← Product Code [LLZ](/productcode/LLZ) · K170841

# BiopSee, Mobile US (K170841)

_Medcom GmbH · LLZ · Jul 26, 2017 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K170841

## Device Facts

- **Applicant:** Medcom GmbH
- **Product Code:** [LLZ](/productcode/LLZ.md)
- **Decision Date:** Jul 26, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The BiopSee bundled system is a combination of MobileUS providing the "Medical Image Processing Workstation" which acquires and displays ultrasound images from external ultrasound scanners and the BiopSee "Image Fusion Navigation Software" which is a stereotaxic accessory for Ultrasound (US) devices. The BiopSee system is intended to be used in clinical interventions and for anatomical structures in a clinical setting where US imaging is currently used for visualizing such procedures. It displays the simulated insertion tool such as a biopsy needle on a computer monitor that shows images of the target organs and the projected future path of the interventional instrument. The patient population is mainly people who need prostate biopsy. It is intended to be used mainly in the Urological departments for prostate biopsies. The BiopSee system is not intended to be used as a standalone diagnostic image device, since it represents information of a patient that could not be congruent with the current (actual) patient position and shall therefore always be seen as an additional source of information. The BiopSee system when connected with an electromagnetic tracker should not be used on or around persons with a cardiac pacemaker, and should not be used around life supporting equipment. BiopSee is a prescription device.

## Device Story

BiopSee is a medical image processing workstation and stereotaxic navigation accessory for ultrasound (US) procedures; primarily used in urology for prostate biopsies. System inputs include live 2D US images from external scanners, 3D DICOM datasets (MRI, CT, PET), and tracking data from electromagnetic sensors or mechanical steppers. Software reconstructs 3D volumes from 2D sequences; performs image fusion of multi-modal datasets; allows organ delineation and planning of instrument access paths. During intervention, system displays live US images overlaid with planned targets and projected needle paths. Operated by physicians in clinical settings to guide biopsy needle placement. Output provides visual guidance for instrument navigation; does not perform automated diagnosis. Benefits include improved visualization of target organs and instrument positioning relative to pre-operative imaging. System is not a standalone diagnostic device; requires physician verification of registration accuracy.

## Clinical Evidence

Bench testing only. Integration testing using CIRS phantoms confirmed software-projected virtual needle position accuracy with an overall error < 5 mm. Electromagnetic guidance system (Ascension 3D Guidance trakSTAR) accuracy stated as 1.4 mm RMS and 0.5 degrees RMS. Length measurement accuracy reported as 5% or better for US images and 5-10% for secondary modality images.

## Technological Characteristics

Workstation hardware (MobileUS) includes CPU, monitor, keyboard, mouse, and medical trolley. Connectivity via DICOM 3.0. Tracking via electromagnetic measurement system (Ascension 3D Guidance trakSTAR) or mechanical stepper. Software runs on standard PCs with Windows OS. Complies with ISO 14971, IEC 62304, IEC 60601-1, and IEC 60601-1-2.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- MIM 3.5 (CIRCA) ([K052379](/device/K052379.md))
- UroNav Fusion Biopsy System ([K153073](/device/K153073.md))
- NaviSuite SSI Edition ([K163119](/device/K163119.md))
- Oncentra Prostate 4.0 ([K112420](/device/K112420.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 26, 2017

MedCom GmbH % Mr. Johannes Messow Quality Manager Dolivostrasse 11 Darmstadt, Hessen 64293 GERMANY

Re: K170841

Trade/Device Name: BiopSee, Mobile US Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: June 21, 2017 Received: June 26, 2017

Dear Mr. Messow:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Michael D. O'Hara For

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K170841

Device Name BiopSee, Mobile US

#### Indications for Use (Describe)

The BiopSee bundled system is a combination of MobileUS providing the "Medical Image Processing Workstation" which acquires and displays ultrasound images from external ultrasound scanners and the BiopSee "Image Fusion Navigation Software" which is a stereotaxic accessory for Ultrasound (US) devices.

The BiopSee system is intended to be used in clinical interventions and for anatomical structures in a clinical setting where US imaging is currently used for visualizing such procedures. It displays the simulated insertion tool such as a biopsy needle on a computer monitor that shows images of the target organs and the projected future path of the interventional instrument.

The patient population is mainly people who need prostate biopsy.

It is intended to be used mainly in the Urological departments for prostate biopsies.

The BiopSee system is not intended to be used as a standalone diagnostic image device, since it represents information of a patient that could not be congruent with the current (actual) patient position and shall therefore always be seen as an additional source of information.

