K170858 · Fujifilm Medical Systems U.S.A, Inc. · IZL · Apr 24, 2017 · Radiology
Device Facts
Record ID
K170858
Device Name
FDR AQRO (DR-XD 1000)
Applicant
Fujifilm Medical Systems U.S.A, Inc.
Product Code
IZL · Radiology
Decision Date
Apr 24, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1720
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The FDR AQRO (DR-XD1000) is a digital mobile X-ray system intended for use in general purpose radiography for generating radiographic images of human anatomy, including adult, pediatric, and neonatal exams. The FDR AQRO is not intended for mammography.
Device Story
FDR AQRO is a compact, lightweight, non-motorized mobile X-ray system for general radiography. It utilizes an integrated X-ray tube and high-voltage generator (mono-block) to reduce size and eliminate high-voltage cables. The system operates on a Li-ion battery (up to 12 hours) or AC power. It uses Fujifilm's FDR D-EVO2 digital detectors (GOS or CsI scintillators) to capture X-ray images. The system includes a built-in console running FDX Console Software (v10.0) with Virtual Grid 2 (VG2) image processing, which allows grid-less imaging to reduce patient dose. Used in clinical settings (hospital, bedside) by healthcare professionals, the system provides digital radiographic images to aid in clinical diagnosis. The compact design facilitates maneuverability in tight spaces. The device benefits patients by providing high-quality diagnostic imaging with reduced dose requirements and improved portability for bedside exams.
Clinical Evidence
Bench testing only. The submission includes verification and validation activities per risk analysis, compliance with voluntary standards (AAMI/ANSI ES60601-1, IEC 60601-1, IEC 60601-1-2, IEC 62304, IEC 62366, IEC 62494-1, DICOM 3.0, IEC 60601-2-54, IEC 60601-1-3, IEC 60601-1-6), and sample phantom and clinical images.
Indicated for general purpose radiography of human anatomy, including adult, pediatric, and neonatal patients. Not indicated for mammography.
Regulatory Classification
Identification
A mobile x-ray system is a transportable device system intended to be used to generate and control x-ray for diagnostic procedures. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
FDR D-EVO Flat Panel Detector System (DR-ID600) (K132509)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 24, 2017
FUJIFILM Medical Systems U.S.A., Inc. Peter Altman Regulatory and Ouality Consultant 419 West Avenue Stamford, Connecticut 06902
Re: K170858
Trade/Device Name: FDR AORO (DR-XD 1000) Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile X-Ray System Regulatory Class: Class II Product Code: IZL Dated: March 20, 2017 Received: March 22, 2017
Dear Peter Altman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Michael D.'Hara
For
Robert A. Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K170858
Device Name FDR AQRO (DR-XD 1000)
#### Indications for Use (Describe)
The FDR AQRO (DR-XD 1000) is a digital mobile X-ray system intended for use in general purpose radiography for generating radiographic images of human anatomy, including adult, pediatric, and neonatal exams. The FDR AORO is not intended for mammography.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
## FDR AQRO(DR-XD 1000)
{3}------------------------------------------------
# 510(k) Summary
# FDR AQRO (DR-XD1000) Mobile X-ray System
## Date: March 20, 2017
### Submitter's Information:
FUJIFILM Medical Systems U.S.A., Inc. 419 West Avenue Stamford, CT, 06902, USA
### Contact Persons:
| Name: | Peter Altman | Dharaben Desai (alternate) |
|------------|--------------------------------------|-----------------------------------|
| Title: | Regulatory and Quality<br>Consultant | Sr. Regulatory Affairs Specialist |
| Telephone: | (203) 602-3576 | (203) 276-3445 |
| Facsimile: | (203) 602-3785 | (203) 602-3785 |
### ldentification of the Device:
| Proprietary/Trade Name: | FDR AQRO (DR-XD1000) |
|-------------------------|----------------------|
| Classification Name: | Mobile X-ray system |
| Regulations Number: | 21 CFR 892.1720 |
| Product Codes: | IZL |
| Device Class: | Class II |
| Review Panel: | Radiology |
| Common Name: | Mobile X-ray System |
### Identification of the Legally Marketed Predicate Device:
| Sirius Starmobile Tiara, K143537 cleared 3/6/2015 | |
|---------------------------------------------------|---------------------|
| Classification Name: | Mobile X-ray system |
| Regulations Number: | 21 CFR 892.1720 |
| Product Codes: | IZL |
| Device Class: | Class II |
| Review Panel: | Radiology |
| Common Name: | Mobile X-ray System |
#### l. DEVICE DESCRIPTION
FUJIFILM's FDR AQRO (DR-XD1000) is a compact, economical, lightweight, nonmotorized, low power (2.5 kW), mobile X-ray system designed to work with FUJIFILM's GOS and Csl scintillator FDR D-EVO2 (DR-ID 12XXSE) family of digital X-ray detectors coupled. The D-EVO2 detectors received clearance on 7/23/2014 via 510(k) K142003.
