K170842 · Foosin Medical Supplies Inc., Ltd. · NBY · Jul 31, 2017 · General, Plastic Surgery
Device Facts
Record ID
K170842
Device Name
WEGO-PTFE
Applicant
Foosin Medical Supplies Inc., Ltd.
Product Code
NBY · General, Plastic Surgery
Decision Date
Jul 31, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5035
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
WEGO-PTFE is indicated for use in all types of soft tissue approximation and/or ligation, including cardiovascular, dental and general surgeries. The device is not indicated for use in ophthalmic surgery, microsurgery and peripheral neural tissue.
Device Story
WEGO-PTFE is a monofilament, synthetic, non-absorbable surgical suture composed of 100% polytetrafluoroethylene (PTFE). The device consists of the suture strand attached to a stainless steel needle. It is intended for use by surgeons in clinical settings for soft tissue approximation and ligation. The device is provided sterile and is for single use. By providing a secure, non-absorbable closure, the suture facilitates tissue healing and structural integrity in various surgical procedures. The device is evaluated against USP and ISO standards for mechanical performance, biocompatibility, and sterility.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including USP 39-NF 34 performance testing (diameter, needle attachment, tensile strength), biocompatibility testing (ISO 10993 series for genotoxicity, cytotoxicity, implantation, irritation, sensitization, and systemic toxicity), and shelf-life/package integrity testing.
Indicated for soft tissue approximation and/or ligation in cardiovascular, dental, and general surgeries. Contraindicated for ophthalmic surgery, microsurgery, and peripheral neural tissue.
Regulatory Classification
Identification
Nonabsorbable expanded polytetrafluoroethylene (ePTFE) surgical suture is a monofilament, nonabsorbable, sterile, flexible thread prepared from ePTFE and is intended for use in soft tissue approximation and ligation, including cardiovascular surgery. It may be undyed or dyed with an approved color additive and may be provided with or without an attached needle(s).
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 31, 2017
Foosin Medical Supplies Inc., Ltd. % Ms. Diana Hong Mid-Link Consulting Co., Ltd P.O. Box 120-119 Shanghai, 200120 Cn
Re: K170842
Trade/Device Name: WEGO-PTFE Regulation Number: 21 CFR 878.5035 Regulation Name: Nonabsorbable Expanded Polytetrafluoroethylene Surgical Suture Regulatory Class: Class II Product Code: NBY Dated: June 15, 2017 Received: June 20, 2017
Dear Ms. Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# David Krause -S
for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K170842
Device Name WEGO-PTFE
Indications for Use (Describe)
WEGO-PTFE is indicated for use in all types of soft tissue approximation and/or ligation, including cardiovascular, dental and general surgeries. The device is not in ophthalmic surgery, microsurgery and peripheral neural tissue
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Tab #7 510(k) Summary
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.
The assigned 510(k) Number: K170842
- 1. Date of Preparation: 7/23/2017
- 2. Sponsor Identification
Foosin Medical Supplies Inc., Ltd. No.20, Xingshan Road, Weihai Torch Hi-tech Science Park, Weihai, Shandong, 264210, China
Establishment Registration Number: 3006562124
Contact Person: Zhipeng Yang Position: President of Technology Tel: +86-631-5660576 Fax: +86-631-5621680 Email: Foosin-ce@163.com
- 3. Designated Submission Correspondent
Ms. Diana Hong (Primary Contact Person) Ms. Jing Cheng (Alternative Contact Person)
### Mid-Link Consulting Co., Ltd
P.O. Box 120-119, Shanghai, 200120, China
Tel: +86-21-22815850, Fax: 240-238-7587 Email: info@mid-link.net
{4}------------------------------------------------
#### 4. Identification of Proposed Device
Trade Name: WEGO-PTFE Common Name: Non-Absorbable Surgical Suture Size: USP 6-0, USP 5-0, USP 4-0, USP 3-0 and USP 2-0
Regulatory Information Classification Name: Nonabsorbable Expanded Polytetraflouroethylene Surgical Suture Classification: II Product Code: NBY Regulation Number: CFR 878.5035 Review Panel: General & Plastic Surgery
Indications for Use:
WEGO-PTFE is indicated for use in all types of soft tissue approximation and/or ligation, including cardiovascular, dental and general surgeries. The device is not indicated for use in ophthalmic surgery, microsurgery and peripheral neural tissue.
### Device Description
The proposed device, WEGO-PTFE, is monofilament, synthetic, non-absorbable surgical suture composed of 100% polytetrafluoroethylene without any additives. The molecular formula is (C-F2)n. WEGO-PTFE is undyed and uncoated.
The proposed device is composed of suture and needle.
WEGO-PTFE is indicated for use in all types of soft tissue approximation and/or ligation, including cardiovascular, dental and general surgeries. The device is not indicated for use in ophthalmic surgery, microsurgery and peripheral neural tissue.
The sutures are available in a range of gauge sizes and lengths attached to stainless steel needles of varying types and sizes.
