K170914 · Texasmedical Technologies, Inc. · DQY · Nov 21, 2017 · Cardiovascular
Device Facts
Record ID
K170914
Device Name
TXM Support Catheter
Applicant
Texasmedical Technologies, Inc.
Product Code
DQY · Cardiovascular
Decision Date
Nov 21, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Indications for Use
The TXM Support Catheters are percutaneous, single lumen catheters designed for use in the peripheral vascular system. TXM Support Catheters are intended to guide and support a guide wire during access of the vasculature, allow for wire exchanges and provide a conduit for the delivery of saline solutions or diagnostic contrast agents.
Device Story
TXM Support Catheter is an over-the-wire (OTW), single-lumen percutaneous catheter. Device features atraumatic tapered tip, lubricious hydrophilic coating, and 6 radiopaque markers for positioning. Available in 8 models (straight or 30° angled tips) with 4 French outer diameter. Used in peripheral vascular system by clinicians to support guidewire navigation, enable wire exchanges, and deliver fluids/contrast. Operates via manual insertion and manipulation. Benefits include improved trackability and navigation through vasculature. Device is single-use and supplied sterile.
Clinical Evidence
Bench testing only. No clinical data. Performance verified via bench tests including dimensional attributes, trackability, crossability, stiffness (ASTM D747-10), flow rates, burst pressure, tensile strength, radiopacity (ASTM-F640-12), torqueability, kink resistance, and coating integrity. Biocompatibility testing performed per ISO 10993 (cytotoxicity, irritation, sensitization, systemic toxicity, pyrogenicity, hemocompatibility).
Indicated for use in the peripheral vascular system to guide and support a guidewire during vascular access, facilitate wire exchanges, and deliver saline or diagnostic contrast agents.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
Spectranetics Quick-Cross Extreme Support Catheters (K082561)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION".
November 21, 2017
Texas Medical Technologies, Inc. % Mr. E.J. Smith Consultant Smith Associates 1468 Harwell Avenue Crofton, Maryland 21114
Re: K170914
Trade/Device Name: TXM Support Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: October 13, 2017 Received: October 16, 2017
Dear Mr. Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good
{1}------------------------------------------------
manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Kenneth J. Cavanaugh - S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K170914
Device Name TXM Support Catheter
#### Indications for Use (Describe)
The TXM Support Catheters are percutaneous, single lumen catheters designed for use in the peripheral vascular system. TXM Support Catheters are intended to guide and support a guide wire during access of the vasculature, allow for wire exchanges and provide a conduit for the delivery of saline solutions or diagnostic contrast agents.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <svg fill="none" height="16" viewbox="0 0 16 16" width="16" xmlns="http://www.w3.org/2000/svg"> <path d="M13.2222 3.77776C13.5157 4.07123 13.5157 4.54549 13.2222 4.83896L6.28889 11.7723C5.99542 12.0658 5.52115 12.0658 5.22768 11.7723L2.76667 9.31129C2.4732 9.01782 2.4732 8.54355 2.76667 8.83702C3.06014 8.24355 3.5344 8.24355 3.82787 8.53702L5.75833 10.4675L12.1722 4.0536C12.4657 3.76013 12.9399 3.76013 13.2222 3.77776Z" fill="black"></path> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span> <svg fill="none" height="16" viewbox="0 0 16 16" width="16" xmlns="http://www.w3.org/2000/svg"> <rect height="15" rx="2.5" stroke="black" width="15" x="0.5" y="0.5"></rect> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 5.1 - Company Information & Contact Person
| Company Name: | Texas Medical Technologies Inc. |
|------------------|----------------------------------------------------|
| Company Address: | 9005 Montana Ave. Ste. A |
| | El Paso, Texas 79925 |
| Telephone: | (915) 774-4321 |
| Fax: | (915) 774-4323 |
| Contact Person: | Cesar Rios, Quality Assurance & Regulatory Manager |
| Date Prepared: | 02/28/2017 |