The FDR AQRO includes a built-in operation console. The AQRO's console uses Version 10.0 of Fujifilm's FDX Console Software. This software received 510(k)
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Fujifilm logo. The logo is black, except for a small red square with a white triangle cut out of it. Below the logo, the words "Value from Innovation" are written in a smaller, black font.
clearance via K170451 on 3/16/2017. The console software includes Virtual Grid 2 (VG2) Image Processing functionality. The VG2 function allows using the mobile X-ray system without a physical grid, resulting in a dose reduction of up to 50% (when compared to using a physical grid). The Virtual Grid 2 Image Processing software received clearance on 4/8/2016 via K153464.
The reduction in the external dimensions of FDR AQRO enables smooth movement in the hospital and at the bedside because of an integrated X-ray tube and high-voltage generator (mono-block) that eliminates the need for High Voltage cables and utilizes less space.
A high performance Li-ion battery provides up to twelve (12) hours of continuous use (at ~20 exposures/hour) with a quick full charge in four hours. A quick charge of 15 minutes provides one hour of usage. Exposure may also be made when the AC power cord is plugged in.
#### ll. INDICATIONS FOR USE
The FDR AQRO (DR-XD1000) is a digital mobile X-ray system intended for use in general purpose radiography for generating radiographic images of human anatomy. including adult, pediatric, and neonatal exams. The FDR AQRO is not intended for mammography.
| Subject Device<br>FDR AQRO<br>(DR-XD1000) | Predicate Device K143537, cleared<br>Mar 6, 2015<br>Sirius Starmobile Tiara | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | | |
| The FDR AQRO (DR-XD1000) is a digital mobile X-ray system intended for use in general purpose radiography for generating radiographic images of human anatomy, including adult, pediatric, and neonatal exams. The FDR AQRO is not intended for mammography. | The Mobile X-ray Unit Sirius Starmobile Tiara is a general radiography system and is composed of the X-ray high voltage generator, X-ray tube, support unit, and digital radiograph device (DR-ID 800) made by Fujifilm Corporation. This device is designed for pediatric and adult patients. It is intended for use in general radiography of the head, body, or extremities including pediatric exams. The device output can provide an aid to diagnosis when used by a qualified physician. | |
| General Characteristics | | |
| Weight | 90kg | 420kg |
| Dimensions | 600x851x1460 mm | 575x1236x1780 mm |
| Energy Source | Batteries charged by AC line | Batteries charged by AC line |
| | Subject Device<br>FDR AQRO<br>(DR-XD1000) | Predicate Device K143537, cleared<br>Mar 6, 2015<br>Sirius Starmobile Tiara |
| User Interface | Up-Down pushbuttons for kV<br>and mAs.<br>Software Driven Touch Panel<br>LCD. | Up-Down pushbuttons for kV and<br>mAs.<br>Software Driven Touch Panel LCD. |
| X-ray Characteristics | | |
| Radiography<br>ratings | 2.5kW | 32kW |
| Tube voltage | 40 to 100 in 1kV steps | 40 to 130 in 1kV steps |
| Maximum tube<br>current | 35mA<br>(automatically<br>set<br>according to tube voltage) | 400mA (automatically set according<br>to tube voltage) |
| mAs | 0.25-25mAs | 0.5-320mAs |
| X-ray tube focal<br>spot size | 1.7 x 2.4 mm | 0.9x1.3 mm:Small focus<br>1.7x2.4 mm:Large focus |
| X-ray tube<br>maximum anode<br>heat capacity | 35kJ (50kHU) | 100kJ(140kHU) |
| X-ray tube target<br>angle | 16 degree | 17 degrees |
| X-ray tube<br>inherent filtration | 1.3 mmAl equivalent | 1.0 mmAl equivalent |
| X-ray tube Added<br>filter | N/A | 0.3 mmAl equivalent |
| X-ray beam<br>limiting device<br>illuminance of<br>light field | 200 Ix or over (Note: at SID<br>100cm) | 160 Ix or over (Note: at SID 100cm) |
| X-ray beam<br>limiting device<br>timer for light<br>field | 30 seconds(This setting can<br>be changed to 60 sec., 90<br>sec, 120 sec.or 180 sec. in<br>the service setting) | 30 seconds |
| X-ray beam<br>limiting device<br>inherent filtration | 1.2 mm Al equivalent | 1.2 mm Al equivalent |
| Total filtration | 2.5 mm Al equivalent | 2.5 mm Al equivalent |
| Maximum SID to<br>floor | 2000mm | 2010mm |
| Tube arm reach | 1200mm | 1330mm |
| Detector Characteristics | | |
| Sizes | 24x30cm, 17x14in, 17x17in | 24x30cm, 17x14in, 17x17in |
| Family | FUJIFILM Flat Panel Detector<br>"D-EVO2" (DR-ID12xxSE)<br>K142003 | FUJIFILM Flat Panel Detector "D-<br>EVO" (DR-ID6XXSE) included in the<br>DR-ID800 of the Sirius Starmobile<br>Tiara<br>K132509 |
| | Subject Device<br>FDR AQRO<br>(DR-XD1000) | Predicate Device K143537, cleared<br>Mar 6, 2015<br>Sirius Starmobile Tiara |
| Scintillator | GOS and CsI | GOS and CsI |
| DQE (RQA5, 1<br>lp/mm, 1mR) –<br>detector alone,<br>without tabletop | 30% (GOS)<br>54% (CsI) | 29% (GOS)<br>53% (CsI) |
| MTF (RQA5, 2<br>lp/mm) | 32% (GOS)<br>54% (CsI) | 32% (GOS)<br>52% (CsI) |
#### SUMMARY OF TECHNOLOGICAL CHARACTERISTICS lll.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the Fujifilm logo. The logo consists of the word "FUJIFILM" in bold, black letters, with a red square above the "I". Below the company name is the slogan "Value from Innovation". The logo is simple and modern.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the Fujifilm logo. The logo consists of the word "FUJIFILM" in a bold, sans-serif font. The "I" in Fujifilm is red and has a small triangle cut out of the top right corner. Below the logo is the tagline "Value from Innovation" in a smaller font.
