← Product Code [GEX](/productcode/GEX) · K170936

# Cutera enlighten III Laser System (K170936)

_Cutera, Inc. · GEX · Apr 25, 2017 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K170936

## Device Facts

- **Applicant:** Cutera, Inc.
- **Product Code:** [GEX](/productcode/GEX.md)
- **Decision Date:** Apr 25, 2017
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The enlighten III laser system is intended for use in surgical and aesthetic applications in the medical specialties of dermatology and general and plastic surgery. 1064 nm: The 1064 nm wavelength of the enlighten III laser system is indicated for: - treatment of benign pigmented lesions on patients with all skin types (Fitzpatrick I-VI) - tattoo removal for dark colored tattoo inks and for multicolored tattoos containing dark colored tattoo inks on patients with all skin types (Fitzpatrick I-VI) 532 nm: The 532 nm wavelength of the enlighten III laser system is indicated for: - treatment of benign pigmented lesions on patients with Fitzpatrick skin types I-III - tattoo removal for lighter colored tattoo inks, including red and yellow inks, on patients with Fitzpatrick skin types I-III 670 nm: The 670 nm wavelength of the enlighten III laser system is indicated for treatment of benign pigments with Fitzpatrick skin types I-III.

## Device Story

Multi-wavelength, pulsed laser system; console-housed laser, power supplies, control electronics; touchscreen interface for parameter selection; articulated arm with rotational mirror knuckles; optical zoom handpiece; low-power diode aiming beam; footswitch-activated. Used in dermatology, general, and plastic surgery clinics by trained professionals. Device delivers laser energy to skin surface; spot size adjusted via handpiece. Modifications enable access to full parameter range: increased max energy (800 mJ at 1064 nm; 400 mJ at 532 nm), additional spot sizes (9 and 10 mm), and 2 ns pulse duration for 670 nm. Max treatment fluences remain unchanged. Clinical benefit: precise removal of benign pigmented lesions and tattoos.

## Clinical Evidence

Bench testing only. Verification and validation testing (Report V0140 rA1) confirmed compliance with IEC 60601-1, IEC 60601-1-6, IEC 62366, IEC 60601-2-22, IEC 60825-1, AAMI/ANSI ES60601-1, and IEC 60601-1-2. No clinical data provided.

## Technological Characteristics

Multi-wavelength pulsed laser (1064 nm, 532 nm, 670 nm). Pulse widths: 660 ps, 750 ps, 2 ns. Articulated arm delivery with optical zoom handpiece. Touchscreen control. Standards: IEC 60601-1 (Ed 3.1), IEC 60601-1-6 (Ed 3.1), IEC 62366 (Ed 1.1), IEC 60601-2-22 (Ed 3.1), IEC 60825-1 (Ed 2.0), AAMI/ANSI ES60601-1, IEC 60601-1-2 (Ed 3).

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Cutera enlighten III Laser System ([K160488](/device/K160488.md))

## Reference Devices

- Quanta Systems SpA, Discovery Pico Family ([K163222](/device/K163222.md))
- Cynosure, RevLite ([K133254](/device/K133254.md))
- Quanta System Q-Plus T ([K073549](/device/K073549.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an image of three stylized human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 25, 2017

Cutera, Inc. Dr. Bradley Renton Vice President. Regulatory and Medical Affairs, & Compliance Officer 3240 Bayshore Blvd. Brisbane, California 94005

Re: K170936

Trade/Device Name: Cutera enlighten III Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: March 29, 2017 Received: March 30, 2017

Dear Dr. Bradley Renton:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR

{1}------------------------------------------------

Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely, Jennifer R. Stevenson -For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director

Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K170936

Device Name Cutera enlighten III Laser System

#### Indications for Use (Describe)

The enlighten III laser system is intended for use in surgical and aesthetic applications in the medical specialties of dermatology and general and plastic surgery.

1064 nm:

The 1064 nm wavelength of the enlighten III laser system is indicated for:

- · treatment of benign pigmented lesions on patients with all skin types (Fitzpatrick I-VI)
• tattoo removal for dark colored tattoo inks and for multicolored tattoos containing dark colored tattoo inks on patients with all skin types (Fitzpatrick I-VI)

532 nm:

The 532 nm wavelength of the enlighten III laser system is indicated for:

- · treatment of benign pigmented lesions on patients with Fitzpatrick skin types I-III
- tattoo removal for lighter colored tattoo inks, including red and yellow inks, on patients with Fitzpatrick skin types I-III

#### 670 nm:

The 670 nm wavelength of the enlighten III laser system is indicated for treatment of benign pigments with Fitzpatrick skin types I-III.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: 20px;">☑</span> |
|----------------------------------------------|-----------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C)  | <span style="font-size: 20px;">☐</span> |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# Section 5 Special 510(K) Summary – Device Modification

This 510(K) Summary of safety and effectiveness for the modified Cutera enlighten III laser system is submitted in accordance with the requirements of the SMDA 1990 and following guidance concerning the organization and content of a 510(K) summary.

