Monosyn Quick Synthetic Absorbable Surgical Suture is intended for general soft tissue approximation of the skin and mucosa, where only short term wound support (6-7 days) is required. Monosyn Quick suture is not indicated for use in cardiovascular or neurosurgery.
Device Story
Monosyn Quick is a sterile, absorbable, flexible monofilament surgical suture composed of a synthetic polyglycolic acid-based copolymer (glycolide, ε-caprolactone, and trimethylene carbonate). It is supplied in various diameters (USP 6-0 to 1) and lengths, with or without attached stainless steel needles. Used by clinicians for soft tissue approximation, the device provides short-term wound support, maintaining 60-70% tensile strength at 5 days and 25-35% at 10 days, with absorption essentially complete by 56 days. The suture is intended for skin and mucosal closure. It is packaged in a sterile pouch and sterilized via ethylene oxide.
Clinical Evidence
Bench testing only. Performance demonstrated via mechanical testing per USP 39 (diameter, tensile strength, needle attachment) and biocompatibility testing per ISO 10993 (cytotoxicity, local effects, irritation, sensitization, systemic toxicity, EO residuals). Accelerated aging data provided. No clinical data included.
Technological Characteristics
Synthetic absorbable monofilament; material: 72% glycolide, 14% ε-caprolactone, 14% trimethylene carbonate. USP 39 compliant for mechanical properties. Sterilized via Ethylene Oxide (ISO 11135-1). Packaging: PET/aluminum/LDPE or HDPE support in pouches.
Indications for Use
Indicated for general soft tissue approximation of skin and mucosa requiring short-term wound support (6-7 days). Contraindicated for cardiovascular or neurosurgery.
Regulatory Classification
Identification
An absorbable poly(glycolide/l-lactide) surgical suture (PGL suture) is an absorbable sterile, flexible strand as prepared and synthesized from homopolymers of glycolide and copolymers made from 90 percent glycolide and 10 percent l-lactide, and is indicated for use in soft tissue approximation. A PGL suture meets United States Pharmacopeia (U.S.P.) requirements as described in the U.S.P. “Monograph for Absorbable Surgical Sutures;” it may be monofilament or multifilament (braided) in form; it may be uncoated or coated; and it may be undyed or dyed with an FDA-approved color additive. Also, the suture may be provided with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around it. Inside the circle is a stylized image of three human profiles facing right, with flowing lines connecting them.
DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 23, 2017
Aesculap®, Inc. Ms. Kathy Racosky Senior Regulatory Affairs Specialist 3773 Corporate Parkway Center Valley, Pennsylvania 18034
Re: K171001
Trade/Device Name: Monosyn Quick Synthetic Absorbable Surgical Suture Regulation Number: 21 CFR 878.4493 Regulation Name: Absorbable Poly(Glycolide/L-Lactide) Surgical Suture Regulatory Class: Class II Product Code: GAM Dated: September 12, 2017 Received: September 13, 2017
Dear Ms. Racosky:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
# David Krause -S
Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
apple Synggerics Aps 25 25 27 2017 11 20
Expiration Date: January 31, 2017
See PRA Statement below.
Monosyn
### 510(k) Prematic Metifisetion OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
# K171001
#### Device Name
Aesculap Monosyn Quick Synthetic Absorbable Surgical Suture
Indications for Use (Describe)
Monosyn Quick Synthetic Absorbable Surgical Suture is intended for general soft tissue approximation of the skin and mucosa, where only short term wound support (6-7 days) is required. Monosyn Quick suture is not indicated for use in cardiovascular or neurosurgery.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Page 1 of 4
#### B. 510(k) SUMMARY (as required by 21 CFR 807.92)
Aesculap Monosyn Quick Synthetic Absorbable Surgical Suture September 12, 2017
| COMPANY: | Aesculap®, Inc.<br>3773 Corporate Parkway<br>Center Valley, PA 18034<br>Establishment Registration Number: 2916714 |
|----------------------|--------------------------------------------------------------------------------------------------------------------|
| CONTACT: | Kathy A. Racosky<br>610-984-9291 (phone)<br>610-791-6882 (fax)<br>kathy.racosky@aesculap.com |
| TRADE NAME: | Monosyn Quick Synthetic Absorbable Surgical Suture |
| COMMON NAME: | Synthetic Absorbable Surgical Suture |
| CLASSIFICATION: | Class II |
| CLASSIFICATION NAME: | Absorbable poly(glycolide/l-lactide) surgical suture |
| REGULATION NUMBER: | 878.4493 |
| PRODUCT CODE: | GAM |
# PURPOSE FOR PREMARKET NOTIFICATION
The purpose for this submission is to gain marketing clearance for the Monosyn Quick Synthetic Absorbable Surgical Suture.
