← Product Code [POP](/productcode/POP) · K170926

# BrainPulse 1100 (K170926)

_Jan Medical, Inc. · POP · Apr 28, 2017 · Neurology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K170926

## Device Facts

- **Applicant:** Jan Medical, Inc.
- **Product Code:** [POP](/productcode/POP.md)
- **Decision Date:** Apr 28, 2017
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 882.1630
- **Device Class:** Class 2
- **Review Panel:** Neurology

## Indications for Use

The BrainPulse is intended for use on a patient's head to non-invasively detect, amplify and capture the skull motion caused by pulsatile flow from the cardiac cycle. The BrainPulse is not indicated to aid in the diagnosis of neurological conditions, diseases, or disorders.

## Device Story

BrainPulse 1100 captures skull motion synchronized with cardiac cycle; utilizes headset with six piezoelectric accelerometers, one PPG sensor for heart rate/timing, and one SPL sensor for ambient noise. Data collector digitizes analog signals; computer provides user interface, hardware control, and data storage. Used in clinical settings; operated by healthcare personnel. Recording duration typically 2-3 minutes (up to 30 minutes). Device displays real-time signals, post-recording reviews, and calculated signal statistics. Benefits include non-invasive monitoring of skull motion related to cardiac pulsatile flow.

## Clinical Evidence

Bench testing only. Design verification and validation performed per 21 CFR 820.30. Software validation (Level of Concern: Moderate/Level B) and system verification confirmed compliance with design inputs and equivalence to the predicate device.

## Technological Characteristics

System components: headset, data collector, computer. Sensors: 6 piezoelectric accelerometers, 1 PPG, 1 SPL. Materials: LDPE, PE, PEEK, Urethane. Signal processing: <20 Hz (mostly <10 Hz). Connectivity: Wired headset to data collector. Software: User interface, hardware control, signal statistics calculation.

## Regulatory Identification

A cranial motion measurement device is a prescription device that utilizes accelerometers to measure the motion or acceleration of the skull. These measurements are not to be used for diagnostic purposes.

## Special Controls

In combination with the general controls of the FD&C Act, the Cranial Motion Measurement Device is subject to the following special controls:

- 1. The technical parameters of the device, hardware and software, must be fully characterized and include the following information:
	- a. Hardware specifications must be provided. Additionally, verification and validation testing as well as a hazard analysis must be performed.
	- b. Software must be described in detail in the Software Requirements Specification (SRS) and Software Design Specification (SDS). Additionally, software verification and validation testing as well as a hazard analysis must be performed.
- 2. The patient-contacting components of the device must be demonstrated to be biocompatible.
- 3. The device must be designed and tested for electrical, thermal and mechanical safety and electromagnetic compatibility (EMC).
- 4. Clinical performance testing must demonstrate the accuracy, precision, stability, and repeatability of measuring cranial motion per the intended use in the intended use environment.
- 5. The labeling must include:
	- The intended use population and the intended use environment. a.
	- b. Any instructions technicians should convey to patients regarding the collection of cranial acceleration data to ensure device measurement accuracy, precision, stability, and repeatability.
	- c. Information allowing clinicians to understand potential sources of variability in the measurement to help recognize and identify changes in the measurement.

*Classification.* Class II (special controls). The special controls for this device are:(1) The technical parameters of the device, hardware and software, must be fully characterized and include the following information:
(i) Hardware specifications must be provided. Additionally, verification and validation testing as well as a hazard analysis must be performed.
(ii) Software must be described in detail in the Software Requirements Specification (SRS) and Software Design Specification (SDS). Additionally, software verification and validation testing as well as a hazard analysis must be performed.
(2) The device parts that contact the patient must be demonstrated to be biocompatible.
(3) The device must be designed and tested for electrical, thermal, and mechanical safety, and electromagnetic compatibility (EMC).
(4) Clinical performance testing must demonstrate the accuracy, precision, stability, and repeatability of measuring cranial motion per the intended use in the intended use environment.
(5) The labeling must include:
(i) The intended use population and the intended use environment.
(ii) Instructions for technicians to convey to patients regarding the collection of cranial acceleration data to ensure device measurement accuracy, precision, stability, and repeatability.
(iii) Information allowing clinicians to understand potential sources of variability in the measurement to help recognize and identify changes in the measurement.

## Predicate Devices

- BrainPulse 1100 ([DEN140040](/device/DEN140040.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Public Health Service

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other, forming a single, abstract shape.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 28, 2017

Jan Medical, Inc. Prabhu Raghavan VP of Clinical, Quality and Regulatory 110 Pioneer Wav. Suite L Mountain View, California 94041

Re: K170926

Trade/Device Name: BrainPulse 1100 Regulation Number: 21 CFR 882.1630 Regulation Name: Cranial Motion Measurement Device Regulatory Class: Class II Product Code: POP Dated: March 27, 2017 Received: March 29, 2017

Dear Mr. Raghavan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

# Michael J. Hoffmann -S

for

Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K170926

Device Name BrainPulse 1100

Indications for Use (Describe)

The BrainPulse is intended for use on a patient's head to non-invasively detect, amplify and capture the skull motion caused by pulsatile flow from the cardiac cycle. The BrainPulse is not in the diagnosis of neurological conditions, diseases, or disorders.

| Type of Use (Select one or both, as applicable)                                                |                                             |
|------------------------------------------------------------------------------------------------|---------------------------------------------|
| <span style="text-decoration: overline;">x</span> Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) Summary

