K170831 · Epiep, Inc. · DYB · May 9, 2017 · Cardiovascular
Device Facts
Record ID
K170831
Device Name
EpiAccess System
Applicant
Epiep, Inc.
Product Code
DYB · Cardiovascular
Decision Date
May 9, 2017
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1340
Device Class
Class 2
Indications for Use
The EpiAccess System with introducer needle and integrated needle tip pressure transducer is intended to access the epicardial surface of the heart via a subxiphoid approach to facilitate guidewire placement into the pericardial space in electrophysiology procedures in adult patients.
Device Story
System facilitates epicardial access during electrophysiology procedures; utilizes sterile, single-use Tuohy needle with integrated distal pressure transducer; connects to reusable Control Unit (Gen2). Input: pressure signals from needle tip; output: real-time pressure/frequency information displayed on Cath Lab monitor. Used in clinical setting by physicians; needle placed under fluoroscopic guidance. System provides pressure data to assist physician in determining needle tip location based on anatomical pressure differences; does not provide clinical decision guidance or alerts. Modification from predicate involves reduced Control Unit size, removal of dedicated touch screen, and integration with existing lab display monitors. Benefits include improved convenience for epicardial access.
Clinical Evidence
Bench testing only. Verification and validation testing performed to demonstrate compliance with design requirements and equivalence to predicate. Testing included electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), usability (IEC 62366), and software validation (BS EN ISO 62304). No clinical data presented.
Technological Characteristics
System comprises single-use Tuohy needle with integrated pressure transducer and reusable Control Unit (Gen2). Connectivity: video output to external Cath Lab display monitor. Standards: IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 62366 (usability), BS EN ISO 62304 (software). Software: embedded, validated, rule-based algorithm for pressure/frequency processing. Sterilization: single-use needle component.
Indications for Use
Indicated for adult patients undergoing electrophysiology procedures requiring access to the epicardial surface of the heart via a subxiphoid approach to facilitate guidewire placement into the pericardial space.
Regulatory Classification
Identification
A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. The bottom half of the circle contains a stylized image of three human profiles facing to the right, with the profiles overlapping each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 09, 2017
EpiEP, Inc. % Elsa Abruzzo CEO/Founder Cygnus Regulatory 3753 Vineyard Place Cincinnati, Ohio 45226
Re: K170831
Trade/Device Name: EpiAccess System (with A0005 EpiAccess Control Unit (Gen2)) Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class: Class II Product Code: DYB Dated: April 12, 2017 Received: April 19, 2017
Dear Elsa Abruzzo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
M.A. Willemen
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K170831
Device Name
EpiAccess System (with A0005 EpiAccess Control Unit (Gen2))
Indications for Use (Describe)
The EpiAccess System with introducer needle and integrated needle tip pressure transducer is intended to access the epicardial surface of the heart via a subxiphoid approach to facilitate guidewire placement into the pericardial space in electrophysiology procedures in adult patients.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|
| <span style="text-decoration: underline;">Prescription Use (Part 21 CFR 201 Subpart D)</span> | <span style="text-decoration: underline;">Over-The-Counter Use (21 CFR 201.66)</span> |
| <span style="text-decoration: underline;">☑</span> | <span style="text-decoration: underline;">☐</span> |
| X Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary EpiEP, Inc.'s EpiAccess System
## Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
#### EpiEP, Inc.
Phone: 1-434-951-0444 Facsimile: 1-434-951-0478
Contact Person: Elsa Chi Abruzzo, RAC, FRAPS Date Prepared: March 12, 2017
#### Name/Address of Sponsor
EpiEP, Inc. 195 Church Street, 17th Floor,
New Haven, CT 06510
| Trade Name: | EpiAccess System (with A0005 EpiAccess Control Unit (Gen2)) | | |
|---------------------------------|-------------------------------------------------------------|------------------|---------|
| Common or Usual Name | Epicardial Introducer System | | |
| Classification Name | Introducer, Catheter | | |
| Classification: | Class II | | |
| Product Code and<br>Regulation: | DYB, 21 CFR 870.1340 | | |
| Classification Panel: | Cardiovascular | | |
| Predicate Devices: | EpiEP, Inc. | EpiAccess System | K142245 |
#### Purpose of the Special 510(k) notice.
The device is a modification to previously cleared EpiEP, Inc. EpiAccess System.
#### Intended Use / Indications for Use
The EpiAccess System with introducer needle and integrated needle tip pressure transducer is intended to access the epicardial surface of the heart via a subxiphoid approach to facilitate guidewire placement into the pericardial space in electrophysiology procedures in adult patients.
{4}------------------------------------------------
## Technological Characteristics
The modified EpiAccess System consists of the following components and accessories: a sterile, single-use (disposable) Tuohy Type Needle with an integrated, distal, needle tip pressure transducer (sensor), and a reusable integrated Control Unit (Gen2). The EpiAccess Needle component remains the same as with the predicate device.
