The Dolphin/IQ and Dolphin/4D are medical Doppler devices intended for noninvasive measurements of blood flow velocities in arteries and veins in adults and Pediatric. The Dolphin systems can be used in hospitals, clinics and physician offices.
Device Story
Dolphin/IQ and Dolphin/4D are transcranial Doppler (TCD) systems for measuring blood flow velocities in intracranial, extracranial, and peripheral vessels. Dolphin/IQ is a module requiring an external computer/display; Dolphin/4D is an integrated system with a computer and touchscreen. Both use identical Doppler hardware and software. Systems support 1.6 MHz, 2 MHz, 4 MHz, and 8 MHz PW/CW probes. Operated by trained medical personnel in hospitals, clinics, and physician offices. Input consists of ultrasonic Doppler signals; system processes these to calculate parameters like peak/mean velocity, pulsatility index, and resistance index. Output is displayed as spectral waveforms, velocity profiles, and numerical data. Clinicians use this information for vascular assessment and monitoring. Benefits include noninvasive diagnostic capability for vascular conditions.
Clinical Evidence
No clinical study was conducted. Substantial equivalence is supported by bench testing, including verification of acoustic output, temperature rise, and velocity accuracy.
Technological Characteristics
Ultrasonic pulsed Doppler imaging system. Transducers: 1.6 MHz, 2 MHz, 4 MHz, 8 MHz. Modes: PWD, CWD. Connectivity: PACS support, analog I/O. Standards: IEC 60601-1 (3.1 Ed), IEC 60601-1-2 (4 Ed), IEC 60601-2-37 (2.1 Ed), NEMA UD 2, UD 3. Software-based signal processing for velocity calculation and HITS detection.
Indications for Use
Indicated for noninvasive measurement of blood flow velocities in arteries and veins in adult and pediatric populations. Contraindicated for fetal or neonatal applications.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Viasonix Ltd. Mr. Dan Manor, CEO 10 Hamelacha Street Raanana, 4366105 ISRAEL
November 14, 2017
Re: K170859
Trade/Device Name: Dolphin/IQ and Dolphin/4D Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, ITX Dated: October 19, 2017 Received: October 23, 2017
Dear Mr. Manor:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice
(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Michael D. O'Hara For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K170859
Device Name Dolphin/IQ and Dolphin/4D
Indications for Use (Describe)
The Dolphin/IQ and Dolphin/4D are medical Doppler devices intended for noninvasive measurements of blood flow velocities in arteries and veins in adults and Pediatric. The Dolphin systems can be used in hospitals, clinics and physician offices
Contraindications : The Dolphin is not intended to be used in fetal or neonatal applications.
Note : The Dolphin is to be used only by trained medical personnel.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
| System: | Dolphin/IQ, Dolphin/4D |
|-------------|------------------------|
| Transducer: | 1.6 MHz Hand Held |
| Clinical Application | | Mode of Operation | | | | | | |
|-----------------------------|------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|---------------------|
| General | Specific | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| (Track 1 Only) | (Tracks 1 & 3) | | | | | | | |
| Ophthalmic | Ophthalmic | | | x | | | | |
| Fetal<br>Imaging<br>& Other | Fetal | | | | | | | |
| | Abdominal | | | | | | | |
| | Intra-operative (Specify) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ (Specify) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | x | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skeletal<br>(Conventional) | | | | | | | |
| | Musculo-skeletal<br>(Superficial) | | | | | | | |
| | Intravascular | | | | | | | |
| | Other (Specify) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | x | | | | |
| | Other (Specify) | | | | | | | |
{4}------------------------------------------------
| System: | Dolphin/IQ, Dolphin/4D |
|-------------|------------------------|
| Transducer: | 2 MHz Monitoring |
| Clinical Application | | Mode of Operation | | | | | | |
|----------------------|---------------------------|-------------------|---|-----|-----|---------|-----------|-----------|
| General | Specific | B | M | PWD | CWD | Color | Combined | Other* |
| (Track 1 Only) | (Tracks 1 & 3) | | | | | Doppler | (Specify) | (Specify) |
| Ophthalmic | Ophthalmic | | | x | | | | |
| | Fetal | | | | | | | |
| | Abdominal | | | | | | | |
| | Intra-operative (Specify) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| Fetal | Pediatric | | | | | | | |
| Imaging | | | | | | | | |
| & Other | Small Organ (Specify) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | x | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skeletal | | | | | | | |
| | (Conventional) | | | | | | | |
| | Musculo-skeletal | | | | | | | |
| | (Superficial) | | | | | | | |
| | Intravascular | | | | | | | |
| | Other (Specify) | | | | | | | |
