← Product Code [NEU](/productcode/NEU) · K170905

# Star Tissue Marker (K170905)

_Scion Medical Technologies, LLC · NEU · Apr 27, 2017 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K170905

## Device Facts

- **Applicant:** Scion Medical Technologies, LLC
- **Product Code:** [NEU](/productcode/NEU.md)
- **Decision Date:** Apr 27, 2017
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 878.4300
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Indications for Use

Star Tissue Marker is indicated for use to radiographically mark soft tissue at the surgical procedure or for future surgical procedures

## Device Story

Star Tissue Marker is an implantable, sterile, single-use device used to mark soft tissue for radiographic identification. It consists of a PEKK (polyetherketoneketone) marker pre-loaded into a delivery system. The delivery system features a 12 cm, 14-gauge needle with 1 cm depth markings, a cannula with a handle, and a push-rod plunger mechanism. The device is operated by a clinician to deliver the marker through the needle's end port into the target tissue. The marker is visible on standard radiographs, mammography, ultrasound, and MRI (up to 3.0 Tesla). The device aids in localizing tissue sites for subsequent surgical procedures.

## Clinical Evidence

Bench testing only.

## Technological Characteristics

Material: PEKK (polyetherketoneketone). Delivery system: 12 cm, 14-gauge needle with 1 cm depth markings. Imaging compatibility: X-ray, mammography, ultrasound, and MRI (up to 3.0 Tesla). Sterile, single-patient use. Standards: ISO 15223-1:2012, ISO 14971:2007, EN 1041:2008.

## Regulatory Identification

An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.

## Predicate Devices

- Beacon Tissue Marker ([K140835](/device/K140835.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 27, 2017

Scion Medical Technologies, LLC Mr. Louis Li Director of OA/RA 4613 West Chester Pike Newtown Square, Pennsylvania 19073

Re: K170905

Trade/Device Name: Star Tissue Marker Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable clip Regulatory Class: Class II Product Code: NEU Dated: February 28, 2017 Received: March 28, 2017

Dear Mr. Li:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

## David Krause -S

for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K170905

Device Name

Star Tissue Marker

Indications for Use (Describe)

Star Tissue Marker is indicated for use to radiographically mark soft tissue at the surgical procedure or for future surgical procedures

| Type of Use (Select one <i>or both</i> , as applicable) |                                                                                                                                                                                                                          |
|---------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                                         | <span> <span style="text-decoration: underline;"><b>X</b></span> Prescription Use (Part 21 CFR 801 Subpart D)             </span> <span>                 Over-The-Counter Use (21 CFR 801 Subpart C)             </span> |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

-

{3}------------------------------------------------

## Special 510(k) Summarv as required by 21 CFR 807.92(a)

| A) Submitted by:                                  | Scion Medical Technologies, LLC                                     |  |
|---------------------------------------------------|---------------------------------------------------------------------|--|
|                                                   | 4613 West Chester Pike<br>Newtown Square, Pennsylvania 19073<br>USA |  |
|                                                   | TEL: (484)860-2685                                                  |  |
| Official Contact:                                 | Louis Li Director of QA/RA                                          |  |
| B) Common name:                                   | Implantable Clip                                                    |  |
| Proprietary Name:                                 | Star Tissue Marker                                                  |  |
| Device Class:                                     | Class II, 21 CFR 878.4300                                           |  |
| Regulation and<br>Classification name:            | marker, radiographic, implantable                                   |  |
| Product code:                                     | NEU                                                                 |  |
| Classification panel: General and Plastic Surgery |                                                                     |  |
| C) Predicate:                                     | Beacon Tissue Marker K140835 May 20th, 2014                         |  |

D) Date Prepared: April 26, 2017

E) Device Description

The Star Tissue Marker consists of a radiographic soft tissue marker and the delivery system. The Star Tissue Marker includes a sterile, single patient use, PEKK discrete marker that is visible on standard radiographs (x-ray, mammography) as well as ultrasound, and Magnetic Resonance Imaging (MRI) at up to 3.0 Tesla field strength.

This submission is for a new cross sectional profile of the tissue marker. The delivery system is sterile, single patient use, and is pre-loaded incorporating the tissue marker. The delivery system has a 12 cm /14 gauge needle with 1 cm depth marks. The delivery system consists of a cannula with a handle with integral tabs to retain the tissue marker, a push rod with a plunger, and a tip cover. The tissue marker is retained within the delivery system until placement is desired, where it is delivered through the end port by fully depressing the plunger into the handle.

F) Intended Use/Indications For Use:

The Star Tissue Marker is indicated for use to radiographically mark soft tissue during a surgical procedure or for future surgical procedures

{4}------------------------------------------------

- G) Substantial Equivalence Comparison and Discussion
In summary, Scion Medical Technologies, LLC believes that the proposed changes in Star Tissue Marker, as described in this submission, do not raise any new or significant questions of safety and efficacy and is substantially equivalent to the predicate Scion Medical Technologies, LLC Beacon Tissue Marker cleared on May 20th, 2014 (K140835).

H) Compliance with Design Controls

The results of assessment of the change to the new tissue marker conducted under Design Controls, support that the new offering is substantially equivalent to the predicate tissue marker.

Compliance with Standards:

ISO 15223-1:2012, Medical devices - Symbols to be used with medical device labels, labeling and information to be supplied - Part 1: General requirements

ISO 14971: 2007, Medical devices - Application of risk management to medical devices

EN 1041:2008 Information supplied by the manufacture of medical devices

---

**Source:** [https://fda-staging.innolitics.com/device/K170905](https://fda-staging.innolitics.com/device/K170905)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com