K171015 · Di-Chem, Inc. · KPO · Oct 13, 2017 · Gastroenterology, Urology
Device Facts
Record ID
K171015
Device Name
Citryte
Applicant
Di-Chem, Inc.
Product Code
KPO · Gastroenterology, Urology
Decision Date
Oct 13, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5820
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Di-Chem, Inc. Citryte Acid Concentrate for bicarbonate dialysis is indicated for use in the treatment of acute and chronic hemodialysis. It is to be used with the appropriate hemodialysis machine and sodium bicarbonate.
Device Story
Citryte™ Acid Concentrate is a dry powder hemodialysis concentrate; supplied in kits (16.5, 20, 25 gallon volumes). Components include sodium chloride, potassium chloride, calcium chloride, magnesium chloride, dextrose, and citric acid. Used in clinical settings by healthcare professionals; mixed with water and sodium bicarbonate in a three-stream hemodialysis machine to create a final proportioned hemodialysis solution. Solution is heated and pumped through a hemodialyzer; separated from patient blood by a semi-permeable membrane. Functions via concentration gradient to remove waste products (urea, creatinine, potassium) and toxins from blood. Provides necessary electrolyte/solute balance for patients with acute or end-stage renal failure.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing verifying chemical composition and compliance with ANSI/AAMI 13958:2014 standards for hemodialysis concentrates.
Technological Characteristics
Dry powder concentrate; polyethylene packaging. Chemical constituents: NaCl, KCl, CaCl2, MgCl2, dextrose, citric acid. Proportioning ratio 45X. Manufactured/tested per ANSI/AAMI 13958:2014. No electronic components or software.
Indications for Use
Indicated for treatment of acute and chronic hemodialysis in patients requiring renal replacement therapy.
Regulatory Classification
Identification
A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.
Advanced Renal Technologies Inc. DRYalysate (K980659)
Di-Chem Inc. Hemo-Lyte Sodium Bicarbonate (K012328)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 13, 2017
Di-Chem, Inc. Keith Buchholz Compliance Manager 12297 Ensign Avenue North Champlin, MN 55316
Re: K171015
Trade/Device Name: Citryte™ Acid Concentrate Regulation Number: 21 CFR§ 876.5820 Regulation Name: Hemodialysis System and Accessories Regulatory Class: II Product Code: KPO Dated: August 30, 2017 Received: September 7, 2017
Dear Keith Buchholz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevicesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Joyce M. Whang -S
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K171015
Device Name Citryte™ Acid Concentrate
Indications for Use (Describe)
The Di-Chem, Inc. Citryte Acid Concentrate for bicarbonate dialysis is indicated for use in the treatment of acute and chronic hemodialysis. It is to be used with the appropriate hemodialysis machine and sodium bicarbonate.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------|---------------------------------------------|
| × Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| SUBMITTER: | Di-Chem, Inc.<br>12297 Ensign Avenue North<br>Champlin, MN 55311<br>Ph. 763-422-8311 Fax. 763-422-8472 |
|--------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) # | K171015 |
| FDA Registration | 2183415 |
| # Contact: | Keith Buchholz |
| DATE Revised: | August 9, 2017 |
| DEVICE NAME: | Citryte™ Acid Concentrate |
| COMMON NAME: | Acid Concentrate for hemodialysis |
| CLASSIFICATION: | Class II per 21 CFR 876.5820 (Hemodialysis<br>System and Accessories) |
| PRODUCT CODE: | KPO: Gastroenterology/Urology |
| PRIOR SUBMISSIONS: | There are no prior submissions for this device |
| PREDICATE DEVICES: | Primary Predicate Devices:<br>Rockwell Medical Technologies CitraPure (K160847)<br>Diasol Inc. Citrisol (K130511)<br><br>Reference Predicate Devices:<br>Advanced Renal Technologies Inc. (K980659)<br>Di-Chem Inc. (K012328) |
## DEVICE DESCRIPTION:
The Di-Chem, Inc. proposed Citryte hemodialysis acid concentrate is provided in dry form and is for use in sodium bicarbonate dialysis. Citryte is comprised of sodium chloride (100.0mEq/L), potassium chloride (1.0-3.0mEq/L), calcium chloride (2.0-3.0mEq/L), magnesium chloride ( 1.0 mEq/L), dextrose ( 100 mg%), and citric acid (2.4mEq/L). All of the chemical constituents meet USP grade or equivalent. The proposed device will be manufactured, tested and labeled in accordance to ANSI/AAMI 13958:2014 guidelines. Each of the proposed formulations will be offered in three different mix volumes, 16.5 gallon, 20 gallon and 25 gallon to meet the three mix volume