| | Cardiac Adult | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral | Peripheral vessel | | | x | | | | |
| Vessel | Other (Specify) | | | | | | | |
{5}------------------------------------------------
| System: | Dolphin/IQ, Dolphin/4D |
|-------------|------------------------|
| Transducer: | 2 MHz Hand Held |
| Clinical Application | | Mode of Operation | | | | | | |
|-----------------------------|------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|---------------------|
| General | Specific | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| (Track 1 Only) | (Tracks 1 & 3) | | | | | | | |
| Ophthalmic | Ophthalmic | | | x | | | | |
| | Fetal | | | | | | | |
| | Abdominal | | | | | | | |
| | Intra-operative (Specify) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| Fetal<br>Imaging<br>& Other | Pediatric | | | | | | | |
| | Small Organ (Specify) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | x | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skeletal<br>(Conventional) | | | | | | | |
| | Musculo-skeletal<br>(Superficial) | | | | | | | |
| | Intravascular | | | | | | | |
| | Other (Specify) | | | | | | | |
| | Cardiac Adult | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | x | | | | |
| | Other (Specify) | | | | | | | |
{6}------------------------------------------------
| System: | Dolphin/IQ, Dolphin/4D |
|-------------|------------------------|
| Transducer: | 4 MHz Hand Held |
| Clinical Application | | Mode of Operation | | | | | | |
|-----------------------------|------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|---------------------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| Ophthalmic | Ophthalmic | | | X | X | | | |
| | Fetal | | | | | | | |
| | Abdominal | | | | | | | |
| | Intra-operative (Specify) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| Fetal<br>Imaging<br>& Other | Pediatric | | | | | | | |
| | Small Organ (Specify) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skeletal<br>(Conventional) | | | | | | | |
| | Musculo-skeletal<br>(Superficial) | | | | | | | |
| | Intravascular | | | | | | | |
| | Other (Specify) | | | | | | | |
| | Cardiac Adult | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | X | X | | | |
| | Other (Specify) | | | | | | | |
{7}------------------------------------------------
| System: | Dolphin/IQ, Dolphin/4D |
|-------------|------------------------|
| Transducer: | 8 MHz Hand Held |
| Clinical Application | | Mode of Operation | | | | | | |
|-----------------------------|------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|---------------------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| Ophthalmic | Ophthalmic | | | x | x | | | |
| Fetal<br>Imaging<br>& Other | Fetal | | | | | | | |
| | Abdominal | | | | | | | |
| | Intra-operative (Specify) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ (Specify) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skeletal<br>(Conventional) | | | | | | | |
| | Musculo-skeletal<br>(Superficial) | | | | | | | |
| | Intravascular | | | | | | | |
| | Other (Specify) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | x | x | | | |
| | Other (Specify) | | | | | | | |
{8}------------------------------------------------
## SECTION 5 - 510(K) SUMMARY
#### 5.1 ADMINISTRATIVE INFORMATION
| Date: | 14-March-2017 |
|------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Viasonix Ltd.<br>10 Hamelacha Street<br>Raanana , ISRAEL 4366105<br>Phone : 972-9-7441692 |
| Official Correspondent: | Dan Manor, CEO |
| Trade Name: | Dolphin/IQ and Dolphin/4D |
| Classification Name:<br>Classification Number: | Ultrasonic Pulsed Doppler Imaging System<br>21 CFR 892.1550 |
| Product Code:<br>Device Class: | IYN, ITX<br>Class II |
| Predicate Devices: | Primary:<br>Sonara and Sonara/Tek<br>VIASYS Healthcare, Inc,<br>510(k) Number - K060421<br><br>Secondary:<br>Doppler-Box<br>COMPUMEDICS GERMANY GMGH - DWL<br>510(k) Number - K051085 |
#### 5.2 DEVICE DESCRIPTION
Dolphin/IQ and Dolphin/4D systems are part of the Dolphin product family of noninvasive peripheral vascular diagnostic systems. The Dolphin/IQ and the Dolphin/4D are transcranial Doppler (TCD) systems for measurement of blood flow velocity intracranially, extracranially and in the peripheral circulation. Both systems share identical Doppler hardware and software. The Dolphin/IQ is a module, that needs to connect to an external computer and display for its' operation while the Dolphin/4D is a complete integrated system that includes and integrated computer system with hard disk, and touch screen display. The functionality and performance of both systems is identical. Both systems
{9}------------------------------------------------
support the same Doppler probes: 1.6 MHz PW hand held probe, 2 MHz PW hand held probe, 2 MHz PW monitoring probe, 4 MHz CW/PW hand held probe and 8 MHz CW/PW hand held probe. Both support the same accessories: IR wireless remote control, foot switch and monitoring head set.