| Applicant:                             | Cutera, Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address:                               | 3240 Bayshore Blvd., Brisbane, CA 94005                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Contact Person:                        | Bradley Renton                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|                                        | 415-657-5568 — phone<br>415-715-3568 – fax<br>brenton@cutera.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Preparation Date:                      | April 19, 2017                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Device Trade Name:                     | Cutera enlighten III Laser System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Common Name:                           | Dermatology Laser                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Classification Name:                   | Instrument, Surgical, Powered, laser<br>79-GEX, 21 CFR 878.4810                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Legally Marketed<br>Predicate Devices: | Cutera enlighten III Laser System (K160488)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Legally Marketed<br>Reference Devices: | Quanta Systems SpA, Discovery Pico Family (K163222)<br>Cynosure, RevLite (K133254)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Device Description:                    | The modified Cutera enlighten III laser system is a multi-<br>wavelength, pulsed laser system. A key feature of the device is its<br>ability to produce multiple laser wavelengths and pulse widths<br>(1064 and 532 nm: 750 ps and 2 ns; 670 nm: 660 ps and 2 ns).                                                                                                                                                                                                                                                                                                                                                                                    |
|                                        | The laser, power supplies, and control electronics are housed<br>inside a console equipped with a touchscreen control panel. The<br>laser treatment parameters are selected using the control panel.<br>Laser emission is activated by depressing a footswitch. An<br>articulated arm with rotational mirror knuckles delivers the laser<br>beam from a laser head inside the console to a handpiece. The<br>handpiece is equipped with an optical zoom assembly that<br>determines the spot size of the laser beam on the treatment<br>surface. The beam of a low-power diode laser is also transmitted<br>through the arm to provide an aiming beam. |
| Indications for Use:                   | 1064 nm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|                                        | The 1064 nm wavelength of the Cutera enlighten III laser system is<br>indicated for:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |

- treatment of benign pigmented lesions on patients with all . skin types (Fitzpatrick I-VI)

{4}------------------------------------------------

- tattoo removal for dark colored tattoo inks and for . multicolored tattoos containing dark colored tattoo inks on patients with all skin types (Fitzpatrick I-VI)
## 532 nm

The 532 nm wavelength of the Cutera enlighten III laser system is indicated for:

- treatment of benign pigmented lesions on patients with ● Fitzpatrick skin types I-III
- tattoo removal for lighter colored tattoo inks, including red and yellow inks, on patients with Fitzpatrick skin types I-III

# 670 nm

The 670 nm wavelength of the Cutera enlighten III laser system is indicated for treatment of benign pigmented lesions on patients with Fitzpatrick skin types I-III.

- Device Modifications The purpose of this submission is to modify the control software of the previously cleared Cutera enlighten III laser system to enable access to the full parameter range of the enlighten III laser system. The modified and unmodified systems are identical in mechanical, electrical, and thermal design. The modified software will:
	- . Change the maximum delivered energy from 600 mJ to 800 mJ for the 1064 nm wavelength and from 300 mJ to 400 mJ for the 532 nm wavelength, while maintaining the previously cleared maximum treatment fluences of 10 J/cm² and 2.5 J/cm² for the 1064 nm and 532 nm wavelengths, respectively.
	- Add 9 and 10 mm incremental spot sizes for the 1064 and 532 ● nm wavelengths.
	- . Add 2 ns pulse duration for the 670 nm wavelength.

Cutera has determined that no new risks have been introduced due to the above changes, as the maximum treatment fluence for all wavelengths remains unchanged and there are no changes to the intended use or indications for use.

#### Performance Data: enlighten III Software Verification and Validation Testing Report (V0140 rA1)

The modified and unmodified devices are electrically, mechanically, and thermally identical and comply with the following standards:

- IEC 60601-1 Medical Electrical Equipment Part 1: General ● Requirements for Safety (Edition 3.1, 2012), including:
	- IC 60601-1-6 Medical Electrical Equipment Part 1-6: o General Requirements for Safety - Collateral Standard:

{5}------------------------------------------------

Usability (Edition 3.1, 2013)

- IEC 62366 Medical Devices Application of Usability O Engineering to Medical Devices (Edition 1.1, 2014)
- IEC 60601-2-22 Medical Electrical Equipment Part 2: O Particular Requirements for Basic Safety and Essential Performance of Surgical, Cosmetic, Therapeutic and Diagnostic Laser Equipment (Edition 3.1, 2012)
- IEC 60825-1 Safety of Laser Products Part 1: Equipment o Classification and Requirements (Edition 2.0, 2007)
- . The product also fulfills the requirements of AAMI/ANSI ES60601-1:2005+A2 (R2012) + A1.
- . IEC 60601-1-2 Medical Electrical Equipment - Part 1-2: General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility (Edition 3, 2007)

Summarv of Technological Characteristics: The modified Cutera enlighten III laser system has the same intended use and indications for use and the same fundamental scientific technology as the previously cleared unmodified Cutera enlighten III laser system (K160488), Quanta System Q-Plus T (K073549), and Cynosure RevLite (K133254), The device design and components are also very similar to these devices. For each of these devices:

- the user interface is located at the front/top of the console;
- . the treatment handpiece is attached to an articulating arm that is connected to the main system console;
- . the laser output at each wavelength is generated within the laser chassis and delivered to the skin through the handpiece attached to the articulated arm; and
- the handpiece allows the spot size on the skin to be . adjusted according to device specifications.

Each system consists of an articulating arm and attached handpiece, as well as an electrically powered system console that produces the laser energy and houses the software and user interface.

The modified Cutera enlighten III laser system provides similar key design aspects, including the same or similar spot sizes, laser wavelengths, pulse widths, and laser types as its predicate and reference devices. The repetition rate of the enlighten III laser system is the same as or within the repetition rate range of the

{6}------------------------------------------------

predicate device. Further, each of the devices presents a range of spot sizes to allow the user to choose the most appropriate spot size for each patient. Therefore, the minor differences do not raise any new safety or effectiveness questions, because the enlighten III parameters are the same as or within the range of the predicate and reference devices.

- Conclusion: The enlighten III laser system is substantially equivalent to the predicate device in terms of indications for use, technical specifications, operating performance features, and general design.

---

**Source:** [https://fda-staging.innolitics.com/device/K170936](https://fda-staging.innolitics.com/device/K170936)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).