{4}------------------------------------------------
sizes of the current commercially available hemodialysis mixing machines. Each case of product contains the measured amount of each chemical component corresponding to the labeled product formula and listed final solution volume. The product packaging consists of two equal weight bags of sodium chloride, one bag of dextrose and one bag containing the citric acid, potassium chloride, calcium chloride and magnesium chloride. Each product bag is labeled indicating its contents, lot number, use by date, and product formula. The product bags are comprised of polyethylene and are 18"x24"x0.004". The four bags are then packaged into one corrugate box. Each box will contain all of the chemical components to make one of the three available volumes of the labeled formulation. The product is intended to be mixed into solution using water meeting or exceeding ANSI/AAMI Hemodialysis Water Quality Standards utilizing commercially available hemodialysis concentrate mixers in accordance with the mixer manufacturer's directions. These chemicals when in solution are intended to be used as the acid portion of a three part hemodialysis treatment in conventional commercially available hemodialysis machines or monitors in providing a hemodialysis treatment. The hemodialysis formulations presented in this 510(k) notification are intended to be used in a three-stream hemodialysis machine in which an acid concentrate (Citryte) is proportioned into one stream, a sodium bicarbonate concentrate solution is proportioned into another stream and a specified volume of water is proportioned into the remaining stream of the hemodialysis machines proportioning system. These three streams are then mixed by the hemodialysis machine to prepare a final proportioned hemodialysis solution. These proportioned hemodialysis solutions are then heated to body temperature and pumped through the hemodialysis compartment of a hemodialyzer (artificial kidney). These hemodialysis solutions are separated from the patient's blood by a semi-permeable cellulosic or non-cellulosic membrane which functions as a molecular weight selective barrier to the passage of molecules above a certain molecular weight. The semi-permeable membrane in the hemodialyzer permits the passage of both ionized and non-ionized molecules, waste products and toxins including blood urea, nitrogen, creatinine and potassium contained in the patient's blood circulating through the hemodialyzer to pass through the semi-permeable membrane into the hemodialysis solution then circulating back to the hemodialysis machine where the solution is ultimately discarded. The ionic and molecular composition of the hemodialysis solution establishes the concentration gradient between the patient's blood and the hemodialysis solution passing through the hemodialyzer which permits the effective removal of waste products and toxins found in excess in the patient's blood during acute and end-stage renal failure. Since different patients have different requirements for the removal rates and quantities of the various molecules and toxins in acute and end-stage renal failure it is necessary to have a variety of different hemodialysis solution formulations to satisfy the needs of all renal failure patients. For this reason the Citryte is offered in multiple formulations containing varying amounts of potassium chloride and calcium chloride.
{5}------------------------------------------------
## Citryte™ Ingredient Ranges
| Criteria | Specification |
|-----------------------------------------------|------------------------------------------------------|
| Formulary Ingredient Range | Sodium 100.0 mEq/L |
| | Calcium 2.0 - 3.0 mEq/L |
| | Potassium 1.0 - 3.0 mEq/L |
| | Magnesium 1.0 mEq/L |
| | Dextrose 100 mg/dl |
| | Citric Acid 2.4 mEq/L |
| | Acetate 0 - 0.3 mEq/L |
| | |
| Concentrate Type | Dry Powder Concentrate<br>16.5gal, 20gal, 25gal kits |
| Proportioning Ratio<br>(Concentrate to Water) | 45X |
The Citryte™ acid concentrate formulas we plan to market upon acceptance of this 510(k) notification include the following. (Note: Any future new formulas within the previously approved predicate device ranges listed above will be implemented in accordance with our design control and maintained in accordance with our device design history files.)