Wherever the term Dolphin is used in this document, it applies to the Dolphin/IQ and Dolphin/4D products. Otherwise, each product is specifically by name. The Dolphin systems are based on Doppler technology and are designed for standard intended use for Transcranial Doppler systems operated only by experienced medical staff.
### 5.3 INTENDED USE AND INDICATIONS FOR USE
The Dolphin/IQ and Dolphin/4D are medical Doppler devices intended for noninvasive measurements of blood flow velocities in arteries and veins in adults and Pediatric. The Dolphin systems can be used in hospitals, clinics and physician offices.
Contraindications: The Dolphin is not intended to be used in fetal or neonatal applications.
Note - The Dolphin is to be used only by trained medical personnel
### 5.4 SUMMARY OF TECHNICAL CHARACTERISTICS
The Dolphin/IQ and Dolphin/4D are similar to the predicate devices cited above with 1.6MHz, 2MHz, 4MHz and 8MHz transducers intended for transcranial and peripheral Doppler applications.
The technological characteristics, e.g., overall design, materials, mechanism of action, mode of operation, performance characteristics, the intended use, use environment and target patient population of the Dolphin/1Q and Dolphin/4D devices are substantially equivalent to the predicate devices cited above.
{10}------------------------------------------------
## 5.4.1 Summary table of Comparison
| Specification | Dolphin/4D and<br>Dolphin/IQ | Sonara and Sonara/tek | Doppler-Box | Differences discussion |
|--------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) number | Proposed Device | K060421 | K051085 | NA |
| Manufacturer | VIASONIX LTD. | VIASYS HEALTHCARE, INC | COMPUMEDICS GERMANY<br>GMGH - DWL | NA |
| Product regulation<br>and code | 21 CFR 892.1550 | 21 CFR 892.1550 | 21 CFR 892.1550 | Identical to predicates |
| | | | | |
| Indications for use | Code: IYN, ITX<br>The Dolphin/IQ and | Code: IYN, ITX<br>Sonara and Sonara/Tek | Code: IYN, ITX<br>The Doppler-Box is a medical | Equivalent. |
| | Dolphin/4D are medical<br>Doppler devices intended<br>for noninvasive<br>measurements of blood<br>flow velocities in arteries<br>and veins in adults and<br>Pediatric. The Dolphin<br>systems can be used in<br>hospitals, clinics and<br>physician offices. | systems are medical<br>ultrasound Doppler<br>devices for measuring the<br>blood flow velocities in<br>arteries and veins non-<br>invasively. | ultrasound device for<br>measuring the blood flow<br>velocities in arteries and<br>veins mainly subcutaneously.<br>The 16MHz probe can also be<br>used intraoperative. | Intended use, use environment and<br>target patient population is<br>substantially equivalent to the<br>predicate devices. The Dolphin<br>devices don't include the 16MHz<br>option for intraoperative use. |
| Clinical Applications | Intracranial<br>Extracranial<br>Peripheral | Intracranial<br>Extracranial<br>Peripheral | Intracranial<br>Extracranial<br>Peripheral<br>Intraoperative | Identical to Sonara and Sonara/tek |
| Weight (kg) | Dolphin/4D: ~6 Kg<br>Dolphin/IQ: ~2 Kg | Sonara: 10 Kg<br>Sonara/tek: 2 Kg | 1.5 Kg | -Dolphin/4D and the predicate,<br>Sonara have an integrated computer<br>and display.<br>-Dolphin/IQ and the predicate,<br>Sonara/tek to be connected<br>externally to PC and monitor. |
| Dimensions (cm) | Dolphin/4D: 47x30x7<br>Dolphin/IQ: 26.5x20.5x5.5 | Sonara: 39x30x26<br>Sonara/tek: 26.5x22x4 | 27x10.5x9 | -Dolphin/4D and the predicate,<br>Sonara have an integrated computer<br>and display.<br>-Dolphin/IQ and the predicate,<br>Sonara/tek to be connected<br>externally to PC and monitor. |
| Frequency modes / | 1.6MHz PW | 2MHz PW | 1MHz PW | The Dolphin 1.6MHz PW doesn't |