# PREDICATE DEVICE
- Primary predicate: Monosyn Synthetic Absorbable Surgical Suture (K011375) .
- . Reference predicate: Caprosyn Absorbable Surgical Suture (K032586)
### DEVICE DESCRIPTION
Monosyn Quick is an absorbable flexible monofilament suture which is supplied sterile. It is composed of a synthetic polyglycolic acid-based copolymer. The Monosyn Quick suture is undyed and will be offered in diameters ranging from USP size 6-0 through 1. It is uncoated and will be available in a variety of cut lengths with or with out needles attached.
# INDICATIONS FOR USE
Monosyn Quick Synthetic Absorbable Surgical Suture is intended for general soft tissue approximation of the skin and mucosa, where only short term wound support (6-7 days) is required. Monosyn Quick suture is not indicated for use in cardiovascular or neurosurgery.
{4}------------------------------------------------
# TECHNOLOGICAL CHARACTERISTICS(compared to Predicate(s)}
As established in this submission, the Monosyn Quick suture is a synthetic absorbable monofilament surgical suture offered undyed in the same range of diameters and cut lengths that are substantially equivalent to other predicate devices cleared by FDA. The subject device is shown to be substantially equivalent and has the same technological characteristics to its predicate devices through comparison in design, intended use, material composition, function and range of sizes. The device characteristics comparing the Monosyn Quick Synthetic Absorbable Surgical Suture to the predicate devices are summarized below.
| | Aesculap Inc.<br>Monosyn Quick Synthetic<br>Absorbable Surgical<br>Suture | Aesculap, Inc.<br>Monosyn Synthetic<br>Absorbable Surgical<br>Suture | United States Surgical<br>Caprosyn Absorbable<br>Surgical Suture |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K# | Proposed device | K011375 | K032586 |
| Indications | Monosyn Quick Synthetic<br>Absorbable Surgical Suture<br>is intended for general soft<br>tissue approximation of the<br>skin and mucosa, where<br>only short term wound<br>support (6-7 days) is<br>required. Monosyn Quick<br>suture is not indicated for<br>use in cardiovascular or<br>neurosurgery. | Monosyn Synthetic<br>Absorbable Surgical Suture<br>is indicated for use in<br>general soft tissue<br>approximation and/or<br>ligation, but not for use in<br>cardiovascular or<br>neurological surgery. | Caprosyn synthetic<br>absorbable sutures are<br>indicated for use in general<br>soft tissue approximation<br>and/or ligation, including<br>use in ophthalmic<br>procedures, but not for use<br>in cardiovascular or<br>neurological surgery, or<br>microsurgery. |
| Absorption Type | Short term absorbable | Mid term absorbable | Short term absorbable |
| Absorption | Essentially complete by 56<br>days | Essentially completed<br>between 60-90 days | Essentially complete by 56<br>days |
| Remaining Tensile<br>Strength | 5 days 60-70%<br>10 days 25-35% | 14 days 50% | 5 days 60%<br>10 days 20-30% |
| Material | 72% glycolide<br>14% ε-caprolactone<br>14% trimethylene carbonate | 72% glycolide<br>14% ε-caprolactone<br>14% trimethylene carbonate | Glycolide, Caprolactone,<br>Trimethylene Carbonate<br>and Lactide |
| Dyed, Un-dyed | Undyed | Un-dyed and Dyed | Un-dyed and Dyed |
| Structure | Monofilament | Monofilament | Monofilament |
| Size | 6-0 through 1 (various<br>lengths) with or w/out<br>needles attached | 5-0 through 1 (various<br>lengths) with or w/out<br>needles attached | 6-0 through 1 (various<br>lengths) with or w/out<br>needles attached |
| Thread length | -45 cm to 120 cm | -35 cm to 150 cm<br>-ligature reels of longer<br>length | -18 in to 60 in<br>-ligature reels of longer<br>length |
