K171009 · Lutronic Corporation · GEX · Jun 23, 2017 · General, Plastic Surgery
Device Facts
Record ID
K171009
Device Name
LASEMD Laser System
Applicant
Lutronic Corporation
Product Code
GEX · General, Plastic Surgery
Decision Date
Jun 23, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The LASEMD Laser System is indicated for use in dermatological procedures requiring the coagulation of soft tissue, treatment of actinic keratosis, and treatment of benign pigmented lesions such as, but not limited to lentigos (age spots), solar lentigos (sun spots) and ephelides (freckles).
Device Story
LASEMD Laser System is a thulium laser (1927 nm) for dermatological procedures. Device uses a footswitch-activated pulsed coherent near-infrared light beam delivered via an optical fiber and handpiece to the treatment zone. An integrated LED touch screen allows user control of laser parameters; a 658 nm aiming beam assists targeting. Operated by clinicians in a clinical setting. Device produces thermal coagulation of soft tissue to treat actinic keratosis and benign pigmented lesions. Benefits include precise tissue coagulation for dermatological conditions.
Clinical Evidence
No clinical data provided. Evidence consists of bench testing (electrical safety per IEC 60601-1, EMC per IEC 60601-1-2) and animal histology confirming equivalent treatment effects to the predicate.
Technological Characteristics
Thulium laser, 1927 nm wavelength. 658 nm aiming beam (≤ 5 mW). Fiber-coupled handpiece delivery. Footswitch activation. Integrated LED touch screen interface. Power: 5 W. Max pulse energy: 20 mJ. Max pulse width: 20 ms. Pulse repetition rate: 43.5 - 307.7 Hz. Spot size: 100-200 μm. Electrical: 100-240V, 50/60 Hz. Standards: ISO 13485:2003, IEC 60601-1, IEC 60601-1-2.
Indications for Use
Indicated for dermatological procedures requiring soft tissue coagulation, treatment of actinic keratosis, and benign pigmented lesions (e.g., lentigos, solar lentigos, ephelides) in patients requiring such dermatological care.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the logo is a stylized image of three human profiles facing to the right, with a symbol resembling a caduceus or staff with a serpent intertwined.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 23, 2017
Lutronic Corporation Jhung Vojir VP, Quality and Regulatory Affairs 6 Neshaminy Interplex. Suite 100 Trevose, Pennsylvania 19053
Re: K171009
Trade/Device Name: LASEMD Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: April 3, 2017 Received: April 4, 2017
Dear Jhung Vojir:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Actinclude requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
# Jennifer R. Stevenson -S3
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K171009
Device Name LASEMD Laser System
Indications for Use (Describe)
The LASEMD Laser System is indicated for use in dermatological procedures requiring the coagulation of soft tissue, treatment of actinic keratosis, and treatment of benign pigmented lesions such as, but not limited to lentigos (age spots), solar lentigos (sun spots) and ephelides (freckles).
Type of Use (Select one or both, as applicable)
| <span style="font-size:100%;">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:100%;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|-------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
|-------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Confidential and Proprietary Information of Lutronic Corporation
{3}------------------------------------------------
# 6. 510(k) SUMMARY OR STATEMENT
The Company's 510(k) Summary for the LASEMD Laser System is as follows:
{4}------------------------------------------------
# 510(k) Summary Lutronic Corporation LASEMD Laser System
This 510(k) Summary is being submitted in accordance with 21 CFR § 807.92.
# 1. General Information
| Applicant: | Lutronic Corporation |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|
| | 219 Sowon-ro<br>Haengsin-dong, Deogyang-gu<br>Goyang-si, Gyeonggi-do 410-220<br>Republic of Korea<br>Tel: (82) 31-908-3440<br>FAX: (82) 31-907-3440 |
| Contact Person: | Jhung Won Vojir, Ph.D.<br>VP of Quality and Regulatory Affairs |
| | Lutronic Corporation<br>6 Neshaminy Interplex, Suite 100<br>Trevose, PA 19053<br>Tel: 215-205-2219<br>FAX: 609-488-6958<br>Email: jvojir@lutronic.com |
| Summary Preparation Date: | June 20, 2017 |
| 2. Names | |
| Trade Name: | LASEMD Laser System |
| Common Name: | Laser Surgical Instrument |
| Classification Name: | Powered Laser Surgical Instrument<br>Product Code: GEX<br>21 CFR § 878.4810<br>Panel: General & Plastic Surgery |
{5}------------------------------------------------
#### 3. Predicate Device
The Fraxel DUAL Laser System is a surgical instrument for performing dermatological procedures in the treatment of actinic keratosis and treatment of pigmented lesions.
| 510(K) Number | K130193 |
|---------------------------|------------------------------------|
| Company Name | Solta Medical, Inc. |
| Device Name | Fraxel DUAL 1550/1929 Laser System |
| Classification Regulation | 21 CFR § 878.4810 |
| Classification Name | Powered Laser Surgical Instrument |
| Product Code | GEX |
| Device Panel | General & Plastic Surgery |
#### 4. Device Description
The LASEMD Laser System is a thulium laser, producing a pulsed beam of coherent near-infrared light (1927 nm) upon activation by a footswitch. The beam is then directed to the treatment zone by means of an optical fiber coupled to a handpiece. An integrated LED touch screen gives the user control over the necessary laser system parameters. The LASEMD Laser System is equipped with a 658 nm aiming beam.
