K171039 · Assut Europe S.P.A. · GAW · Aug 29, 2017 · General, Plastic Surgery
Device Facts
Record ID
K171039
Device Name
Assut Filbloc Permanent Sutures
Applicant
Assut Europe S.P.A.
Product Code
GAW · General, Plastic Surgery
Decision Date
Aug 29, 2017
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 878.5010
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
FilBloc Permanent sutures are intended for general soft tissue approximation, excluding closure of the epidermis, where use of a non-absorbable suture is appropriate.
Device Story
Assut FilBloc Permanent Sutures are non-absorbable polypropylene monofilament sutures used for soft tissue approximation. Devices are supplied in various lengths and gauge sizes (USP 2 to 4/0), with or without stainless steel needles, and with optional unidirectional or bidirectional barbs or a suture-material 'block' for anchoring. Used by surgeons in clinical settings to close wounds securely without knot-tying. The device remains in the patient at the site of suturing. Benefits include rapid, secure wound closure and reduced need for manual knotting. No complex processing or software involved.
Clinical Evidence
Bench testing only. Compliance with USP 29 standards for suture diameter, attachment, and tensile strength. Biocompatibility testing performed per ISO 10993-1 (cytotoxicity, sensitization, irritation, acute/subchronic toxicity, genotoxicity, implantation, material-mediated pyrogenicity). Sterility and shelf-life validated.
Indicated for general soft tissue approximation in patients requiring non-absorbable sutures; excludes closure of the epidermis.
Regulatory Classification
Identification
Nonabsorbable polypropylene surgical suture is a monofilament, nonabsorbable, sterile, flexible thread prepared from long-chain polyolefin polymer known as polypropylene and is indicated for use in soft tissue approximation. The polypropylene surgical suture meets United States Pharmacopeia (U.S.P.) requirements as described in the U.S.P. Monograph for Nonabsorbable Surgical Sutures; it may be undyed or dyed with an FDA approved color additive; and the suture may be provided with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 29, 2017
Assut Europe S.p.A. % Mr. Roger Gray Donawa Lifescience Consulting Srl Piazza Albania, 10 Rome, 00153 Italy
Re: K171039
Trade/Device Name: Assut Filbloc Permanent Sutures Regulation Number: 21 CFR 878.5010 Regulation Name: Nonabsorbable polypropylene surgical suture Regulatory Class: Class II Product Code: GAW Dated: July 18, 2017 Received: July 20, 2017
Dear Mr. Gray:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# David Krause -S
for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K171039
Device Name Assut FilBloc Permanent sutures
Indications for Use (Describe)
FilBloc Permanent sutures are intended for general soft tissue approximation, excluding closure of the epidermis, where use of a non-absorbable suture is appropriate.
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
|----------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
|----------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human ServicesFood and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Type of Use (Select one or both, as applicable)
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
# 510(k) Summary
| Device Name: | Assut FilBloc Permanent Sutures |
|-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Type of 510(k) submission: | Abbreviated |
| Date of submission: | 17 July 2017 |
| Manufacturer: | Assut Europe SpA<br>Zona Industriale<br>67062 Magliano dei Marsi (AQ)<br>Italy |
| Phone:<br>Fax: | +39-06-72677348<br>+39-06-72675380 |
| FDA Registration Number: | 9617547 |
| 510(k) Owner and Submitter: | Assut Europe SpA<br>Via G.Gregoraci, 12<br>00173 Rome<br>Italy |
| Owner/Operator Number: | 9044488 |
| 510(k) Application Correspondent: | Mr Roger Gray<br>VP Quality and Regulatory<br>Donawa Lifescience Consulting<br>Piazza Albania 10<br>00153 Rome<br>Italy |
| Phone:<br>Fax:<br>Email: | +39 06 578 2665<br>+39 06 574 3786<br>rgray@donawa.com |
| FDA Product Code: | GAW |
| FDA Regulation Number: | 21 CFR 878.5010 |
| FDA Classification Name: | Nonabsorbable Polypropylene Surgical Suture |
| Classification Panel: | General and Plastic Surgery |
| Common Name: | Nonabsorbable Polypropylene Surgical Suture |
| FDA Classification: | Class II |
| Submission Type: | 510(k) |
| Indication for Use: | FilBloc Permanent sutures are intended for general soft tissue<br>approximation, excluding closure of the epidermis, where use of<br>a non-absorbable suture is appropriate. |
| Predicate Device: | Quill Nonabsorbable Polypropylene barbed suture, K052373.<br>From information on the FDA website, there is no indication that<br>this device has been subject to a design-related recall. |
{4}------------------------------------------------
Reference Device:
Assut FilBloc Absorbable Polydioxanone Surqical Suture, K150553. This device has not been subject to a design-related recall.