## 5.2 – Device Name & Classification
| Proprietary Name: | TXM Support Catheter |
|----------------------|------------------------|
| Common Name: | Catheter, Support |
| Classification Name: | Catheter, Percutaneous |
| Regulation Number: | 21 CFR 870.1250 |
| Product Code: | DQY |
| Device Class: | II |
# 5.3 – Predicate Device
Legally Marketed Substantially Equivalent Predicate Device
| Proprietary Name:<br>Company Name:<br>Common Name:<br>Classification Name:<br>Regulation Number:<br>Product Code:<br>Device Class:<br>510(k) Number<br>Reference Devices | TrailBlazer™ Support Catheter<br>ev3 Inc.<br>Catheter, Support<br>Catheter, Percutaneous<br>21 CFR 870.1250<br>DQY<br>II<br>K092299 |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| Proprietary Name: | TrailBlazer™ Angled Support Catheter |
| Company Name: | ev3 Inc. |
| Common Name: | Catheter, Support |
| Classification Name: | Catheter, Percutaneous |
| Regulation Number: | 21 CFR 870.1250 |
| Product Code: | DQY |
| Device Class: | II |
| 510(k) Number | K162384 |
{4}------------------------------------------------
| Proprietary Name: | Spectranetics Quick-Cross Extreme Support Catheters |
|----------------------|-----------------------------------------------------|
| Company Name: | Spectranetics Corp. |
| Common Name: | Catheter, Support |
| Classification Name: | Catheter, Percutaneous |
| Regulation Number: | 21 CFR 870.1330 |
| Product Code: | DQY |
| Device Class: | II |
| 510(k) Number | K082561 |
Note: Reference devices are only to support the substantial equivalence of the subject device in regards to the angled tip.
#### 5.4 - Device Description
The TXM Support Catheter is an over the wire (OTW) single lumen catheter with atraumatic tapered tip. The distal tip catheter shaft has 6 radiopaque markers (including the tip) that works as an aid to estimate positioning within the vasculature. It consists of a lubricious inner liner made from Teflon, and stainless steel braid over the liner and an outer layer that consists of clear Nylon 12 polymer. One gold marker band is positioned on the proximal side of the tip. A polycarbonate hub is attached to the proximal end of the TXM Support Catheter. The TXM Support Catheter system is offered in 8 models, four with a straight tip and four with an angled tip, and the angled tip configuration has an angle of 30°. All models will have an outer diameter of the 4 French size (0.052") and an inner diameter of 0.039" tapered down to 0.0355" towards the distal tip so that is compatible with a 0.035" guidewires. The device is available in lengths of 65cm, 90cm. 135cm, and 150cm. The catheter is coated with a lubricious hydrophilic coating. The device is supplied sterile and is intended for single use only.
The following table lists the models and sizes available for TXM Support Catheter.
| Product<br>Code | French<br>size | Shaft<br>Length<br>(cm) | Outer<br>Diameter<br>(inches) | Inner Diameter (inches) | Marker<br>Band<br>Material | Tip<br>Shape |
|-----------------|----------------|-------------------------|-------------------------------|-------------------------|----------------------------|--------------|
| SC-35065-S | 4 | 65 | 0.052" | 0.039" - 0.0355" | Gold | Straight |
| SC-35090-S | 4 | 90 | 0.052" | 0.039" - 0.0355" | Gold | Straight |
| SC-35135-S | 4 | 135 | 0.052" | 0.039" - 0.0355" | Gold | Straight |
| SC-35150-S | 4 | 150 | 0.052" | 0.039" - 0.0355" | Gold | Straight |
| SC-35065-A | 4 | 65 | 0.052" | 0.039" - 0.0355" | Gold | Angled |
| SC-35090-A | 4 | 90 | 0.052" | 0.039" - 0.0355" | Gold | Angled |
| SC-35135-A | 4 | 135 | 0.052" | 0.039" - 0.0355" | Gold | Angled |
| SC-35150-A | 4 | 150 | 0.052" | 0.039" - 0.0355" | Gold | Angled |
Table 5.4 TXM Support Catheter Models and Sizes
{5}------------------------------------------------
### 5.5 - Indications for Use
The TXM Support Catheters are percutaneous, single lumen catheters designed for use in the peripheral vascular system. TXM Support Catheters are intended to guide and support a guide wire during access of the vasculature, allow for wire exchanges and provide a conduit for the delivery of saline solutions or diagnostic contrast agents.