The reusable Control Unit (Gen2) has been modified for a reduction in overall size, which eliminates the need for a custom cart. The modified Control Unit component also allows for connection to an existing display monitor in the Cath Lab rather than having a dedicated touch screen LCD display monitor as with the predicate device.
Both Control Unit components are for use with the cleared EpiAccess Needle component and have an input for a commercial A-Line component and video output to a display monitor. Both the modified and the predicate device use embedded, validated software with the same algorithm to provide the same pressure/frequency information to the user via the same graphical user interface screen for the same intended use.
The EpiAccess System displays the pressure measurement information, which physicians may be able to use to determine needle tip location based on known anatomical pressure differences. This additional information regarding needle tip location is an added convenience feature over standard Tuohy introducer needles for epicardial access. The system does not alert the user to tip location or provide clinical decision guidance. The EpiAccess Needle is placed under visualization with fluoroscopic imaging standard for electrophysiology procedures.
Both control unit components are reusable and not patient contacting. Both control units have been tested and passed for electrical safety and EMC, reliability, and transit testing.
#### Performance Data
Bench tests were conducted on the modified EpiAccess System with the Control Unit (Gen2) to demonstrate that it meets defined design requirements and can perform in a manner equivalent to EpiAccess System predicate device (K142245). Testing included verification and validation testing.
Manufacturing and traceability of devices tested were conducted in accordance with 21 CFR 820 Good Manufacturing Practices and BS EN ISO 13485:2016 Medical Devices – Quality Management Systems - Requirements for Regulatory Purposes. In all instances, the modified EpiAccess System functioned as intended and results observed were as expected. These test results confirm that EpiAccess System is safe, meets the design inputs, and raises no new safety or efficacy concerns. A summary of the EpiAccess Systems design control activities with regards to risk analysis and verification and validation activities is provided in this 510 (k) submission.
{5}------------------------------------------------
Nonclinical testing included:
Performance testing per the applicable sections of:
| BS EN ISO 14971:2012 | Medical Devices – Application of Risk |
|----------------------|---------------------------------------|
| | Management to Medical Devices |
Electrical Safety and Electromagnetic Compatibility (EMC) per the applicable sections of:
| IEC 60601-1 Edition 3.1 | Medical Electrical Equipment – Part 1: General<br>Requirements for Basic Safety and Essential<br>Performance |
|-------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ANSI/AAMI/IEC 60601-1-2 Edition 4.0 | Medical Electrical Equipment – Part 1-2: General<br>Requirements for Basic Safety and Essential<br>Performance - Collateral Standard: Electromagnetic<br>Disturbances - Requirements and Tests |
| IEC 60601-1-6 Edition 3.1 | Medical Electrical Equipment - Part 1-6: General<br>Requirements for Basic Safety and Essential<br>Performance - Collateral Standard: Usability |
Packaging validation in compliance with applicable sections of:
| ASTM D4169-16 | Standard Practice for Performance Testing of |
|---------------|----------------------------------------------|
| | Shipping containers and Systems |
Software validation in compliance with applicable sections of:
| BS EN ISO 62304:2015 | Medical Device Software – Software Life Cycle<br>Processes |
|-----------------------------|--------------------------------------------------------------------------------------------------|
| FDA Guidance (Jan 11, 2002) | General Principles of Software Validation; Final<br>Guidance for Industry and FDA Staff |
| FDA Guidance (May 11, 2005) | Guidance for the Content of Premarket<br>Submission for Software Contained in Medical<br>Devices |
| FDA Guidance (Sep 9, 1999) | Off-The-Shelf Software Use in Medical Devices |
Usability and Human Factors Testing in compliance with applicable sections of:
| IEC 62366:2015 | Medical Devices - Application of Usability |
|----------------|--------------------------------------------|
| | Engineering to Medical Devices |
# Substantial Equivalence
The EpiAccess System has the same intended use and indications for use, and similar technological characteristics as the predicate device. The technological characteristics of the EpiAccess System with the Control Unit (Gen2) modification are substantially equivalent to the predicate device EpiAccess System (K142245). Where differences exist between the proposed device and the predicate device, performance testing demonstrated that these differences do not adversely affect the safety and effectiveness of the proposed device or different question of safety and efficacy.
{6}------------------------------------------------
# Conclusions
The EpiAccess System modified with the Control Unit (Gen2) has the same intended use, indications, principles of operation, and technological characteristics as the EpiEP, Inc.'s previously cleared EpiAccess System (K142245). The minor differences in dimensions and design of the Control Unit (Gen2) component for the modified EpiAccess System do not raise any new questions of safety or effectiveness. Thus, EpiAccess System modified with the Control Unit (Gen2) is substantially equivalent to its predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.