#### Device Description:
The Assut FilBloc Permanent Suture range of devices comprises a variety of gauge sizes and lengths, supplied with or without stainless steel needles, which are also available in a variety of different sizes and shapes. The sutures may have a 'block' at one end, which allows surgeons to close wounds quickly and securely without tying knots or changing suturing techniques. The block is made from the same material as the suture, and is used to anchor the suture.
Assut FilBloc Permanent Sutures contact the patient and remain attached to the patient at the point suturing and may remain in place for significant periods.
The sutures with needles do not need to be used with any other devices, but the needleless sutures need to be used together with legally available separate needles.
#### Technological Characteristics:
The sutures are manufactured in polypropylene monofilament thread, white natural (undyed) or dyed blue colour (phthalocyanine copper - C.I. 74160; 21 CFR§ 74.3045). The suture thread can be smooth or can have unidirectional or bidirectional barbs along the axis of the monofilament surface, either convergent or divergent.
The chemical formula of polypropylene is -(CH(CH3)-CH2)n. The chemical formula of the blue dye is C32H16CuN8. The suture content of the blue dye is below 0.5 wt%.
Assut FilBloc Permanent Sutures are available sterile (ethylene oxide, with SAL 10°) for single use in a wide range of lengths and in sizes from USP 2 to 4/0, (sizes USP 4/0, USP 3/0, USP 2/0, USP 0, USP 1, USP 2), with or without needles (single or double), with or without single-end blocks, and with or without barbs. The barbs can be unidirectional or bidirectional, and if bidirectional, the barbs can be opposing each other or not, at the purchaser's choice.
#### Performance Data:
Physical testing was performed to establish compliance with USP 29, including <861> suture diameter, <871> suture attachment, <881> tensile strength. Tests were also undertaken to establish sterility, shelf life and packaging integrity.
#### Biocompatibility:
In accordance with the recommendations of ISO 10993-1 and FDA quidance 'Use of International Standard ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process', 16 June 2016, for devices described as implants in contact with tissuelbone for over 30 days, the following tests were carried out to establish the biocompatibility of the FilBloc Permanent Sutures:
- . Cytotoxicity (ISO 10993-5)
- . Sensitization (ISO 10993-10)
- Irritation or Intracutaneous Reactivity (ISO 10993-10) .
- Acute Systemic Toxicity (ISO 10993-11) ●
- . Subchronic Toxicity (ISO 10993-11)
- Genotoxicity (ISO 10993-3) ●
- Implantation (ISO 10993-6)
- . Material Mediated Pyrogenicity
Bacterial endotoxin testing (LAL test) is carried out on every batch of sutures.
{5}------------------------------------------------
#### Substantial Equivalence:
The features and characteristics of predicate device and reference device selected for an evaluation of substantial equivalence with the Filloc Permanent Sutures are provided in Table 1, together with a comparison with the same features and charcteristics of the subject device.