The Indications for Use are identical, the devices are both intended for peripheral vascular use and the Indications for Use do not change the intended use of the TXM Support Catheter when compared to the predicate device.
### 5.6 - Summary of Technological Characteristics Comparison
Based on a comparison of the indications for use, fundamental design, technology and principles of operation, materials, performance, sterilization, and packaging, it is determined that the TXM Support Catheter is substantially equivalent to the predicate device. Table 5.6 below provides a comparison of the TXM Support Catheter and the predicate.
| Technical Characteristics /<br>Principle of Operation | TXM Support Catheter | TrailBlazer ™ Support<br>Catheter | Substantially<br>Equivalent? |
|-------------------------------------------------------|----------------------------------------------------------------------------------------|-----------------------------------|------------------------------|
| Length | 65cm, 90cm, 135 cm,150cm | 65cm, 90cm, 135 cm,150cm | Yes |
| Outer Diameter | 0.052" | 0.063" | Yes |
| Inner Diameter | 0.039"- 0.0355" | 0.042"-0.036" | Yes |
| Shape | Straight and Angled | Straight | Yes |
| Inner Liner Material | Polytetrafluoroethylene<br>(PTFE)* | Unknown | Yes |
| Braid Reinforcement Material | Stainless Steel* | Not Applicable | Yes |
| Radiopaque Markers | A Gold marked band and<br>radiopaque markers of<br>Nylon 12 loaded with<br>Tungsten. * | Platinum/Iridium | Yes |
| Outer Shaft Material | Nylon 12* | Unknown | Yes |
| Luer Material | Polycarbonate* | Unknown | Yes |
| Luer Connector | Female Luer Connector* | Female Luer Connector | Yes |
| Anatomical Site Use | Peripheral Vascular system | Peripheral Vascular system | Yes |
| Delivery to Site | Over-the-wire | Over-the-wire | Yes |
| Guidewire Compatibility | Maximum 0.035" | Maximum 0.035" | Yes |
| Packaging | Tyvek Pouch | Tyvek Pouch | Y es |
| Sterilization | EtO Gas | EtO Gas | Yes |
Table 5.6 Comparison of the TXM Support Catheter and the Predicate Device
* Denotes a patient-contacting material.
{6}------------------------------------------------
## 5.7 - Performance Testing Summary
The following bench tests were performed to evaluate the design elements and performance characteristics of the TXM Support Catheter and to demonstrate substantial equivalence to the predicate device. The TXM Support Catheter met the predetermined acceptance criteria. Testing was performed on non-aged devices (T=0) as well as on devices subject to 2 years of accelerated aging (T=2). Tests results show that the TXM Support Catheter is substantially equivalent to the predicate device.
# 5.7.1- Bench Testing Summary
Table 5.7.1 below provides a summary of the bench testing performed on the TXM Support Catheter.
