eCareCoordinator is software intended for use in data aggregation, patient interaction facilitation, storage and clinical information management with independent physiological devices and ancillary systems that are connected either directly or through networks. The software is intended to provide patient information from the patient location through networking technology to a remote care team. eCareCoordinator does not send any real time alarms and is not intended to provide automated treatment decisions. This software is an informational tool only and is not to be used as a substitute for professional judgment of healthcare providers in diagnosing and treating patients.
Device Story
eCareCoordinator 1.5 is a software-only telemedicine system for home-based patient monitoring. It facilitates data aggregation, patient interaction, and clinical information management. The system connects to independent physiological devices (via Bluetooth or manual entry) and ancillary systems to transmit patient data to a remote care team. Clinicians use the eCC interface to view population data, measurement trends, and survey responses, while patients use the eCP interface on Android tablets or legacy Telestations. Features include two-way video calls, educational videos, and customizable intervention rules that trigger clinician notifications. The device does not provide real-time alarms or automated treatment decisions; it serves as an informational tool to support clinical decision-making. It is intended for use by care teams and patients capable of engaging with the software.
Clinical Evidence
No clinical testing was performed. Substantial equivalence was supported by non-clinical verification and validation activities, including risk analysis, product specifications, design reviews, and functional/usability testing.
Technological Characteristics
Software-only telemedicine system. Operates on Android tablets and legacy Telestations. Connectivity via network and Bluetooth for physiological device data. Features include user authentication, role-based access control, audit logging, and locked-down tablet configuration. No patient-contacting components.
Indications for Use
Indicated for use by patients and by care teams for collecting and reviewing patient data from patients who are capable and willing to engage in use of this software, to transmit medical and non-medical information through integrated technologies.
Regulatory Classification
Identification
A radiofrequency physiological signal transmitter and receiver is a device used to condition a physiological signal so that it can be transmitted via radiofrequency from one location to another, e.g., a central monitoring station. The received signal is reconditioned by the device into its original format so that it can be displayed.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 14, 2017
Visicu Inc. Milind Gramopadhye Director of Quality and Regulatory 217 East Redwood St. Suite 1900 Baltimore, Maryland 21202
Re: K171029
Trade/Device Name: eCareCoordinator 1.5 Regulation Number: 21 CFR 870.2910 Regulation Name: Radiofrequency Physiological Signal Transmitter and Receiver Regulatory Class: Class II Product Code: DRG Dated: June 15, 2017 Received: June 16, 2017
Dear Milind Gramopadhye:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
M. G. Hillemann
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K171029
Device Name eCareCoordinator 1.5
#### Indications for Use (Describe)
eCareCoordinator and its accessories are indicated for use by patients and by care teams for collecting and reviewing patient data from patients who are capable and willing to engage in use of this software, to transmit medical and non-medical information through integrated technologies.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------|---------------------------------------------------------------------------|
| <div> <span>☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the Philips logo in blue, followed by the text "eCareCoordinator 1.5" and "Traditional 510(k)" in black. The Philips logo is in a bold, sans-serif font. The text below the logo is in a smaller, regular font. The image is likely a product label or advertisement.
# 510(k) Summary eCareCoordinator 1.5
This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR 807.92(c).
Date Prepared: April 5, 2017
#### l. Submitter's name and address
| Manufacturer: | Visicu, Inc.<br>217 East Redwood Street<br>Suite 1900<br>Baltimore, MD 21202 |
|-----------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Daniel R. Plonski<br>Senior Manager, Quality and Regulatory<br>Phone: 410-246-5357<br>Fax: 410-276-1970<br>E-mail: daniel.plonski@philips.com |
#### II. Device information
| Device Name: | eCareCoordinator 1.5 |
|-----------------------|----------------------|
| Common Name: | Telemedicine System |
| Classification panel: | Cardiovascular |
| Classification | ProCode | Description |
|----------------|---------|------------------------------------------------------------------|
| 870.2910 | DRG | Radiofrequency physiological signal transmitter<br>and receiver. |
#### III. Predicate device information
Trade name: eCareCoordinator Visicu, Inc. Manufacturer: 510(k) clearance: K141706 Classification name Radiofrequency physiological signal transmitter and receiver. Device class: Class II Classification regulation: 21 CFR 892.2910 Classification panel: Cardiovascular Product code: DRG
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows the Philips logo in blue, followed by the text "eCareCoordinator 1.5" and "Traditional 510(k)" in black. The Philips logo is a stylized wordmark with a distinctive design. The text below the logo appears to be related to the product or service offered by Philips.
#### IV.Device Description
eCareCoordinator (eCC) is a software-only telemedicine system, designed to enable the support of patients in the home setting. eCC is intended to support the clinician with monitoring of remote patients. Clinicians use eCC to manage populations of ambulatory care patients, while keeping primary care physicians informed of patient status. eCC is a software-only device and does not contain any patient-contacting components.
