← Product Code [LOX](/productcode/LOX) · K170941

# RX NC Takeru PTCA Balloon Dilatation Catheter (K170941)

_KANEKA Corporation · LOX · Jul 7, 2017 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K170941

## Device Facts

- **Applicant:** KANEKA Corporation
- **Product Code:** [LOX](/productcode/LOX.md)
- **Decision Date:** Jul 7, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.5100
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The RX NC Takeru PTCA Balloon Dilatation Catheter is indicated for balloon dilatation of the stenotic portion in the coronary artery or bypass graft stenosis for the purpose of myocardial perfusion. This product is also indicated for the post-delivery expansion of balloon expandable stents.

## Device Story

RX NC Takeru is a rapid exchange PTCA balloon dilatation catheter used by interventional cardiologists in a catheterization lab. The device consists of a distal tube, guidewire transition tube, balloon, radiopaque markers, mid/proximal tubes, core wire, hub, and strain relief. The physician inserts a 0.014-inch guidewire through the catheter lumen to the target lesion. An external inflation device connected to the hub delivers pressurized inflation medium to expand the balloon, dilating the stenotic vessel or expanding a stent. The device facilitates myocardial perfusion by restoring blood flow. It is compatible with 5 or 6 Fr guiding catheters. The catheter is provided sterile for single use.

## Clinical Evidence

Bench testing only. No clinical data. Performance evaluated via dimensional verification, balloon fatigue, burst pressure, compliance, inflation/deflation time, bond strength, flexibility, kink, torque, radiopacity, coating integrity, and particulate evaluation. Biocompatibility testing (ISO 10993-1) leveraged from reference device.

## Technological Characteristics

Rapid exchange PTCA catheter; 1450 mm working length; 2.0-5.0 mm balloon diameters; 6-30 mm working lengths. Materials include hydrophilic coating. Sterilization via Ethylene oxide. Compatible with 0.014" guidewires and 5/6 Fr guiding catheters. Operates via hydraulic pressurization.

## Regulatory Identification

Standard PTCA Catheter: A PTCA catheter is a device that operates on the principle of hydraulic pressurization applied through an inflatable balloon attached to the distal end. A PTCA balloon catheter has a single or double lumen shaft. The catheter features a balloon of appropriate compliance for the clinical application, constructed from a polymer. The balloon is designed to uniformly expand to a specified diameter and length at a specific pressure as labeled, with well characterized rates of inflation and deflation and a defined burst pressure. The device generally features a type of radiographic marker to facilitate fluoroscopic visualization of the balloon during use. A PTCA catheter is intended for balloon dilatation of a hemodynamically significant coronary artery or bypass graft stenosis in patients evidencing coronary ischemia for the purpose of improving myocardial perfusion. A PTCA catheter may also be intended for the treatment of acute myocardial infarction; treatment of in-stent restenosis (ISR) and/or post-deployment stent expansion. Cutting/scoring PTCA Catheter: A cutting/scoring PTCA catheter is a balloon-tipped catheter with cutting/scoring elements attached, which is used in those circumstances where a high pressure balloon resistant lesion is encountered. A cutting/scoring PTCA catheter is intended for the treatment of hemodynamically significant coronary artery stenosis for the purpose of improving myocardial perfusion. A cutting/scoring PTCA catheter may also be indicated for use in complex type C lesions or for the treatment of in-stent restenosis.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is “Class II Special Controls Guidance Document for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters.” See § 870.1(e) for the availability of this guidance document.(b)
*Cutting/scoring PTCA Catheter* —(1)*Identification.* A cutting/scoring PTCA catheter is a balloon-tipped catheter with cutting/scoring elements attached, which is used in those circumstances where a high pressure balloon resistant lesion is encountered. A cutting/scoring PTCA catheter is intended for the treatment of hemodynamically significant coronary artery stenosis for the purpose of improving myocardial perfusion. A cutting/scoring PTCA catheter may also be indicated for use in complex type C lesions or for the treatment of in-stent restenosis.(2)
*Classification.* Class III (premarket approval). As of May 28, 1976, an approval under section 515 of the act is required before this device may be commercially distributed. See § 870.3.