| Transducers<br>(MHz) | 2MHz PW<br>4MHz PW/CW<br>8MHz PW/CW | 4MHz PW/CW<br>8MHz PW/CW | 2MHz PW<br>4MHz PW/CW<br>8MHz PW/CW<br>16MHz PW | adversely impact the substantial<br>equivalent to the predicate, the<br>Doppler-Box. |
| Patient surface<br>contact materials | Compatible | compatible | compatible | Patient surface contact probes<br>materials are identical to all<br>predicate devices from the same<br>material/assembly manufacturer |
| 2 MHz Monitoring<br>Probe | available | available | available | Identical to the predicate Sonara<br>device |
| Monitoring headset | available | available | available | Similar to the predicate devices. All<br>have same look, function and probe<br>handling. |
| User controls | Remote control, foot<br>switch, touch screen, key<br>board, mouse | Remote control, foot<br>switch, touch screen, key<br>board, mouse | Remote control, foot switch,<br>touch screen, key board,<br>mouse | Identical to the predicate devices |
| Display modes | Unilateral, bilateral,<br>monitoring, external<br>channels, HITS | Unilateral, bilateral,<br>monitoring, external<br>HITS | Unilateral, bilateral,<br>monitoring, external<br>channels, HITS | Identical to Doppler-Box |
| Specification | Dolphin/4D and<br>Dolphin/IQ | Sonara and Sonara/tek | Doppler-Box | Differences discussion |
| Sample Volume (2 MHz) | 1-20 mm | 1-20 mm | Available, range unknown | Identical to Sonara and Sonara/tek |
| Scale (2 MHz) | Up to 32 KHz depth dependent | Up to 16 KHz depth dependent | Up to 32 KHz depth dependent | Identical to Doppler-Box |
| Power control | 0-100 % of maximal derated Ispta within FDA guidelines | 0-100 % of maximal derated Ispta within FDA guidelines | 0-100 % of maximal derated Ispta within FDA guidelines | Identical to the predicate devices |
| Maximal Acoustic Ispta.3 (mW/cm²) | Below maximal FDA guideline limits | Below maximal FDA guideline limits | Below maximal FDA guideline limits | Equivalent to predicate devices |
| | Comply with FDA limits:<br>Ispta.3 ≤ 720 mW/cm²<br>MI ≤ 1.9 or the global maximum derated ISPPA ≤ 190 W/cm². | Comply with FDA limits:<br>Ispta.3 ≤ 720 mW/cm²<br>MI ≤ 1.9 or the global maximum derated ISPPA ≤ 190 W/cm² | Comply with FDA limits:<br>Ispta.3 ≤ 720 mW/cm²<br>MI ≤ 1.9 or the global maximum derated ISPPA ≤ 190 W/cm² | |
| M-mode display | available | available | available | Identical to the predicate devices |
| Multi-gate windows | Up to 8 | Up to 8 | Available, # not known | Identical to Sonara and Sonara/tek |
| HITS detection | Available | Available | Available | Similar to predicate devices |
| Velocity profile display | Available | Available | Not available | Similar to Sonara and Sonara/tek |
| Cursors | Available | Available | Available | Identical to the predicate devices |
| Audio replay | Available | Available | Available | Identical to the predicate devices |
| Sweep time display | Up to 3 minutes | Up to 2 minutes | Available, sweep time not known | The sweep time display is only a display option with no impact on intended use |
| Parameters | Peak velocity, mean velocity, end diastolic velocity, pulsatility Index, resistance index, systolic to diastolic ratio, rise time, heart rate | Peak velocity, mean velocity, end diastolic velocity, pulsatility Index, resistance index, systolic to diastolic ratio, heart rate | Peak velocity, mean velocity, end diastolic velocity, pulsatility Index, resistance index, systolic to diastolic ratio, rise time, heart rate | Identical to the predicate devices |
| Measurement accuracy | ± 10% accuracy | ± 10% accuracy | ± 10% accuracy | Identical to the predicate devices |