#### V. Intended use/ Indications for Use
#### Intended Use:
eCareCoordinator is software intended for use in data aggregation, patient interaction facilitation, storage and clinical information management physiological devices and ancillary systems that are connected either directly or through networks. The software is intended to provide patient information from the pation through networking technology to a remote care team. eCareCoordinator does not send any real time alarms and is not intended to provide automated treatment decisions. This software is an informational tool only and is not to be used as a substitute for professional judgment of healthcare providers in diagnosing and treating patients.
#### Indication for Use:
eCareCoordinator and its accessories are indicated for use by patients and by care teams for collecting and reviewing patients who are capable and willing to engage in use of this software, to transmit medical and non-medical information through integrated technologies.
### VI.Comparison of Technological Characteristics with the Predicate Device
eCareCoordinator 1.5 is an enhanced version of eCareCoordinator 1.0, originally cleared under K141706.
A comparison matrix (Table 5-1) shows the similarities and differences. The eCareCoordinator 1.5 software with the listed enhancements is substantially equivalent to the previously cleared eCareCoordinator 1.0. The Intended Use has been modified to clarify and modernize the wording, however, the concept has not changed. Technological characteristics and principles of operation remain the same, Differences in the available features, as discussed below, do not present any new questions of safety or effectiveness.
{5}------------------------------------------------
| Specification /<br>Feature | eCareCoordinator 1.5<br>(Subject Device) | eCareCoordinator<br>(Predicate device –<br>K141706) | Comparison |
|---------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------|
| Intended Use | eCareCoordinator is<br>software intended for use in<br>data aggregation, patient<br>interaction facilitation,<br>storage and clinical<br>information management<br>with independent<br>physiological devices and<br>ancillary systems that are<br>connected either directly or<br>through networks. The<br>software is intended to<br>provide patient information<br>from the patient location<br>through networking<br>technology to a remote care<br>team. eCareCoordinator<br>does not send any real time<br>alarms and is not intended<br>to provide automated<br>treatment decisions. This<br>software is an informational<br>tool only and is not to be<br>used as a substitute for<br>professional judgment of<br>healthcare providers in<br>diagnosing and treating<br>patients. | eCareCoordinator is<br>software intended for use in<br>data aggregation, patient<br>interaction facilitation,<br>storage and clinical<br>information management<br>with independent<br>physiological devices and<br>ancillary systems that are<br>connected either directly or<br>through networks. The<br>software is intended to<br>provide patient information<br>from the patient location<br>through networking<br>technology to a remote care<br>team. eCareCoordinator<br>does not send any real time<br>alarms and is not intended<br>to provide automated<br>treatment decisions. This<br>software is an informational<br>tool only and is not to be<br>used as a substitute for<br>professional judgment of<br>healthcare providers in<br>diagnosing and treating<br>patients. | Same |
| Indications for<br>Use | eCareCoordinator and its<br>accessories are indicated<br>for use by patients and by<br>care teams for collecting<br>and reviewing patient data<br>from patients who are<br>capable and willing to<br>engage in use of this<br>software, to transmit<br>medical and non-medical<br>information through<br>integrated technologies. | eCareCoordinator and its<br>accessories are indicated<br>for use by patients and by<br>care teams for managing<br>patients who are capable<br>and willing to engage in use<br>of this software, to transmit<br>medical and non-medical<br>information through<br>integrated technologies. | Same – wording updated for<br>clarity. |
| Target<br>population | At home patients | At home patients | Same |
| Technological Characteristics | | | |
| System<br>components | Software Only | Software Only | Same |
| User Interfaces | Clinical (eCC) and<br>Patient (eCP) | Clinical (eCC) and<br>Patient (eCP) | Enhanced user interfaces |
| Clinician-Patient<br>Communications | Two-way video | None | Added video communications |
| Measurement<br>Features | None | None | Same |
| Specification /<br>Feature | eCareCoordinator 1.5<br>(Subject Device) | eCareCoordinator<br>(Predicate device –<br>K141706) | Comparison |
| System Alarms | None | None | Same |
| Waveform<br>Transmission | None | None | Same |
| User Access and<br>Patient Data<br>Security | User Authentication<br>services, roles-based data<br>access, logging for audit<br>trail, locked down tablet | User Authentication<br>services, roles-based data<br>access, logging for audit<br>trail, locked down tablet | Same |
| Clinical User Interface (eCC) Features | | | |
| Population View | Population screen with<br>measurement and survey<br>flags and other display<br>fields | Population screen with<br>measurement and survey<br>flags and other display<br>fields | Same |