## Predicate Devices

- NC Quantum Apex™ Monorail PTCA Dilatation Catheter ([P860019](/device/P860019.md)/S241)

## Reference Devices

- RX Takeru PTCA Balloon Dilatation Catheter ([K163372](/device/K163372.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the logo.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 7, 2017

Kaneka Corporation c/o Christopher M. Sloan Quintiles Consulting 1801 Rockville Pike, Suite 300, Rockville, MD 20852 US

Re: K170941

Trade/Device Name: RX NC Takeru Regulation Number: 21 CFR 870.5100 Regulation Name: Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheter Regulatory Class: Class II Product Code: LOX Dated: June 19, 2017 Received: June 20, 2017

Dear Christopher Sloan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA mav publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Sincerely,

Fernando Aguel

Fernando Aguel
-S

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

#### Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) K170941

Device Name

RX NC Takeru PTCA Balloon Dilatation Catheter

Indications for Use (Describe)

The RX NC Takeru PTCA Balloon Dilatation Catheter is indication of the stenotic portion in the coronary artery or bypass graft stenosis for the purpose of myocardial perfusion.

This product is also indicated for the post-delivery expansion of balloon expandable stents.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/14)

{3}------------------------------------------------

#### 5.0 510(K) SUMMARY

# RX NC Takeru PTCA Balloon Dilatation Catheter

#### 510(k) Submitter

Kaneka Corporation Medical Device Division 1-12-32, Akasaka, Minato-ku Tokyo, Japan Contact Person: Toshihiko Motomine Telephone: +81-3-5574-8023 Email:

### Official Correspondent

Christopher M. Sloan Principal Consultant Quintiles Consulting 1801 Rockville Pike, Suite 300 Rockville, MD 20852 Phone: (301) 272-3114 Email: chris.sloan@quintilesims.com

Date Prepared: June 9, 2017

{4}------------------------------------------------

## Subject Device Name:

| Trade Name           | RX NC Takeru                                                                                             |
|----------------------|----------------------------------------------------------------------------------------------------------|
| Common or usual name | Percutaneous Transluminal Coronary<br>Angioplasty (PTCA) catheter                                        |
| Classification name  | Percutaneous Transluminal Coronary<br>Angioplasty (PTCA) catheter [21 CFR<br>870.5100; product code LOX] |
| Class                | II                                                                                                       |
| Classification Panel | Cardiovascular (74)                                                                                      |

#### Predicate Device:

- NC Quantum Apex™ Monorail PTCA Dilatation Catheter or "NC Quantum Apex" ● [P860019/S241 (Boston Scientific Corporation)]
#### Reference Device:

- RX Takeru PTCA Balloon Dilatation Catheter or "RX Takeru" [K163372 (Kaneka Corporation)]
## Device Description:

RX NC Takeru PTCA Balloon Dilatation Catheter (RX NC Takeru) is a rapid exchange type of balloon dilation catheter, which consists of a distal tube, guidewire transition tube, balloon, radiopaque markers, mid tube, proximal tube, core wire, hub, and strain relief. A balloon is attached to the distal end of the catheter, and it can be inflated and deflated using the inflation device connected to the hub at the proximal end. RX NC Takeru has a guidewire lumen at the distal end of the catheter through which a guidewire can be inserted, and also an opening along the guidewire transition tube to the guidewire port for the exit of a guidewire.

The maximum compatible diameter of a guidewire used together with RX NC Takeru in a PTCA procedure is 0.014 inches. Additionally, guiding catheters with a diameter of 5 or 6 Fr have been deemed to be compatible with RX NC Takeru. The nominal inflated balloon diameters range from 2.0 mm to 5.0 mm with balloon working lengths of 6 mm to 30 mm. The catheter working length is 1450 mm.