| Table 1: Predicate device comparison table | | | |
|------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------|
| Item | Predicate Device | Subject device | Similarity |
| Device name | Quill Nonabsorbable Polypropylene<br>Barbed Suture | FilBloc Permanent Sutures | N/A |
| Device Manufacturer | Angiotech, Reading PA, US | Assut Europe, Italy | N/A |
| 510(k) Reference | K052373 | Not yet assigned | N/A |
| FDA Product Code | GAW | GAW | Same |
| FDA Classification<br>Name | Nonabsorbable polypropylene surgical<br>suture | Nonabsorbable polypropylene surgical<br>suture | Same |
| FDA Regulation<br>Number | 21 CFR 878.5010 | 21 CFR 878.5010 | Same |
| Device description | Nonabsorbable polypropylene surgical<br>suture in various thread sizes and needle<br>shapes/sizes, with bi-directional thread<br>barbs | Nonabsorbable polypropylene surgical<br>suture in various thread sizes with or<br>without unidirectional or bidirectional<br>barbs, and with or without various<br>needle shapes/sizes | Similar |
| Indications for Use /<br>Intended use | Quill® Nonabsorbable Polypropylene<br>Barbed Sutures are indicated for use in<br>soft tissue approximation excluding<br>closure of the epidermis. | FilBloc Permanent sutures are<br>intended for general soft tissue<br>approximation, excluding closure of<br>the epidermis, where use of a non-<br>absorbable suture is appropriate. | Equivalent to<br>predicate. Very<br>similar to<br>reference<br>device |
| Rx only? | Yes | Yes | Same |
| Monofilament? | Yes | Yes | Same |
| USP sizes | 4/0 to 2 | 4/0 to 1 | Different |
| Suture barbs | Bi-directional | None<br>Bidirectional<br>Unidirectional | Different from<br>predicate but<br>same as<br>reference<br>device |
| Locking block? | Not available | Optional | Different from<br>predicate but<br>same as<br>reference<br>device |
| Thread color | White<br>Blue<br>Copper | White<br>Blue | Similar |
| Sterility | Sterile by EO, SAL 10-6 | Sterile by EO, SAL 10-6 | Same |
| Single use? | Yes | Yes | Same |
| Complies with<br>applicable USP<br>monographs? | Yes | Yes | Same |
{6}------------------------------------------------
| Table 1: Predicate device comparison table | | | | | | |
|--------------------------------------------|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|--|--|--|
| Item | Predicate Device | Subject device | Similarity | | | |
| Needles attached? | Yes | Yes<br>Also available without needles | Different from<br>predicate but<br>same as<br>reference<br>device | | | |
| Needle shapes | 1/2 circle<br>3/8 circle | 1/2 circle<br>3/8 circle<br>5/8 circle<br>"ski" needle<br>Straight | Different from<br>predicate but<br>same as<br>reference<br>device | | | |
| Needle cross-<br>section | Unknown | Cylindrical<br>Triangular<br>Trapezoidal | Same as<br>reference<br>device | | | |
| Needle point | Taper<br>Diamond | Cylindrical<br>Triangular<br>Diamond (Taper-cut)<br>Blunt point<br>Spatula | Different from<br>predicate but<br>same as<br>reference<br>device | | | |
| Needle eye | Unknown | Open<br>Closed | Same as<br>reference<br>device | | | |
| Needle materials | Surgical stainless steel:<br>AISI 470 | Surgical stainless steel:<br>AISI 470<br>AISI 455<br>AISI 301<br>AISI 302<br>AISI 304<br>AISI 304L<br>AISI 316<br>AISI 320<br>AISI 321<br>Cobalt chromium | Similar, same<br>as reference<br>device | | | |
| Needle coating | Siliconized | Siliconized | Same | | | |
| Biocompatibility | Biocompatible | Biocompatible | Same | | | |
The subject device and the predicate device have many identical or similar properties and features. In summary, the differencesare:
- USP sizes available: The subject device has one USP thread size less in its range. .
- Suture barb orientation: The predicate device has bidirectional barbs on all versions in the . range, whereas the subject device has options with or without barbs, and when barbs are included, these can be either bidirectional or unidirectional. The reference device has the same options of no barbs, bidirectional barbs and unidirectional barbs.
- Locking block: The predicate device does not offer a locking block option, whereas the subject device range does, which is the same as the reference device.
- . Needle sizes, shapes, points, etc: The subject device includes a wider variety of needle sizes, shapes, points, etc, than the predicate device. The reference device has a similarly wide range of needle sizes, shapes, points, etc.
{7}------------------------------------------------
### Conclusion:
Based on the information contained within this submission, it is concluded that the FilBloc Permanent Sutures are substantially equivalent to the identified predicate device already in interstate commerce within the USA.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.