# Dimensional Bioceramics Calcium Sulfate Device (DB-CSD) (K171161)

_Dimensional Bioceramics, LLC · MQV · Sep 25, 2017 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K171161

## Device Facts

- **Applicant:** Dimensional Bioceramics, LLC
- **Product Code:** [MQV](/productcode/MQV.md)
- **Decision Date:** Sep 25, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3045
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

Dimensional Bioceramics™ (DB) Calcium Sulfate Device (CSD) is a bone graft substitute that is intended for use as a bone void filler for voids and gaps and pelvis that are not intrinsic to the bony structure. These osseous defects may be surgically created or created from a traumatic injury to the bone. DB-CSD is intended to be gently packed into bony voids or gaps of the skeletal system (i.e., extremities and pelvis). DB-CSD is a bone graft substitute that resorbs and replaced with bone during the healing process. DB-CSD is biodegradable and may be used in an infected site.

## Device Story

Bone graft substitute kit; contains medical grade calcium sulfate powder and mixing solution. Clinician mixes components in sterile bowl to create paste; paste injected, digitally implanted, or molded into pellets. Applied directly to bony voids or gaps in extremities and pelvis. Resorbs and is replaced by bone during healing process. Biodegradable; suitable for use in infected sites. Benefits patient by filling voids to facilitate natural bone healing.

## Clinical Evidence

No clinical data. Evidence includes bench testing (FTIR, XRD, SEM, Mercury Intrusion Porosimetry, in vitro dissolution) and an ovine critical-size cancellous bone defect model. Animal study confirmed biocompatibility, resorption, and bone formation at 12 weeks, showing no significant differences from the predicate.

## Technological Characteristics

Resorbable calcium sulfate bone void filler. Supplied as powder and mixing solution kit. Critical specs: chemistry, crystallinity, physical form, porosity, solubility. Sterilized via radiation (25-40 kGy) per ISO 11137-2006 to SAL 10^-6. Biocompatibility per ISO 10993-1.

## Regulatory Identification

A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.

## Predicate Devices

- Stimulan Rapid Cure ([K141830](/device/K141830.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

### DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Dimensional Bioceramics™ LLC % Patsy J. Trisler, JD, RAC Regulatory Consultant Trisler Consulting 5600 Wisconsin Avenue #509 Chevy Chase, Maryland 20815

September 25, 2017

Re: K171161

Trade/Device Name: Dimensional Bioceramics Calcium Sulfate Device (DB-CSD) Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: Class II Product Code: MOV Dated: August 14, 2017 Received: August 17, 2017

Dear Ms. Trisler:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR

{1}------------------------------------------------

Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

Katherine D. Kavlock -S

for

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K171161

Device Name Dimensional Bioceramics Calcium Sulfate Device (DB - CSD)

Indications for Use (Describe)

Dimensional Bioceramics™ (DB) Calcium Sulfate Device (CSD) is a bone graft substitute that is intended for use as a bone void filler for voids and gaps and pelvis that are not intrinsic to the bony structure. These osseous defects may be surgically created or created from a traumatic injury to the bone. DB-CSD is intended to be gently packed into bony voids or gaps of the skeletal system (i.e., extremities and pelvis). DB-CSD is a bone graft substitute that resorbs and replaced with bone during the healing process. DB-CSD is biodegradable and may be used in an infected site.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{3}------------------------------------------------