{5}------------------------------------------------

## Indications for Use

The RX NC Takeru PTCA Balloon Dilatation Catheter is indicated for balloon dilatation of the stenotic portion in the coronary artery or bypass graft stenosis for the purpose of myocardial perfusion.

This product is also indicated for the post-delivery expansion of balloon expandable stents.

## Comparison of Indications for Use to Predicate Device

The RX NC Takeru has the same intended use (percutaneous transluminal coronary angioplasty) as the NC Quantum Apex. The indications for use of the RX NC Takeru is similar to those of the NC Quantum Apex. T Therefore, the subject device, RX NC Takeru, may be considered substantially equivalent to the predicate device.

## Comparison of Technological Characteristics to Predicate Device

Percutaneous Transluminal Coronary Angioplasty (PTCA) is the technological principle for both RX NC Takeru and the predicate device. PTCA is based on the use of Percutaneous Coronary Intervention (PCI) devices for the purpose of myocardial perfusion.

The RX NC Takeru and predicate device have following same technological elements:

- Operating principle balloon dilatation of stenotic portion by pressurization of . inflation medium
- . Fundamental catheter design - balloon, shaft, radiopaque marker, hub, hydrophilic coating
- Shaft type - rapid exchange
- Concomitantly used devices guidewire, guiding catheter, inflation device ●
- . Sterilization - Ethylene oxide

There are following minor technological differences between RX NC Takeru and predicate device:

- Combination of balloon diameter and balloon length .
- Shaft diameters
- Catheter effective length
- . Nominal Pressure and Rated Burst Pressure of certain balloon sizes

## Performance Testing

To demonstrate substantial equivalence of RX NC Takeru to the predicate device, the technological characteristics and performance criteria were evaluated using the bench testing recommendations outlined in the FDA Guidance Document "Class II Special Controls Guidance Document for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA)

{6}------------------------------------------------

Catheters" dated September 8, 2010. The following in vitro tests were performed on the subject device:

Dimensional Verification, Balloon Preparation, Deployment and Retraction, Balloon Rated Burst Pressure, Balloon Fatigue (Repeat Balloon Inflations), Balloon Compliance, Balloon Inflation and Deflation Time, Catheter Bond Strength, Flexibility and Kink Test, Torque Strength, Radiopacity, Coating Integrity, Particulate Evaluation, Balloon Rated Burst Pressure (in Stent), and Balloon Fatigue (Repeat Balloon Inflations; in Stent), Transportation and Shelf Life Testing.

The results from these tests demonstrate that the technological characteristics and performance criteria of the RX NC Takeru are adequate for the intended use of the device and that the device can perform in a manner equivalent to devices currently on the market with the same intended use.

## Biocompatibility:

To demonstrate the biological safety of the body-contacting materials and substantial equivalence of the RX NC Takeru to the predicate device, the following biocompatibility testing, which was performed on the reference device (RX Takeru) in accordance with "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process"; Guidance for Industry and Food and Drug Administration Staff" (dated June 16, 2016), was leveraged:

Cytotoxicity, Sensitization, Intracutaneous Reactivity, Acute Systemic Toxicity, Pyrogen, Hemocompatibility (in vivo thromboresistance, hemolysis, and complement), and Genotoxicity (mouse lymphoma, bacterial reverse mutation and in vivo cytogenetics assay)

The results from these tests demonstrate that the RX NC Takeru is biocompatible for its intended use similar to the predicate device.

## Conclusions:

The RX NC Takeru met all the predetermined acceptance criteria of design verification and validation as specified by applicable standards, guidance, test protocols and/or customer inputs. The RX NC Takeru PTCA Balloon Dilatation Catheter is substantially equivalent to legally marketed predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K170941](https://fda-staging.innolitics.com/device/K170941)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
