical devices - Quality management systems - Requirements for regulatory purposes (ISO 13485:2003)
- . EN 60601-1:2006 + A1:2013 Medical electrical equipment - Part 1: General requirements for safety (IEC 60601-1:2005, IEC 60601-1:2005/A1:2012)
- . EN 60601-1-1:2001 + A1:2006 Medical electrical equipment - Part 1-1: General requirements for safety - Collateral standard: Safety requirements for medical electrical systems (IEC 60601-1-1:2000)
- . EN 60601-1-2:2007 + AC:2010 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests (IEC 60601-1-2:2007 (Modified))
- EN 60601-1-6:2010Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability (IEC 60601-1-6:2010)
- . EN 60601-2-37:2008 Medical electrical equipment - Part 2-37: Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment (IEC 60601-2-37:2007)
- . EN ISO 14971:2012 Medical devices - Application of risk management to medical devices (ISO 14971:2007)
- . EN 62304:2006 Medical device software - Software life-cycle processes (IEC 62304:2006)
- . EN 62366:2008 Medical devices - Application of usability engineering to medical devices (IEC 62366:2007)
- EN ISO 10993-1:2009 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process(ISO 10993-1:2009)
- . EN ISO 10993-5:2009 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity (ISO 10993-5:2009)
- . EN ISO 10993-10: 2010 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization (ISO 10993-10: 2010)
- EN ISO 10993-11:2009 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity (ISO 10993-11:2006)
{14}------------------------------------------------
### Clinical Study:
No clinical study was conducted.
#### Conclusion:
Based upon the testing and comparison to the predicate device, the Sonoscanner U-Lite EXP Diagnostic Ultrasound Device has similar intended use, identical technological characteristics and operating principle. The U-Lite EXP is substantially equivalent to the predicate U-Lite and does not raise any new safety and effectiveness issues.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.