| Physical:<br>- Diameter<br>- Length<br>- Needle Attachment<br>- Tensile Strength | All characteristics meet<br>USP Requirements, except<br>for diameter. | All characteristics meet<br>USP Requirements, except<br>for diameter. | All characteristics meet<br>USP Requirements, except<br>for diameter. |
| Needle material | 300 or 400 series stainless<br>steel | 300 or 400 series stainless<br>steel | unknown |
3
{5}------------------------------------------------
## Page 3 of 4
| | Aesculap Inc.<br>Monosyn Quick Synthetic<br>Absorbable Surgical<br>Suture | Aesculap, Inc.<br>Monosyn Synthetic<br>Absorbable Surgical<br>Suture | United States Surgical<br>Caprosyn Absorbable<br>Surgical Suture |
|---------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------|
| K# | Pending | K011375 | K032589 |
| Packaging | Coated cardboard support<br>in double peel pouch. Inner<br>pouch composed of PET,<br>aluminum foil and LDPE<br>layers. Medical grade<br>paper and PET/LDPE<br>laminate outer pouch.<br>or<br>Oval HDPE support with a<br>coated cardboard lid in a<br>single pouch composed of<br>polyethylene terephthalate<br>polyester (PETP).<br>or<br>cardboard support fold card<br>in a single pouch composed<br>of polyethylene<br>terephthalate polyester<br>(PETP). | Coated cardboard support<br>in double peel pouch. Inner<br>pouch composed of PET.<br>aluminum foil and LDPE<br>layers. Medical grade<br>paper and PET/LDPE<br>laminate outer pouch. | unknown |
| Sterilization | Ethylene Oxide (EO) | Ethylene Oxide (EO) | unknown |
# PERFORMANCE DATA
As recommended by the FDA's Class II Special Control Guidance Document for Surgical Sutures, including mechanical testing in accordance to USP 39 for synthetic absorbable suture, biocompatibility testing in accordance to ISO 10993-1, and in vitro as well as in vivo resorption testing has been performed to demonstrate that the Monosyn Quick Synthetic Absorbable Surgical Suture meets current performance requirements for synthetic absorbable sutures unless otherwise labeled, and that Monosyn Quick is substantially equivalent to other predicate devices. The test results demonstrate that the Monosyn Quick complies with the following standards:
USP 39 Monograph for Absorbable Surgical Sutures
USP 39 <861> Sutures - Diameter (where diameter deviates minimally from USP requirements, it will be labeled as such)
USP 39 <881> Tensile Strength
USP 39 <871> Sutures - Needle Attachment
ISO 10993-1:2009, Biological evaluation of medical devices -- Part 1: Evaluation and testing within a risk management process. (Biocompatibility)
ISO 10993-5, Biological evaluation of medical devices -- Part 5: Tests for in vitro cytotoxicity ISO 10993-6 2007, Biological Evaluation Of Medical Devices - Part 6: Tests For Local Effects After Implantation
ISO 10993-7, Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals
ISO 10993-10 2010, Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization
{6}------------------------------------------------
Page 4 of 4
ISO 10993-11 2006, Biological Evaluation Of Medical Devices - Part 11: Tests For Systemic Toxicity
ISO 11135-1:2007, Sterilization of health care products- Ethylene oxide- Part 1: Requirements for development, validation and routing control of a sterilization process for medical devices
The Monosyn Quick Synthetic Absorbable Surgical Suture is blister packed and sterilized by Ethylene Oxide. Accelerated aging data for the Monosyn Quick has been generated to support this submission.
#### CONCLUSION:
Non clinical testing demonstrated that the device is as safe, as effective, and performs as well as the predicate devices.
In addition, residual strength and absorption rate studies were performed. The results of this testing demonstrates that the Monosyn Quick is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.