K171131 · Mazza Healthcare, LLC · FMF · Nov 8, 2017 · General Hospital
Device Facts
Record ID
K171131
Device Name
MHC Standard and NRFit Tip Syringes
Applicant
Mazza Healthcare, LLC
Product Code
FMF · General Hospital
Decision Date
Nov 8, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The MHC Standard and NRFit™ Tip Syringes are intended to inject fluids into or withdraw fluids from the body.
Device Story
Single-use piston syringes; rigid polycarbonate barrels; plungers with synthetic rubber gaskets; manual operation. Input: manual force applied to plunger to express or withdraw fluids. Output: fluid delivery or aspiration. Neuraxial models feature ISO 80369-6 compliant tips to prevent misconnections. Used in clinical settings by healthcare professionals. Graduated markings on barrel allow volume measurement. Benefits include standardized neuraxial connections to reduce misconnection risks.
Clinical Evidence
No clinical or animal testing performed. Substantial equivalence demonstrated via bench testing, including ISO 80369-6/20 compliance (visual inspection, stress cracking, fluid leakage, subatmospheric pressure air leakage, axial load resistance) and biocompatibility per ISO 10993-1.
Technological Characteristics
Materials: Polycarbonate barrel, ABS plunger, synthetic rubber/elastomer gasket, medical grade oil lubricant. Dimensions: 5cc and 10cc volumes. Connectivity: ISO 80369-6 compliant neuraxial tips or standard luer slip/lock. Sterilization: EtO. No software or electronic components.
Indications for Use
Indicated for the injection of fluids into or withdrawal of fluids from the body. No specific patient population, age, or gender restrictions are defined.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
DMC Medical Single-Use Polycarbonate Syringe (K103736)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 8, 2017
Mazza Healthcare, LLC. % Jon Gilbert Regulatory Consultant to Mazza Healthcare, LLC Jon Gilbert 1641 Jeurissen Lane Chanhassen, Minnesota 55317
Re: K171131
Trade/Device Name: MHC Standard and NRFit™ Tip Syringes Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: FMF Dated: September 29, 2017 Received: October 2, 2017
Dear Jon Gilbert:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Tina
Kiang-S
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K171131
#### Device Name
MHC Standard and NRFit™ Tip Syringes
Indications for Use (Describe)
The MHC Standard and NRFitM Tip Syringes are intended to inject fluids into or withdraw fluids from the body.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Section 5. 510(k) Summary
Submitter
| Sponsor: | Mazza Healthcare, LLC<br>2101 Waukegan Road, Suite 208<br>Bannockburn, IL 60015<br>Jon Gilbert, Consultant<br>+01 (906) 361-3237<br>igilb.raca@gmail.com | |
|-------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Contact: | | |
| Date of Preparation: | September 29, 2017 | |
| Subject Device:<br>Common Name:<br>Trade Name:<br>Requlation Number:<br>Regulation Name:<br>Review Panel:<br>Regulatory Class:<br>Product Code: | Standard and Neuraxial Tip Syringes<br>MHC Standard and NRFit™ Tip Syringes<br>21 CFR 880.5860<br>Piston syringe<br>General Hospital<br>Class II<br>FMF | |
| Primary Predicate:<br>510(k): | Vesco Medical NRFit™ Tip Syringes<br>K170218 cleared February 24, 2017 | |
| Predicate Device:<br>510(k): | DMC Medical Single-Use Polycarbonate Syringe<br>K103736 cleared June 10, 2011 | |
Device Description:
The MHC Standard and NRFit™ Tip Syringes are provided sterile or in bulk non-sterile for further processing (e.g., sterilization). They are single use devices consisting of rigid polycarbonate barrels and plungers with a synthetic rubber gasket. The syringe barrels with neuraxial tips are ISO 80369 - 6 compliant and allow for connections of neuraxial specific applications while reducing the likelihood of misconnections to non-neuraxial devices. The syringe barrels are printed with graduated markings in cc (millliters) indicating the volume of liquid inside the syringe barrel. The proposed models for the standard and neuraxial devices are listed below in Table 1.
| MHC Model # | Item Description |
|-------------|----------------------------------------------------------|
| 09-050-WH | 5cc Polycarbonate Luer Slip Tip Syringe W/Silicone |
| 09-060-WH | 5cc Polycarbonate Luer Lock Tip Syringe W/Silicone |
| 09-055-WHN | 5cc Polycarbonate Neuraxial Slip Tip Syringe W/Silicone |
| 09-010-WH | 10cc Polycarbonate Luer Slip Tip Syringe w/Silicone |
| 09-120-WH | 10cc Polycarbonate Luer Lock Tip Syringe w/Silicone |
| 09-015-WHN | 10cc Polycarbonate Neuraxial Slip Tip Syringe w/Silicone |
#### Table 1. MCH Models and Description
{4}------------------------------------------------
A difference between the predicate syringe (K103736) without a neuraxial tip and the subject syringe with a neuraxial tip is the design of the needle s hub. While the predicate devices (K103736) are designed with a luer slip tip (taper) or luer lock connector the subject neuraxial devices are equipped with a NRFit™ neuraxial connection as standardized in ISO 80369-6. Additionally, the NRFit™ Tip Syringe piston is yellow.