| Sodium<br>100.0mEq/L | Potassium<br>1.0-3.0mEq/L | Calcium<br>2.0-3.0 mEq/L | Magnesium<br>1.0 mEq/L | Citric Acid<br>2.4mEq/L | Dextrose<br>100 mg% |
|----------------------|---------------------------|--------------------------|------------------------|-------------------------|---------------------|
| 100 | 1.0 | 2.0 | 1.0 | 2.4 | 100 |
| 100 | 1.0 | 2.5 | 1.0 | 2.4 | 100 |
| 100 | 2.0 | 2.0 | 1.0 | 2.4 | 100 |
| 100 | 2.0 | 2.25 | 1.0 | 2.4 | 100 |
| 100 | 2.0 | 2.5 | 1.0 | 2.4 | 100 |
| 100 | 2.0 | 3.0 | 1.0 | 2.4 | 100 |
| 100 | 3.0 | 2.0 | 1.0 | 2.4 | 100 |
| 100 | 3.0 | 2.5 | 1.0 | 2.4 | 100 |
| 100 | 3.0 | 3.0 | 1.0 | 2.4 | 100 |
Citryte™ Dry Acid Concentrate Formulas
{6}------------------------------------------------
#### PREDICATE DEVICES (PRIMARY AND REFERENCE):
The Di-Chem. Inc. Citryte™ hemodialysis dry concentrate is substantially equivalent to the Rockwell Medical CitraPure® acid concentrate, the Diasol Citrisol acid concentrate, and the Advanced Renal Technologies DRYalysate acid concentrate.
#### INDICATIONS FOR USE:
Di-Chem, Inc. Citryte™ Acid Concentrate for Bicarbonate dialysis is indicated for use in the treatment of acute and chronic hemodialysis. It is to be used with the appropriate hemodialysis machine and sodium bicarbonate monitor.
This indications for use statement is essentially the same as the indications for use statement for the predicate devices.
### TECHNOLOGICAL CHARACTERISTICS:
#### (PRIMARY PREDICATE DEVICES)
Comparing the proposed device to the primary predicate devices shows that they utilize the exact same range of chemical components, exact same packaging materials, and the exact same chemical formulations.
There are no significant differences.
#### (REFERENCE PREDICATE DEVICE DRYalysate)
Comparing the proposed device to the reference predicate device shows that they utilize a similar range of chemical components, exact same packaging materials, and similar chemical formulations. The only difference is the proposed Citryte device does not contain the 0.3mEq/L of sodium acetate that the DRYalysate reference predicate device does.
#### (REFERENCE PREDICATE DEVICE Hemo-Lyte Sodium Bicarbonate)
The Di-Chem Inc. Hemo-Lyte Sodium Bicarbonate for hemodialysis is being included as a reference predicate device because it is the other necessary component required to perform a hemodialysis treatment and shares the identical product packaging material in both size and polyethylene material and is sourced from the same suppliers as the proposed Citryte device.
#### SUMMARY OF NON-CLINICAL TESTS:
In vitro testing was performed to verify the chemical composition of the proposed device was identical to that of the predicate devices and within the ranges set forth by ANSI/AAMI 13958:2014. Testing was performed in accordance with
{7}------------------------------------------------
our standard operating procedures utilizing validated equipment and analytical methods. The results of the testing met the requirements of ANSI/AAMI 13958:2014 (Concentrates for Hemodialysis and Related Therapies) which specifies that all electrolytes identified on the device label shall be present within ±5% or ±0.1mEg/L and glucose within ±5% or ±0.05g/L (expressed as dialysis fluid concentrations), whichever is greater of the stated concentration, with the exception of sodium, which shall be present within ±2.5% of the labeled concentration. The results of these tests confirmed the proposed Citryte device met the listed range requirements stated in ANSI/AAMI 13958:2014 and is chemically equivalent to the predicate devices for all the listed chemical formulations.
## CLINICAL TEST RESULTS:
Clinical testing was not performed.
## SUBSTANTIAL EQUIVALENCE:
#### (PRIMARY PREDICATE DEVICES)
The proposed Di-Chem Citryte device is manufactured utilizing the same chemicals, formula composition ranges, packaging materials and intended use as the Rockwell Medical CitraPure and Diasol Citrisol primary predicate devices.
#### (REFERENCE PREDICATE DEVICE DRYalvsate)
The proposed Di-Chem Citryte device is manufactured utilizing the same chemicals, formulation composition ranges, packaging materials and intended use as the Advanced Renal Technologies DRYalysate reference predicate devices with the only difference being the proposed Citryte device does not contain the 0.3mEq/L of sodium acetate that the DRYalysate reference predicate device does.
#### (REFERENCE PREDICATE DEVICE Hemo-Lyte Sodium Bicarbonate)
Sodium bicarbonate is the other necessary chemical component required to perform a hemodialysis dialysis treatment. The Hemo-Lyte sodium bicarbonate is manufactured in the same facility and uses the exact same polyethylene bag packaging material in the exact same size from the exact same suppliers as the proposed Citryte device.
#### CONCLUSIONS:
Comparing the proposed Citryte acid concentrate device to the predicate devices shows they are substantially equivalent in intended use, chemical composition, chemical formulations, packaging materials and device labeling.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.