| Test<br>No. | Test Name | Applicable Standard or<br>Internal Test Method | Test Results | |
|-------------|-------------------------------------|------------------------------------------------|--------------|------|
| 1 | Dimensional And Physical Attributes | Internal Test Method | T=0<br>T=2 | Pass |
| 2 | Catheter Compatibility | Internal Test Method | T=0<br>T=2 | Pass |
| 3 | Air Leakage | ISO 10555-1 | T=0<br>T=2 | Pass |
| 4 | Trackability | Internal Test Method | T=0<br>T=2 | Pass |
| 5 | Crossability | Internal Test Method | T=0<br>T=2 | Pass |
| 6 | Catheter Stiffness | ASTM D747-10 | T=0<br>T=2 | Pass |
| 7 | Gravity Flow Rate | ISO 10555-1 | T=0<br>T=2 | Pass |
| 8 | Power Injection for Flow Rate | ISO 10555-1 | T=0<br>T=2 | Pass |
| 9 | Lubricity of Hydrophilic Coating | Internal Test Method | T=0<br>T=2 | Pass |
| 10 | Durability of Hydrophilic Coating | Internal Test Method | T=0<br>T=2 | Pass |
| 11 | Liquid Leakage | ISO 10555-1 | T=0<br>T=2 | Pass |
| 12 | Static Burst Pressure | ISO 10555-1 | T=0<br>T=2 | Pass |
| 13 | Tensile Strength | ISO 10555-1 | T=0<br>T=2 | Pass |
| 14 | Radiopacity | ASTM-F640-12 | T=0<br>T=2 | Pass |
| 15 | Torqueability | Internal Test Method | T=0<br>T=2 | Pass |
| 16 | Torque Strength | Internal Test Method | T=0<br>T=2 | Pass |
Table 5.7.1 Bench Testing TXM Support Catheter
{7}------------------------------------------------
| 17 | Kink Resistance | Internal Test Method | T=0<br>T=2 | Pass |
|----|------------------------------------------------|----------------------------|------------|------|
| 18 | Corrosion Resistance | ISO 10555-1 | T=0<br>T=2 | Pass |
| 19 | Coating Integrity and Particulates | Internal Test Method | T=0<br>T=2 | Pass |
| 20 | Packing Integrity (Seal Strength) | ASTM F88 / F88M - 15 | T=0<br>T=2 | Pass |
| 21 | Packing Integrity (Packing Dye<br>Penetration) | ASTM F1929-15<br>/F1886-09 | T=0<br>T=2 | Pass |
| 22 | Shipping and Transit | ISTA 3A | T=0<br>T=2 | Pass |
| 23 | Accelerated Aging | ASTM F1980-07 | Pass | |
| 24 | Female Luer Verification | ISO 594 | Pass | |
# 5.7.2 - Biocompatibility Testing Summary
The TXM Support Catheter is classified as an Externally Communicating Device, Circulating Blood, Limited Contact (≤24 hours). Biocompatibility testing was performed in accordance with ISO 10993 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process" (2009). Table 5.7.2 below describes the testing performed to determine biocompatibility.
| Biological Effect | Test | Compliance Standard |
|-------------------|--------------------------------------------------|---------------------|
| Cytotoxicity | L929 MEM ELUTION | ISO10993-5 |
| Irritation | Intracutaneous Injection | ISO10993-10 |
| Sensitization | Kligman Maximization<br>Murine Local Lymph Assay | ISO10993-10 |
| Systemic Toxicity | Systemic Injection | ISO10993-11 |
| Material Mediated | Rabbit Pyrogen | USP<151> |
| Pyrogenicity | Limulus Amebocyte Lysate | ISO10993-11 |
| Hemocompatibility | Hemolysis Complete (Direct and Indirect) | ISO10993-4 |
| | Complement Activation | ISO10993-4 |
| | Thromboresistance Evaluation <i>In vivo</i> | ISO10993-4 |
Table 5.7.2. Summary of Biocompatibility Testing for the TXM Support Catheter
# 5.8 - Sterilization Testing Summary
Table 5.8 Summary of Sterilization Testing
| Validation Sterilization<br>Process | Compliance<br>Standard | Sterility Assurance Level<br>(SAL) | Validation<br>Result |
|-------------------------------------|------------------------|------------------------------------|----------------------|
| Ethylene Oxide Gas | ISO 11135 | $10^{-6}$ | Pass |
{8}------------------------------------------------
# 5.9 – Conclusion
The TXM Support Catheter is substantially equivalent in intended use, fundamental design, technology and principles of operation, materials, performance, sterilization, and packaging to the predicate device. Differences between the devices do not raise any new issues of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.