K170830 · Pioneer Surgical Technology, Inc., Dba Rti Surgical, Inc. · OVE · Jun 28, 2017 · Orthopedic
Device Facts
Record ID
K170830
Device Name
Unison-C Anterior Cervical Fixation System
Applicant
Pioneer Surgical Technology, Inc., Dba Rti Surgical, Inc.
Product Code
OVE · Orthopedic
Decision Date
Jun 28, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Unison-C Anterior Cervical Fixation System is indicated for stand-alone anterior cervical interbody fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one level from C2 to T1. Degenerative disc disease is defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The system is intended to be used with autograft bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion, and is implanted via an anterior approach. Implants must be used with two of the provided bone screws. This system is to be used in patients who have had six weeks of non-operative treatment.
Device Story
Intervertebral body fusion system for stand-alone anterior cervical procedures; provides structural stability post-discectomy. System comprises PEEK-OPTIMA LT1 polymer interbody implants with tantalum radiographic markers and titanium alloy screws/locking mechanism. Implanted via anterior approach; requires two bone screws for fixation; no supplemental fixation needed. Used by surgeons in clinical settings to facilitate fusion using autograft/allogenic bone graft. Radiographic pins allow visualization of fusion status. Benefits include structural support and stabilization of cervical spine segments.
Clinical Evidence
No clinical data required; substantial equivalence supported by non-clinical bench testing including static/dynamic axial compression, compressive shear, torsion, subsidence, locking mechanism disassociation, screw push-out, and torque-to-failure testing per ASTM F2077-14 and ASTM F2267-04.
Technological Characteristics
Materials: PEEK-OPTIMA LT1 (ASTM F2026), Tantalum (ASTM F560), Titanium alloy (ASTM F136). Principle: Intervertebral body fusion with integrated locking mechanism. Form factor: Cervical interbody implant with screws. Sterilization: Gamma irradiation. No software/electronics.
Indications for Use
Indicated for skeletally mature patients with degenerative disc disease (DDD) at one level from C2 to T1, defined as discogenic neck pain with radiographic confirmation, following six weeks of failed non-operative treatment.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
Predicate Devices
Unison-C Anterior Cervical Fixation System (K152793)
{0}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Pioneer Surgical Technology, Inc. DBA RTI Surgical, Inc. Sarah Pleaugh Regulatory Affairs Specialist 375 River Park Circle Marquette, Michigan 49855
June 28, 2017
Re: K170830
Trade/Device Name: Unison®-C Anterior Cervical Fixation System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: OVE Dated: June 1, 2017 Received: June 2, 2017
Dear Ms. Pleaugh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K170830
Device Name
Unison®-C Anterior Cervical Fixation System
#### Indications for Use (Describe)
The Unison-C Anterior Cervical Fixation System is indicated for stand-alone anterior cervical interbody fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one level from C2 to T1. Degenerative disc disease is defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The system is intended to be used with autograft bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion, and is implanted via an anterior approach. Implants must be used with two of the provided bone screws. This system is to be used in patients who have had six weeks of non-operative treatment.
Type of Use (Select one or both, as applicable)|X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for RTI Surgical. The logo consists of the letters "rti" in a blue square, followed by the word "surgical" in gray. The "i" in "rti" has a small green dot above it. The logo is simple and modern, and the colors are clean and professional.
| 510(k) Owner / Submitter: | Pioneer Surgical Technology, Inc.<br>dba RTI Surgical, Inc.<br>375 River Park Circle<br>Marquette, MI 49855<br>Telephone: 906.226.9909<br>Fax: 906.225.5868 |
|----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Official Contact Person: | Sarah Pleaugh, RAC<br>Regulatory Affairs Specialist |
| Date Prepared: | June 1, 2017 |
| Trade or Proprietary Name: | Unison®-C Anterior Cervical Fixation System |
| Common or Usual Name: | Intervertebral Fusion Device With Bone Graft, Cervical |
| Classification: | Class II per 21 CFR §888.3080 Device Classification |
| Product Code: | OVE |
| Classification Panel: | Division of Orthopedic Devices |
| Predicate Devices: | Primary: K152793 Unison-C Anterior Cervical Fixation System<br>(Pioneer Surgical Technology, Inc. DBA RTI Surgical, Inc.)<br><br>Reference: K141314 SCARLET®AC-T Secured Anterior<br>Cervical Cage (SPINEART) |
## DESCRIPTION OF THE DEVICE
510(k) SUMMARY
The Unison-C Anterior Cervical Fixation system is intended for stand-alone cervical interbody fusion procedures to provide structural stability in skeletally mature individuals following discectomy. The system includes intervertebral body implants (with an integrated locking mechanism and radiographic pins) and screws, and does not require the use of supplemental fixation.
These implants are manufactured from a radiolucent polymer (PEEK-OPTIMA® LT1 polymer from INVIBIO® Biomaterial Solutions (polyether ether ketone) (ASTM F2026)), to allow radiographic imaging inside the implant to evaluate fusion status. The implants are assembled with radiographic markers composed of tantalum (ASTM F560) to facilitate proper implant positioning. The screws and locking mechanism are manufactured from of titanium alloy (ASTM F136). Sterile implants are provided sterile by gamma irradiation.
Implants are supplied with instrumentation necessary to facilitate the insertion and removal of the implants, as well as general manual surgical instruments. Cases and caddies are supplied for sterilization and transport of the non-sterile implants and instruments.
The purpose of this submission is to obtain clearance of modifications to the predicate system (K152793).
{4}------------------------------------------------
## INDICATIONS FOR USE
The Unison-C Anterior Cervical Fixation System is indicated for stand-alone anterior cervical interbody fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one level from C2 to T1. Degenerative disc disease is defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The system is intended to be used with autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion, and is implanted via an anterior approach. Implants must be used with two of the provided bone screws. This system is to be used in patients who have had six weeks of non-operative treatment.
## TECHNOLOGICAL CHARACTERISTICS
The subject and predicate devices have the same or similar technological characteristics, indications for use, materials of manufacture and principles of operation. The minor design modifications do not raise any new issues of safety and effectiveness.
## NON-CLINICAL PERFORMANCE
The subject Unison-C Anterior Cervical Fixation System has been tested in the following test modes:
- Static axial compression per ASTM F2077-14 .
- .. Static compressive shear per ASTM F2077-14
- Static torsion per ASTM F2077-14 "
- Static Subsidence per ASTM F2267-04 =
- Dynamic axial compression fatique per ASTM F2077-14 =
- Dynamic compressive shear per ASTM F2077-14 "
- Dynamic torsion per ASTM F2077-14 י
- Locking mechanism disassociation testing =
- Screw push-out 트
- Screw-interbody torque-to-failure testing "
- Locking mechanism torque-to-failure testing ..
The results of this non-clinical testing demonstrate that the subject Unison-C Anterior Cervical Fixation System is substantially equivalent to the predicate system.
Pyrogenicity of the sterile devices was evaluated using the Limulus amebocyte lysate (LAL) assay. The device will be tested to ensure the endotoxin level meets the requirements of maximum endotoxin limit for implantable medical devices [20 EU per device].
#### CLINICAL PERFORMANCE
No clinical performance data was required for a determination of substantial equivalence.
#### CONCLUSION OF SUBSTANTIAL EQUIVALENCE
The subject Unison-C Anterior Cervical Fixation System is substantially equivalent to the predicates based on a comparison of overall technological characteristics, principles of operation, indications for use, materials, sterilization, packaging, and mechanical performance.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.