Skan-C Mobile C-Arm X-Ray System - 230V Variant, Skan-C Mobile C-Arm X-Ray System - 110V Variant
Applicant
Skanray Technologies Private Limited
Product Code
OXO · Radiology
Decision Date
Nov 24, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1650
Device Class
Class 2
Indications for Use
The Skan-C, a Mobile Surgical C-Arm X-Ray System, is intended to provide Fluoroscopic and Radiographic images of the patient during Diagnostic, Surgical and Interventional procedures. Examples of Clinical Applications may include Orthopaedic. GI Procedure like Endoscopy and Cholenography. Neurology, Urology Procedures, Vascular, Critical Care and Emergency Room Procedures. Skan-C is not recommended for Cardiac Applications. Skan-C Surgical C-Arm is indication in real time and/or recording of surgical region of interest and anatomy, using X-ray imaging technique.
Device Story
Mobile surgical C-arm X-ray system; provides real-time fluoroscopic and radiographic imaging for surgical/interventional guidance. System comprises mobile C-arm unit with X-ray generator/collimator and separate workstation for image display, storage, and manipulation. Operated by clinicians in OR, clinic, or emergency settings. Inputs: X-ray radiation passing through patient anatomy. Transforms inputs via image intensifier and CCD camera into digital images. Output: Real-time visualization and recording of surgical region of interest. Healthcare providers use output to guide surgical instruments, verify anatomical placement, and monitor procedures. Benefits include improved visualization of anatomy during minimally invasive interventions.
Clinical Evidence
No clinical trials performed. Evidence consists of usability testing with independent participants and qualitative assessment of imaging performance by radiologists. Usability study confirmed no discomfort or complex interfaces. Radiologist review confirmed image quality is adequate for indicated clinical applications. Bench testing included compliance with IEC 60601-1, IEC 60601-2-54, IEC 60601-2-43, IEC 60601-1-3, and IEC 60601-1-2, alongside FDA performance standards 21 CFR 1020.30-1020.32.
Technological Characteristics
Mobile C-arm X-ray system. Stationary anode X-ray tube (0.6/1.8 focal spot, 45KHU heat capacity). Triple-mode Thales image intensifier (9", 6", 4.5"). CCD camera (1k x 1k resolution). Max radiographic power 2.2 kW. Connectivity: Workstation for image display/storage. Power: 100-110 VAC, 50/60 Hz. Operating environment: 10-40°C, 20-80% humidity. Complies with IEC 60601 series and 21 CFR 1020.30-1020.32.
Indications for Use
Indicated for patients requiring fluoroscopic and radiographic imaging during diagnostic, surgical, and interventional procedures, including orthopaedic, GI (endoscopy, cholangiography), neurology, urology, vascular, critical care, and emergency room procedures. Not indicated for cardiac applications.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
November 24, 2017
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Skanray Technologies Private Limited % Parul Chansoria Regulatory Consultant Elexes Medical Consulting Pvt Ltd. #6494. Tralee Village Dr. Dr Dublin, California 94568
## Re: K170946
Trade/Device Name: Skan C Mobile C-Arm X-Ray system - 230V Variant (303-000187-1), Skan C Mobile C-Arm X-Ray system - 110V Variant (303-000187-2) Regulation Number: 21 CFR 892.1650 Regulation Name: Image-Intensified Fluoroscopic X-Ray System Regulatory Class: Class II Product Code: OXO, OWB Dated: October 17, 2017 Received: November 9, 2017
Dear Parul Chansoria:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Robert Oals
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K170946
Device Name
Skan C Mobile C-Arm X-Ray system - 230V Variant (303-000187-1), Skan C Mobile C-Arm X-Ray system - 110V (303000187-2)
Indications for Use (Describe)
The Skan-C, a Mobile Surgical C-Arm X-Ray System, is intended to provide Fluoroscopic and Radiographic images of the patient during Diagnostic, Surgical and Interventional procedures.
Examples of Clinical Applications may include Orthopaedic. GI Procedure like Endoscopy and Cholenography. Neurology, Urology Procedures, Vascular, Critical Care and Emergency Room Procedures.
Skan-C is not recommended for Cardiac Applications.
