← Product Code [DQA](/productcode/DQA) · K170965

# Fingertip Pulse Oximeter (K170965)

_Shenzhen Jumper Medical Equipment Co., Ltd. · DQA · Nov 3, 2017 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K170965

## Device Facts

- **Applicant:** Shenzhen Jumper Medical Equipment Co., Ltd.
- **Product Code:** [DQA](/productcode/DQA.md)
- **Decision Date:** Nov 3, 2017
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.2700
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Pediatric

## Indications for Use

The Fingertip Pulse Oximeter JPD-500F is non-invasive device intended for spotchecking of functional oxygen saturation of arterial hemoglobin (Spo2) and pulse rate. The portable fingertip device is indicated for adult and pediatric patients in home and hospital environments (including clinical use in internist/ surgery, anesthesia, intensive care, etc).

## Device Story

Fingertip Pulse Oximeter JPD-500F; non-invasive, portable device for spot-checking SpO2 and pulse rate. Operates via optical sensing (photoplethysmography) on fingertip; processes light absorption signals to calculate oxygen saturation and pulse rate. Output displayed on integrated screen for immediate review by clinicians or patients. Used in home and hospital environments (e.g., surgery, anesthesia, intensive care). Provides real-time physiological data to assist in monitoring patient status.

## Clinical Evidence

No clinical data provided; substantial equivalence supported by bench testing.

## Technological Characteristics

Portable fingertip form factor; optical sensing (LED/photodiode) for SpO2 and pulse rate measurement; battery-powered; integrated digital display. No specific material standards or complex software algorithms described.

## Regulatory Identification

An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue box. To the right of the blue box is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 3. 2017

Shenzhen Jumper Medical Equipment Co., Ltd. % Migo Yang Consultant Shenzhen Joyantech Consulting Co., Ltd. 1122#, International Mayor Communication Center Baishizhong Road 55# Nanshan District, Shenzhen, 518000 CHINA

Re: K170965

Trade/Device Name: Fingertip Pulse Oximeter (Model JPD-500F) Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DOA Dated: September 15, 2017 Received: October 5, 2017

Dear Migo Yang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Tina Kiang
-s

Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

510(k) Number (if known)

K170965

Device Name

Fingertip Pulse Oximeter (Model: JPD-500F)

Indications for Use (Describe)

The Fingertip Pulse Oximeter JPD-500F is non-invasive device intended for spotchecking of functional oxygen saturation of arterial hemoglobin (Spo2) and pulse rate. The portable fingertip device is indicated for adult and pediatric patients in home and hospital environments (including clinical use in internist/ surgery, anesthesia, intensive care, etc).

| Type of Use (Select one or both, as applicable)   |                                             |
|---------------------------------------------------|---------------------------------------------|
| \[X] Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

 The burden time for this collection of information is estimated to average 79 hours per response, including the
 time to review instructions, search existing data sources, gather and maintain the data needed and complete
 and review the collection of information. Send comments regarding this burden estimate or any other aspect
 of this information collection, including suggestions for reducing this burden, to:

 Department of Health and Human Services
   
Food and Drug Administration
   
Office of Chief Information Officer
   
Paperwork Reduction Act (PRA) Staff
   
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

EF

PSC Publishing Services (301) 443-6740

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

---

**Source:** [https://fda-staging.innolitics.com/device/K170965](https://fda-staging.innolitics.com/device/K170965)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