Even though the subject device is small and compact, it still provides the ability to maneuver and perform all the typical functions required of a mobile x-ray system. The beam coverage of the subject device is equivalent to the predicate device because the focal spot size is the same as predicate. Anode heat capacity of the subiect device is less but the maximum mAs is also small so it will not impact the ability to perform exams. Acceptable image quality can be obtained with the FDR AQRO despite the smaller values of kV and mAs because of the highly sensitive detector system and the VG2 software.
The D-EVO II and D-EVO detectors, DR-ID12xxSE and DR-ID6xxSE used in the FDR AQRO and Sirius Starmobile Tiara respectively, have the same Indications for Use. FDR D-EVO II detectors' scintillator materials (GOS or Csl), indirect conversion method (a-Si). Fuiifilm's unique ISS technology, readout properties. 150um pixel pitch, and 16 bit-depth remain the same as the D-EVO detectors. The MTF and DQE measurements are very similar between FDR D-EVO II and D-EVO detectors. While DR-ID12xxSE and DR-ID6xxSE are wireless detectors, the wireless feature in DR-ID12xxSE has been improved by expanding operating frequency options, and adding new wireless components. The active area size and pixel matrix of the DR-ID12xxSE detectors are only slightly different from the predicate device (K132509). . In general the detectors used with both systems have the same physical and technical characteristics and their performance is substantially equivalent (note that the FDR D-EVOII Flat Panel Detector System (DR-ID1200) received clearance using the FDR D-EVO Flat Panel Detector System (DR-ID600) as the predicate device).
#### SUBSTANTIAL EQUIVALENCE IV.
The FDR AQRO (DR-XD1000) is substantially equivalent to the following legally marketed device:
| Legally Marketed Device | 510(k) # | Clearance Date |
|-------------------------|----------|----------------|
| Sirius Starmobile Tiara | K143537 | 3/6/2015 |
Both systems, the subject device and the predicate (K143537), are intended for use in general purpose radiography for generating radiographic images.
Both the subject device, FDR AQRO (DR-XD 1000), and the predicate device, Sirius Starmobile Tiara (K143537), are mobile X-ray systems and have similar Indications For Use, i.e., qeneral purpose radiography of adult and pediatric patients, although the FDR AQRO is also indicated for use with neonatal patients. The kev technological characteristics of the subject device and the predicate device are
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the Fujifilm logo. The logo is black, except for a small red square with a white triangle cut out of it. Below the logo is the text "Value from Innovation".
similar and the differences do not affect the IFU. Both systems use detectors that are substantially equivalent.
#### > SUMMARY OF STUDIES
Non-clinical Performance Data: The FDR AQRO conformity includes the voluntary standards such as AAMI/ANSI ES60601-1, IEC 60601-1, IEC 60601-1-2, IEC 62304, IEC 62366, IEC 62494-1, DICOM 3.0, IEC 60601-2-54, IEC 60601-1-3, and IEC 60601-1-6. In addition the device complies with the requirements of 21 CFR Subchapter J, Electronic Product Radiation Control. As required by the risk analysis, all verification and validation activities for the FDR AQRO were performed and the results were satisfactory. The submission contains sample phantom images.
Clinical Performance Data: The submission contains sample clinical images.
#### VI. CONCLUSION
Based upon the supporting data summarized above, we concluded the FDR AQRO is as safe and effective as the legally marketed device K143537 and does not raise different questions of safety and effectiveness than K143537.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.