Skan-C Surgical C-Arm is indication in real time and/or recording of surgical region of interest and anatomy, using X-ray imaging technique.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------|-----------------------------------------------|
| <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### Section.5- 510K Summary 1
# 5.1. Sponsor
| Date: | September 28, 2017 |
|-----------------|-------------------------------------------------------------------------------------------------------|
| Applicant: | Skanray Technologies Private Limited,<br>Plot # 15-17, Hebbal Industrial Area<br>Mysore-570016, India |
| Contact Person: | Parul Chansoria, Regulatory Consultant, Elexes |
| Email Id: | parul@elexes.com |
| Phone: | 650-528-2445 |
# 5.2. Establishment Registration Number
| FDA | 3009001657 |
|-------------------------|------------|
| establishment | |
| registration<br>number: | |
# 5.3. Device Name
| Trade Name: | Skan C |
|--------------------|-----------------------------------------------------------------|
| Variants: | Skan C Mobile C-Arm X-Ray system-230V Variant<br>(303-000187-1) |
| | Skan C Mobile C-Arm X-Ray system-110V Variant<br>(303-000187-2) |
| Regulation Name: | lmage-Intensified Fluoroscopic X-Ray System |
| Regulation Number: | 892.1650 |
| Product Code: | OXO -Primary & OWB-Secondary |
# 5.4. Predicate Device
| GE | OEC Fluorostar - K043076 |
|---------------------------|---------------------------------------------|
| Regulation Name: | Image-intensified fluoroscopic x-ray system |
| Regulation number: | 892.1650 |
| Product Code: | OWB |
| Subsequent Product Codes: | JAA, OXO |
## 5.5. Reference Device
| Genoray | ZEN-2090 Pro - K091918 |
|------------------|---------------------------------------------|
| Regulation Name: | Image-intensified fluoroscopic x-ray system |
{4}------------------------------------------------
| Regulation number: | 892.1650 |
|---------------------------|----------|
| Product Code: | OWB |
| Subsequent Product Codes: | JAA, OXO |
5.6. Product Description
SKAN-C, is a mobile X-Ray C-Arm fluoroscopic device to assist in quiding medical intervention surgical procedures. The device can also be used for radiographic applications. The device is designed in such a way that it can be moved around and can be positioned for the required anatomical/clinical/procedural position.
SKAN-C, a Mobile Surgical C-Arm consists of two units, namely, Mobile Image Intensified C-Arm unit with generator, and a Work-Station for Image display, store and manipulation. C-Arm unit with generator is capable of movements which are essential for patient positioning, like horizontal travel, orbital movement, wig-wag movement and C rotation. The X-ray generator, X-Ray control system and collimator controls are housed in the C-Arm unit.
### 5.7. Indications for Use
The Skan-C, a Mobile Surgical C-Arm X-Ray System, is intended to provide Fluoroscopic and Radiographic images of the patient during diagnostic, surgical and interventional procedures.
Examples of clinical applications may include orthopaedic, GI procedures like endoscopy and cholenography, neurology, urology, vascular, critical care and emergency room procedures.
Skan-C is not recommended for cardiac applications.
Skan-C Surgical C-Arm is indicated for visualization in real time and/or recording of surgical region of interest and anatomy, using X-ray imaging technique.
### 5.8. Technology Characteristics
Major similarities between subject and predicate device.
