K171111 · Cynosure, Inc. · PKT · Jun 13, 2017 · General, Plastic Surgery
Device Facts
Record ID
K171111
Device Name
Sculpsure
Applicant
Cynosure, Inc.
Product Code
PKT · General, Plastic Surgery
Decision Date
Jun 13, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Cynosure SculpSure™ is intended for non-invasive lipolysis of the abdomen, flanks, back, and thighs in individuals with a Body Mass Index (BMI) of 30 or less. The device is intended to affect the appearance of visible fat bulges in the abdomen, flanks, back, and thighs.
Device Story
SculpSure is a diode laser system for non-invasive lipolysis; utilizes 1060 nm infrared laser energy to heat subcutaneous adipose tissue; console-based system with four applicators secured to patient via belt; operated by clinicians in aesthetic settings; laser energy disrupts adipocyte cells; healthcare providers use device to reduce visible fat bulges; patient benefits include non-surgical body contouring; minor software updates from predicate; no changes to biocompatibility or electrical safety.
Clinical Evidence
Prospective, controlled study at 3 centers; 168 subjects (214 treatment areas: back, outer thigh, inner thigh). Primary endpoints: blind evaluation of pre/post photos, ultrasound-measured adipose tissue thickness change, and 6-point Likert patient satisfaction scale at 12 weeks. Results: 86% of post-treatment photos correctly identified; 8.6% mean normalized fat reduction (p<0.01); 83% patient satisfaction rate. Safety analysis showed all adverse events were transient and mild; no serious adverse events reported.
Technological Characteristics
Diode laser system; 1060 ±20 nm wavelength; CW pulse width; max power density 1.4 W/cm²; 4x6 cm² spot size per applicator; belt-based patient attachment; 200-240V power; software-controlled; non-invasive.
Indications for Use
Indicated for non-invasive lipolysis of abdomen, flanks, back, and thighs in individuals with BMI ≤ 30 to reduce visible fat bulges.
Regulatory Classification
Identification
A Low Level Laser System for Aesthetic Use is a device using low level laser energy for the disruption of adipocyte cells within the fat layer for the release of fat and lipids from these cells for noninvasive aesthetic use.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Industry and Food and Drug Administration Staff; Class II Special Controls Guidance Document: Low Level Laser System for Aesthetic Use.” See § 878.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 13, 2017
Cynosure, Inc Ms. Amy Tannenbaum Regulatory Affairs Specialist 5 Carlisle Road Westford, Massachusetts 01886
Re: K171111
Trade/Device Name: Sculpsure Regulation Number: 21 CFR 878.5400 Regulation Name: Low level laser system for aesthetic use Regulatory Class: Class II Product Code: PKT Dated: April 13, 2017 Received: April 14, 2017
Dear Ms. Tannenbaum:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Actinclude requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Jennifer R. Stevenson -S3
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K171111
Device Name SculpSure™
Indications for Use (Describe)
The Cynosure SculpSure™ is intended for non-invasive lipolysis of the abdomen, flanks, back, and thighs in individuals with a Body Mass Index (BMI) of 30 or less. The device is intended to affect the appearance of visible fat bulges in the abdomen, flanks, back, and thighs.
Type of Use (*Select one or both, as applicable*)
| <div> <span> <svg height="16" width="16"> <rect fill="white" height="16" stroke="black" stroke-width="1" width="16" x="0" y="0"></rect> <line stroke="black" stroke-width="2" x1="2" x2="14" y1="2" y2="14"></line> <line stroke="black" stroke-width="2" x1="14" x2="2" y1="2" y2="14"></line> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <svg height="16" width="16"> <rect fill="white" height="16" stroke="black" stroke-width="1" width="16" x="0" y="0"></rect> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Section 5
510(k) Summary
## 510(k) Summary for SculpSure Laser System
A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92.