# K171161 510(k) SUMMARY DB-CSD

| I. SUBMITTER                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|-------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter Name:                                                                     | Dimensional Bioceramics™ LLC                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Submitter Address:                                                                  | 163 McKenzie Creek Road<br>Scotts Valley, CA 95066                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Contact Person:<br>Telephone #:                                                     | David Delaney, Director<br>831.234.4892                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Date Prepared:                                                                      | August 14, 2017                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| II. DEVICE                                                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Device Trade Name:                                                                  | Dimensional Bioceramics Calcium Sulfate Device<br>(DB-CSD)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Regulatory Name(s):                                                                 | Resorbable Calcium Sulfate Bone Void Filler                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Classification #:                                                                   | 21 CFR 888.3045                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Product Code:                                                                       | MQV                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| III. PREDICATE DEVICE(s)                                                            | K141830, Stimulan Rapid Cure, Biocomposites, Ltd.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| IV. DEVICE DESCRIPTION                                                              |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Device Identification,<br>Characteristics,<br>Mechanism of Action:                  | Dimensional Bioceramics (DB) Calcium Sulfate Device<br>(CSD) [DB-CSD] Bone Void Filler Kit is provided sterile<br>for single patient use. The Kit contains medical grade<br>calcium sulfate powder and mixing solution in pre-<br>measured quantities. When mixed together in a sterile<br>mixing bowl, the resultant paste may be injected, digitally<br>implanted, or applied to a mold to produce pellets.<br><br>DB-CSD resorbs and is replaced with bone during the<br>healing process.<br><br>DB-CSD is a moldable, biocompatible, biodegradable,<br>and resorbable calcium sulfate material that is to be<br>applied directly to the intended sites. It may be used<br>within an infected site.<br><br>The critical specifications are chemistry, crystallinity,<br>physical form, porosity, and solubility. |
| V. INDICATIONS FOR USE                                                              | Dimensional Bioceramics™ (DB) Calcium Sulfate Device<br>(CSD) is a bone graft substitute that is intended for use<br>as a bone void filler for voids and gaps and pelvis that<br>are not intrinsic to the stability of the bony structure.<br>These osseous defects may be surgically created or<br>created from a traumatic injury to the bone. DB-CSD is<br>intended to be gently packed into bony voids or gaps of<br>the skeletal system (i.e., extremities and pelvis). DB-<br>CSD is a bone graft substitute that resorbs and replaced<br>with bone during the healing process. DB-CSD is<br>biodegradable and biocompatible and may be used in an<br>infected site.                                                                                                                                          |
| VI. SUMMARY OF TESTING<br>[PERFORMANCE DATA]                                        | Performance, Safety and Biocompatibility test data are<br>provided in the 510(k) application. They include the<br>following:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Laboratory Testing                                                                  | Critical specifications (chemistry, crystallinity, physical<br>form, porosity, and solubility) of DB-CSD were compared<br>with those of the predicate device. Chemistry was<br>determined by Fourier Transformed Infrared<br>Spectroscopy (FTIR) and X-ray Diffraction (XRD)<br>techniques. Crystallinity was determined by X-ray<br>Diffraction. Physical form was determined by Scanning<br>Electron Microscopy. Porosity was determined by<br>Mercury Intrusion Porosimetry. Solubility was measured<br><i>in vitro</i> dissolution method measuring Ca2+ ion<br>concentration in solution using Inductively Coupled<br>Plasma - Atomic Emission Spectroscopy.                                                                                                                                                   |
| Biocompatibility Testing:                                                           | Biocompatibility tests, according to ISO 10993-1 were<br>performed and are provided in the 510(k)<br>documentation. They demonstrated DB-CSD met the<br>requirements of the ISO standards.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Pyrogenicity Testing:                                                               | ISO Materials Mediated Rabbit Pyrogen testing was<br>performed and provided in the 510(k). The results of the<br>testing indicate the DB-CSD is non-pyrogenic.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Sterilization and Shelf-Life<br>Testing                                             | DB-CSD will be provided as a single use, sterile medical<br>device. The radiation dose of 25k Gy-40 kGy has been<br>validated in accordance with ISO 11137-2006,<br>Sterilization of Health Care Products - Radiation to<br>Sterility Assurance Level (SAL) 10-6.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Animal Testing:                                                                     | An animal study was performed to analyze the<br>biocompatibility, implant resorption, bone formation, and<br>surgical handling properties following metaphyseal<br>implantation in an ovine, critical size, cancellous bone<br>defect model. No abnormal acute hematologic and<br>major organ toxicity was observed in both subject and<br>predicate devices. Histological and radiographic data<br>demonstrated that both materials appeared to be                                                                                                                                                                                                                                                                                                                                                                 |
|                                                                                     | biocompatible, and complete resorption of the implanted<br>regions, followed by normal bone healing occurred for<br>both subject and predicate devices by 12 weeks.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Clinical Testing:                                                                   | N/A: This product type does not require clinical testing.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| VII. COMPARISON OF<br>TECHNOLOGICAL<br>CHARACTERISTICS WITH<br>THE PREDICATE DEVICE | DB-CSD intended use and critical specifications are<br>substantially equivalent to the predicate device, Stimulan<br>Rapid Cure of Biocomposites Ltd (K141830).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                                                                     | Both medical devices are composed of a hydrated<br>calcium sulfate salt that is implanted following mixing<br>with a solution component that yields 5, 10, or 20 cc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|                                                                                     | Performance test results demonstrated that DB-CSD has<br>substantially equivalent critical specifications (chemistry,<br>crystallinity, physical form, porosity, and solubility) as the<br>predicate device, Stimulan Bone Void Filler.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| VIII. CONCLUSIONS                                                                   | Based on the comparison provided and the data<br>submitted in the 510(k), it can be concluded the DB-CSD<br>is substantially equivalent to the predicate device,<br>Stimulan - Rapid Cure Bone Void Filler (K141830).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |

{4}------------------------------------------------

{5}------------------------------------------------

---

**Source:** [https://fda-staging.innolitics.com/device/K171161](https://fda-staging.innolitics.com/device/K171161)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).