The clinical technique, indications for use, technical specifications, materials used, and sterility status (validation and sterility assurance level) are identical between the subject and predicate Standard syringes. As stated above, the NRFit™ tip syringes provides connections of neuraxial specific applications while reducing the likelihood of misconnections to non-neuraxial devices
## Indications for Use:
The MHC Standard and NRFit™ Tip Syringes are intended to inject fluids into or withdraw fluids from the body.
## Performance Testing:
The performance of the subject MHC Standard Tip Syringes is identical to the predicate devices as no material (with the exception of the yellow piston), manufacturing or technological changes have occurred. The performance of the subject MHC NRFit™ Tip Syringes is similar to the predicate neuraxial syringes (K170218) as both designs are based upon ISO 80369-6, Small bore connectors for liquids and gases in healthcare applications - Part 6: Connectors for neuraxial applications.
## Non-Clinical Tests
Verification and validation testing was performed with the subject neuraxial (NRFit) Tip Syringes. It was found that subject neuraxial (NRFit) Tip Syringes are in compliance with design and performance requirements according to ISO 80369-6:2016 "Small-bore connectors for liquids and gases in healthcare applications - Part 6: Connectors for neuraxial applications," and ISO 80369-20:2016 "Small-bore connectors for liquids and gases in healthcare applications - Part 20: Common Test Methods, including Visual Inspection (burrs, hooks, cracks, foreign contamination, missing components), Stress Cracking per ISO 80369-6 (sec. 6.3) and ISO 80369-20 (Annex E), Fluid Leakage by Pressure Decay per ISO 80369-20 (Annex B), Subatmospheric Pressue Air Leakage per ISO 80369-20 (Annex D), and, Resistance to Separation from Axial Load per ISO 80369-20 (Annex F). Biocompatibility testing has demonstrated the subject devices meet the guidelines as described in the FDA guidance document entitled, "Use of International Standard ISO10993-1, " Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process".
## Clinical and Animal Tests
Clinical tests were not required to demonstrate performance of between the subject and predicate devices. Product functionality has been adequately assessed by non-clinical tests.
{5}------------------------------------------------
| Component: | Vesco Medical<br>K170218 | DMC Medical,<br>Ltd, K103736 | MHC Standard<br>and Neuraxial* |
|--------------------------------------|---------------------------------------------|------------------------------|--------------------------------|
| Syringe Barrel: | Polypropylene | Polycarbonate | Polycarbonate |
| Plunger Tip: | Synthetic Rubber | Elastomer | Elastomer |
| Plunger | Polypropylene | ABS polymer | ABS polymer |
| Syringe Piston Lube<br>Coating | Polydimethylsiloxane,<br>Silanol Terminated | Medical Grade<br>Oil | Medical Grade<br>Oil |
| Calibrated Barrel<br>Volume: | Yes | Yes | Yes |
| Recommended<br>Sterilization method: | EtO | EtO | EtO |
| 510(k) Approval | K170218 | K103736 | K171131 |
Table 2 Technological Characteristics & Substantial Equivalence
*Subiect device Standard and NRFit components are comprised of same materials and manufactured using same manufacturing methods. K103736 best represents predicate materials and components while K170218 and ISO 80369-6 represents requisite neuraxial tip dimensional configurations and technological characteristics and standards.
The operational characteristics, when compared to the predicate device, are identical. The user connects the syringe via a standard slip tip or threaded luer lock, then manually advances or withdraws the plunger internal to the barrel to express or withdraw fluids. Fluids are measured via the printed external housing of the barrel: measurements are indicated in cc (milliliters). Operation is similar for most all piston svringes whether fitted with a threaded luer end, a slip-fit tip or neuraxial tip.
The subject device standard syringes are similar in materials and performance to the Sponsor's currently distributed standard svringes cleared in K103736. The technological characteristics of the subject device neuraxial syringes are consistent with ISO standard and are equivalent to the predicate device cleared in K170218.
## Conclusion:
The MHC Standard and NRFit™ Tip Syringes are intended to inject fluids into or
withdraw fluids from the body. The materials, intended use, technological and operational characteristics of the subject devices are substantially equivalent to the predicate Standard Syringes of K103736 and the neuraxial syringes cleared in K170218.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.