# 5. Indications for Use
The LASEMD Laser System is indicated for use in dermatological procedures requiring the coagulation of soft tissue, treatment of actinic keratosis, and treatment of benign pigmented lesions such as, but not limited to lentigos (age spots), solar lentigos (sun spots) and ephelides (freckles).
#### 6. Substantial Equivalence
The LASEMD Laser System is substantially equivalent to the legally marketed Solta Medical Fraxel DUAL Laser System that is the subject of 510(k) K130193.
| Manufacturer | Lutronic Corporation | Solta Medical, Inc. |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device | LASEMD Laser System | Fraxel DUAL 1550/1927<br>Laser System |
| 510(k) Number | K171009 | K130193 |
| Indications for Use | The LASEMD Laser System is<br>indicated for use in<br>dermatological procedures<br>requiring the coagulation of<br>soft tissue, treatment of actinic<br>keratosis, and treatment of<br>benign pigmented lesions such<br>as, but not limited to lentigos<br>(age spots), solar lentigos (sun<br>spots) and ephelides (freckles). | 1927nm: The Fraxel 1927 nm<br>laser is indicated for use in<br>dermatological procedures<br>requiring the coagulation of<br>soft tissue, treatment of actinic<br>keratosis, and treatment of<br>pigmented lesions such as, but<br>not limited to lentigos (age<br>spots), solar lentigos (sun<br>spots) and ephelides<br>(freckles). |
{6}------------------------------------------------
| Classification | GEX, 21 CFR § 878.4810 | GEX, 21 CFR § 878.4810 |
|-----------------------------|---------------------------------------------|---------------------------------------------|
| Laser Type | Thulium laser | Thulium laser |
| Laser Wavelength | 1927 nm | 1927 nm |
| Aiming Beam | 658 nm ≤ 5 mW | 658 nm ≤ 5 mW |
| Beam Delivery | Fiber and Handpiece | Fiber and Handpiece |
| Emission Control | Footswitch | Footswitch |
| Display Screen | Yes | Yes |
| Power | 5 W | 12 W (1927 nm) |
| Max Pulse Energy | 20 mJ | 20 mJ (1927 nm) |
| Max Pulse Width | 20 ms | 10 ms |
| Pulse Repetition Rate | 43.5 - 307.7 Hz | 0 - 3 kHz |
| Tip Size | 4 mm x 10 mm | Small tip: 7 mm<br>Large tip: 15 mm |
| Dimensions of Laser Console | 413 mm (W) x 324 mm (L) x<br>255.5 mm (D) | 489 mm (W) x 451 mm (L) x<br>444.5 mm (D) |
| Spot Size | 100 $ μ $ m, 200 $ μ $ m | 200 $ μ $ m |
| Weight of Laser System | 33.1 kg | 68.0 kg |
| Electrical Rating | AC 100-240V, 50/60 Hz,<br>Power Consumption | AC 100-240V, 50/60 Hz,<br>Power Consumption |
#### 7. Performance Data
The Company's Performance Data for the LASEMD Laser System is as follows:
# Bench Testing
The LASEMD Laser System complies with all applicable standards, including ISO 13485:2003, ISO 60601-1 for electrical safety and IEC 60601-1-2 for electromagnetic compatibility.
# Animal Testing
Histology confirms that the treatment skin effect of the LASEMD Laser System is the same as the predicate Fraxel DUAL 1927 nm Laser System when used at the same energy settings.
# Clinical Testing
No performance data has been provided since the LASEMD Laser System is equivalent to the previously cleared predicate device with no new issues regarding safety and effectiveness.
#### 8. Conclusion
The LASEMD Laser System and the legally marketed Fraxel DUAL Laser System cleared under 510(k) number K130193, have the same intended use and Indications for Use statement. While the technological characteristics differ between the two systems, the differences have been established to be minor. Performance testing data established that the LASEMD Laser System is as safe and effective as the legally marked predicate
{7}------------------------------------------------
device and that the LASEMD Laser System does not raise any different questions of safety and effectiveness than the predicate. On this basis and in accordance with 21 CFR \$ 807.100(b), the LASEMD Laser System is substantially equivalent to the Fraxel DUAL Laser System and can be legally marketed in the U.S.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.