| accessories | Remote control, foot switch, monitoring head set | Remote control, foot switch, monitoring head set | Remote control, foot switch, monitoring head set | Equivalent |
| Velocity units | Cm/sec or KHz | Cm/sec or KHz | Cm/sec or KHz | Identical to the predicate devices |
| Summary screens | available | available | available | Identical to the predicate devices |
| Patient database | available | available | available | Identical to the predicate devices |
| Patient search options | available | available | available | Identical to the predicate devices |
| Spectrum color palette selection | available | available | Not available | Identical to Sonara and Sonara/tek |
| Insonation angle | User defined | User defined | User defined | Identical to the predicate devices |
| Database backup options | available | available | available | Identical to the predicate devices |
| Database statistics | available | available | Not available | Identical to Sonara and Sonara/tek |
| Export | Multiple formats | Multiple formats | Multiple formats | Identical to the predicate devices |
| Analog input channels | 8 | Not available | 8 | Identical to Doppler-Box |
| Analog output channels | 4 | Not available | 4 | Identical to Doppler-Box |
| Configurable protocols | available | available | available | Identical to the predicate devices |
| Specialty tests | available | available | available | Identical to the predicate devices |
| Connectivity to PACS systems | available | available | available | Identical to the predicate devices |
| Printer support | available | available | available | Identical to the predicate devices |
| Standards Compliance | IEC 60601-1 , 3.1 Ed.<br>IEC 60601-1-2. 4 Ed.<br>IEC 60601-2-37. 2.1 Ed. | IEC 60601-1 , 3.0 Ed.<br>IEC 60601-1-2. 3 Ed.<br>IEC 60601-2-37. 2.0 Ed. | IEC 60601-1 , 3.0 Ed.<br>IEC 60601-1-2. 2.0 Ed.<br>IEC 60601-2-37. 2.0 Ed. | Dolphin/4D and Dolphin/IQ complies with the most recent standards editions |
{11}------------------------------------------------
{12}------------------------------------------------
### 5.5 UTILIZATION OF STANDARDS AND GUIDANCE'S:
The Dolphin/IQ and Dolphin/4D meets the following standards and guidance's:
- 1. IEC 60601-1:2005+A1:2012 Medical electrical equipment Part 1: General requirements for basic safety and essential performance
- 2. IEC 60601-1-2:2014 Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- 3. IEC 60601-2-37: 2007(AMD1:2015) [Edition 2.1] Medical Electrical Equipment -Part 2-37: Particular Requirements for The Basic Safety and Essential Performance of Ultrasonic Medical Diagnostic and Monitoring Equipment
- 4. NEMA UD 2-2004 (R2009), Acoustic Output Measurement Standard For Diagnostic Ultrasound Equipment Revision 3
- 5. UD 3-2004 (R2009) Standard for Real-Time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment. Revision 2.
- 6. Guidance for Industry and FDA Staff Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers: September 9, 2008
### 5.6 Summary of Non-Clinical Performance Testing
Summary of Non-Clinical Tests:
The Dolphin/IQ and Dolphin/4D devices have been thoroughly tested through verification of specifications and validation, including software validation. The following performance verification testing were applied to the development of the system: Acoustic output Measurement, temperature rise and velocity accuracy testing.
{13}------------------------------------------------
### 5.7 Summary of Clinical Performance Data
No clinical study was conducted to support this application.
### 5.8 CONCLUSIONS
Based on its underlying technology and bench tests performed, the Dolphin/IQ and Dolphin/4D are substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.