The BiopSee system when connected with an electromagnetic tracker should not be used on or around persons with a cardiac pacemaker, and should not be used around life supporting equipment. BiopSee is a prescription device.

| Type of Use (Select one or both, as applicable)                                  |                                                                                 |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span>      Prescription Use (Part 21 CFR 801 Subpart D)    </span> </div> | <div> <span>      Over-The-Counter Use (21 CFR 801 Subpart C)    </span> </div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# III. 510(k) Summary of Safety and Effectiveness

### A. Submitter

| Name:               | MedCom GmbH                                     |
|---------------------|-------------------------------------------------|
| Address:            | Dolivostr. 11<br>Darmstadt, HE 64293<br>Germany |
| Establishment Reg.: | 3006579682                                      |
| Telephone:          | +49 (6151) 95147-285                            |
| Fax:                | +49 (6151) 95147-20                             |
| Contact:            | Mr. Johannes Messow<br>jmessow@medcom-online.de |
| Date:               | March 27, 2017                                  |

- B. Device

| Trade Name:     | BiopSee                 | MobileUS                                                                |
|-----------------|-------------------------|-------------------------------------------------------------------------|
| Common name:    | BiopSee, BiopSee System | MobileUS, MobileUS Stand-<br>Alone                                      |
| Classification: | Regulatory Class:       | II                                                                      |
|                 | Product Code:           | LLZ / Classification Name:<br>System, Image Processing,<br>Radiological |
|                 | CFR Section:            | 892.2050                                                                |
|                 | Panel:                  | Radiology                                                               |

{4}------------------------------------------------

# C. Predicate Devices

# BiopSee:

| Device trade name:     | MIM 3.5 (CIRCA)                          |
|------------------------|------------------------------------------|
| 510(k) number:         | K052379                                  |
| Company name:          | MIMVISTA CORP.                           |
| Classification Number: | 892.2050                                 |
| Classification:        | Class II                                 |
| Product code:          | LLZ                                      |
| Device trade name:     | UroNav Fusion Biopsy System              |
| 510(k) number:         | K153073                                  |
| Company name:          | Philips Medical Systems International BV |
| Classification Number: | 892.2050                                 |
| Classification:        | Class II                                 |
| Product code:          | LLZ                                      |

| Device trade name:     | NaviSuite SSI Edition |
|------------------------|-----------------------|
| 510(k) number:         | K163119               |
| Company name:          | MedCom GmbH.          |
| Classification Number: | 892.2050              |
| Classification:        | Class II              |
| Product code:          | LLZ                   |

#### MobileUS:

| Device trade name:     | UroNav Fusion Biopsy System              |
|------------------------|------------------------------------------|
| 510(k) number:         | K153073                                  |
| Company name:          | Philips Medical Systems International BV |
| Classification Number: | 892.2050                                 |
| Classification:        | Class II                                 |
| Product code:          | LLZ                                      |
| Device trade name:     | Oncentra Prostate 4.0                    |
| 510(k) number:         | K112420                                  |
| Company name:          | Nucletron Corporation                    |
| Classification Number: | 892.5700                                 |
| Classification:        | Class II                                 |
| Product code:          | JAQ                                      |
| Device trade name:     | NaviSuite SSI Edition                    |
| 510(k) number:         | K163119                                  |
| Company name:          | MedCom GmbH.                             |
| Classification Number: | 892.2050                                 |
| Classification:        | Class II                                 |
| Product code:          | LLZ                                      |

{5}------------------------------------------------

#### D. Reason for Submission

New device application

#### E. Standards

1. ISO 14971:2007, Medical devices - Application of risk management to medical devices. (General I (QS/RM))

2. IEC 62304:2006, Medical Device Software - Software Life Cycle Processes. (Software/Informatics)

3. IEC 60601-1:2005/A1:2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance

4. IEC 60601-1-2 Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements And Tests

#### F. Description

The BiopSee bundled system (also called BiopSee System) is a medical image processing workstation that provides image quided instrument navigation. The hardware part (MobileUS) consists of the system unit which includes a CPU, a monitor, medical-grade keyboard, mouse and power supply and a medical trolley. It is designed to display ultrasound images from different sources. Optionally it can be equipped with an electromagnetic positioning system to track the ultrasound probe or other medical instruments.