| | | Skanray<br>Skan-C | GE OEC<br>Fluorostar | SE<br>Remark/Comment |
|-------------------------------|--------------------------------------------------------------|-------------------------------|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|
| X-Ray Tube | Type | Stationary | Stationary | Same |
| | Focal Spot | 0.6/1.8 | 0.5/1.5 | Similar, slightly higher focal spot is to improve the thermal performance of the generator. |
| | Anode Heat capacity | 45KHU | 46KHU | Similar, Is the characteristic of X-Ray Tube. Meets the criteria of 10min continuous |
| | | Skanray<br>Skan-C | GE OEC<br>Fluorostar | SE<br>Remark/Comment |
| | | | | exposure |
| Image<br>Intensifier | Make | Thales | Thales/Thomson | |
| | Modes | Triple Mode,<br>9", 6" & 4.5" | Triple Mode, 9",<br>6" & 4.5" | Same |
| | DQE (%) | 65 | 65 | Same |
| Camera | Technology | CCD | CCD | Same |
| | Resolution | 1k x 1k | 1k x 1k | Same |
| Max<br>Radiographi<br>c Power | | 2.2 kW | 2.2 kW | Same |
| Loading<br>Factors | Radiography<br>KV Range | 40-110 kV | 36-110 kV | Better Stability within the<br>range specified. |
| | mAs Range | 200 mAs | 52 mAs | Better mAs range for<br>dense anatomy due to<br>higher Focal Spot size. |
| | Fluoroscopy kV<br>Range | 40-110 kV | 36-110 kV | Better Stability within the<br>range specified |
| | Fluoroscopy<br>mA Range<br>(Normal and<br>boost Mode) | 0.2-8 mA | 0.2-8 mA | Same |
| Physical<br>Parameters | Free Space<br>Available<br>between<br>Source and<br>Detector | 75 Cm | 76 Cm | The minor difference in<br>the physical parameters<br>in terms of dimensions<br>does not have any impact<br>on the claimed indications<br>for use. |
| | SID | 95 Cm | 98 Cm | |
| | C-Arm Depth | 65 Cm | 66 Cm | |
| | Horizontal<br>Travel | 20 Cm | 20 Cm | |
| | Vertical<br>Movement | 40 Cm | 43 Cm | |
| | Orbital Rotation | (125°)90°/- | (120°)90°/- | |
| | | Skanray<br>Skan-C | GE OEC<br>Fluorostar | SE<br>Remark/Comment |
| | | 35° | 30° | |
| | Angulation | $180° (+/- 180°)$ | $450° (+225°/- 225°)$ | |
| | Wig-Wag | +-12,5° | +-10° | |
| Operating<br>Environment | Input Energy<br>Source | 100-110 VAC,<br>50/60 Hz | 100-110 VAC,<br>50/60 Hz | Same |
| | Operating<br>Temperature<br>Range | 10deg C to<br>40deg C | 10deg C to<br>40deg C | Same |
| | Operating<br>Humidity<br>Range | 20- 80 % Non-<br>Condensing | 20- 80 % Non-<br>Condensing | Same |
{5}------------------------------------------------
{6}------------------------------------------------
#### Table 1: Technological characteristics
With the above comparison, the subject device, Skan C, is equivalent in technological and other characteristics to the predicate device. GE OEC Fluorostar.
#### 5.9. FDA Guidance Documents
The following guidance documents were referred during development of the subject device:
- 1. Content of Premarket Submissions for Management of Cybersecurity in Medical Devices.
- 2. Off-The-Shelf Software Use in Medical Devices.
- 3. Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.
- 4. Draft Guidance for Industry and Food and Drug Administration Staff -Paediatric Information for X-ray Imaging Device Premarket Notifications
- 5. Guidance for the Submission of 510(k)s for Solid State X-ray Imaging Devices - Guidance for Industry and Food and Drug Administration Staff.
#### 5.10. Non-Clinical Tests:
The subject device has been evaluated and found to be compliant to safety and essential performance as per IEC 60601-1, accuracy of loading factors, reproducibility
{7}------------------------------------------------
of radiation output and imaging performance as per IEC 60601-2-54, recovery management, patient data, last image hold, image measuring functions as per IEC 60601-2-43, half value layer, leakage radiation in loading and normal state and strav radiation as per IEC 60601-1-3, conducted emission, radiated emission, harmonics, voltage fluctuations, electrostatic discharge (ESD), electrical fast transients (EFT), radiated RF electromagnetic field, continuous conducted RF, surges, power frequency magnetic field, voltage dips and short interruption as per IEC 60601-1-2, detective quantum Efficiency (DQE), spatial resolution, dynamic range, beam alignment, recovery time, dose requirements, stability of device characteristics with time, frame rate, reuse rate as per FDA guidance document for solid state X-ray imaging devices. The device also complies with FDA performance standards 21CFR 1020.30-1020.32. The verification and validation testing performed demonstrated compliance to the design requirements and intended use of SKAN C.
#### 5.11. Clinical Tests:
Usability aspects of the device were tested by the users and independent participants including the setting up device and the post imaging processes. The results of the study did not reveal any discomfort or complex user interfaces for the intended purpose of the device.
Independent views of Radiologists were obtained on the imaging performances and the acquired images were of adequate quality for the indicated use. The results of the validation activities confirmed that the device is safe and effective for its intended application.
#### 5.12. Conclusion
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the non-clinical tests, clinical tests and relative information provided in this premarket notification, we conclude that SKAN C Mobile C-arm X-ray system is substantially equivalent to predicate device with regard to safety and performance.
{8}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.