| 807.92(a)(1) Submitter Information | |
|--------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant | Cynosure, Inc. |
| Address | 5 Carlisle Road<br>Westford, MA 01886 |
| Phone Number | (781)-993-2454 |
| Fax Number | (978) 256-6556 |
| Establishment Registration<br>Number | 1222993 |
| Contact Person | Amy Tannenbaum |
| Preparation Date | April 13, 2017 |
| 807.92(a)(2) Name of Device | |
| Trade or Proprietary Name | SculpSure |
| Common or Usual Name | Laser |
| Classification Name | Laser for disruption of adipocyte cells for aesthetic use |
| Classification Panel | General & Plastic Surgery |
| Regulation | 21 CFR 878.5400 |
| Regulatory Class | II |
| Product Code(s) | PKT |
| 807.92 (a)(3) Legally marketed device(s) to which equivalence is claimed | |
| Predicate Devices | Cynosure SculpSure K160470<br>Zeltiq Coolsculpting K162050<br>The predicate devices have not been subject to a design-related<br>recall. |
| 807.92(a)(4) Device Description | |
| | The Cynosure SculpSure is a diode laser system. The main<br>components of SculpSure are a console and four applicators that<br>deliver the laser energy to the patient. Electrically efficient<br>semiconductors generate optical radiation (1060 nm) which is<br>used to deliver laser energy to subcutaneous tissue layers. |
| 807.92(a)(5) Intended Use of the Device | |
| | The SculpSure laser system is intended for non-invasive lipolysis<br>of the abdomen, flanks, back, and thighs, in individuals with a |
{4}------------------------------------------------
Section 5
510(k) Summary
| 807.92(a)(6) Summary of the Technological Characteristics of the Device Compared to the Predicate | | | |
|---------------------------------------------------------------------------------------------------|----------------------------------------------|----------------------------------------------|---------------------------------|
| | Proposed SculpSure Laser System | Cynosure SculpSure Laser System | Zeltiq Coolsculpting Device |
| 510(k) Number | Pending | K160470 | K162050 |
| Manufacturer | Cynosure, Inc. | Cynosure, Inc. | Zeltiq Aesthetics, Inc. |
| Lipolysis Method | Heat-assisted | Heat-assisted | Cold-assisted |
| Device Type | Diode Laser | Diode Laser | N/A |
| Wavelength | 1060 ±20 nm<br>(infrared) | 1060 ±20 nm<br>(infrared) | N/A |
| Spot Size | 4 x 6 cm² on each of<br>the Applicator heads | 4 x 6 cm² on each of<br>the Applicator heads | 4.5 x 7 cm² (2x)<br>approximate |
| Pulse Width (laser ON time) | CW | CW | N/A |
| Power Density | Up to 1.4 W/cm² | Up to 1.4 W/cm² | N/A |
| Attachment to patient | Belt | Belt | Belt |
| Voltage | 200-240V~, Single Phase | 200-240V~, Single Phase | N/A |
| Current | 20A | 20A | N/A |
807.92(b)(1) Non-clinical tests submitted
There have been minor changes to the software from the previously cleared SculpSure in K160470. These changes have been addressed and additional information is available in Section 16. There have been no changes to the biocompatibility or electrical safety testing from the previously cleared SculpSure device. Additional information is available in Section 15 and Section 17, respectively.
807.92(b)(2) Clinical tests submitted
Prospective, controlled studies were conducted at 3 study centers to evaluate the safety and efficacy of Cynosure SculpSure laser for the non-invasive fat reduction of the inner thigh, outer thigh and back. The study was conducted in the same manner as the previous abdomen and flanks studies. The treatment techniques used and subsequent follow up schedule for the study were kept consistent between the studies. A total of 168 subjects, making up 214 treatment areas, were enrolled in the study. This is broken down into 55 subjects/55 treatment areas for back, 52 subjects/52 treatment areas for outer thigh, and 61 subjects/107 treatment areas for inner thighs
{5}------------------------------------------------
Each subject received 1 treatment with the SculpSure. Treatment effectiveness was assessed through blind evaluation of pre and post treatment (12-week) photographs; percentage change in the adipose tissue thickness from baseline to the12-week follow up measured through ultrasound imaging; and patient satisfaction measured at the 12-week post treatment follow up visit through a 6-point Likert scale. All subjects that were treated in the study were included in the safety analysis and adverse events were assessed at all visits. On an average, blind evaluators were able to identify 86% of the post treatment photographs (back 91%, outer thigh 87% and inner thigh 83%); ultrasound imaging at 12-week post follow up visit showed 8.6% normalized fat reduction (back 10.6%, outer thigh 7.2%, and inner thigh 8.0%). All treatments had a p-value of <0.01 indicating that results were statistically significant. The patient satisfaction survey at 12 week post follow up visit showed 83% satisfaction rate (back 81%, outer thigh 88% and inner thigh 80%). All168 subjects were included in the safety analysis and adverse events reporting. All events were transient and majority was mild in nature. There were no serious adverse events. The study concluded that the SculpSure is safe and effective for non-invasive fat reduction for the back and thighs.
## 807.92(b)(3) Conclusions drawn from clinical and non-clinical tests submitted
Cynosure believes that SculpSure is substantially equivalent to its predicate devices with same intended use and similar technological characteristics. The non-clinical data for SculpSure supports the safety of the device and the biocompatibility, hardware and software verification and validation demonstrate that the SculpSure should perform as intended in the specified use conditions. Additionally, the successful clinical performance of SculpSure as documented in the clinical study demonstrate that the SculpSure has a safety and effectiveness profile that is similar to its predicate devices with a successful treatment of the back and thighs.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.