WEGO-PTFE complies with the requirements of the European Pharmacopoeia for Sterile Non-Absorbable Strands and the requirements of the United States Pharmacopoeia for Non-Absorbable Surgical Sutures.
- ર : Identification of Predicate Device
510(k) Number: K072076 Product Name: Cytoplast PTFE Suture Manufacturer: Osteogenics Biomedical, Inc.
{5}------------------------------------------------
#### 6. Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- USP 39-NF 34:2016 Nonabsorbable Surgical Suture; >
- USP 39-NF 34:2016 <861> Sutures Diameter; >
- USP 39-NF 34:2016 <871> Sutures Needle Attachment; >
- > USP 39-NF 34:2016 <881> Tensile Strength;
- > ISO 10993-3:2014 Biological Evaluation of Medical Devices, Part 3: Tests for Genotoxicity, Carcinogenicity and Reproductive Toxicity;
- > ISO 10993-5:2009 Biological evaluation of medical devices -- Part 5: Tests for in vitro cytotoxicity:
- > ISO 10993-6:2007 Biological evaluation of medical devices -- Part 6: Tests for local effects after implantation;
- A ISO 10993-10: 2010 Biological evaluation of medical devices -- Part 10: Tests for irritation and skin sensitization ;
- > ISO 10993-11:2006 Biological evaluation of medical devices -- Part 11: Tests for systemic toxicity;
- A USP <85> Bacterial Endotoxins Limit;
- ♪ ASTM F88/F88M-09 Standard Test Method for Seal Strength of Flexible Barrier Materials;
- > ASTM F1140/F1140M-13. Standard Test Methods for Internal Pressurization Failure Resistance of Unrestrained Packages;
- A ASTM F1929-12 Standard Test Method for Detecting Seal Leaks in Porous Medical Package by Dye Penetration;
- > ISO 10993-7:2008 Biological evaluation of medical devices -- Part 7: Ethylene oxide sterilization residuals;
- ♪ ISO 11737-2:2009, Sterilization of medical devices-Microbiological methods-Part 2: Tests of sterility performed in the definition, validation and maintenance of a sterilization process
The following stability testing was performed to support the proposed shelf life: Product performance test reports (three years, two years, one year and six months) Package integrity test reports (three years, two years, one year and six months)
{6}------------------------------------------------
#### 7. Clinical Test Conclusion
No clinical study is included in this submission.
- 8. Substantially Equivalent (SE) Comparison
| Item | Proposed Device<br>WEGO-PTFE | Predicate Device<br>Cytoplast PTFE Suture<br>K072076 |
|--------------------------------|------------------------------------------------|------------------------------------------------------|
| Product Code | NBY | NBY |
| Regulation Number | CFR 878.5035 | CFR 878.5035 |
| Class | II | II |
| Intended Use | WEGO-PTFE is indicated for use in | The Cytoplast PTFE suture is indicated |
| | all types of soft tissue approximation | for use in all types of soft tissue |
| | and/or ligation, including | approximation and/or ligation, including |
| | cardiovascular, dental and general | cardiovascular, dental and general |
| | surgeries. The device is not indicated | surgeries, as well as repair of the dura |
| | for use in ophthalmic surgery, | mater. The device is not indicated for use |
| | microsurgery and peripheral neural | in ophthalmic surgery, microsurgery and |
| | tissue. | peripheral neural tissue. |
| Material | Polytetrafluoroethylene | Polytetrafluoroethylene |
| Color | Undyed | Undyed |
| Absorbable /<br>Non-absorbable | Non-absorbable | Non-absorbable |
| Braided /<br>Monofilament | Monofilament | Monofilament |
| Sterility | EO Sterilized | EO Sterilized |
| Suture Size | USP 6-0, USP 5-0, USP 4-0, USP 3-0,<br>USP 2-0 | USP 4-0, USP 3-0, USP 2-0 |
| Single Use | Yes | Yes |
| Performance | Comply with: | Comply with: |
| | USP 39 <861> | USP 30 <861> |
| | USP 39 <871> | USP 30 <871> |
| | USP 39 <881> | USP 30 <881> |
| Biocompatibility | Conforms to the requirements of ISO | Conforms to the requirements of ISO |
| | 10993 series Standards | 10993 series Standards |
- Table 1 Comparison of Proposed Device and Predicated Device (K072076)
{7}------------------------------------------------
Ref.:M0232015
#### 9. Substantially Equivalent (SE) Conclusion
Based on the comparison, the intended use, characteristic, materials, sterility of proposed device is determined to be Substantially Equivalent (SE) to the predicate device.
In addition, the results of performance tests performed on the proposed device can also demonstrate the proposed device is complied with FDA recognized standards, which the predicate device was also complied with. The results of biocompatibility studies performed on the proposed device demonstrate that the patient materials used in proposed device are biocompatible.
Based on the comparison above, the proposed device, WEGO-PTFE, is determined to be Substantially Equivalent (SE) to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.