BrainPulse™ 1100

| Date Prepared                        | March 27, 2017                                                                                                                                                                                                                                                                                  |
|--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submission number                    | K170926                                                                                                                                                                                                                                                                                         |
| Company Name and Address             | Jan Medical Inc.<br>110 Pioneer Way, Suite L<br>Mountain View, California 94041<br>USA                                                                                                                                                                                                          |
| Contact Person                       | Prabhu Raghavan<br>Vice President of Clinical, Quality and Regulatory<br>Phone: 1-650-316-8813<br>Fax:     1 650-316-8812                                                                                                                                                                       |
| Device:                              | BrainPulse™ 1100                                                                                                                                                                                                                                                                                |
| Trade Name:                          | BrainPulse                                                                                                                                                                                                                                                                                      |
| Predicate Device:                    | BrainPulse 1100 (de novo granted under DEN140040)                                                                                                                                                                                                                                               |
| Requlation Number:                   | 21 CFR 882.1630                                                                                                                                                                                                                                                                                 |
| Product Code:                        | POP                                                                                                                                                                                                                                                                                             |
| Common/Classification Name:          | Cranial Motion Measurement Device                                                                                                                                                                                                                                                               |
| Classification Advisory<br>Committee | Neurology                                                                                                                                                                                                                                                                                       |
| Review Advisory Committee            | Neurology                                                                                                                                                                                                                                                                                       |
| Requlatory Class:                    | -                                                                                                                                                                                                                                                                                               |
| Intended Use:                        | The BrainPulse is intended for use on a patient's head to non-<br>invasively detect, amplify and capture the skull motion caused<br>by pulsatile flow from the cardiac cycle. The BrainPulse is not<br>indicated to aid in the diagnosis of neurological conditions,<br>diseases, or disorders. |

{4}------------------------------------------------

all are non-sterile and reusable: . Headset with detachable cable that is patient contacting and includes: One Photoplethysmograph (PPG) sensor for o detecting the heart rate and timing; One Sound Pressure Level (SPL) sensor for o detecting ambient environment noise; and Six Accelerometer sensors to detect the o acceleration at six selected locations. ● Data Collector, which digitizes the analog signals from the headset . Computer, which incorporates the device Software and space to store the BrainPulse recording data and, . Device Software, which provides the user interface, hardware control software libraries The BrainPulse collects and stores skull motion caused by pulsatile flow from the cardiac cycle. The normal brain structure produces a motion that is driven by and synchronized with the heart rate, and is manifested by slight acceleration of the skull. The device uses piezoelectric-based accelerometer sensors that measure skull motion rather than brain sounds. The typical frequencies that are employed with the BrainPulse are below 20 Hz and mostly below 10 Hz, well below the lower limit of audible sound. The headset senses the motion and the Data Collector digitizes the signal. The computer mainly provides the user interface and stores the data for further processing. Users place the device's headset on patients and setup the user interface to perform a BrainPulse recording. Typically, a recording is about 2-3 minutes long, though users may obtain recordings up to 30 minutes long depending on their applications. Recordings may be obtained at any time. Patient Contacting LDPE (Low Density Poly Ethylene), PE (Poly Ethylene), Materials of Use PEEK (Polyether ether ketone) and Urethane Substantial Equivalence: The current submission is for the same device and model that was cleared by the agency in de novo, DEN140040. The only change is to the software user interface. The hardware is

The BrainPulse™ 1100 is a non-invasive device having three

basic components that are provided as an entire system and

Device Description:

{5}------------------------------------------------

|                              | unchanged; it is the same hardware that was cleared in DEN140040. Software validation as well as system verification was performed to show equivalence to the previous software that was cleared in DEN140040. This validation and verification testing demonstrated that the device output is equivalent to the output of the previous device version.                                                                                                                                                                                                                                                                                                                                                          |
|------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Changes to Predicate Device: | The device hardware elements are unchanged, i.e., there are no changes to the headset, data collector, tablet or cables. The device software was updated to provide the following: <ul><li>An improved user interface that includes options to add hospital and patient details such as metadata to the signal, options to view the signal in real-time, options to review the signal after completing a recording, and automatic stopping of recording based on specified time and heartbeat intervals.</li><li>Calculate and display statistics from the signal that are helpful for signal evaluation.</li></ul>                                                                                              |
| Testing Summary:             | Design Verification and Validation was performed in accordance with approved internal design controls procedures, and in accordance with 21 CFR 820.30. Risk analysis was conducted to review the updated software and to confirm that all software-based risk controls were implemented and verified. An overall system verification was also performed to ensure that the system operates as per design inputs. Software validation was performed with a Level of Concern determined to be moderate, or level B (the same level as the software that was reviewed in DEN140040). The device met all acceptance criteria during verification and validation tests and demonstrated compliance to design inputs. |
| Biocompatibility Summary     | The changes in this submission were limited to software and therefore, no additional biocompatibility was deemed necessary. Jan Medical submitted the evaluation of biocompatibility of the BrainPulse 1100 device components for direct and indirect patient contact and for biocompatibility in compliance to ISO 10993 Part 1 Biological Evaluation of Medical Devices as part of DEN140040.                                                                                                                                                                                                                                                                                                                  |

---

**Source:** [https://fda-staging.innolitics.com/device/K170926](https://fda-staging.innolitics.com/device/K170926)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).