The system is able to load different image modalities such as Magnetic Resonance Imaging (MRI), Computed Tomography (CT), Positron emission tomography (PET) and allows organ delineation and planning instrument access paths within the 3D dataset. It is designed to display 2D images from commercially available ultrasound systems and either reconstruct a 3D ultrasound image or fuse the images with pre-loaded other datasets.

The BiopSee software can be installed on standard PCs under Microsoft Windows operating systems which follow the minimum requirements described in the service manual, similar to MobileUS hardware. BiopSee is designed to work in combination with an ultrasound scanner (the Ultrasound System is used in order to display the live US images) and a probe, a positioning system (to track the position of the probe) and medical navigation instruments (to track the position of the medical instruments like needles). The ultrasound scanners connected must be tested with the BiopSee System and shall be legally marketed in the US.

{6}------------------------------------------------

Examples of compatible devices are:

| Device                                            | Type                 | 510(k) / registration number |
|---------------------------------------------------|----------------------|------------------------------|
| System, Imaging,<br>Pulsed Doppler,<br>Ultrasonic | HITACHI Noblus       | K160559                      |
|                                                   | BK 2300              | K151910                      |
|                                                   | Supersonic AIXPLORER | K142100                      |
|                                                   | ESAOTE 6200 My Lab   | K153277                      |

It can also be connected to a commercially available mechanical stepper in order to acquire the precise position of the ultrasound probe. Alternatively an Electromagnetic Measurement System (EMMS) can be integrated in the MobileUS hardware (or equivalent) to accurately track the position of the ultrasound probe.

Within BiopSee, one is able to: acquire, process and display ultrasound images and create 3D volumes out of 2D image sequences. The software is also capable of loading and displaying 3D DICOM 3 image series.

In more details the system allows organ delineation and planning instrument access paths within a 3D dataset. To enhance the procedure planning the user can load multiple datasets and perform image fusion.

During the navigation procedure, the system displays live ultrasound image overlaid by the planning objects, thus guiding the user through medical instrument placement at the intended position. The planned access paths and targets can be updated reflecting the real situation for proper documentation. The physician can attach a commercially available biopsy needle guide to the ultrasound probe and use it in combination with a prostate biopsy needle to perform tissue biopsy. Similarly the physician can attach a template needle quide to the stepper in order to insert prostate biopsy needles to the desired positions.

{7}------------------------------------------------

Image /page/7/Figure/2 description: The image is a system overview diagram for the BiopSee application. It illustrates the connections and interactions between different modules and components. The diagram includes sections for Foreign Devices, External Applications, BiopSee Application, File System, 3rd Party Modules, and External Modules, showing how they are interconnected and how data flows between them.

System Overview

#### G. Intended Use

The BiopSee bundled system is a medical imaging workstation for medical professionals, which is designated to support image guided instrument navigation. Additionally, the BiopSee software can be used for a general ultrasound examination.

Within BiopSee, one is able to: acquire, process and display ultrasound images and create 3D volumes out of 2D image sequences. The software is also capable of loading and displaying 3D DICOM 3 image series.

The system allows organ delineation and planning instrument access paths within a 3D dataset. To enhance the procedure planning the user can load multiple datasets and perform image fusion.

During the navigation procedure, the system displays live ultrasound image overlaid by the planning objects, thus guiding the user through medical instrument placement at the intended position. The planned access paths and targets can be updated reflecting the real situation for proper documentation.

{8}------------------------------------------------

BiopSee runs on standard PCs under Microsoft Windows operating systems. It is designed to work with compatible commercially available ultrasound scanners and probes, positioning systems and medical navigation instruments.

#### H. Indications for Use

The BiopSee bundled system is a combination of MobileUS providing the "Medical Image Processing Workstation" which acquires and displays ultrasound images from external ultrasound scanners and the BiopSee "Image Fusion Navigation Software" which is a stereotaxic accessory for Ultrasound (US) devices.

The BiopSee system is intended to be used in clinical interventions and for anatomical structures in a clinical setting where US imaging is currently used for visualizing such procedures. It displays the simulated image of a tracked insertion tool such as a biopsy needle on a computer monitor that shows images of the target organs and the current and the projected future path of the interventional instrument.

The patient population is mainly people who need prostate biopsy.

It is intended to be used mainly in the Urological departments for prostate biopsies.

The BiopSee system is not intended to be used as a standalone diagnostic image device, since it represents information of a patient that could not be congruent with the current (actual) patient position and shall therefore always be seen as an additional source of information.

The BiopSee system when connected with an electromagnetic tracker should not be used on or around persons with a cardiac pacemaker, and should not be used around life supporting equipment.