| Overall Score | Sum of weighted scores for:<br>Measurements, survey<br>responses, issues,<br>readmission risk and days<br>since discharge | Sum of weighted scores for:<br>Measurements, survey<br>responses, issues,<br>readmission risk and days<br>since discharge | Same |
| Measurements<br>View | Display of patient<br>measurements and survey<br>responses | Display of patient<br>measurements and survey<br>responses | Same |
| Patient Chart | Graphical and tabular<br>displays of patient<br>information | Graphical and tabular<br>displays of patient<br>information | Same |
| Intervention<br>Rules | Used to trigger notifications<br>(flags) to clinician.<br>Customizable by institution<br>or clinician | Used to trigger notifications<br>(flags) to clinician.<br>Customizable by institution<br>or clinician | Same |
| Surveys | Clinical users have<br>option to send<br>surveys to patients | Clinical users have<br>option to send<br>surveys to patients | Same |
| Protocols | Used to set patient tasks,<br>surveys and intervention<br>rules. Customizable by<br>institution or clinician | Used to set patient tasks,<br>surveys and intervention<br>rules. Customizable by<br>institution or clinician | Same |
| Clinician Tasks | Communication and<br>tracking of clinical care<br>tasks | Communication and<br>tracking of clinical care<br>tasks | Same |
| Sticky Notes | Optionally added to patient<br>calendar | Optionally added to patient<br>calendar | Same |
| Reports | Hard copy patient record<br>report and system<br>administration reports | System administration<br>reports | Added patient record report |
| Patient User Interface (eCP) Features | | | |
| Platform | Android Tablet and legacy<br>Telestation | Android Tablet and legacy<br>Telestation | Same |
| Measurements<br>from home<br>devices | Wireless Bluetooth and<br>manual entry | Manual entry | Bluetooth connectivity<br>implemented |
| Appointments | Scheduled by care provider,<br>appear on patient's | Scheduled by care provider,<br>appear on patient's | Same |
| Specification /<br>Feature | eCareCoordinator 1.5<br>(Subject Device) | eCareCoordinator<br>(Predicate device –<br>K141706) | Comparison |
| | calendar | calendar | |
| Activities and<br>Reminders | Scheduled by care provider,<br>appear on patient's<br>calendar | Scheduled by care provider,<br>appear on patient's<br>calendar | Same |
| Surveys | Created and assigned to<br>patients by the care<br>provider | Created and assigned to<br>patients by the care<br>provider | Same |
| Video Calls | Capability to schedule and<br>conduct video calls with<br>care providers | None | Added video call capability |
| Educational<br>Videos | Access to pre-installed<br>educational videos | None | Educational videos added |
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for Philips, followed by the text "eCareCoordinator 1.5" and "Traditional 510(k)". The Philips logo is in blue, and the text is in black. The text "eCareCoordinator 1.5" appears to be the name of a software product, and "Traditional 510(k)" refers to a type of medical device clearance from the FDA.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the Philips logo in blue, followed by the text "eCareCoordinator 1.5" and "Traditional 510(k)" in black. The Philips logo is a stylized version of the company's name. The text below the logo indicates that the image is related to the eCareCoordinator 1.5 software and the Traditional 510(k) regulatory pathway. The text is left-aligned and the font is sans-serif.
Table 5-1 Comparison Table
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for Philips, followed by the text "eCareCoordinator 1.5" and "Traditional 510(k)". The Philips logo is in blue, and the text is in black. The text is aligned to the left, and the logo is above the text. The image is likely a product label or advertisement.
#### Performance Data VII.
The following performance data were provided in support of the substantial equivalence determination:
### Summary of Non-clinical testing
No performance standards for telehealth systems or components have been issued under the authority of Section 514. eCareCoordinator was tested in accordance with Philips verification and validation processes. Quality Assurance measures were applied to the system design and development, including:
- Risk Analysis
- Product Specifications
- Design Reviews
- Verification & Validations
### Summary of Clinical Testing
Clinical Performance testing for Philips eCareCoordinator 1.5 was not performed, as there were no new clinical applications that had hazards or risk mitigations that required clinical performance testing to support equivalence.
### Conclusions drawn from the Non-clinical and Clinical testing
Verification validation activities have been conducted to establish the performance, functionality, and usability characteristics of the new device with respect to the predicate, intended use and defined requirements. Testing included detailed functional, system level and usability testing. Test results demonstrated that eCareCoordinator 1.5 meets all specifications and user needs.
#### VIII. Conclusion
eCareCoordinator 1.5 is substantially equivalent to the previously cleared eCareCoordinator. Both devices have the same intended use, indications for use, technological characteristics and principles of operation. Differences in the available features, as discussed above, do not change the intended use or present any new questions of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.