BiopSee is a prescription device.

#### 1. Technological Comparison to Predicate Devices

#### The device BiopSee is substantially equivalent to the following predicate devices:

| K052379 | MIM 3.5 (CIRCA)             |
|---------|-----------------------------|
| K153073 | UroNav Fusion Biopsy System |
| K163119 | NaviSuite SSI Edition       |

The predicate device MIM 3.5 (CIRCA) has the same main functionalities and characteristics as BiopSee for both transrectal and transperineal approaches and uses also commercially available electronic stepper and electromagnetic tracking devices.

The predicate device UroNav (Version 2.0) and BiopSee are adequately similar, with the exception of the transperineal version of BiopSee, having the same intended use as well as the same main functionalities and characteristics. The difference in the transperineal version of BiopSee is that an electronically tracked stepper is used whereas UroNav uses an electromagnetically tracked stepper.

The software of the predicate device NaviSuite is also developed by MedCom, and includes the main functionalities for tracked needles version.

{9}------------------------------------------------

Refer to section X for a detailed predicate device comparison.

| devices: |                             |
|----------|-----------------------------|
| K153073  | UroNav Fusion Biopsy System |
| K112420  | Oncentra Prostate 4.0       |
| K163119  | NaviSuite SSI Edition       |

# The device MobileUS is substantially equivalent to the following predicate

The predicate device UroNav Fusion Biopsy System is substantially equivalent with MobileUS since it is a mobile computer workstation, a commercially available navigation device and software with functions that are commonly found in various medical imaging applications. The MobileUS is similar to UroNav since it does not directly contact the patient, nor does it control any life sustaining devices. Diagnosis is not performed by the MobileUS system but by Radiologists, Urologists, Clinicians and referring Physicians.

The predicate device Oncentra Prostate 4.0 encompasses the majority of the MobileUS functionality. Actually MobileUS is a small subset of Oncentra prostate, in particular: ultrasound image display, encoder status and position display and grid overlay display. Therefore they are adequately similar. The exception is support for electromagnetic tracking which Oncentra Prostate doesn't have.

The predicate device NaviSuite SSI Edition has similar functionality with respect to displaying ultrasound image and processing data from the electromagnetic tracking system. The functionality of MobileUS is a subset of that of NaviSuite SSI Edition. Both systems use similar tracking hardware from the same manufacturer.

Refer to section X for a detailed predicate device comparison.

{10}------------------------------------------------

#### J. Non-clinical Performance Data

Non-clinical verification and validation software tests were conducted to confirm that the BiopSee System meets its intended use and is safe and effective.

#### System Accuracy Test:

The software cannot determine the accuracy of the treatment procedure. The Therapist or Physician is responsible for verifying the correctness of the registration. For this means the software offers side by side and overlay views (with a real-time US image).

Integration testing for needle navigation using CIRS Phantoms to confirm the software projected virtual needle position over the real time US image, showed accuracy with an overall error better than 5 mm.

#### Tracking Quality Test:

The system can detect low sensor tracking quality and is able to show if the tracking is distorted (for example by metallic objects etc.).

Tests for verifying the tracking accuracy and sensor in range status have been carried out during Integration tests.

The BiopSee system uses an integrated "Ascension 3D Guidance trakSTAR" as electromagnetic guidance system.

Statement: The accuracy for the electromagnetic quidance system "Ascension 3D Guidance trakSTAR" is stated by it manufacturer NDI Europe with: Accuracy: 1.4 mm RMS, 0.5 degrees RMS

# System Accuracy

#### 2nd Modality Image Display

The system accuracy concerning the matching between the ultrasound image and the 20d modality image depends highly on the registration process and the condition of the patient compared to the situation when the 2nd modality (i.e. the CT scan) has been acquired.

However, from the technical point of view an "overlapping" accuracy of at least 5mm can be achieved under the presumption that the registration has been done with an overall error better than 5mm.

#### Length measurement (Ruler Function)

In the ultrasound image: 5% error, or better. In the 2nd modality image: 5%-10%, or better. Please note that this value depends on the quality of the 2nd modality dataset.

{11}------------------------------------------------

# K. Conclusion

Based on the information provided in this Premarket Notification MedCom concludes that BiopSee bundled system is as safe and effective and substantially equivalent to the predicate devices described herein.

---

**Source:** [https://fda-staging.innolitics.com/device/K170841](https://fda-staging.innolitics.com/device/K170841)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
